Library › Devices and mechanical support › Cardiac resynchronization therapy
Cardiac resynchronization therapy
Biventricular and conduction system resynchronization in adults: patient selection by guideline class, the randomized evidence, implantation and lead position, programming, response and nonresponse, and follow-up.
Cardiac resynchronization therapy (CRT) paces the left ventricle so that a ventricle activated late by left bundle branch block, or by right ventricular pacing, contracts in step with the septum. Biventricular CRT is a Class 1 recommendation for adults with a left ventricular ejection fraction (LVEF) of 35% or less, sinus rhythm, left bundle branch block (LBBB) with a QRS duration of 150 ms or more, and New York Heart Association (NYHA) class II to ambulatory IV symptoms on guideline-directed medical therapy (GDMT). It reduces death and heart failure hospitalization. Benefit narrows with shorter QRS duration and non-LBBB morphology, and CRT increased mortality when the QRS was narrow. Effective biventricular capture of nearly 100% of beats is required. Conduction system pacing is an established alternative when coronary sinus pacing fails, and its randomized comparison with biventricular pacing has given mixed results.
Definition, device types, and terminology of resynchronization
Cardiac resynchronization therapy is pacing intended to restore coordinated ventricular activation in a patient whose ventricles are activated in a delayed, sequential pattern. The 2023 Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society (HRS/APHRS/LAHRS) guideline groups all pacing strategies with this aim under the term cardiac physiologic pacing (CPP). CPP includes biventricular pacing, which captures the left ventricle from its epicardial surface through a coronary vein, and conduction system pacing (CSP), which captures the His bundle or the left bundle branch and spreads activation through the Purkinje network. Unqualified "CRT" means biventricular or left ventricular-only pacing.
A conventional biventricular system has a right atrial lead for sensing and atrial pacing, a right ventricular lead in the apex or septum, and a left ventricular lead advanced through the coronary sinus into a lateral or posterolateral tributary vein. The left ventricular lead can also be placed on the epicardium surgically or on the endocardium. The generator delivers right and left ventricular stimuli simultaneously or with a programmed interventricular offset after a sensed or paced atrial event. Patients with permanent atrial fibrillation may receive a system without a functioning atrial lead.
| Configuration | Components | Principal use |
|---|---|---|
| CRT-P (resynchronization pacemaker) | Atrial, right ventricular, and coronary sinus leads; no high-voltage capacitor | CRT indication without a separate defibrillator indication, or when competing risks and patient goals favor pacing alone |
| CRT-D (resynchronization defibrillator) | As CRT-P, with a right ventricular shock coil and defibrillator generator | CRT indication with a primary or secondary prevention implantable cardioverter-defibrillator (ICD) indication |
| Epicardial CRT | Surgically placed left ventricular lead via thoracotomy or thoracoscopy, connected to a transvenous system | Failed coronary sinus implant; concomitant cardiac surgery |
| Wireless endocardial CRT | Leadless left ventricular endocardial electrode stimulated by ultrasound from a subcutaneous transmitter and battery | Failed or high-risk coronary sinus implant, nonresponse to conventional CRT |
| His bundle pacing CRT | Lead at the His bundle; correction of LBBB restores activation through both bundles | Rescue after failed coronary sinus implant; selected first-line use |
| Left bundle branch area pacing CRT | Lead screwed across the interventricular septum to the left bundle branch or its fascicles | Rescue after failed coronary sinus implant; alternative to biventricular CRT in selected patients |
| HOT-CRT and LOT-CRT | His bundle-optimized (HOT-CRT) or left bundle-optimized (LOT-CRT) CRT: conduction system lead plus a coronary sinus lead | Incomplete electrical resynchronization with conduction system pacing alone |
Resynchronization systems and their usual roles. CRT-P, cardiac resynchronization therapy pacemaker; CRT-D, cardiac resynchronization therapy defibrillator; LBBB, left bundle branch block.
Electrical dyssynchrony in left bundle branch block
In LBBB, activation enters the ventricles through the right bundle. The septum depolarizes from right to left, and the impulse then crosses slowly from cell to cell to the lateral and posterolateral left ventricular wall, which is activated last. The early-activated septum contracts while the lateral wall is still relaxed and stretches it; the lateral wall then contracts late against a septum that has begun to relax and moves away from it. This sequence produces the septal flash and apical rocking seen on echocardiography and wastes part of left ventricular work in internal shortening of opposing walls. It also prolongs isovolumic contraction, shortens the filling period, and worsens functional mitral regurgitation through dyssynchronous papillary muscle contraction. Myocardium in the early-activated septum does little work, and the late-activated lateral wall is loaded and hypertrophies.
Chronic right ventricular pacing produces the same sequence, because the impulse spreads from the right ventricle to the left ventricular free wall through working myocardium. The 2023 HRS/APHRS/LAHRS guideline treats LBBB-associated and pacing-induced cardiomyopathy as two forms of dyssynchrony-induced ventricular dysfunction, and both can reverse when synchronous activation is restored. A systematic review cited in the guideline estimated the pooled prevalence of pacing-induced cardiomyopathy at about 12%, using varied definitions. Most studies define it as a fall in LVEF of 10 percentage points or more from a baseline above 50%, with ventricular pacing of 20% or more and no other explanation.
Biventricular pacing does not restore normal activation. It creates two activation wavefronts, one from the right ventricular lead and one from the epicardial left ventricular site, which collide in the left ventricular free wall. The paced QRS is often narrower than the intrinsic LBBB complex but rarely normal. Benefit depends on a resynchronizable substrate: a conduction delay in the left bundle system with latest activation in the lateral wall. Right bundle branch block and nonspecific intraventricular delay that reflect diffuse myocardial disease or scar do not provide that substrate in most patients. The hemodynamic effect is immediate, with a higher rate of rise of left ventricular pressure and stroke volume at lower myocardial oxygen cost. Reverse remodeling, with smaller end-systolic volume, higher LVEF, and less mitral regurgitation, develops over months and was sustained at 5 years in the REVERSE follow-up cited by the guideline.
Defining left bundle branch block for CRT selection
QRS morphology predicts response to CRT, so the definition of LBBB used to select patients matters. The 2023 HRS/APHRS/LAHRS guideline uses the 2009 American Heart Association/American College of Cardiology Foundation/Heart Rhythm Society (AHA/ACCF/HRS) electrocardiography standard: a QRS duration of 120 ms or more with broad, notched, or slurred R waves in leads I, aVL, V5, and V6. Strict criteria were proposed because about one-third of patients diagnosed with LBBB by conventional criteria may instead have left ventricular hypertrophy combined with left anterior fascicular block. The Strauss criteria require a QRS duration of 140 ms or more in men or 130 ms or more in women, with mid-QRS notching or slurring in at least 2 contiguous leads. Conduction system pacing trials such as LEFT-BUNDLE-CRT enrolled patients by the Strauss criteria. The surface electrocardiogram (ECG) of bundle branch and fascicular block, and the localization of block, are covered in the bundle branch and fascicular block section of the bradycardia and atrioventricular block hub.
| Pattern | ECG criteria | Relevance to CRT |
|---|---|---|
| LBBB, 2009 AHA/ACCF/HRS | QRS ≥120 ms; broad, notched, or slurred R in I, aVL, V5, V6 | Definition used by the 2023 HRS/APHRS/LAHRS guideline |
| LBBB, Strauss | QRS ≥140 ms (men) or ≥130 ms (women); mid-QRS notching or slurring in ≥2 contiguous leads | Stricter; identifies true complete LBBB and was used in randomized trials of conduction system pacing |
| Right bundle branch block | QRS ≥120 ms; rsR′ in V1–V2; wide S in I and V6 | Little or no benefit from biventricular CRT in patient-level analysis |
| Nonspecific intraventricular conduction delay | QRS ≥120 ms without criteria for either bundle branch block | Benefit at QRS ≥150 ms in patient-level analysis; none below 150 ms |
| Right ventricular paced QRS | Wide, LBBB-like complex after a ventricular stimulus | Substrate for upgrade when pacing-induced cardiomyopathy develops |
ECG patterns in the selection of patients for resynchronization. The 2009 statement defines LBBB from a QRS duration of 120 ms; the Strauss criteria were proposed to exclude left ventricular hypertrophy with left anterior fascicular block.
QRS duration is measured as the widest QRS across all 12 leads, from the earliest onset to the latest offset, at a standard paper speed or with digital calipers. Automated measurements vary between ECG systems by several milliseconds, and the 150-ms and 120-ms thresholds are close enough to many patients' values that manual confirmation on more than one tracing is standard before implantation. QRS duration also changes with heart rate, drugs that slow conduction, and electrolyte disorders.
Epidemiology and burden of conduction delay in heart failure
The prevalence of LBBB in the general population ranges from 0.2% to 1.1%, rising with age and with structural heart disease. About 30% of patients with dilated cardiomyopathy have an intraventricular conduction delay, most often LBBB. In a Spanish cohort of 2254 patients with NYHA class II to IV heart failure cited by the 2023 guideline, 30.2% had LBBB, 7.6% right bundle branch block, and 8.7% a nonspecific intraventricular conduction delay. Wide QRS in heart failure is associated with higher mortality and with more advanced ventricular remodeling, and non-LBBB patterns carry higher mortality than LBBB in several cohorts.
Women made up roughly 20% to 30% of enrollees in the pivotal CRT trials, and patients aged 70 years or older were also underrepresented. In a patient-level meta-analysis of 8 randomized trials (6369 patients), 38% were 70 years or older; CRT delayed the composite of heart failure hospitalization or death (adjusted hazard ratio 0.75) and death alone (0.80), and the benefit was at least as great in older patients. Women are less likely than men to receive CRT despite similar or greater relative benefit.
Randomized trials in moderate to severe heart failure: MIRACLE, COMPANION, and CARE-HF
MIRACLE. The Multicenter InSync Randomized Clinical Evaluation randomized 453 patients with moderate to severe symptoms, LVEF of 35% or less, and QRS of 130 ms or more to active CRT or no pacing for 6 months after implantation. CRT improved 6-minute walk distance (+39 m vs +10 m), NYHA class, quality-of-life score (−18.0 vs −9.0 points), treadmill time, and LVEF (+4.6 vs −0.2 percentage points). Fewer patients required hospitalization (8% vs 15%) or intravenous medication (7% vs 15%) for heart failure. Implantation failed in 8% of patients, and coronary sinus perforation requiring pericardiocentesis occurred in 2 patients.
