CardioBoardExam.com

Library › Devices and mechanical support › Implantable cardioverter-defibrillators

Implantable cardioverter-defibrillators: transvenous, subcutaneous, extravascular, and wearable

Implantable and wearable cardioverter-defibrillators in adults, from system design and device choice through detection and therapy programming, shocks, complications, follow-up, shared decision-making, and deactivation.

An implantable cardioverter-defibrillator (ICD) detects ventricular tachycardia (VT) and ventricular fibrillation (VF) by rate and duration and terminates them with antitachycardia pacing or a shock. Transvenous systems also provide bradycardia pacing and resynchronization. Subcutaneous and extravascular systems keep all leads outside the veins and heart, and the wearable cardioverter-defibrillator protects for weeks to months while a permanent decision is made. In primary prevention, a lowest therapy zone of 185–200 bpm and detection that lasts 6–12 seconds or 30 intervals reduce inappropriate and unnecessary therapy. In the Multicenter Automatic Defibrillator Implantation Trial–Reduce Inappropriate Therapy (MADIT-RIT), high-rate programming also lowered all-cause mortality.

Defibrillator systems and what they provide

An ICD is a cardiac implantable electronic device that continuously senses ventricular activation, classifies rhythms by rate, duration, and morphology, and delivers tiered therapy: antitachycardia pacing (ATP), synchronized cardioversion, and unsynchronized defibrillation. All transvenous ICDs also function as pacemakers. The device family includes transvenous, subcutaneous, and extravascular implanted systems and one external system. The choice among them depends on the need for pacing, the need for ATP or resynchronization, vascular access, infection risk, and expected years of device dependence.

SystemLead and generator positionBradycardia pacingATPResynchronizationPrincipal liabilities
Single-chamber transvenous ICDRight ventricular lead with one or two shock coils; left pectoral generatorYes (ventricular)YesNoLead fracture or insulation failure, venous occlusion, tricuspid regurgitation, and endovascular infection
Dual-chamber transvenous ICDAdds a right atrial leadYes (atrial and ventricular)YesNoOne more lead to fail; higher early complication rate when the atrial lead is not needed for pacing
Cardiac resynchronization therapy defibrillator (CRT-D)Adds a coronary sinus lead or a conduction system pacing leadYesYesYesLonger procedure, coronary sinus lead dislodgement, phrenic nerve stimulation
Subcutaneous ICDParasternal subcutaneous lead; generator over the left lateral chest wall near the midaxillary lineNo (brief post-shock pacing only)NoNoInappropriate shocks from oversensing; ECG screening required; no pacing
Extravascular (substernal) ICDLead in the anterior mediastinum behind the sternum; generator in the left midaxillary regionPause-prevention pacing onlyYesNoSubsternal tunneling, lead dislodgement, oversensing; shorter follow-up
Wearable cardioverter-defibrillatorGarment with dry sensing electrodes and gel-releasing defibrillation pads; monitor worn at the waist or on a strapNoNoNoProtection only while worn; discomfort; temporary use

Defibrillator systems used in adults.

Pulse generator, capacitors, battery, and leads

The pulse generator is a hermetically sealed titanium can that houses the battery, one or more high-voltage capacitors, sensing amplifiers, the charging circuit, the output switching network, telemetry, and memory for stored electrograms. In most transvenous systems the can itself is an active electrode, the "active can," and forms part of the shock vector. The header holds the lead connectors; contemporary transvenous systems use a single four-pole high-voltage connector (DF4). It carries the pace-sense and shock conductors in one pin and removes the separate shock-coil set screws of older DF-1 connectors.

Capacitors and charging. A battery cannot deliver the current a defibrillation shock requires, so the device charges one or two capacitors over several seconds and then discharges them through the shock electrodes. Charge time therefore adds to the interval between onset of VF and therapy, and it lengthens as the battery depletes. Capacitors are periodically reformed by automatic charging to maintain their dielectric layer; prolonged charge times on interrogation indicate battery depletion or capacitor deterioration. When the battery reaches the elective replacement indicator, the generator is replaced within the period the manufacturer specifies, because charge times and remaining therapies become unreliable after end of service.

Transvenous leads. A defibrillation lead combines a pace-sense tip and ring, one or two shock coils, insulation, and a fixation mechanism. In the integrated bipolar configuration, the lead senses between the tip and the distal coil instead of a separate ring. A dual-coil lead places one coil in the right ventricle and one in the superior vena cava; a single-coil lead relies on the right ventricular coil and the can. A superior vena cava coil can lower the defibrillation threshold in some patients, but it becomes adherent to the vein wall and makes extraction more difficult and more hazardous. Active-fixation leads anchor with an extendable helix and passive-fixation leads with tines. True bipolar sensing (tip to ring) is less susceptible to far-field signals than integrated bipolar sensing.

Frontal chest radiograph showing an ICD generator in the left upper chest below the clavicle. A single lead passes through the left subclavian region and superior vena cava into the right ventricle, with a short dense shock coil in the superior vena cava and a second dense coil in the right ventricle ending at the apex.
Single-chamber transvenous ICD with a dual-coil lead. The generator lies in a left infraclavicular pocket, and the lead enters the right ventricle with its tip near the apex. The two radiopaque segments are the shock coils, one in the superior vena cava and one in the right ventricle. Credit: Gregory Marcus, Wikimedia Commons, CC BY 3.0.

Nonvascular leads. The subcutaneous ICD lead runs parallel to the left sternal border, with a shock coil between two sensing electrodes, and the generator sits over the lateral chest wall. The extravascular ICD lead is tunneled from the xiphoid into the substernal space so that its two coils and two ring electrodes lie close to the anterior heart. This position permits pacing and lower shock energies than a subcutaneous lead.

Defibrillation physics: biphasic waveforms, shock vectors, and the defibrillation threshold

Defibrillation succeeds when the shock depolarizes or prolongs refractoriness in a critical mass of myocardium so that the wandering wavefronts of VF have no excitable tissue to enter. ICDs deliver a truncated exponential biphasic waveform: the capacitor discharges for a first phase, the output bridge reverses polarity, and a shorter second phase follows. Biphasic waveforms defibrillate at lower energy than monophasic waveforms, which made pectoral transvenous systems practical. The shock vector runs between the right ventricular coil and the active can, with the superior vena cava coil added in dual-coil systems. The subcutaneous ICD shocks between the parasternal coil and the lateral can, and the extravascular ICD between substernal coils and a lateral can.

The defibrillation threshold is the lowest energy that terminates VF, but it is probabilistic: the same energy succeeds on some attempts and fails on others. Device efficacy is therefore judged by the margin between the energy that reliably terminates induced VF and the maximum output. Contemporary transvenous generators deliver shocks of up to about 40 joules (J), and all shocks were programmed to 40 J in the NORDIC ICD trial. The subcutaneous ICD delivers a nonprogrammable 80 J for spontaneous episodes, and the extravascular ICD was designed to deliver defibrillation energy similar to that of transvenous systems. The higher subcutaneous energy compensates for current that passes through chest wall tissue before it reaches the heart.