COMPANION. The Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure trial randomized 1520 patients with NYHA class III or IV heart failure from ischemic or nonischemic cardiomyopathy and QRS of 120 ms or more, in a 1:2:2 ratio, to optimal pharmacologic therapy, CRT-P, or CRT-D. Both devices reduced the primary end point of death or hospitalization from any cause (hazard ratio 0.81 with CRT-P and 0.80 with CRT-D). Death or heart failure hospitalization fell by 34% with CRT-P and 40% with CRT-D. All-cause mortality fell by 24% with CRT-P (P = 0.059) and by 36% with CRT-D (P = 0.003).
CARE-HF. The Cardiac Resynchronization-Heart Failure trial randomized 813 patients with NYHA class III or IV heart failure, LVEF of 35% or less, and markers of cardiac dyssynchrony to medical therapy alone or with CRT-P. Over a mean of 29.4 months, death or unplanned hospitalization for a major cardiovascular event occurred in 39% of the CRT group and 55% of the control group (hazard ratio 0.63). Death occurred in 20% and 30% (hazard ratio 0.64). CRT reduced interventricular mechanical delay, end-systolic volume index, and mitral regurgitant jet area and increased LVEF. In the extension phase (mean 37.4 months), mortality was 24.7% with CRT and 38.1% with medical therapy (hazard ratio 0.60), with fewer deaths from both worsening heart failure (hazard ratio 0.55) and sudden death (hazard ratio 0.54). CARE-HF is the trial that established a mortality benefit for CRT without a defibrillator.
Randomized trials in mild heart failure: REVERSE, MADIT-CRT, and RAFT
REVERSE. The Resynchronization Reverses Remodeling in Systolic Left Ventricular Dysfunction trial implanted CRT devices in 610 patients with NYHA class I or II symptoms (class I patients had prior symptoms), QRS of 120 ms or more, and LVEF of 40% or less, and randomized them 2:1 to CRT on or off for 12 months. The primary clinical composite end point did not differ significantly (16% vs 21% worsened). Left ventricular end-systolic volume index fell by 18.4 mL/m² with CRT on and by 1.3 mL/m² with CRT off, and time to first heart failure hospitalization was prolonged (hazard ratio 0.47).
MADIT-CRT. The Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy randomized 1820 patients with ischemic or nonischemic cardiomyopathy, LVEF of 30% or less, QRS of 130 ms or more, and NYHA class I or II symptoms 3:2 to CRT-D or ICD alone. Over an average of 2.4 years, death or a heart failure event occurred in 17.2% with CRT-D and 25.3% with ICD alone (hazard ratio 0.66). The result was driven by a 41% reduction in heart failure events, and annual mortality was 3% in both groups. In the prespecified QRS morphology analysis, the hazard ratio for the primary end point was 0.47 with LBBB and 1.24 without LBBB. In post-trial follow-up to 7 years, all-cause mortality among patients with LBBB was 18% with CRT-D and 29% with ICD alone (adjusted hazard ratio 0.59). In patients without LBBB, CRT-D showed no benefit and possible harm (adjusted hazard ratio for death 1.57).
RAFT. The Resynchronization-Defibrillation for Ambulatory Heart Failure Trial randomized 1798 patients with NYHA class II or III heart failure, LVEF of 30% or less, and intrinsic QRS of 120 ms or more (or paced QRS of 200 ms or more) to ICD or CRT-D. Over a mean of 40 months, death or heart failure hospitalization occurred in 33.2% with CRT-D and 40.3% with ICD (hazard ratio 0.75). CRT-D reduced death (hazard ratio 0.75) and heart failure hospitalization (hazard ratio 0.68). In long-term follow-up of 1050 patients at the 8 highest-enrolling sites (median 7.7 years overall and 13.9 years among survivors), death occurred in 71.2% with CRT-D and 76.4% with ICD, and time to death was longer with CRT-D (acceleration factor 0.80).
Landmark randomized trials of resynchronization
| Trial | Population | Comparison | Principal result |
|---|---|---|---|
| MIRACLE (2002) | NYHA III–IV, LVEF ≤35%, QRS ≥130 ms; n = 453 | CRT on vs off, 6 months | Better walk distance, NYHA class, quality of life, and LVEF |
| COMPANION (2004) | NYHA III–IV, QRS ≥120 ms; n = 1520 | CRT-P vs CRT-D vs medical therapy | Death or any hospitalization: HR 0.81 (CRT-P), 0.80 (CRT-D); mortality −36% with CRT-D |
| CARE-HF (2005) | NYHA III–IV, LVEF ≤35%, dyssynchrony; n = 813 | CRT-P vs medical therapy | Death or CV hospitalization 39% vs 55% (HR 0.63); death 20% vs 30% (HR 0.64) |
| REVERSE (2008) | NYHA I–II, LVEF ≤40%, QRS ≥120 ms; n = 610 | CRT on vs off, 12 months | Composite NS; LVESVi −18.4 vs −1.3 mL/m²; HF hospitalization delayed (HR 0.47) |
| MADIT-CRT (2009) | NYHA I–II, LVEF ≤30%, QRS ≥130 ms; n = 1820 | CRT-D vs ICD | Death or HF event 17.2% vs 25.3% (HR 0.66); at 7 years, lower mortality in LBBB only (HR 0.59) |
| RAFT (2010) | NYHA II–III, LVEF ≤30%, QRS ≥120 ms; n = 1798 | CRT-D vs ICD | Death or HF hospitalization 33.2% vs 40.3% (HR 0.75); death HR 0.75 |
| EchoCRT (2013) | NYHA III–IV, LVEF ≤35%, QRS <130 ms with echocardiographic dyssynchrony; n = 809 | CRT on vs off | Stopped for futility; death 11.1% vs 6.4% (HR 1.81) |
| BLOCK-HF (2013) | AV block, NYHA I–III, LVEF ≤50%; n = 691 | Biventricular vs RV pacing | Death, urgent HF care, or LVESVi rise ≥15%: 45.8% vs 55.6% (HR 0.74) |
| APAF-CRT mortality (2021) | Permanent AF, QRS ≤110 ms, recent HF hospitalization; n = 133 | AV junction ablation with CRT vs drug rate control | All-cause death 11% vs 29% (HR 0.26) |
| BUDAPEST-CRT (2023) | LVEF ≤35%, RV pacing ≥20%, paced QRS ≥150 ms; n = 360 | CRT-D upgrade vs ICD | Death or HF hospitalization 10% vs 32% (HR 0.27) |
Results are CRT versus comparator. AF, atrial fibrillation; AV, atrioventricular; CV, cardiovascular; HF, heart failure; HR, hazard ratio; LVESVi, left ventricular end-systolic volume index; NS, not significant; RV, right ventricular.
QRS duration, morphology, sex, and height as determinants of benefit
QRS duration. In an individual patient meta-analysis of 3782 patients in sinus rhythm from 5 randomized trials, QRS duration was the only baseline variable that predicted the magnitude of CRT benefit on death and on death or heart failure hospitalization. The estimated benefit increased with QRS duration, and the benefit was statistically significant from about 140 ms. In that analysis, morphology added no information once duration was known, partly because 78% of the patients had LBBB.
QRS morphology. A patient-level analysis of 6264 patients from 8 trials, with separate estimates for 4549 with LBBB, 691 with right bundle branch block, and 1024 with nonspecific intraventricular conduction delay, found that CRT reduced heart failure hospitalization or death with QRS of 150 ms or more and either LBBB (hazard ratio 0.56) or intraventricular conduction delay (0.59). It found no association with benefit in right bundle branch block (0.97) or with any morphology below 150 ms. MADIT-CRT, its long-term follow-up, and the subgroups of RAFT point in the same direction: LBBB identifies patients who benefit, and non-LBBB patients with mild symptoms and moderate QRS prolongation do not.
Sex and height. In a US Food and Drug Administration meta-analysis of 4076 patients from 3 CRT-D versus ICD trials with predominantly NYHA class II symptoms, women with LBBB and QRS of 130 to 149 ms had a 76% reduction in heart failure or death (hazard ratio 0.24). Men in the same range had no significant benefit (0.85). Neither sex benefited below 130 ms, and both benefited at 150 ms or more. A later individual patient analysis found that height and QRS duration independently predicted benefit for the composite outcome and that sex itself did not. Shorter patients, many of them women, benefit at shorter QRS durations, consistent with a QRS duration that reflects conduction delay relative to ventricular size. These data are the basis of the 2023 guideline's Class 1 recommendation for women and other selected patients with LBBB and QRS of 120 to 149 ms.
Cause of cardiomyopathy, scar, and ventricular size. Female sex, nonischemic cardiomyopathy, and LBBB are associated with greater reverse remodeling. Scar burden on late gadolinium enhancement cardiac magnetic resonance correlates with less reduction in end-systolic volume, and pacing in scar lengthens the paced QRS and raises capture thresholds. Ischemic cardiomyopathy is not a reason to withhold CRT: in MADIT-CRT the benefit did not differ significantly by cause.
Narrow QRS and mechanical dyssynchrony: EchoCRT and PROSPECT
Mechanical dyssynchrony on echocardiography is present in many patients with heart failure and a narrow QRS, which led to trials of CRT selected by echocardiographic criteria. The Predictors of Response to CRT (PROSPECT) study evaluated 12 echocardiographic dyssynchrony parameters in 498 patients with standard CRT indications at 53 centers. Sensitivity ranged from 6% to 77% and specificity from 31% to 93%, all areas under the receiver operating characteristic curve were 0.62 or less, and interobserver variability was large. No single parameter could be recommended to select patients.
EchoCRT randomized 809 patients with NYHA class III or IV heart failure, LVEF of 35% or less, QRS under 130 ms, and echocardiographic dyssynchrony to CRT on or off after implantation. The trial was stopped for futility after a mean of 19.4 months. Death or first heart failure hospitalization occurred in 28.7% with CRT and 25.2% without (hazard ratio 1.20, P = 0.15), and death occurred in 11.1% and 6.4% (hazard ratio 1.81, P = 0.02). Biventricular pacing in a heart without an electrical delay creates dyssynchrony instead of correcting it. The 2023 guideline therefore assigns Class 3: No Benefit both to CRT for QRS under 120 ms and to preprocedural echocardiographic assessment of dyssynchrony to predict outcome.
Guideline indications for CRT and cardiac physiologic pacing by class
The 2023 HRS/APHRS/LAHRS guideline grades the indications for CRT and for conduction system pacing. The 2022 American Heart Association/American College of Cardiology/Heart Failure Society of America (AHA/ACC/HFSA) heart failure guideline carries the core recommendations for biventricular CRT in heart failure with reduced ejection fraction (HFrEF). Its recommendation table, with the ICD criteria that determine CRT-D, is summarized in the cardiac resynchronization section of the HFrEF hub. All recommendations apply to patients whose symptoms persist on GDMT.
| Clinical situation | Recommendation | Class (LOE) |
|---|---|---|
| LVEF ≤35%, sinus rhythm, on GDMT | ||
| LBBB, QRS ≥150 ms, NYHA II–IV | Biventricular CRT to improve symptoms and reduce morbidity and mortality | 1 (A) |
| LBBB, QRS ≥150 ms, NYHA II–IV, when effective biventricular CRT cannot be achieved | His bundle pacing with LBBB correction or left bundle branch area pacing | 2a (C-LD) |
| LBBB, QRS ≥150 ms, NYHA II–IV | Conduction system pacing as an alternative to biventricular CRT | 2b (C-LD) |
| LVEF ≤30%, LBBB, QRS ≥150 ms, NYHA I | Biventricular CRT to reduce worsening heart failure and improve remodeling | 2b (B-R) |
| LBBB, QRS 120–149 ms, NYHA II–IV, select characteristics such as female sex | Biventricular CRT | 1 (A) |
| LBBB, QRS 120–149 ms, NYHA II–IV | Biventricular CRT | 2a (B-R) |
| Non-LBBB, QRS ≥150 ms, NYHA III–ambulatory IV | Biventricular CRT to improve functional class, structure, and LVEF | 2a (A) |
| Non-LBBB, QRS ≥150 ms, NYHA II | Cardiac physiologic pacing | 2b (B-R; C-LD for CSP) |
| Non-LBBB, QRS ≥150 ms, NYHA II–IV, biventricular CRT not achievable | His bundle or left bundle branch area pacing | 2b (C-LD) |
| Non-LBBB, QRS 120–149 ms, NYHA III–IV | Cardiac physiologic pacing; usefulness not well established | 2b (B-NR) |
| Non-LBBB, QRS <150 ms, NYHA I–II | Biventricular CRT not recommended | 3: No Benefit (B-R) |
| QRS <120 ms, NYHA II–IV | Biventricular CRT not recommended | 3: No Benefit (B-R) |
| LVEF 36%–50% | ||
| LBBB, QRS ≥150 ms, sinus rhythm, NYHA II–IV | Cardiac physiologic pacing to maintain or improve LVEF | 2b (C-LD) |
| Pacing indication, substantial ventricular pacing expected (20%–40% or more) | Cardiac physiologic pacing to reduce pacing-induced cardiomyopathy | 2a (B-R) |
| Pacing indication, less than substantial pacing expected | Conventional right ventricular lead with minimized pacing (Class 2a); conduction system pacing lead as an alternative (Class 2b); cardiac physiologic pacing when LBBB is present (Class 2b) | 2a / 2b |
| Atrial fibrillation | ||
| AV junction ablation, LVEF ≤50% | Biventricular CRT | 2a (B-R) |
| Otherwise meets CRT criteria | Biventricular CRT | 2a (B-NR) |
| AV junction ablation | His bundle pacing with or without a backup lead; left bundle branch area pacing | 2b (C-LD) |
| Existing pacemaker or ICD | ||
| Decline in left ventricular function or worsening heart failure attributed to substantial ventricular pacing | Upgrade to biventricular CRT | 1 (B-NR) |
| Same | Revision to conduction system pacing | 2a (B-NR) |
| Normal LVEF | ||
| Pacing indication, substantial ventricular pacing expected | Cardiac physiologic pacing | 2b (B-NR) |
| Pacing indication, less than substantial pacing expected | Biventricular CRT not indicated | 3: No Benefit (B-R) |
| All candidates | ||
| Life expectancy under 1 year | Shared decision-making weighing quality of life against procedural risk | 1 (C-EO) |
Recommendations of the 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing. CPP, cardiac physiologic pacing (biventricular CRT, His bundle pacing, or left bundle branch area pacing); CSP, conduction system pacing; LOE, level of evidence (A, high quality; B-R, randomized; B-NR, nonrandomized; C-LD, limited data; C-EO, expert opinion); RV, right ventricular.
How European guidance differs. As reported in the 2025 European Society of Cardiology (ESC) clinical consensus statement on conduction system pacing, the 2021 ESC pacing and CRT guideline uses 130 ms as the lower QRS limit. CRT should be considered for non-LBBB QRS above 150 ms (Class IIa) and may be considered at 130 to 150 ms (Class IIb). For patients with LVEF of 40% or less who have atrioventricular block and need ventricular pacing, the guideline recommends CRT over right ventricular pacing regardless of NYHA class or QRS duration. The 2025 ESC consensus advises conduction system pacing as rescue when coronary sinus implantation fails and considers it possibly appropriate as an alternative to biventricular pacing in LBBB with QRS of 130 ms or more.
CRT and physiologic pacing indication finder
The finder applies the 2023 HRS/APHRS/LAHRS recommendations to the LVEF band, QRS duration and morphology, NYHA class, rhythm, and expected ventricular pacing. It returns the class of recommendation for the best-supported strategy and names the alternatives the guideline grades.
CRT and physiologic pacing indication finder
2023 HRS/APHRS/LAHRSClasses are from the 2023 HRS/APHRS/LAHRS guideline. For LVEF of 35% or less with a pacing indication and substantial expected ventricular pacing, the finder also applies the 2021 ESC recommendation, as reported in the 2025 ESC consensus. The defibrillator decision follows separate ICD criteria. All CRT recommendations assume persistent symptoms on GDMT and expected survival with good functional status beyond 1 year, or shared decision-making when survival is shorter.
Guideline-directed medical therapy and timing of implantation
All CRT trials enrolled patients on contemporary medical therapy. LVEF frequently rises with an angiotensin receptor-neprilysin inhibitor, a beta blocker, a mineralocorticoid receptor antagonist, and a sodium-glucose cotransporter 2 inhibitor, and patients whose LVEF recovers above 35% no longer meet the main indications. The sequence of drug initiation, titration, and the interval before LVEF is remeasured for device decisions are covered in the GDMT titration section of the HFrEF hub. CRT and GDMT are complementary, and CRT does not replace any drug class. The 2023 guideline recommends continued GDMT after implantation, including in patients whose LVEF improves (Class 2a), and continued optimization of drugs in patients with an unfavorable response (Class 1). In a dedicated CRT clinic cited by the guideline, 74% of patients labeled nonresponders had opportunities for substantial up-titration or addition of heart failure drugs.
Two circumstances justify implantation without a full drug-optimization period. A patient who needs a pacemaker for atrioventricular block, or who is undergoing atrioventricular junction ablation, receives a resynchronization-capable or conduction system pacing device at the first procedure when LVEF is reduced, because right ventricular pacing worsens dyssynchrony. Bradycardia that prevents beta-blocker titration is a pacing indication in its own right, and the device chosen should not itself impose right ventricular dyssynchrony.
Choice between CRT-P and CRT-D
The defibrillator component of a CRT-D device follows the indications for an ICD, which are covered in the primary-prevention ICD section of the ventricular arrhythmia and sudden death hub. Resynchronization itself lowers sudden death: in the CARE-HF extension, CRT-P reduced sudden death (hazard ratio 0.54) as well as death from heart failure. COMPANION, the only trial to randomize CRT-P and CRT-D against medical therapy, found a significant reduction in mortality with CRT-D (36%) and a smaller, nonsignificant reduction with CRT-P (24%), without a direct comparison powered for mortality between the 2 devices.
The case for adding a defibrillator is strongest in ischemic cardiomyopathy, in younger patients, and after sustained ventricular arrhythmia. In nonischemic cardiomyopathy, the DANISH trial randomized 1116 patients with LVEF of 35% or less, 58% of whom received CRT in both groups, to an ICD or usual care. After a median of 67.6 months, all-cause death occurred in 21.6% with an ICD and 23.4% without (hazard ratio 0.87, not significant), although sudden death was halved (4.3% vs 8.2%). CRT-P is commonly chosen for older patients, for those with major comorbidity or frailty, and for many patients with nonischemic cardiomyopathy after shared decision-making. A CRT-D generator is larger, has a shorter battery life, and adds the risks of inappropriate shocks and high-voltage lead failure.
At elective generator replacement, the 2023 guideline recommends that the choice between replacing a CRT-D and revising to CRT-P be individualized (Class 1). The decision weighs previous response to CRT, prior appropriate therapies, continuing arrhythmic risk, inappropriate therapies, lead performance, and goals of care. Arrhythmic risk falls but persists when LVEF recovers, and the guideline judges the data inadequate to support routine discontinuation of defibrillator therapy at replacement. Revision from CRT-D to CRT-P may require an adapter when a DF-4 defibrillation lead is in place.
Patients who need ventricular pacing: BLOCK-HF and pacing-induced cardiomyopathy
BLOCK-HF enrolled 918 patients with atrioventricular block requiring pacing, NYHA class I to III symptoms, and LVEF of 50% or less; 691 were randomized to right ventricular or biventricular pacing and followed for an average of 37 months. The primary outcome of death, urgent care for heart failure requiring intravenous therapy, or an increase of 15% or more in left ventricular end-systolic volume index occurred in 45.8% with biventricular pacing and 55.6% with right ventricular pacing (hazard ratio 0.74, 95% credible interval 0.60 to 0.90). Results were similar in patients receiving pacemakers and ICDs, and left ventricular lead-related complications occurred in 6.4%. BLOCK-HF shows that the harm of dyssynchronous pacing begins before LVEF falls into the range of conventional CRT indications.
The 2023 guideline defines substantial ventricular pacing as an anticipated or measured burden of 20% to 40% or more and recommends periodic assessment of ventricular function when such pacing cannot be reduced by programming (Class 1). The physiologic pacing section of the bradycardia and atrioventricular block hub covers the choice among biventricular, His bundle, and left bundle branch area pacing in bradycardia and the evidence comparing them with right ventricular pacing. The CRT side of the decision, patients with reduced LVEF who need pacing and patients whose LVEF falls after right ventricular pacing, is covered here.
At the time of cardiac surgery in patients likely to require future CRT, intraoperative placement of a permanent epicardial left ventricular lead is useful (Class 2a). In the RESCUE trial cited by the guideline, 178 patients undergoing coronary bypass surgery with LVEF of 35% or less, NYHA class III or IV symptoms, and a wide QRS or echocardiographic dyssynchrony were randomized to an epicardial CRT system at surgery or to surgery alone. CRT reduced all-cause mortality (hazard ratio 0.43) and readmission (9.9% vs 28.7%). An epicardial lead that is not connected to a generator can preclude magnetic resonance imaging at many institutions.