Factors that raise the defibrillation threshold include severe left ventricular hypertrophy or dilation, amiodarone, a right-sided transvenous implant, and a lead positioned away from the right ventricular apex or septum. In subcutaneous systems, fat or air between the device and the chest wall and a generator placed too anteriorly also raise it. For the subcutaneous ICD, the PRAETORIAN score (named for the Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter-Defibrillator Therapy trial) grades implant position on the postoperative chest radiograph as low, intermediate, or high risk. It uses the amount of subcutaneous fat beneath the coil and the generator and the position of the generator. In 1,063 patients in the iSUSI registry, the composite of failed defibrillation testing and ineffective spontaneous shocks occurred in 4.5%, 6.7%, and 16.8% of patients with low, intermediate, and high PRAETORIAN scores, respectively. Subcutaneous fat beneath the coil, anterior generator position, high body mass index, and high post-shock impedance independently predicted failure. Options when testing shows an inadequate margin include reversing shock polarity, adding or removing the superior vena cava coil from the vector, repositioning the lead, and adding a subcutaneous array. In subcutaneous systems, the generator or coil is repositioned.

Indications in summary: who receives an ICD

ICD indications by disease are covered in detail on the ventricular tachycardia and sudden cardiac death page. All ICD recommendations in the 2017 American Heart Association/American College of Cardiology/Heart Rhythm Society (AHA/ACC/HRS) guideline assume meaningful survival of more than 1 year with acceptable functional status.

Secondary prevention. Survivors of cardiac arrest due to VF or hemodynamically unstable VT, and patients with sustained VT and structural heart disease, receive an ICD when no completely reversible cause is found (Class 1). The basis is three randomized comparisons with amiodarone: the Antiarrhythmics Versus Implantable Defibrillators trial (AVID), the Cardiac Arrest Study Hamburg (CASH), and the Canadian Implantable Defibrillator Study (CIDS). In the individual-patient meta-analysis, the ICD reduced all-cause mortality (hazard ratio [HR] 0.72; 95% confidence interval [CI] 0.60 to 0.87) and arrhythmic death (HR 0.50), extending mean survival by 4.4 months over 6 years, with greater benefit when the ejection fraction was 35% or less. Survivor evaluation and the exclusion of reversible causes are covered on the survivors of sudden cardiac arrest section and the post-cardiac arrest care page.

Primary prevention in ischemic cardiomyopathy. The Multicenter Automatic Defibrillator Implantation Trial II (MADIT-II) randomized 1,232 patients with prior myocardial infarction (MI) and an ejection fraction of 30% or less, without electrophysiologic testing, to an ICD or conventional therapy. Mortality over an average 20 months was 14.2% with the ICD and 19.8% without (HR 0.69; P=0.016). The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) randomized 2,521 patients with ejection fraction 35% or less and New York Heart Association (NYHA) class II or III heart failure, half with nonischemic disease, to a single-chamber, shock-only ICD, amiodarone, or placebo. The ICD reduced mortality (HR 0.77), and amiodarone did not. The resulting Class 1 indications require guideline-directed medical therapy and a wait of at least 40 days after MI and 90 days after revascularization. The qualifying ejection fraction is 35% or less with NYHA class II or III symptoms, or 30% or less with NYHA class I symptoms.

Primary prevention in nonischemic cardiomyopathy. The Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) trial and SCD-HeFT supported the ICD in nonischemic cardiomyopathy. The Danish Study to Assess the Efficacy of ICDs in Patients with Non-ischemic Systolic Heart Failure on Mortality (DANISH) randomized 1,116 patients, 58% with cardiac resynchronization therapy (CRT) in both groups, and found no difference in all-cause death over a median 67.6 months (21.6% vs 23.4%; HR 0.87; P=0.28). Sudden cardiac death was nonetheless halved (4.3% vs 8.2%; HR 0.50). The 2017 guideline recommendation remains Class 1 for ejection fraction 35% or less with NYHA class II or III symptoms despite guideline-directed therapy. Age, comorbidity, scar on cardiac magnetic resonance, high-risk genotypes such as LMNA, and the patient's preferences shape the decision, as set out on the heart failure with reduced ejection fraction page.

Inherited and infiltrative disease. In hypertrophic cardiomyopathy, arrhythmogenic right ventricular cardiomyopathy, long QT syndrome, Brugada syndrome, catecholaminergic polymorphic VT, cardiac sarcoidosis, and lamin A/C cardiomyopathy, ICD decisions follow disease-specific risk markers and are made with the ejection fraction as only one input. In catecholaminergic polymorphic VT, the pain and catecholamine surge of a shock can provoke further arrhythmia and a shock-induced storm, and shocks delivered for the initiating triggered rhythm rarely terminate it. The ICD is therefore combined with nonselective beta-blockade, flecainide, or left cardiac sympathetic denervation and programmed with long detection and a high rate cutoff. In cardiac sarcoidosis, the 2017 guideline supports an ICD with sustained ventricular arrhythmia or an ejection fraction of 35% or less. An ICD is reasonable with an ejection fraction above 35% when syncope, myocardial scar on magnetic resonance or positron emission tomography, inducible sustained VT, or a pacing indication is present (Class 2a). These decisions are covered on the hypertrophic cardiomyopathy, arrhythmogenic right ventricular cardiomyopathy, long QT syndrome, Brugada syndrome, and nonischemic cardiomyopathy and sarcoidosis sections.

TrialPopulationComparisonResult
AVID, CASH, CIDS (meta-analysis)Survivors of VF or sustained VTICD vs amiodaroneAll-cause death HR 0.72; arrhythmic death HR 0.50
MADIT-IIPrior MI, ejection fraction ≤30%ICD vs conventional therapyAll-cause death 14.2% vs 19.8%; HR 0.69
SCD-HeFTEjection fraction ≤35%, NYHA II–III, ischemic or nonischemicShock-only ICD vs amiodarone vs placeboICD HR 0.77 vs placebo; amiodarone no benefit
DEFINITENonischemic cardiomyopathy, ejection fraction ≤35%, ventricular ectopy or nonsustained VTICD vs standard therapyAll-cause death HR 0.65 (P=0.08); arrhythmic death HR 0.20
DANISHNonischemic cardiomyopathy, ejection fraction ≤35%, contemporary therapyICD vs usual careAll-cause death HR 0.87 (P=0.28); sudden death HR 0.50
Defibrillator in Acute Myocardial Infarction Trial (DINAMIT)6 to 40 days after MI, ejection fraction ≤35%, impaired autonomic markersICD vs no ICDNo mortality difference (HR 1.08); fewer arrhythmic and more nonarrhythmic deaths

Randomized trials that define the benefit of the ICD. The early post-MI trials (DINAMIT and the later Immediate Risk-Stratification Improves Survival [IRIS] trial) explain the 40-day waiting period. HR, hazard ratio.