Upgrade from a pacemaker or ICD to CRT
Upgrade of a right ventricular pacing system to biventricular CRT is recommended when LV function declines or heart failure worsens because of substantial ventricular pacing (Class 1), and revision to conduction system pacing can be beneficial (Class 2a). BUDAPEST-CRT enrolled 360 patients with NYHA class II to ambulatory IV symptoms, a pacemaker or ICD, right ventricular pacing of 20% or more, and a paced QRS of 150 ms or more. They were randomized 3:2 to upgrade to CRT-D or to an ICD. The primary composite of death, heart failure hospitalization, or less than 15% reduction in end-systolic volume at 12 months occurred in 32.4% with CRT-D and 78.9% with ICD (odds ratio 0.11). Death or heart failure hospitalization occurred in 10% and 32% (hazard ratio 0.27), and procedure- or device-related complications were similar (12.3% vs 7.8%).
Upgrades carry more risk than de novo implants. In a meta-analysis of biventricular upgrade cited by the 2023 guideline, infection averaged 3.7%, pneumothorax 2.0%, cardiac perforation or tamponade 1.4%, and lead-related complications 3.3%. Existing leads can occlude the subclavian or innominate vein, so venography precedes the procedure; a stenosed vein is treated with venoplasty, lead extraction to create access, or a contralateral approach with tunneling. The pocket is revised to accept the larger generator, and the previous generator's remaining battery life and the age of existing leads determine whether leads are reused. Revision to left bundle branch area pacing avoids a coronary sinus lead: in a series of 20 patients with pacing-induced cardiomyopathy and infranodal block, upgrade succeeded in 19, and LVEF rose from 36.3% to 51.9%.
Atrial fibrillation in CRT candidates and recipients
Atrial fibrillation does not exclude CRT. It does threaten the mechanism of benefit: conducted beats that arrive before the programmed pacing interval inhibit or fuse with biventricular stimulation. In the permanent atrial fibrillation subgroup of RAFT, CRT showed no benefit over an ICD, and only one-third of those patients achieved biventricular pacing above 95% although they appeared rate-controlled at enrollment. In patients with atrial fibrillation who otherwise meet CRT criteria, biventricular CRT can be beneficial to improve quality of life, functional capacity, and LVEF (Class 2a), on the condition that near-complete biventricular capture is achieved with drugs or ablation.
Atrioventricular junction ablation with resynchronization is the most reliable way to guarantee capture. In the morbidity phase of APAF-CRT, 102 patients with severely symptomatic permanent atrial fibrillation, QRS of 110 ms or less, and a heart failure hospitalization in the previous year were randomized to ablation with CRT or to drug rate control. Death from heart failure, heart failure hospitalization, or worsening heart failure occurred in 20% and 38% (hazard ratio 0.38). In the mortality phase (133 patients), the trial was stopped for efficacy after a median of 29 months. All-cause death occurred in 11% with ablation and CRT and 29% with drugs (hazard ratio 0.26), with similar benefit in patients with LVEF of 35% or less and above 35%. The 2023 guideline gives biventricular CRT a Class 2a recommendation for patients with LVEF of 50% or less undergoing atrioventricular junction ablation, and His bundle or left bundle branch area pacing a Class 2b recommendation. The decision to ablate the atrioventricular junction, and rate and rhythm control in heart failure, are covered in the atrioventricular node ablation section of the atrial fibrillation hub.
CONDUCT-AF compared left bundle branch area pacing with biventricular pacing after atrioventricular junction ablation in 82 patients with refractory atrial fibrillation, LVEF below 50%, QRS of 120 ms or less, and N-terminal pro-B-type natriuretic peptide (NT-proBNP) above 600 ng/L. At 6 months LVEF rose from 36% to 46% with left bundle branch area pacing and from 34% to 46% with biventricular pacing. The trial reported a between-group difference of 0.4 percentage points (95% CI, −3.1 to 3.8; P = .002 for noninferiority), which is an estimate and not the simple difference between the two crude gains. Paced QRS, procedure time, and fluoroscopy time were shorter with conduction system pacing.
In CRT recipients in sinus rhythm, new atrial fibrillation lowers biventricular pacing and is the most common reason for its loss. The guideline cites an analysis in which a mean ventricular rate above 80 beats per minute and a maximum above 100 beats per minute during atrial fibrillation were associated with biventricular pacing below 95% and with more heart failure hospitalization and death. Rhythm control, faster rate control, or atrioventricular junction ablation is reasonable when atrial fibrillation lowers biventricular pacing in a patient with an unfavorable response (Class 2a).
Percentage of biventricular pacing and effective capture
Benefit from CRT rises with the proportion of beats that are biventricularly paced. In 36,935 patients followed in a remote monitoring network, the greatest reduction in mortality was seen with biventricular pacing above 98% of ventricular beats, with incremental benefit as pacing percentage rose. An analysis cited in the 2023 guideline found 44% fewer clinical events with biventricular pacing above 92% than at or below 92% (hazard ratio 0.56). The programming target is therefore as close to 100% as possible, and 98% is the usual threshold below which the cause is sought.
The device counter overstates resynchronization. It records a paced event whenever a ventricular stimulus is delivered, including beats in which intrinsic conduction has already depolarized the ventricles (pseudofusion) and beats that are partly intrinsic and partly paced (fusion). The counter also cannot detect loss of left ventricular capture, because the right ventricular stimulus is still delivered. Effective resynchronization is confirmed by the paced morphology on a 12-lead ECG, by Holter or extended monitoring with morphology analysis, and by device algorithms that assess left ventricular capture on a beat-to-beat basis.
| Cause of lost resynchronization | Mechanism | Correction |
|---|---|---|
| Atrial fibrillation with rapid conduction | Conducted beats preempt or fuse with pacing | Rate control, rhythm control, or atrioventricular junction ablation |
| Frequent premature ventricular complexes | Ectopic beats are not resynchronized; surrounding beats are often fused | Drug suppression or catheter ablation (Class 2a when pacing is reduced); triggered pacing after sensed ectopy |
| Atrioventricular delay longer than intrinsic conduction | Intrinsic activation precedes the stimulus | Shorter sensed and paced atrioventricular delays, reset when intrinsic conduction changes |
| Sinus or atrial tachycardia above the upper tracking rate | Pacing falls behind intrinsic conduction during exertion | Higher upper tracking rate, beta blockade, treatment of the tachycardia |
| Left ventricular exit block or threshold rise | Right ventricular-only pacing produces an LBBB-like paced QRS | Higher output, alternative vector on a quadripolar lead, lead revision |
| Left ventricular lead dislodgement | Loss of capture, phrenic stimulation, or change of paced morphology | Radiograph and interrogation; repositioning |
| Phrenic nerve stimulation limiting output | Output reduced below reliable capture to avoid diaphragmatic pacing | Vector change, pulse-width adjustment, lead revision |
| Ventricular sensing after a paced atrial beat (crosstalk, far-field) | Inappropriate inhibition of ventricular output | Blanking, sensitivity, and vector programming |
Causes of loss of effective biventricular pacing and their correction. A reported biventricular pacing percentage near 100% does not exclude pseudofusion, fusion, or loss of left ventricular capture.
Implantation: venous access, coronary sinus anatomy, and lead delivery
Preprocedure assessment. A 12-lead ECG to define rhythm, atrioventricular conduction, and QRS duration and morphology (Class 1) and echocardiography to measure LVEF (Class 1) are recommended before any cardiac physiologic pacing device. Imaging by echocardiography, cardiac magnetic resonance, or CT to target the left ventricular lead may be considered (Class 2b). Shared decision-making that covers the evidence for the individual pacing approaches and the patient's goals is recommended (Class 1). Venous patency is assessed by venography when prior devices, central lines, or dialysis access are present, and anticoagulation and antiplatelet therapy are managed to limit pocket hematoma.
Access and coronary sinus cannulation. Leads are introduced through the cephalic, axillary, or subclavian vein; extrathoracic access avoids subclavian crush of leads between the clavicle and first rib. The coronary sinus ostium lies posterior and inferior in the right atrium, guarded by a variably developed Thebesian valve. A preshaped or deflectable long sheath is advanced into the coronary sinus, and an occlusive balloon venogram in left and right anterior oblique projections maps the tributaries: the posterior (middle cardiac) vein, the posterolateral and lateral marginal veins, the anterolateral veins, and the anterior interventricular vein. A persistent left superior vena cava drains into an enlarged coronary sinus and changes the approach.
Lead delivery and testing. The left ventricular lead is delivered over a 0.014-inch guidewire or with a stylet into a lateral or posterolateral vein at a mid or basal level. All available pacing vectors are tested for capture threshold, sensing, impedance, and phrenic nerve capture at high output, and the electrical delay from QRS onset to the local left ventricular electrogram (QLV) is measured. The paced 12-lead ECG at a candidate site confirms left ventricular capture and narrowing. A quadripolar left ventricular lead is recommended (Class 1). In the randomized MORE-CRT trial of 1078 patients, freedom from intraoperative and postoperative left ventricular lead events at 6 months was 83.0% with a quadripolar lead and 74.4% with a bipolar lead, driven by fewer intraoperative events (6.26% vs 12.1%). Lead positioning and programming to achieve the narrowest QRS duration can be beneficial (Class 2a).
When the coronary sinus approach fails. Coronary sinus lead placement fails in about 5% to 10% of attempts. The 2023 guideline lists the reasons for abandoning or crossing over: venous or coronary sinus occlusion, dissection or perforation, small or tortuous tributaries, a nonlateral target vein, unstable lead position, capture thresholds above 5 V at 1 ms in all vectors, phrenic stimulation in all vectors, a QLV under 90 ms, and failure to narrow the paced QRS. Crossover to His bundle or left bundle branch area pacing is reasonable (Class 2a), and crossover to a surgical epicardial lead might be reasonable (Class 2b). A biventricular generator implanted after a failed coronary sinus attempt allows a later epicardial lead to be connected.
Left ventricular lead position: lateral, nonapical, latest activation, and scar
Apical versus nonapical position. In 799 patients in the CRT-D arm of MADIT-CRT, the benefit of CRT was similar with anterior, lateral, and posterior lead positions in the short axis. An apical lead was associated with more heart failure or death (hazard ratio 1.72) and more death (hazard ratio 2.91) than a basal or midventricular lead. Pacing from a nonapical position is reasonable to improve clinical and structural response (Class 2a). Posteroanterior and lateral chest radiographs locate the lead and are recommended in patients with an unfavorable response (Class 1). Septal lead positions, including the anterior interventricular vein, respond less favorably than lateral positions.
Targeting the latest-activated segment. In TARGET, 220 patients underwent speckle-tracking radial strain imaging and were randomized to a lead placed at the latest-contracting segment without scar or to standard placement. The targeted group had more echocardiographic responders at 6 months (70% vs 55%), more clinical responders (83% vs 65%), and fewer deaths or heart failure hospitalizations. Electrical delay at the lead tip measured as QLV predicts remodeling: in a substudy of SMART-AV, reverse remodeling rose from 38.7% in the shortest QLV quartile to 68.4% in the longest, and the longest quartile carried a 3.21-fold odds of response after adjustment for QRS duration and morphology. Targeting the site of latest activation may be considered (Class 2b).
Targeting the latest electrical activation site in DANISH-CRT. The national DANISH-CRT trial, published in 2026, randomized 1001 patients to a left ventricular lead at the latest electrically activated site within the coronary sinus branches or in a posterolateral nonapical branch. Activation at the lead tip was only 9 ms later in the targeted group. Over a median of 45.8 months, death or unplanned heart failure hospitalization occurred in 28% of the targeted group and 26% of the standard group (hazard ratio 1.10, P = 0.45), and lead-related complications were more frequent with targeting. The trial shows that, once the lead is posterolateral and nonapical, searching the available veins for the latest electrical site does not improve outcomes.
Scar. Pacing in transmural scar raises the capture threshold, widens the paced QRS, and is associated with less remodeling. Cardiac magnetic resonance with late gadolinium enhancement, overlaid on the coronary venous anatomy, can direct the lead away from scar, an approach the guideline describes without a separate recommendation.
Multipoint pacing. Pacing from 2 electrodes on a quadripolar lead (multipoint pacing) or from 2 left ventricular leads (multisite pacing) activates more of the left ventricle. In MORE-CRT MPP, 1111 patients who were echocardiographic nonresponders after 6 months of biventricular pacing were randomized to multipoint pacing or continued biventricular pacing; 29.4% with multipoint pacing and 30.4% with biventricular pacing became responders over the next 6 months, with no difference even with wide spacing between the pacing sites.
Epicardial, endocardial, and leadless left ventricular pacing
Surgical epicardial leads. A left ventricular lead can be placed on the lateral or posterolateral epicardium through a limited thoracotomy or by video-assisted thoracoscopy, and mapping the latest local activation during surgery can guide the site. A large observational study cited by the 2023 guideline found survival and LVEF improvement similar with surgical epicardial and transvenous coronary sinus leads over a mean of 5.1 years, and 2 small randomized trials showed equivalent clinical and structural outcomes. Thresholds of epicardial leads tend to rise over time, and general anesthesia and thoracic access add risk in advanced heart failure.
Endocardial left ventricular leads. A conventional lead placed in the left ventricular endocardium by transseptal or transapical routes activates the ventricle from inside, with less dependence on venous anatomy. All endocardial lead techniques require lifelong anticoagulation with an international normalized ratio target of about 2.5 to 3.5, and thromboembolic events occur with lapses in anticoagulation, which confines the approach to selected patients.
Wireless endocardial pacing. The WiSE-CRT system places a leadless receiver electrode on the left ventricular endocardium, stimulated by ultrasound energy from a subcutaneously implanted transmitter and battery. SOLVE-CRT enrolled 183 patients who were nonresponders, previously untreatable by coronary sinus pacing, or at high risk for an upgrade. In the 100 patients of the primary efficacy analysis, mean end-systolic volume fell by 16.4%. Freedom from type I complications was 80.9%, with device-system events in 6.6%, vascular events in 2.7%, stroke in 1.6%, and cardiac perforation in 3.8%, mostly early in the study.
Procedural complications of CRT implantation
CRT carries more risk than single- or dual-chamber pacing because the procedure is longer, a third lead is placed in the coronary venous system, and patients have advanced heart failure. In RAFT, adverse events within 30 days occurred in 124 patients assigned to CRT-D and 58 assigned to an ICD. The general complications of transvenous devices, including pneumothorax, pocket hematoma, infection, tricuspid regurgitation, and lead extraction, are covered in the implantation and complications section of the bradycardia and atrioventricular block hub.
| Complication | Features | Management |
|---|---|---|
| Coronary sinus dissection | Contrast staining of the vessel wall during venography or sheath advancement; usually contained by the pericardium around the atrioventricular groove | Stop advancing; most heal without sequelae; echocardiography for effusion; lead placement may be completed if the true lumen is secure or deferred |
| Coronary sinus or venous perforation | Wire or lead exits the vein; low-pressure venous bleeding | Echocardiographic monitoring; pericardiocentesis for tamponade (2 cases in MIRACLE) |
| Phrenic nerve stimulation | Diaphragmatic twitching, often positional and not present supine at implant | Alternative vector, lower output with adequate margin, pulse-width change, repositioning |
| Left ventricular lead dislodgement | Rising threshold, loss of capture, change in paced QRS, often within weeks | Reprogramming of a quadripolar lead; repositioning; active-fixation leads in unstable anatomy |
| High left ventricular threshold | Scar, small vein, epicardial fat; rapid battery depletion | Vector change; new lead at generator replacement (Class 2b) |
| Contrast nephropathy | Venography in patients with chronic kidney disease | Limited contrast volume, hydration as tolerated |
| Lead malposition in the left heart | Lead crosses a patent foramen ovale or septal defect, or enters an artery; tall R in V1 during right-sided pacing | Lateral radiograph and echocardiography; repositioning and anticoagulation decisions |
| Ventricular proarrhythmia | Epicardial left ventricular pacing reverses the transmural activation sequence | Uncommon; reprogramming or endocardial or conduction system pacing in recurrent cases |
Complications specific to CRT implantation and their management.
The paced ECG after CRT
A 12-lead ECG at follow-up visits confirms left ventricular capture and is recommended for optimizing pacing configurations (Class 1). With the left ventricular lead in a lateral vein, left ventricular capture activates the ventricles from left to right and from the lateral wall toward the septum. The left ventricular contribution therefore produces a positive deflection in V1 and a negative deflection in lead I. Biventricular pacing therefore typically shows a dominant R or an RS complex in V1, a Q or QS in lead I, and a rightward or superior frontal axis. Right ventricular pacing alone produces an LBBB-like paced complex with a QS or rS in V1 and a broad positive R in leads I and aVL. A paced QRS that changes from the implant pattern to an LBBB-like morphology indicates loss of left ventricular capture. Anodal capture at the right ventricular ring during left ventricular-to-right ventricular vectors adds a right ventricular component and changes the morphology at high outputs.
The paced QRS after biventricular pacing is narrower than the intrinsic complex in most responders. Failure of the QRS to narrow, or prolongation of the QRS, with biventricular pacing occurs in a substantial minority. The 2025 ESC consensus describes it in about one-third of patients and associates it with a worse prognosis. A dominant R wave in V1 is consistent with left ventricular capture, but it can also appear when a lead intended for the right ventricle has crossed into the left heart, which is why the lateral chest radiograph is part of postimplant review.
| Paced pattern | V1 | Lead I | Interpretation |
|---|---|---|---|
| Biventricular capture, lateral left ventricular lead | Dominant R or RS | Q, QS, or small r | Expected pattern; QRS usually narrower than baseline |
| Right ventricular pacing only | QS or rS | Broad positive R | Loss of left ventricular capture, dislodgement, or programming to right ventricular pacing |
| Left ventricular pacing only | Tall R, RBBB-like | QS or negative | Programmed left ventricular-only pacing, or right ventricular lead failure |
| Fusion with intrinsic conduction | Intermediate; varies with atrioventricular delay | Intermediate | Intended in adaptive left ventricular pacing; unintended when the atrioventricular delay is too long |
| Left bundle branch area pacing | qR or rSR′ (right bundle delay pattern) | Positive R | Left conduction system capture; narrow QRS with short V6 R-wave peak time |
Paced ECG patterns after resynchronization. RBBB, right bundle branch block. Pacemaker malfunction in general (failure to capture, failure to sense, and oversensing) is covered in the paced ECG and malfunction section of the bradycardia and atrioventricular block hub.
Device programming and atrioventricular and interventricular optimization
Nominal programming. A sensed atrioventricular delay is programmed short enough to ensure ventricular capture before intrinsic conduction and long enough to allow complete atrial contribution to left ventricular filling. The paced atrioventricular delay is longer than the sensed delay because atrial pacing adds the interval from stimulus to atrial depolarization. Interventricular timing is usually simultaneous or with modest left ventricular pre-excitation. The lower rate, rate response for chronotropic incompetence, and upper tracking rate are programmed so that pacing continues during exertion.
Atrioventricular optimization trials. SMART-AV randomized 980 CRT-D recipients to a fixed atrioventricular delay of 120 ms, echocardiographically optimized delays, or an electrogram-based algorithm. Median reductions in end-systolic volume at 6 months were 21 mL with the algorithm, 19 mL with echocardiography, and 15 mL with the fixed delay, with no significant difference between the arms, and functional end points did not differ. Routine echocardiographic or algorithmic optimization at implantation is therefore not required; optimization is reserved for patients with an unfavorable response, with very long or very short intrinsic atrioventricular conduction, or with long interatrial conduction delay.
Echocardiographic methods. The iterative method records transmitral Doppler while the atrioventricular delay is shortened stepwise from a long value until the A wave is truncated. The delay is then lengthened to the shortest value that preserves a complete A wave separate from the E wave. Output-based methods select the delay or interventricular offset that maximizes the left ventricular outflow tract velocity-time integral. Electrocardiographic optimization selects the timing that produces the narrowest paced QRS, an approach the 2023 guideline supports for lead positioning and programming (Class 2a).
Adaptive algorithms. In patients with normal atrioventricular conduction and LBBB, pacing only the left ventricle timed to fuse with intrinsic conduction down the right bundle can resynchronize with fewer right ventricular stimuli. In the Adaptive CRT trial (522 patients), an algorithm switched between synchronized left ventricular pacing and biventricular pacing and adjusted delays automatically. It was noninferior to echocardiographically optimized biventricular pacing for the clinical composite score (73.6% vs 72.5% improved) and reduced right ventricular pacing by 44%. AdaptResponse randomized 3617 patients with LBBB, PR of 200 ms or less, and LVEF of 35% or less to the adaptive algorithm or conventional CRT. The trial stopped for futility after a median of 59.0 months; death or intervention for heart failure decompensation occurred in 23.5% with the adaptive algorithm and 25.7% with conventional CRT at 60 months (hazard ratio 0.89, P = 0.077). A second type of algorithm (SyncAV) adjusts the atrioventricular delay continuously to the measured intrinsic conduction so that biventricular pacing fuses with native activation; the 2023 guideline reports greater QRS narrowing than with conventional CRT pacing. Device algorithms do not replace confirmation of capture and QRS narrowing on the 12-lead ECG.