When an ICD is not indicated or offers no net benefit

The ICD prevents arrhythmic death; it does not treat the substrate or the competing causes of death. Net benefit depends on the share of deaths that are sudden and arrhythmic, which falls as heart failure advances and as noncardiac illness accumulates. The 2017 guideline lists the situations in which an ICD is not indicated:

The early post-MI period is the clearest example of competing risk. In DINAMIT (6 to 40 days after MI), the ICD reduced arrhythmic death (HR 0.42) but nonarrhythmic death increased (HR 1.75), and all-cause mortality did not differ (HR 1.08). The IRIS trial produced the same pattern. Patients early after MI or revascularization, or with newly diagnosed nonischemic cardiomyopathy, are therefore reassessed after a period of guideline-directed medical therapy; the wearable cardioverter-defibrillator can bridge selected patients through this period.

Choosing the system: transvenous, subcutaneous, or extravascular

Once an ICD is indicated, four considerations determine the system: the present or foreseeable need for bradycardia pacing, an indication for resynchronization, the likelihood that monomorphic VT will be terminated by ATP, and the lifetime burden of transvenous lead risk.

Patient featureFavorsBasis
Bradycardia pacing indication, or high-grade conduction disease likelyTransvenous ICD (or CRT-D)Subcutaneous ICD cannot pace (Class 3: Harm when pacing is needed); extravascular ICD provides only pause prevention
Ejection fraction ≤35%, sinus rhythm, left bundle branch block, QRS ≥150 ms, NYHA II, III, or ambulatory IV on guideline-directed therapyCRT-DResynchronization indication; see the cardiac resynchronization therapy section
Recurrent monomorphic VT known to respond to pacingTransvenous or extravascular ICDATP terminates most monomorphic VT without a shock; the subcutaneous ICD has no ATP
Inadequate venous access (occlusion, congenital anatomy, hemodialysis access preservation)Subcutaneous or extravascular ICDSubcutaneous ICD recommended when pacing is not needed (Class 1)
High infection risk or prior device infectionSubcutaneous or extravascular ICDNo intravascular lead; lower risk of bloodstream infection and lead-associated endocarditis
Young patient expecting decades of device therapySubcutaneous or extravascular ICDAvoids cumulative transvenous lead failure and extraction risk
Failed subcutaneous ICD ECG screeningTransvenous or extravascular ICDScreening predicts T-wave oversensing
Any ICD candidate without a pacing, ATP, or CRT needSubcutaneous ICDReasonable per 2017 AHA/ACC/HRS (Class 2a); PRAETORIAN noninferiority

Device selection by patient feature. Classes from the 2017 AHA/ACC/HRS guideline. A modular system that pairs a subcutaneous ICD with a communicating leadless pacemaker adds ATP and bradycardia pacing without transvenous leads.

When CRT is indicated, the choice between CRT-D and a CRT pacemaker follows the same primary-prevention considerations as a standalone ICD, with added weight on age and comorbidity after DANISH.

Single- versus dual-chamber transvenous ICDs and bradycardia programming

An atrial lead is placed when the patient needs atrial pacing or atrioventricular synchrony: sinus node dysfunction with a pacing indication, atrioventricular block in a patient with preserved ventricular function, or pacemaker syndrome. An atrial lead added solely to improve discrimination of supraventricular tachycardia (SVT) has not shown clinical benefit. A propensity-matched analysis in the National Cardiovascular Data Registry included 32,034 Medicare beneficiaries without a pacing indication who received a primary-prevention ICD. Dual-chamber devices had more complications within 90 days (4.72% vs 3.51%; P<0.001) and similar 1-year mortality (HR 0.99) and hospitalization. The 2015 consensus statement of the Heart Rhythm Society, European Heart Rhythm Association, Asia Pacific Heart Rhythm Society, and Latin American Society of Cardiac Stimulation and Electrophysiology (HRS/EHRA/APHRS/SOLAECE) states that a single-chamber ICD is reasonable in preference to a dual-chamber device when the only reason for the atrial lead is SVT discrimination, unless a known SVT may enter the VT zone (Class 2a).

The Dual Chamber and VVI Implantable Defibrillator (DAVID) trial established the principle of minimizing ventricular pacing. In 506 patients with ejection fraction 40% or less and no pacing indication, dual-chamber rate-responsive pacing at 70 bpm produced frequent right ventricular pacing. It reduced 1-year survival free of death or heart failure hospitalization compared with ventricular backup pacing at 40 bpm (73.3% vs 83.9%; relative hazard 1.61).

SituationProgrammingClass
No bradycardia pacing indicationPacing parameters adjusted so that ventricular stimulation is minimized1
Sinus node disease with a pacing indicationDual-chamber pacing to reduce atrial fibrillation and stroke, avoid pacemaker syndrome, and improve quality of life1
Atrioventricular block, sinus rhythm, no or mild left ventricular dysfunctionDual-chamber in preference to single-chamber ventricular pacing2a
Atrioventricular block, mild to moderate left ventricular dysfunction, ventricular pacing expectedCRT in preference to dual-chamber pacing2a
Dual-chamber ICD with native PR interval ≤230 msMode, automatic mode change, and rate response programmed so that native conduction minimizes ventricular pacing2a
Chronotropic incompetenceSensor-driven rate response, especially in young and active patients2a
Biventricular ICDHighest achievable percentage of biventricular pacing, preferably above 98%2a

Bradycardia mode and rate programming in ICD recipients, 2015 HRS/EHRA/APHRS/SOLAECE consensus. Pacing-induced cardiomyopathy and conduction system pacing are covered on the bradycardia and atrioventricular block page.

Transvenous implantation: access, lead position, and periprocedural anticoagulation

Transvenous ICDs are implanted under local anesthesia with sedation through the left cephalic, axillary, or subclavian vein, with the generator in a prepectoral or subpectoral pocket. Left-sided implantation places the heart within the shock vector between the right ventricular coil and the can; right-sided implants have higher defibrillation thresholds. Extrathoracic axillary access, guided by fluoroscopy or ultrasound, avoids the costoclavicular space where leads are crushed between the clavicle and first rib and reduces pneumothorax compared with an intrathoracic subclavian puncture. The right ventricular lead is positioned at the apex or on the septum; at implant, the R wave is ideally above 5 to 7 mV, the pacing threshold low, and pacing and shock impedances within the manufacturer's ranges. Fluoroscopy in two views confirms septal position and excludes placement in the coronary sinus or the left ventricle through an atrial septal defect or patent foramen ovale.

Anticoagulation. In the Bridge or Continue Coumadin for Device Surgery Randomized Controlled Trial (BRUISE CONTROL), among 681 patients with an annual thromboembolic risk of 5% or more, continued warfarin reduced clinically significant pocket hematoma compared with heparin bridging (3.5% vs 16.0%; relative risk 0.19). Heparin bridging is therefore avoided. In BRUISE CONTROL-2, continuing or interrupting a direct oral anticoagulant produced the same hematoma rate (2.1% in both arms). Antiplatelet management, especially with dual antiplatelet therapy after recent coronary stenting, is individualized. Periprocedural anticoagulant decisions are covered on the anticoagulant reversal and periprocedural management page.