Programming for conduction system pacing. The 2023 guideline's interrogation table advises atrioventricular delays shorter than conventional settings because of the latency from stimulus to QRS onset. The reduction is about 30 to 50 ms for His bundle pacing, 20 to 30 ms for left bundle branch area pacing, and 10 to 20 ms for biventricular CRT. His bundle capture is confirmed separately from right ventricular myocardial capture, and outputs are programmed at least 1 V above the His capture threshold. Automatic capture algorithms are avoided in His bundle pacing unless His and myocardial thresholds are known to be similar.
Defining response: clinical composite score and reverse remodeling
No single definition of response to CRT exists. Clinical definitions include survival, freedom from heart failure hospitalization, improvement in NYHA class, quality of life, 6-minute walk distance, peak oxygen uptake, and the clinical composite score. Echocardiographic definitions include a reduction in left ventricular end-systolic volume of 15% or more, an absolute rise in LVEF of 5 percentage points or more, smaller ventricular dimensions, higher stroke volume, and less mitral regurgitation. The 2023 guideline lists these criteria in its Table 5 and adds the concepts of the "stabilizer" or "nonprogressor," whose ventricle neither remodels nor deteriorates, and the "favorable" and "unfavorable" responder.
The clinical composite score classifies patients as improved, unchanged, or worsened by combining NYHA class, patient global assessment, and major clinical events. Patients who die, are hospitalized for worsening heart failure, or report a worse NYHA class or a moderately or markedly worse global assessment are classified as worsened. Patients who have not worsened and whose NYHA class or global assessment improves are classified as improved. The score was the primary end point of REVERSE and the Adaptive CRT trial.
| Measure | Usual response threshold | Observed response |
|---|---|---|
| Clinical composite score | Improved (NYHA or global assessment better; no death, heart failure hospitalization, or withdrawal for heart failure) | 69% of 426 patients at 6 months in PROSPECT |
| Left ventricular end-systolic volume | Reduction of 15% or more | 56% of 286 patients at 6 months in PROSPECT; 70% vs 55% with targeted vs standard placement in TARGET |
| LVEF | Absolute increase of 5 percentage points or more | 81% to 86% in HeartSync-LBBP |
| Super-response | LVEF rise of 15 points or more, or LVEF reaching 50% (definitions vary) | 36% with biventricular pacing and 55% with left bundle branch pacing in HeartSync-LBBP |
| Late response | Remodeling first evident after 6 months | About 30% of 6-month nonresponders in MORE-CRT MPP |
Definitions of response to CRT and response rates in selected studies. Rates depend on the population, the core laboratory, and the time of assessment.
Roughly one-third of patients do not improve by clinical or echocardiographic criteria, a proportion that has changed little over 2 decades. The proportion is lower in patients with typical LBBB and long QRS, in women, and in nonischemic cardiomyopathy, and higher with non-LBBB morphology, QRS under 150 ms, and extensive scar. A follow-up echocardiogram 3 to 12 months after implantation is recommended to assess remodeling and prognosis (Class 1). Because remodeling can continue for 2 years, a patient without remodeling at 6 months is not yet a confirmed nonresponder, and stabilization of a progressive disease is itself a benefit.
Evaluation and management of the patient with an unfavorable response
An unfavorable response is evaluated systematically. Many causes are correctable, and the 2023 guideline recommends continued optimization of medical and device therapy in all such patients (Class 1).
| Step | Finding sought | Action |
|---|---|---|
| Substrate confirmation | Non-LBBB or QRS under 150 ms; review of the pre-implant ECG by strict criteria | Counseling about the expected benefit; emphasis on medical therapy and on other causes |
| Medical therapy review | Doses below target; missing drug classes; diuretic dose not matched to volume status | Titration of all four foundational drug classes |
| Device interrogation | Biventricular pacing under 98%; atrial fibrillation burden; ectopy; atrioventricular delay; thresholds; phrenic stimulation | Reprogramming; rate or rhythm control; PVC suppression or ablation (Class 2a) |
| 12-lead ECG | Loss of left ventricular capture; pseudofusion; wide paced QRS | Output or vector change; atrioventricular delay shortening |
| Posteroanterior and lateral chest radiographs | Apical or septal left ventricular lead; dislodgement | Radiographs recommended (Class 1); lead revision when position is unfavorable |
| Repeat echocardiogram | Persistent dyssynchrony, mitral regurgitation, right ventricular dysfunction, new ischemia | Atrioventricular or interventricular optimization; treatment of mitral regurgitation |
| Other causes of poor response | Ischemia, valve disease, uncontrolled hypertension, anemia, iron deficiency, sleep apnea | Specific treatment |
| Alternative pacing | Suboptimal resynchronization despite the steps above | His bundle or left bundle branch area pacing, or surgical epicardial lead (Class 2a) |
| Advanced therapies | Progressive stage D heart failure | Referral for transplantation or durable left ventricular assist device evaluation |
Stepwise evaluation of an unfavorable response to biventricular CRT, after the 2023 HRS/APHRS/LAHRS guideline. PVC, premature ventricular complex.
Secondary mitral regurgitation often lessens after resynchronization, and transcatheter edge-to-edge repair is considered only after GDMT and CRT have been optimized; the trial evidence and criteria are covered in the transcatheter repair section of the HFrEF hub. The 2023 guideline recommends against managing congestion by thoracic impedance alone (Class 3: No Benefit), because the DOT-HF trial of 335 patients showed no reduction in death or hospitalization when impedance data were available to clinicians.
Conduction system pacing as an alternative to biventricular CRT
His bundle pacing captures the conduction system proximal to the bundle branches. When LBBB results from a proximal lesion in the left-sided fibers of the His bundle, pacing distal to the block or at high output can correct it, and the paced QRS then shows near-normal activation. His bundle pacing achieves the most physiologic activation, but it has higher capture thresholds that rise over time in a proportion of patients, small sensed R waves, a risk of atrial oversensing, and no correction of block distal to the pacing site. Left bundle branch area pacing places the lead tip in the left side of the septum near the left bundle branch or its fascicles. It bypasses proximal and many intra-Hisian lesions and has lower and more stable thresholds.
Criteria for capture. The 2023 guideline defines left ventricular septal pacing as deep septal lead placement with a right bundle delay pattern in V1. Left bundle branch capture requires, in addition, at least one of 5 findings. The first is a transition between nonselective and selective capture, or between nonselective capture and septal capture, during threshold testing. The second is an abrupt shortening of the V6 R-wave peak time by 10 ms or more at high output. The third is a V6 R-wave peak time under 74 ms without LBBB or under 80 ms with LBBB. The fourth is a V6–V1 interpeak interval above 44 ms. The fifth is a physiology-based criterion such as a left bundle potential with injury current. During implantation and follow-up, a 12-lead ECG is used to document conduction system capture and bundle branch block correction (Class 1).
| Trial | Population | Comparison | Primary end point and result |
|---|---|---|---|
| His-SYNC (2019) | 41; CRT candidates | His bundle vs biventricular pacing | Greater QRS narrowing with His pacing and a trend to greater LVEF gain; 48% crossed over from His to biventricular pacing |
| His-Alternative (2021) | 50; LVEF ≤35%, LBBB | His bundle vs biventricular pacing | LVEF +16 vs +13 points (NS) by intention to treat; His correction in 72%; higher thresholds with His pacing |
| LBBP-RESYNC (2022) | 40; nonischemic cardiomyopathy, LBBB | Left bundle branch vs biventricular pacing | LVEF gain 5.6 points greater with left bundle branch pacing at 6 months |
| LEVEL-AT (2022) | 70; CRT indication with LBBB >130 ms or non-LBBB >150 ms | CSP (mostly LBBAP) vs biventricular pacing | Left ventricular activation time shortening noninferior (−28 vs −21 ms); 23% crossed over from CSP |
| HOT-CRT (2023) | 100; LVEF <50%, CRT indication | CSP with coronary sinus lead if needed vs biventricular pacing | LVEF +12.4 vs +8.0 points at 6 months; complications 6% vs 20% |
| LEFT-BUNDLE-CRT (2026) | 176; Strauss LBBB, CRT indication | LBBAP vs biventricular pacing | Response 89.7% vs 94.6%; noninferiority not met; similar heart failure hospitalization |
| HeartSync-LBBP (2026) | 200; LVEF ≤35%, LBBB | Left bundle branch vs biventricular pacing | Death or HF hospitalization 8% vs 28% (HR 0.26) over a median of 36 months; mortality not different |
| PhysioSync-HF (2026) | 173; LVEF ≤35%, LBBB ≥130 ms, NYHA II–III | CSP vs biventricular pacing | Hierarchical composite worse with CSP (odds ratio 2.36); LVEF 35% vs 39% at 12 months |
| CONDUCT-AF (2026) | 82; AV junction ablation, LVEF <50%, QRS ≤120 ms | LBBAP vs biventricular pacing | LVEF gain noninferior (reported difference 0.4 points; 95% CI, −3.1 to 3.8) |
Published randomized trials comparing conduction system pacing with biventricular pacing for resynchronization. CSP, conduction system pacing; LBBAP, left bundle branch area pacing; NS, not significant.
Early trials. The first randomized comparisons were small and measured electrical or echocardiographic end points. His-SYNC randomized 41 patients and showed greater QRS narrowing with His bundle pacing, but nearly half of the patients assigned to His pacing crossed over, mostly because nonspecific intraventricular delay could not be corrected. In His-Alternative, 50 patients with LVEF of 35% or less and LBBB were randomized; His correction was achieved in 72% of the His group, LVEF rose by 16 and 13 points (not significant), and thresholds were higher with His pacing at implantation (1.8 V vs 1.2 V) and 6 months (2.3 V vs 1.4 V). LBBP-RESYNC randomized 40 patients with nonischemic cardiomyopathy and LBBB and found a 5.6-point greater LVEF improvement with left bundle branch pacing. LEVEL-AT randomized 70 patients and found conduction system pacing noninferior for shortening of left ventricular activation time measured by electrocardiographic imaging, end-systolic volume, and the combined end point of death or heart failure hospitalization. HOT-CRT randomized 100 patients with LVEF below 50%, adding a coronary sinus lead when conduction system pacing alone did not resynchronize; LVEF rose more with this strategy and complications were fewer.