Infection prevention. A single preprocedural dose of an antistaphylococcal antibiotic, usually cefazolin, is standard. In the cluster-randomized Prevention of Arrhythmia Device Infection Trial (PADIT) of 19,603 patients, adding vancomycin, a bacitracin pocket wash, and 2 days of oral cephalexin did not significantly reduce hospitalization for infection. In the World-wide Randomized Antibiotic Envelope Infection Prevention Trial (WRAP-IT), an absorbable antibiotic-eluting envelope reduced major infection at 12 months in patients undergoing generator replacement, upgrade, revision, or initial CRT-D implantation (0.7% vs 1.2%; HR 0.60; P=0.04), without more complications. Hematoma, early reintervention, generator replacement, kidney failure, diabetes, and corticosteroid use raise the risk of infection.

Defibrillation testing at implantation

Defibrillation testing induces VF and confirms that the system senses, detects, and terminates it with a margin below maximum output. Testing requires sedation and an external defibrillator, and it carries rare risks of failed resuscitation, stroke, and pulmonary edema. Two randomized trials showed that routine testing at a first left-sided transvenous implant does not improve outcomes.

In the Shockless IMPLant Evaluation (SIMPLE), 2,500 patients were randomized to testing or no testing. The primary outcome of arrhythmic death or failed appropriate shock occurred in 7% per year without testing and 8% per year with it (HR 0.86; 95% CI 0.65 to 1.14; P<0.0001 for noninferiority), and 30-day safety outcomes were similar. In NORDIC ICD, 1,077 patients were randomized with all shocks programmed to 40 J, and first-shock efficacy for spontaneous VT and VF during follow-up was noninferior without testing.

SettingRecommendationClass
Subcutaneous ICD implantationTesting recommended1
Initial left pectoral transvenous implant with satisfactory sensing, pacing, impedance, and fluoroscopic lead positionOmitting testing reasonable2a
Right pectoral transvenous implant or pulse generator changeTesting reasonable2a
Nonchronic intracardiac thrombus, atrial fibrillation or flutter without adequate anticoagulation, critical aortic stenosis, unstable coronary disease, recent stroke or transient ischemic attack, hemodynamic instabilityTesting should not be performed3: Harm

Intraprocedural defibrillation testing, 2015 HRS/EHRA/APHRS/SOLAECE consensus. When testing is omitted, the initial shock energy in the highest zone is programmed to the maximum available output (Class 2a).

Testing is also performed during extravascular ICD implantation, as in the pivotal study, and it is considered when hypertrophic cardiomyopathy or a high-threshold substrate is present or the lead position is in doubt.

Subcutaneous ICD: design, sensing vectors, and ECG screening

The subcutaneous ICD (S-ICD) consists of a generator implanted over the left lateral chest wall near the midaxillary line and a single lead tunneled subcutaneously from a small incision at the xiphoid along the left parasternal margin. The lead carries a shock coil between a proximal and a distal sensing electrode. Three sensing vectors are available: proximal electrode to can, distal electrode to can, and distal to proximal electrode. All vectors record a far-field signal similar to a surface ECG, so the device senses both the QRS complex and the T wave from outside the chest. T-wave oversensing is the dominant cause of inappropriate shocks. Implantation is guided by anatomic landmarks with limited fluoroscopy, and the two-incision technique omits the superior parasternal incision. An intermuscular generator position between the serratus anterior and latissimus dorsi is widely used and keeps the can posterior.

ECG screening. Before implantation, a surface ECG recorded with electrodes in the positions of the subcutaneous sensing electrodes and can simulates the three sensing vectors. A manufacturer template tests whether the QRS and T-wave amplitudes fit in the sensing vectors, at rest in the supine and standing positions and, when relevant, during exercise. The 2015 consensus states that an S-ICD is not implanted unless at least one vector is satisfactory in both the supine and the standing position. A prolonged QRS duration predicts screening failure. Screening fails more often with bundle branch block, hypertrophic cardiomyopathy with large T waves, Brugada syndrome with dynamic ST changes, and congenital heart disease with right ventricular hypertrophy.

Detection. The subcutaneous ICD uses a shock zone, in which rate alone triggers therapy, and an optional conditional shock zone. In the conditional zone, rhythms are analyzed by a static morphology comparison with a stored normal QRS-T template, a dynamic beat-to-beat comparison, and QRS width. Supraventricular rhythms and oversensing are rejected. Detection uses a nonprogrammable 18-of-24-interval criterion that extends automatically after nonsustained episodes, and the device confirms the arrhythmia at the end of charging before the shock. The 2015 consensus considers it reasonable to program two zones: a conditional zone with discrimination starting at 200 bpm or less and a shock zone without discrimination at 230 bpm or more (Class 2a). A high-pass sensing filter reduces T-wave oversensing and inappropriate shocks without reducing appropriate shocks.

Subcutaneous ICD evidence: IDE, EFFORTLESS, UNTOUCHED, PRAETORIAN, and ATLAS

The subcutaneous ICD has been evaluated in a nonrandomized approval study, a large registry, a performance-goal study in a typical primary-prevention population, and two randomized comparisons with transvenous ICDs.

StudyDesign and populationMain findings
S-ICD Investigational Device Exemption (IDE) study (2013)Prospective, nonrandomized; 330 enrolled, 314 implanted; mean ejection fraction 36%180-day complication-free rate 99%; all 38 spontaneous VT or VF episodes in 21 patients converted; inappropriate shock 13.1% at mean 11 months
EFFORTLESS (2017)Global registry; 985 patients, mean age 48, 65% primary preventionComplications 4.1% at 30 days and 8.4% at 1 year; inappropriate shock 8.1% at 1 year and 11.7% at 3.1 years; spontaneous conversion 97.4%
UNTOUCHED (2021)1,111 primary-prevention patients, ejection fraction ≤35% (mean 26.4%), no pacing need; conditional zone 200–249 bpm and shock zone ≥250 bpm18-month freedom from inappropriate shock 95.9%; all-cause shock-free 90.6%; complication-free 92.7%; conversion of appropriate discrete episodes 98.4%
PRAETORIAN (2020)Randomized, 849 patients without pacing need; subcutaneous vs transvenous; median follow-up 49.1 monthsComposite of complications and inappropriate shocks 15.1% vs 15.7% at 48 months (HR 0.99; noninferior); complications 31 vs 44 patients (HR 0.69); inappropriate shocks 41 vs 29 patients (HR 1.43)
ATLAS (2022)Randomized, 503 patients younger than 60, with an inherited phenotype, or with lead-complication risk factorsPerioperative major lead-related complications 0.4% vs 4.8%; trend toward more inappropriate shocks (HR 2.37; 95% CI 0.98 to 5.77); more pain on the day of implant

Studies of the subcutaneous ICD. In PRAETORIAN, appropriate shocks were more frequent in the subcutaneous group (83 vs 57 patients) and deaths did not differ significantly (83 vs 68; HR 1.23; 95% CI 0.89 to 1.70).

The trials show that the subcutaneous ICD exchanges lead-related and intravascular complications for a higher, though falling, rate of inappropriate shocks, with equivalent protection against VF in patients who do not need pacing. The 2017 guideline recommends it in ICD candidates with inadequate venous access or high infection risk when pacing is not needed or anticipated (Class 1). It considers the device reasonable in any candidate without a pacing, ATP, or CRT need (Class 2a) and classes it as harmful when bradycardia pacing, ATP, or CRT is required (Class 3: Harm).

Limitations of the subcutaneous ICD and the modular pacing-defibrillator system

The subcutaneous ICD provides only brief post-shock pacing through the chest wall and no ATP. Patients who later develop atrioventricular block, symptomatic sinus node dysfunction, recurrent monomorphic VT requiring frequent shocks, or a CRT indication need an additional or replacement system. In EFFORTLESS, few systems were extracted for a new need for ATP, biventricular pacing, or bradycardia pacing (5, 4, and 1 patients, respectively). The generator is larger than a transvenous generator. The interval from onset to shock is longer because of the longer detection and charging times, and obesity and anterior generator placement raise the defibrillation threshold.

A modular system pairs a subcutaneous ICD with a leadless right ventricular pacemaker that receives commands by conducted communication. In the single-group study of 293 enrolled patients, 97.5% were free from leadless pacemaker-related major complications at 6 months, and wireless communication succeeded in 98.8% of tests. Pacing thresholds were 2.0 V or less in 97.4%, and ATP terminated 61.3% of treated episodes. The system addresses the absence of ATP and bradycardia pacing without transvenous leads; leadless pacing is covered on the leadless pacemaker section.

Extravascular (substernal) ICD

The extravascular ICD places a single lead in the anterior mediastinum, tunneled from a subxiphoid incision along the posterior surface of the sternum, with the generator in the left midaxillary region. Because the electrodes lie close to the right ventricle, the lead can sense ventricular activation with near-field quality, deliver ATP and pause-prevention pacing, and defibrillate at energies similar to transvenous systems, without entering the veins or the heart.

Two chest radiographs after extravascular ICD implantation. The frontal view shows a single substernal lead running vertically behind the lower sternum, with two shock coils labeled coil 1 and coil 2 and two ring electrodes labeled ring 1 and ring 2, and the generator in the left lateral chest wall. The lateral view shows the lead lying immediately behind the sternum, anterior to the heart, and the generator posterior and inferior.
Extravascular ICD on chest radiography. (A) Frontal view: the substernal lead with its two shock coils facing rightward and its two ring (sensing and pacing) electrodes facing leftward; the generator lies over the left lateral chest wall. (B) Lateral view: the lead runs behind the sternum, anterior to the heart, without entering the vascular system. Credit: Breitenstein A, Delaite JY, Dayal N, J Innov Card Rhythm Manag 2024;15:6062–6065, Figure 2, CC BY 4.0.

Pivotal study. In a prospective, nonrandomized premarket study, 356 patients with a Class 1 or 2a ICD indication were enrolled and 316 underwent an implantation attempt. Among 302 patients who completed defibrillation testing, defibrillation succeeded in 98.7%, and 299 of 316 (94.6%) were discharged with a working system. Freedom from major system- or procedure-related complications at 6 months was 92.6%, with 25 major complications in 23 patients (7.3%) and no major intraprocedural complications. ATP succeeded in 50.8% of treated episodes. Twenty-nine patients received 118 inappropriate shocks for 81 episodes, and 8 systems were explanted without replacement over a mean follow-up of 10.6 months.

Real-world registry. In the Enlighten post-approval registry of 787 patients (mean age 49 years, 34.1% secondary prevention) treated by 168 implanters, substernal tunneling and lead placement succeeded in 99.2%, and 96.3% remained implanted after testing. One-year freedom from major complications was 89.1%. Shocks terminated all 61 discrete spontaneous episodes, ATP terminated 74.2% of monomorphic VT, and the 1-year rate of first inappropriate shock was 7.1%, with noncardiac oversensing the leading cause.

The extravascular ICD suits patients who want to avoid transvenous leads and benefit from ATP or occasional pause prevention. Prior sternotomy or mediastinal surgery can obliterate the substernal space and complicate tunneling, and the system does not suit patients who need sustained bradycardia pacing or resynchronization. There are no randomized comparisons with transvenous or subcutaneous systems.

Wearable cardioverter-defibrillator: design and function

The wearable cardioverter-defibrillator (WCD) is an external device worn under clothing. The most widely used system consists of a chest garment with dry, nonadhesive sensing and defibrillation electrodes, connected to a monitor worn at the waist or on a shoulder strap. It records surface ECG signals and detects VT and VF by rate and morphology within programmable zones; the default thresholds are 150 bpm for VT and 200 bpm for VF. When an arrhythmia is detected, the device vibrates and sounds alarms. A conscious patient presses two response buttons to withhold the shock, which prevents shocks for tolerated rhythms and artifact. If the buttons are not pressed within the programmed response time, the electrodes release conductive gel, bystanders are warned by voice prompts, and the device shocks. Up to five shocks of 75 to 150 J (default 150 J) can be delivered per episode, and the sequence from detection to treatment usually takes less than 1 minute. Recorded ECGs and wear-time data are transmitted for remote review.

The WCD provides no pacing and no ATP. Protection exists only while the garment is worn, so wear time determines effectiveness. Additional systems, including one with adhesive patch electrodes, have been developed, but the clinical evidence comes almost entirely from the vest system.

Wearable defibrillator indications, evidence, and wear time

The WCD is used when the risk of sudden death is transiently elevated or when an indicated ICD cannot be implanted. The 2017 AHA/ACC/HRS guideline considers it reasonable after explantation of an ICD for infection in a patient with prior sudden cardiac arrest or sustained ventricular arrhythmia (Class 2a). It may be reasonable (Class 2b) when risk is transiently increased and an ICD is not yet appropriate, as in the settings in the table below and during systemic infection. European guidance places more weight on the secondary-prevention bridge: the 2022 European Society of Cardiology (ESC) ventricular arrhythmia guideline gives a Class 2a recommendation for adults with a secondary-prevention indication who are temporarily not candidates for an ICD, and Class 2b in selected patients early after MI.

SettingRationaleUS guideline class
ICD explanted for infection, prior cardiac arrest or sustained VTIndication persists while the infection is treated and before reimplantation2a
Within 40 days of MI, ejection fraction ≤35%High early sudden death risk; ICD not indicated (DINAMIT, IRIS); the Vest Prevention of Early Sudden Death Trial (VEST) did not reduce arrhythmic death2b
Within 90 days of revascularization, ejection fraction ≤35%Ejection fraction may recover2b
Newly diagnosed nonischemic cardiomyopathyEjection fraction often improves with guideline-directed therapy2b
Myocarditis or secondary cardiomyopathyTransient risk with possible recovery2b
Awaiting heart transplantationBridge for patients who do not otherwise receive an ICD2b

Wearable cardioverter-defibrillator indications, 2017 AHA/ACC/HRS guideline.