Trials published in 2026. Three larger randomized trials gave discordant results. HeartSync-LBBP randomized 200 patients with LVEF of 35% or less and LBBB at 6 centers in China; implantation succeeded in 98% with left bundle branch pacing and 94% with biventricular pacing. Over a median of 36 months, death or heart failure hospitalization occurred in 8% with left bundle branch pacing and 28% with biventricular pacing (hazard ratio 0.26), driven by heart failure hospitalization (7.0% vs 28.0%); mortality did not differ (2.0% vs 5.0%), echocardiographic response was similar (86.0% vs 81.0%), and super-response was more frequent (55.0% vs 36.0%). PhysioSync-HF randomized 173 patients with NYHA class II or III symptoms, LVEF of 35% or less, and LBBB with QRS of 130 ms or more at 14 hospitals in Brazil. Conduction system pacing, mostly left bundle branch area pacing, failed noninferiority and was inferior to biventricular pacing for a hierarchical composite of death, heart failure hospitalization, urgent heart failure visits, and change in LVEF at 12 months (odds ratio 2.36); LVEF reached 35% with conduction system pacing and 39% with biventricular pacing, and direct medical cost was lower with conduction system pacing. LEFT-BUNDLE-CRT randomized 176 patients with Strauss LBBB. The 6-month response, defined by clinical composite score or a 15% reduction in end-systolic volume, was 89.7% with left bundle branch area pacing and 94.6% with biventricular pacing. Noninferiority was not met, although response rates were high and clinical outcomes were similar in both arms.
Where conduction system pacing stands. Biventricular CRT remains the reference therapy for HFrEF with LBBB and a Class 1 indication, supported by large randomized trials with mortality end points and long-term follow-up. Conduction system pacing is reasonable when biventricular CRT cannot be achieved (Class 2a in LBBB) and may be considered as a first-line alternative (Class 2b). The 2026 trials did not establish superiority of either approach: the largest effect favoring left bundle branch pacing came from the 6-center trial in China with a low event rate in the conduction system arm, and the trial with a hierarchical end point favored biventricular pacing. The 2025 ESC consensus lists frailty or limited life expectancy (favoring simpler devices), failed coronary sinus implantation, and nonresponse to biventricular CRT as settings in which conduction system pacing may be preferred, and judges its role in non-LBBB uncertain. In nonspecific intraventricular delay and in very dilated ventricles, left bundle capture may not shorten lateral wall activation, and adding a coronary sinus lead (LOT-CRT) is an option when the paced QRS remains wide.
Complications of conduction system pacing. In the MELOS registry of 2533 left bundle branch area pacing implants cited by the 2023 guideline, lead complications occurred in 8.3%, including acute perforation into the left ventricle in 3.7%, usually managed by repositioning at the procedure; threshold rise occurred in 0.7% and lead dislodgement in 1.5%. Acute chest pain occurred in 1% and acute coronary syndrome in 0.4%, and the guideline lists extraction of deep septal leads and long-term consequences of delayed right ventricular activation among concerns without long-term data. His bundle leads have higher revision rates than right ventricular leads, and a backup ventricular lead may be reasonable in pacing-dependent patients who receive His bundle pacing (Class 2b). In a CRT-D, a His bundle lead is not used for tachycardia detection, because small R waves and atrial oversensing compromise arrhythmia discrimination.
Follow-up, remote monitoring, and device-based heart failure diagnostics
Remote monitoring is recommended in all patients with a cardiac physiologic pacing device for device and arrhythmia management (Class 1), and multidisciplinary follow-up in combined heart failure and device clinics can be useful (Class 2a). In the CONNECT trial cited by the guideline, remote alerts shortened the median time from a clinical event to a clinical decision from 22 days to 4.6 days. Transmissions report biventricular pacing percentage, atrial arrhythmia burden, ventricular arrhythmias, lead parameters, battery status, activity, heart rate variability, and thoracic impedance.
In IN-TIME, 664 patients with NYHA class II or III heart failure, LVEF of 35% or less, and a recently implanted dual-chamber ICD or CRT-D (58.7% CRT-D) were randomized to daily automatic multiparameter telemonitoring or standard care. At 1 year the clinical composite score worsened in 18.9% with telemonitoring and 27.2% without (odds ratio 0.63), with 10 deaths in the telemonitoring group and 27 in the standard care group. Benefit from monitoring depends on alerts that trigger defined clinical responses: diuretic and GDMT adjustment, atrial fibrillation management, and programming changes. Device follow-up schedules, magnetic resonance imaging conditions, and perioperative management of pacing systems are covered in the device follow-up section of the bradycardia and atrioventricular block hub, and ICD programming and shocks in the ICD programming section of the ventricular arrhythmia hub.
Heart failure with improved ejection fraction after CRT. Biventricular pacing continues at generator replacement in all patients thought to have benefited, including those whose LVEF normalizes and those whose disease stabilized (Class 1). In a small randomized study cited by the guideline, super-responders whose pacing was switched off deteriorated clinically and echocardiographically. GDMT is continued (Class 2a). When a high left ventricular threshold drains the battery, implantation of a new lead may be considered at generator replacement (Class 2b).
Special populations: women, older adults, kidney disease, and end of life
Women. Women have smaller ventricles and shorter stature on average and benefit from CRT at shorter QRS durations than men. The 2023 guideline's Class 1 recommendation for select patients with LBBB and QRS of 120 to 149 ms names female sex as an example.
Older adults. In the patient-level analysis of 8 trials, CRT benefit was as great or greater in patients aged 70 years or older, and age alone does not limit the indication. Frailty, cognitive impairment, and competing illness weigh more than age in the choice between CRT-P and CRT-D and in the decision to implant at all.
Chronic kidney disease. Few patients with advanced kidney disease were enrolled in the CRT trials. Contrast volume during coronary sinus venography is limited, and the arm needed for current or future dialysis access is spared when the implant side is chosen.
Life expectancy under 1 year. Trials excluded patients with limited survival. The 2023 guideline recommends shared decision-making for these patients (Class 1), because CRT can improve symptoms and quality of life even when it does not prolong life. Deactivation of the defibrillator function at the end of life is ethically and legally supported, while continued pacing in a pacing-dependent patient is decided case by case.
Congenital heart disease and surgery. Adults with congenital heart disease and a systemic right ventricle or single ventricle can develop dyssynchrony-related dysfunction, and resynchronization in these patients depends on individual anatomy and venous access. In patients undergoing cardiac surgery, the epicardial lead options described above apply. New LBBB after transcatheter aortic valve implantation occurs in about one quarter of procedures (26.4% of 6,357 patients in a 2025 meta-analysis of 17 studies). It is more frequent with self-expanding valves and deeper implantation. The 2023 guideline makes no recommendation on the type of device in these patients beyond the general rules for atrioventricular block and LBBB.
Prognosis after cardiac resynchronization
The mortality benefit of CRT persists for years in CARE-HF (CRT-P), in MADIT-CRT among patients with LBBB (CRT-D), and in RAFT, where the advantage of CRT-D over ICD alone persisted to a median of nearly 14 years among survivors although most patients in both groups had died by then. Patients with reverse remodeling have lower rates of death, heart failure hospitalization, and ventricular arrhythmia than those without, but arrhythmic risk persists after LVEF improves. Worse prognosis after CRT is associated with non-LBBB morphology, atrial fibrillation with incomplete biventricular pacing, a paced QRS wider than the intrinsic QRS, low biventricular pacing percentage, and absence of reverse remodeling at follow-up.
Sources of error in CRT selection, implantation, and follow-up
- Acceptance of an automated QRS duration near a threshold: automated values differ between systems by several milliseconds, and the indication class changes at 120 and 150 ms. The widest QRS in any lead is measured manually on more than one tracing.
- Classification of nonspecific intraventricular delay or left ventricular hypertrophy with fascicular block as LBBB: the expected benefit is overestimated. Strict LBBB criteria include mid-QRS notching or slurring and, in the Strauss definition, longer QRS thresholds.
- Implantation before LVEF is reassessed on optimized GDMT: patients whose LVEF recovers above 35% receive a device they no longer need, unless a pacing indication justifies a physiologic pacing device at the first procedure.
- Selection of narrow-QRS patients by echocardiographic dyssynchrony: EchoCRT showed no benefit and higher mortality, and PROSPECT showed that no dyssynchrony parameter predicts response.
- Reliance on the device-reported biventricular pacing percentage: pseudofusion, fusion, and loss of left ventricular capture are counted as paced beats. The 12-lead ECG and capture-confirming algorithms establish effective resynchronization.
- Acceptance of an apical or septal left ventricular lead without review of the lateral radiograph: apical position was associated with more heart failure and death in MADIT-CRT, and a posterior course of a "right ventricular" lead on the lateral film can indicate passage into the left heart.
- Interpretation of a dominant R in V1 as proof of correct biventricular capture: left ventricular-only pacing, a malpositioned right ventricular lead, and left bundle branch area pacing all produce a right bundle delay pattern.
- Labeling a patient a nonresponder at 6 months: about 30% of early nonresponders remodel in the following 6 months, and stabilization of a progressive cardiomyopathy is a benefit that remodeling criteria do not capture.
- Routine atrioventricular optimization by echocardiography in all patients: SMART-AV showed no advantage over a fixed 120-ms delay, and the effort is better spent on nonresponders and on patients with abnormal atrioventricular conduction.
- Assumption that atrial fibrillation excludes CRT, or that rate control at rest ensures capture: CRT benefits patients with atrial fibrillation only when biventricular pacing approaches 100%, which often requires atrioventricular junction ablation.
- Generalization of conduction system pacing results across morphologies: randomized evidence concerns LBBB, and in nonspecific intraventricular delay and very dilated ventricles left bundle capture can leave lateral wall activation delayed.
- Deactivation of CRT or downgrade after LVEF normalizes: withdrawal of pacing in super-responders led to deterioration, and the 2023 guideline recommends continuing biventricular pacing at generator replacement.