VEST: wearable cardioverter-defibrillator after myocardial infarction. VEST randomized 2,302 patients with acute MI and an ejection fraction of 35% or less, in a 2:1 ratio, to a WCD plus guideline-directed therapy or therapy alone. Arrhythmic death at 90 days, the primary outcome, was 1.6% with the device and 2.4% without (relative risk 0.67; 95% CI 0.37 to 1.21; P=0.18). Death from any cause was 3.1% and 4.9% (relative risk 0.64; uncorrected P=0.04). Median wear was 18.0 hours per day, and only 12 of the 48 patients in the device group who died were wearing it at the time. Twenty patients (1.3%) received an appropriate shock and 9 (0.6%) an inappropriate shock. Analyses restricted to patients who wore the device suggest benefit, but they are vulnerable to confounding by the characteristics of patients who adhere.

Registries. In the Prospective Registry of Patients Using the Wearable Cardioverter Defibrillator (WEARIT-II), 2,000 patients with ischemic or nonischemic cardiomyopathy or inherited disease wore the device for a median 90 days and 22.5 hours per day. Forty-one patients had 120 sustained ventricular tachyarrhythmias, and 54% of these patients received an appropriate shock. Ten patients (0.5%) received an inappropriate shock. Sustained ventricular tachyarrhythmia occurred in 3% of ischemic and inherited-disease patients and 1% of nonischemic patients by 3 months. At the end of use, 42% received an ICD; improvement in ejection fraction was the most common reason not to implant one.

Wear time and practical use. Patients are trained to remove the vest only to bathe, and wear time and alarms are reviewed remotely. Guideline-directed medical therapy is titrated during the wear period so that the ejection fraction is reassessed on optimal therapy, commonly after about 3 months. Acute myocarditis and peripartum cardiomyopathy are other transient conditions in which the device is used while recovery is awaited. Skin irritation, discomfort, sleep disturbance, and false alarms limit wear time.

How an ICD detects ventricular arrhythmia: sensing, zones, and duration

Sensing. Ventricular electrograms during VF vary widely in amplitude, so ICDs use automatic adjustment of sensitivity. After a sensed event, the sensing threshold starts at a proportion of the preceding electrogram amplitude and decays toward a programmed minimum, so that the T wave falls below the threshold while low-amplitude VF is still detected. Blanking periods after sensed and paced events prevent double counting of the same depolarization. Programming the minimum sensitivity too high risks undersensing VF; programming it too low, or a large T wave relative to the R wave, risks T-wave oversensing.

Rate zones. Detection is primarily by rate. Sensed intervals are compared with programmed cycle-length boundaries that define a VF zone, an optional fast VT zone, an optional VT zone, and an optional monitor zone that records but does not treat. A rate is converted to a cycle length by dividing 60,000 by the rate in beats per minute: 200 bpm is a cycle length of 300 ms, and 250 bpm is 240 ms. Multiple zones allow different detection durations, discriminators, and therapies at different rates, with the shortest delay and the most aggressive therapy at the fastest rates.

Duration. Detection is complete when a programmed number of intervals, or a programmed proportion of a sliding window (for example, 30 of 40 intervals), or a programmed time falls in the zone, depending on the manufacturer. Longer duration lets nonsustained VT end on its own, which accounts for a large share of the therapy avoided by modern programming. Before a shock is delivered, the device reconfirms the arrhythmia during or after charging so that a rhythm that ended during charging is not shocked; the shock is then synchronized to a sensed R wave when possible.

Redetection and episode termination. After therapy, the device redetects with shorter criteria if the arrhythmia persists and escalates to the next programmed therapy. An episode ends when a programmed number of consecutive intervals are slower than the lowest zone.

The 2015 HRS/EHRA/APHRS/SOLAECE consensus specifies, for primary-prevention patients, detection lasting at least 6 to 12 seconds or 30 intervals (Class 1) and a slowest therapy zone between 185 and 200 bpm (Class 1). For secondary-prevention patients, the same duration applies (Class 1), and when the clinical VT rate is known, it is reasonable to program the slowest therapy zone at least 10 bpm below the documented VT rate but not faster than 200 bpm (Class 2a). Higher minimum rates are appropriate in young patients and in those in whom discriminators perform poorly, provided there is no clinical VT below the cutoff.

ICD detection zone and timing calculator

The calculator converts rate to cycle length, determines whether a tachycardia falls within a zone, and estimates the time needed to complete detection when all intervals fall in the zone.

Tachycardia cycle length—
Zone cutoff cycle length—
Time to complete detection—
Converted rate—
—

Cycle length (ms) = 60,000 ÷ rate (bpm). Time to detection = intervals required × tachycardia cycle length. A tachycardia at 200 bpm (300 ms) detected over 30 intervals takes 9.0 seconds; at 250 bpm (240 ms), 30 intervals take 7.2 seconds. Charging adds several seconds before a shock.

Discriminating supraventricular tachycardia from VT

Sinus tachycardia, atrial fibrillation, atrial flutter, and other SVTs overlap with VT in rate, especially in young and active patients. Discriminators withhold therapy when a rhythm in the VT zone has the features of SVT. They are applied in the VT and fast VT zones; in the fastest zone, rate alone triggers therapy because withholding therapy for VF is unacceptable. Single-chamber devices rely on onset, stability, and electrogram morphology. Dual-chamber devices add the relationship between atrial and ventricular rates and timing.

DiscriminatorPrincipleClassified as VT whenFailure modes
MorphologyCompares the shock-channel electrogram during tachycardia with a template stored during conducted rhythmA programmed proportion of beats fail to match the templateRate-related bundle branch block, outdated template after a change in conduction, electrogram clipping, myopotential and noise contamination
Sudden onsetCompares the cycle length at onset with the preceding cycle lengthsAbrupt shortening of cycle lengthVT that starts during sinus tachycardia (gradual onset, therapy withheld) and SVT with abrupt onset (classified as VT)
StabilityMeasures variability of RR intervalsIntervals are regularAtrial fibrillation becomes more regular at fast ventricular rates; VT can be irregular early or with antiarrhythmic drugs
Atrial-ventricular rate comparisonCounts atrial and ventricular ratesVentricular rate exceeds atrial rateAtrial undersensing, atrial lead dislodgement, far-field R-wave oversensing on the atrial channel
Atrioventricular association and pattern analysisAnalyzes the timing and sequence of atrial and ventricular eventsDissociation or ventricular-led patternVT with 1:1 retrograde conduction; atrial flutter with fixed conduction

SVT-VT discriminators in transvenous ICDs. Manufacturer algorithms combine these elements in different sequences.

The 2015 consensus recommends that discriminators be programmed to include rhythms faster than 200 bpm and potentially up to 230 bpm (Class 1). It considers it reasonable to reacquire the morphology template when the match is unsatisfactory (Class 2a). Disabling the SVT discriminator time-out, which otherwise delivers therapy when a rhythm classified as SVT persists for a set time, may be reasonable (Class 2b). Discrimination algorithms are contraindicated in complete heart block and when known to be unreliable in the individual patient. Dual-chamber discriminators are contraindicated with known atrial lead dislodgement, atrial undersensing or far-field R-wave oversensing, and permanent atrial fibrillation. They are programmed only when the atrial lead is chronic or atrial sensing is reliable, as corrected in the consensus erratum.

Antitachycardia pacing and shock programming

ATP delivers a train of pacing stimuli faster than the tachycardia. The stimuli enter the excitable gap of a re-entrant circuit, and when a wavefront collides with the circuit's own wavefront and meets refractory tissue, the tachycardia stops. Burst ATP uses a fixed cycle length within a train; ramp ATP shortens the cycle length within a train. ATP terminates most monomorphic scar-related VT, rarely terminates polymorphic VT or VF, and can accelerate VT, which is why a shock is always available behind it.

In the Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial, 634 patients were randomized to empirical ATP or a shock as the first therapy for fast VT at 188 to 250 bpm. Fast VT accounted for 32% of ventricular tachyarrhythmias and 76% of the episodes that conventional programming would have detected as VF and shocked. ATP, given as 8 pulses at 88% of the tachycardia cycle length, terminated 81% of episodes (72% after adjustment), with similar rates of acceleration, syncope, and sudden death, and quality of life improved more in the ATP arm. Many devices also deliver one ATP sequence while the capacitor charges and abort the charge if ATP succeeds, so that fast VT in the VF zone is treated without delaying a needed shock.

RecommendationClass
In structural heart disease with an ATP-capable device, ATP active in all ventricular tachyarrhythmia zones for arrhythmias up to 230 bpm, except when ATP is documented to be ineffective or proarrhythmic1
At least 1 ATP attempt with a minimum of 8 stimuli at a cycle length of 84% to 88% of the tachycardia cycle length1
Burst ATP programmed in preference to ramp ATP1
Shock therapy available in all ventricular tachyarrhythmia zones (rarely omitted for tolerated slow VT)2a
Initial shock in the highest rate zone programmed to maximum energy unless testing documents efficacy at lower energy2a

Tachycardia therapy programming, 2015 HRS/EHRA/APHRS/SOLAECE consensus.

Shock polarity, the shock vector (inclusion of the superior vena cava coil), and the number of shocks per episode are programmable. Shocks for VT are synchronized; shocks for VF are delivered on the confirmation that VF persists after charging.

Evidence-based programming trials and the consensus they produced

Early ICDs were programmed with low rate cutoffs and short detection times, inherited from secondary-prevention practice. Randomized trials in primary-prevention populations then showed that higher cutoffs and longer detection reduce inappropriate and unnecessary therapy without more syncope or arrhythmic death.

TrialPopulation and armsMain results
MADIT-RIT (2012)1,500 primary-prevention patients. Conventional: 2.5-second delay at 170–199 bpm, 1.0-second delay at ≥200 bpm. High-rate: therapy only at ≥200 bpm with a 2.5-second delay. Delayed: 60 seconds at 170–199, 12 seconds at 200–249, 2.5 seconds at ≥250 bpmFirst inappropriate therapy: HR 0.21 (high-rate) and 0.24 (delayed). All-cause mortality: HR 0.45 (P=0.01) and 0.56 (P=0.06). No difference in procedure-related adverse events. Mean follow-up 1.4 years
ADVANCE III (2013)1,902 patients (75% primary prevention). Detection over 30 of 40 vs 18 of 24 intervalsTherapies 42 vs 67 per 100 patient-years (rate ratio 0.63). First inappropriate shock 5.1 vs 11.6 per 100 patient-years (rate ratio 0.55). No significant difference in mortality or arrhythmic syncope
PROVIDE (2014)1,670 primary-prevention patients. Higher detection rates, longer detection, empiric ATP, and optimized discriminators vs conventional parameters2-year shock rate 12.4% vs 19.4%. Time to first shock prolonged (HR 0.62). All-cause mortality HR 0.70 (P=0.036). No increase in arrhythmic syncope
PainFREE Rx II (2004)634 patients. ATP vs shock as first therapy for fast VT at 188–250 bpmATP terminated 81% of fast VT (72% adjusted); acceleration, syncope, and sudden death similar

Randomized programming trials. The 2015 consensus translated these results into recommendations for all manufacturers' devices.

All-cause mortality was lower with high-rate programming in MADIT-RIT (HR 0.45) and with PROVIDE programming (HR 0.70). The delayed-therapy arm of MADIT-RIT showed a smaller, nonsignificant difference (HR 0.56, P=0.06). These findings suggest that avoidable therapy is not benign. Shocks are painful and can injure myocardium, and both shocks and ATP can provoke arrhythmia. The detection duration and lowest therapy zone recommended in the 2015 consensus (Class 1) follow from these trials. The consensus also considers it useful to program more than one zone so that tiered therapy and discriminators can be used (Class 2a). Programming a monitor zone without therapy to record untreated arrhythmias may be reasonable (Class 2b). Lead-failure alerts are recommended (Class 1), and noise-rejection and T-wave oversensing algorithms may be reasonable (Class 2b). Patients with documented slow VT, those with a low-rate VT history after ablation, and those taking antiarrhythmic drugs that slow VT may need a lower zone, which is placed below the documented VT rate.

Appropriate shocks and what they signify

An appropriate shock is a shock delivered for VT or VF. In SCD-HeFT, 269 of 811 patients (33.2%) with an implanted ICD received at least one shock over a median 45.5 months: 128 received only appropriate shocks, 87 received only inappropriate shocks, and 54 received both. After adjustment, an appropriate shock was associated with a 5.68-fold higher risk of subsequent death and an inappropriate shock with a 1.98-fold higher risk. The risk remained elevated in patients who survived more than 24 hours after an appropriate shock (HR 2.99), and the most common mode of death after any shock was progressive heart failure.

An appropriate shock marks a sicker patient and a changing substrate; successful termination does not end the evaluation. A first appropriate shock leads to review of heart failure therapy, ischemia, electrolytes, thyroid function, drug adherence, and recently started drugs. It also leads to reprogramming of ATP when the VT was monomorphic and pace-terminable. Antiarrhythmic drugs or catheter ablation are considered for recurrent monomorphic VT, as covered on the catheter ablation of VT section and the antiarrhythmic drugs for ventricular arrhythmias section. Unnecessary therapy is therapy for true VT that would have ended spontaneously; it is avoided by longer detection; drugs are not the remedy.

Inappropriate shocks: atrial arrhythmias, oversensing, and lead failure

An inappropriate shock is a shock for any rhythm other than VT or VF. In MADIT-II, 83 of 719 ICD recipients (11.5%) received one or more inappropriate shocks, which made up 31.2% of all shock episodes. Atrial fibrillation triggered 44%, other SVT 36%, and abnormal sensing 20%. Smoking, prior atrial fibrillation, diastolic hypertension, and a previous appropriate shock predicted inappropriate shocks, and the stability discriminator was programmed less often in patients who received them (17% vs 36%). Patients with inappropriate shocks had higher subsequent mortality (HR 2.29). Contemporary programming has lowered these rates considerably.