References
- Chung MK, Patton KK, Lau CP, et al. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm. 2023;20:e17–e91. doi:10.1016/j.hrthm.2023.03.1538
- Glikson M, Burri H, Abdin A, et al. European Society of Cardiology (ESC) clinical consensus statement on indications for conduction system pacing, with special contribution of the European Heart Rhythm Association of the ESC and endorsed by the Asia Pacific Heart Rhythm Society, the Canadian Heart Rhythm Society, the Heart Rhythm Society, and the Latin American Heart Rhythm Society. Europace. 2025;27:euaf050. doi:10.1093/europace/euaf050
- Strauss DG, Selvester RH, Wagner GS. Defining left bundle branch block in the era of cardiac resynchronization therapy. Am J Cardiol. 2011;107:927–934. doi:10.1016/j.amjcard.2010.11.010
- Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure (MIRACLE). N Engl J Med. 2002;346:1845–1853. doi:10.1056/NEJMoa013168
- Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure (COMPANION). N Engl J Med. 2004;350:2140–2150. doi:10.1056/NEJMoa032423
- Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure (CARE-HF). N Engl J Med. 2005;352:1539–1549. doi:10.1056/NEJMoa050496
- Cleland JG, Daubert JC, Erdmann E, et al. Longer-term effects of cardiac resynchronization therapy on mortality in heart failure (the CARE-HF trial extension phase). Eur Heart J. 2006;27:1928–1932. doi:10.1093/eurheartj/ehl099
- Linde C, Abraham WT, Gold MR, et al. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms (REVERSE). J Am Coll Cardiol. 2008;52:1834–1843. doi:10.1016/j.jacc.2008.08.027
- Moss AJ, Hall WJ, Cannom DS, et al. Cardiac-resynchronization therapy for the prevention of heart-failure events (MADIT-CRT). N Engl J Med. 2009;361:1329–1338. doi:10.1056/NEJMoa0906431
- Zareba W, Klein H, Cygankiewicz I, et al. Effectiveness of cardiac resynchronization therapy by QRS morphology in the Multicenter Automatic Defibrillator Implantation Trial-Cardiac Resynchronization Therapy (MADIT-CRT). Circulation. 2011;123:1061–1072. doi:10.1161/CIRCULATIONAHA.110.960898
- Goldenberg I, Kutyifa V, Klein HU, et al. Survival with cardiac-resynchronization therapy in mild heart failure. N Engl J Med. 2014;370:1694–1701. doi:10.1056/NEJMoa1401426
- Tang AS, Wells GA, Talajic M, et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure (RAFT). N Engl J Med. 2010;363:2385–2395. doi:10.1056/NEJMoa1009540
- Sapp JL, Sivakumaran S, Redpath CJ, et al. Long-term outcomes of resynchronization-defibrillation for heart failure. N Engl J Med. 2024;390:212–220. doi:10.1056/NEJMoa2304542
- Ruschitzka F, Abraham WT, Singh JP, et al. Cardiac-resynchronization therapy in heart failure with a narrow QRS complex (EchoCRT). N Engl J Med. 2013;369:1395–1405. doi:10.1056/NEJMoa1306687
- Chung ES, Leon AR, Tavazzi L, et al. Results of the Predictors of Response to CRT (PROSPECT) trial. Circulation. 2008;117:2608–2616. doi:10.1161/CIRCULATIONAHA.107.743120
- Curtis AB, Worley SJ, Adamson PB, et al. Biventricular pacing for atrioventricular block and systolic dysfunction (BLOCK-HF). N Engl J Med. 2013;368:1585–1593. doi:10.1056/NEJMoa1210356
- Merkely B, Hatala R, Wranicz JK, et al. Upgrade of right ventricular pacing to cardiac resynchronization therapy in heart failure: a randomized trial (BUDAPEST-CRT). Eur Heart J. 2023;44:4259–4269. doi:10.1093/eurheartj/ehad591
- Cleland JG, Abraham WT, Linde C, et al. An individual patient meta-analysis of five randomized trials assessing the effects of cardiac resynchronization therapy on morbidity and mortality in patients with symptomatic heart failure. Eur Heart J. 2013;34:3547–3556. doi:10.1093/eurheartj/eht290
- Friedman DJ, Al-Khatib SM, Dalgaard F, et al. Cardiac resynchronization therapy improves outcomes in patients with intraventricular conduction delay but not right bundle branch block: a patient-level meta-analysis of randomized controlled trials. Circulation. 2023;147:812–823. doi:10.1161/CIRCULATIONAHA.122.062124
- Zusterzeel R, Selzman KA, Sanders WE, et al. Cardiac resynchronization therapy in women: US Food and Drug Administration meta-analysis of patient-level data. JAMA Intern Med. 2014;174:1340–1348. doi:10.1001/jamainternmed.2014.2717
- Zeitler EP, Dalgaard F, Abraham WT, et al. Benefit of cardiac resynchronization therapy among older patients: a patient-level meta-analysis. Am Heart J. 2024;267:81–90. doi:10.1016/j.ahj.2023.11.002
- Køber L, Thune JJ, Nielsen JC, et al. Defibrillator implantation in patients with nonischemic systolic heart failure (DANISH). N Engl J Med. 2016;375:1221–1230. doi:10.1056/NEJMoa1608029
- Brignole M, Pokushalov E, Pentimalli F, et al. A randomized controlled trial of atrioventricular junction ablation and cardiac resynchronization therapy in patients with permanent atrial fibrillation and narrow QRS (APAF-CRT). Eur Heart J. 2018;39:3999–4008. doi:10.1093/eurheartj/ehy555
- Brignole M, Pentimalli F, Palmisano P, et al. AV junction ablation and cardiac resynchronization for patients with permanent atrial fibrillation and narrow QRS: the APAF-CRT mortality trial. Eur Heart J. 2021;42:4731–4739. doi:10.1093/eurheartj/ehab569
- Hayes DL, Boehmer JP, Day JD, et al. Cardiac resynchronization therapy and the relationship of percent biventricular pacing to symptoms and survival. Heart Rhythm. 2011;8:1469–1475. doi:10.1016/j.hrthm.2011.04.015
- Boriani G, Connors S, Kalarus Z, et al. Cardiac resynchronization therapy with a quadripolar electrode lead decreases complications at 6 months: results of the MORE-CRT randomized trial. JACC Clin Electrophysiol. 2016;2:212–220. doi:10.1016/j.jacep.2015.10.004
- Singh JP, Klein HU, Huang DT, et al. Left ventricular lead position and clinical outcome in the Multicenter Automatic Defibrillator Implantation Trial-Cardiac Resynchronization Therapy (MADIT-CRT) trial. Circulation. 2011;123:1159–1166. doi:10.1161/CIRCULATIONAHA.110.000646
- Khan FZ, Virdee MS, Palmer CR, et al. Targeted left ventricular lead placement to guide cardiac resynchronization therapy: the TARGET study: a randomized, controlled trial. J Am Coll Cardiol. 2012;59:1509–1518. doi:10.1016/j.jacc.2011.12.030
- Gold MR, Birgersdotter-Green U, Singh JP, et al. The relationship between ventricular electrical delay and left ventricular remodelling with cardiac resynchronization therapy. Eur Heart J. 2011;32:2516–2524. doi:10.1093/eurheartj/ehr329
- Nielsen JC, Svendsen JH, Johansen JB, et al. Targeted left ventricular lead placement in biventricular pacing for heart failure: a national, multicentre, double-blind, randomised controlled trial in Denmark. Lancet. 2026;408:1175–1185. doi:10.1016/S0140-6736(26)01597-7
- Singh JP, Rinaldi CA, Sanders P, et al. Leadless ultrasound-based cardiac resynchronization system in heart failure (SOLVE-CRT). JAMA Cardiol. 2024;9:871–879. doi:10.1001/jamacardio.2024.2050
- Ellenbogen KA, Gold MR, Meyer TE, et al. Primary results from the SmartDelay determined AV optimization: a comparison to other AV delay methods used in cardiac resynchronization therapy (SMART-AV) trial. Circulation. 2010;122:2660–2668. doi:10.1161/CIRCULATIONAHA.110.992552
- Wilkoff BL, Filippatos G, Leclercq C, et al. Adaptive versus conventional cardiac resynchronisation therapy in patients with heart failure (AdaptResponse): a global, prospective, randomised controlled trial. Lancet. 2023;402:1147–1157. doi:10.1016/S0140-6736(23)00912-1
- Leclercq C, Burri H, Delnoy PP, et al. Cardiac resynchronization therapy non-responder to responder conversion rate in the MORE-CRT MPP trial. Europace. 2023;25:euad294. doi:10.1093/europace/euad294
- Packer M. Proposal for a new clinical end point to evaluate the efficacy of drugs and devices in the treatment of chronic heart failure. J Card Fail. 2001;7:176–182. doi:10.1054/jcaf.2001.25652
- Upadhyay GA, Vijayaraman P, Nayak HM, et al. His corrective pacing or biventricular pacing for cardiac resynchronization in heart failure (His-SYNC). J Am Coll Cardiol. 2019;74:157–159. doi:10.1016/j.jacc.2019.04.026
- Vinther M, Risum N, Svendsen JH, et al. A randomized trial of His pacing versus biventricular pacing in symptomatic HF patients with left bundle branch block (His-Alternative). JACC Clin Electrophysiol. 2021;7:1422–1432. doi:10.1016/j.jacep.2021.04.003
- Wang Y, Zhu H, Hou X, et al. Randomized trial of left bundle branch vs biventricular pacing for cardiac resynchronization therapy (LBBP-RESYNC). J Am Coll Cardiol. 2022;80:1205–1216. doi:10.1016/j.jacc.2022.07.019
- Pujol-Lopez M, Jiménez-Arjona R, Garre P, et al. Conduction system pacing vs biventricular pacing in heart failure and wide QRS patients: LEVEL-AT trial. JACC Clin Electrophysiol. 2022;8:1431–1445. doi:10.1016/j.jacep.2022.08.001
- Vijayaraman P, Pokharel P, Subzposh FA, et al. His-Purkinje conduction system pacing optimized trial of cardiac resynchronization therapy vs biventricular pacing: HOT-CRT clinical trial. JACC Clin Electrophysiol. 2023;9:2628–2638. doi:10.1016/j.jacep.2023.08.003
- Cano Ó, Pérez-Roselló V, Di Marco A, et al. Left bundle branch area vs biventricular pacing for cardiac resynchronization therapy: the LEFT-BUNDLE-CRT trial. Eur Heart J. Published online April 14, 2026. doi:10.1093/eurheartj/ehag225
- Chen X, Liu X, Li R, et al. Long-term outcomes of left bundle-branch pacing vs biventricular pacing in heart failure: the HeartSync-LBBP randomized clinical trial. JAMA Cardiol. 2026;11:352–359. doi:10.1001/jamacardio.2026.0083
- Zimerman A, Dal Forno A, Rohde LE, et al. Conduction system vs biventricular pacing in heart failure: the PhysioSync-HF randomized clinical trial. JAMA Cardiol. 2026;11:360–368. doi:10.1001/jamacardio.2026.0101
- Žižek D, Ivanovski M, Mrak M, et al. Biventricular vs conduction system pacing after atrioventricular node ablation for heart failure with atrial fibrillation: the CONDUCT-AF trial. Eur Heart J. Published online October 6, 2026. doi:10.1093/eurheartj/ehag699
- Hindricks G, Taborsky M, Glikson M, et al. Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet. 2014;384:583–590. doi:10.1016/S0140-6736(14)61176-4
- Alzu'bi H, Rmilah AA, Bahmad HF, et al. Predictors of post-TAVR left bundle branch block: a systematic review and meta-analysis. J Cardiovasc Electrophysiol. 2025;36:2041–2060. doi:10.1111/jce.16725