CauseStored electrogram findingsCorrection
Atrial fibrillation with rapid conductionIrregular ventricular intervals; atrial rate faster than ventricular; ventricular morphology matches templateHigher cutoff, longer detection, discriminators, rate control, rhythm control or ablation of atrial fibrillation; anticoagulation as indicated
Sinus tachycardia, atrial flutter, or atrial tachycardiaGradual onset (sinus), 1:1 or fixed atrioventricular relationship, template matchHigher cutoff, discriminators, beta-blocker, ablation of atrial flutter or tachycardia
T-wave oversensingAlternating ventricular electrogram amplitudes; sensed events at the T wave on the near-field channel; double countingSensitivity or sensing vector reprogramming, T-wave algorithms, correction of hyperkalemia; S-ICD vector change and sensing filter
Lead conductor fracture or insulation breachVery short, nonphysiologic intervals; noise on the near-field channel with a normal far-field channel; abrupt impedance changeTherapy suspended or noise algorithms enabled while monitored; lead revision or replacement
Loose set screw or connector problemNoise early after implant or generator changePocket revision
Myopotential oversensingHigh-frequency signals with arm or trunk effort, mainly on integrated bipolar or subcutaneous channelsSensing vector change; sensitivity adjustment
Electromagnetic interferenceHigh-frequency signals on all channels, related to an external sourceRemoval of the source; avoidance counseling

Causes of inappropriate ICD shocks. Stored electrograms from the near-field (pace-sense) and far-field (shock) channels distinguish physiologic from nonphysiologic signals. Interrogation during storm is covered on the electrical storm page.

T-wave, P-wave, and myopotential oversensing

T-wave oversensing. When the sensed T wave exceeds the decaying sensing threshold, cardiac cycles are counted twice and the measured rate doubles, which can cross the VF cutoff during sinus rhythm. It occurs with low-amplitude R waves, large T waves, hyperkalemia, drug effects, ischemia, inherited channelopathies with large or long T waves, and changes in posture. In the subcutaneous ICD, which senses a surface-like signal, T-wave oversensing is the main cause of inappropriate shocks. Correction includes treating the metabolic cause, reprogramming sensitivity or the sensing vector, enabling a T-wave algorithm, and, in the subcutaneous ICD, selecting another vector or enabling the high-pass sensing filter. Reducing ventricular sensitivity to suppress T waves risks undersensing VF, so detection of induced or spontaneous VF is confirmed when the margin is narrow.

P-wave and far-field R-wave oversensing. In integrated bipolar leads placed near the tricuspid annulus, or in nonvascular systems with electrodes near the atria, atrial signals can be counted on the ventricular channel. Far-field R-wave sensing on the atrial channel misleads dual-chamber discriminators into classifying VT as SVT or SVT as VT.

Myopotential and diaphragmatic oversensing. Pectoral or diaphragmatic muscle activity can produce high-frequency signals, most often on integrated bipolar sensing or in subcutaneous systems, during arm movement, coughing, or straining.

Lead failure, noise, and advisory leads

Transvenous defibrillation leads are the weakest component of the system. Conductor fracture produces intermittent make-break signals, very short and nonphysiologic intervals on the near-field channel, and abrupt rises in pacing impedance; insulation breach produces falling impedance, oversensing, and, when the high-voltage conductor is involved, a short circuit that diverts a shock. Lead integrity algorithms combine counts of very short intervals, nonsustained high rates, and impedance trends and alert the clinic before a shock occurs. When the shock channel shows no corresponding signal, noise algorithms can withhold a shock.

Two lead families illustrate population-level failure. The Sprint Fidelis lead, recalled in 2007, failed mainly by pace-sense conductor fracture; its failure hazard rose with time, and fracture often presented as inappropriate shocks. The Riata family of silicone-insulated leads developed inside-out abrasion with conductors that became visible outside the lead body (externalization) on fluoroscopy, and externalized leads had more electrical dysfunction.

Management of a failing lead depends on whether the pacing or the high-voltage conductor is affected, pacing dependence, the age of the patient, and extraction risk. A new lead is added or the old one extracted; prophylactic replacement of a functioning advisory lead is weighed against extraction risk. Until revision, the device is programmed to minimize inappropriate therapy, for example, by enabling noise algorithms or by switching to integrated bipolar sensing after a ring-conductor fracture. The consensus considers these temporary measures. They are not long-term solutions.

Management after an ICD shock

The response to a shock depends on the number of shocks, symptoms, and the rhythm on the stored electrogram. Device interrogation, in person or by remote transmission, determines whether the shock was appropriate, inappropriate, or unnecessary before treatment changes are made.

SituationResponse
Single shock, patient feels well afterwardDevice clinic notified; remote transmission or prompt in-person interrogation; review of electrolytes, drugs, heart failure status, and adherence
Shock with chest pain, dyspnea, syncope, or persistent symptomsEmergency evaluation: ECG, electrolytes, troponin, and interrogation; treatment of ischemia or decompensated heart failure
Two or more shocks in 24 hours, or repeated shocksEmergency evaluation for electrical storm, inappropriate therapy, or lead failure; continuous monitoring
Inappropriate shocks for SVT or oversensing with ongoing risk of repeat shocksMagnet over the generator to suspend detection while monitored with an external defibrillator available, followed by reprogramming
Appropriate shock for monomorphic VTATP reprogramming, correction of triggers, beta-blocker optimization, antiarrhythmic drug, and evaluation for catheter ablation
Lead noise or abnormal impedanceTherapies modified while monitored; urgent lead evaluation and revision

Management after an ICD shock. Electrical storm, defined as 3 or more episodes of sustained ventricular arrhythmia in 24 hours, is managed on the electrical storm page.

Patients and families are instructed before discharge on what a shock feels like, when to call, and when to seek emergency care. External defibrillation and cardiopulmonary resuscitation proceed normally in a patient with an ICD, with pads placed away from the generator, and the device is interrogated afterward.

Magnet response, electromagnetic interference, and surgery

A clinical magnet placed over an ICD generator suspends tachyarrhythmia detection and therapy while it remains in place in most devices; when it is removed, detection resumes. Unlike a pacemaker, most ICDs do not switch to asynchronous pacing with a magnet, so a magnet does not protect a pacing-dependent patient from inhibition of pacing by electrocautery. In that situation the device is reprogrammed to an asynchronous pacing mode.

ManufacturerBradycardia pacing with magnetTachyarrhythmia function with magnet
MedtronicNo changeDetection and therapy suspended
AbbottNo changeDetection and therapy suspended (programmable)
Boston ScientificNo changeTherapy inhibited at nominal settings; the magnet response can be programmed off or to store an electrogram
BiotronikNo changeDetection and therapy suspended; therapy reactivates after 8 hours of continuous magnet application
MicroPort (formerly Sorin)Pacing rate rises to 96 bpm (80 bpm near replacement); pacing mode unchangedDetection and therapy suspended

Transvenous ICD magnet responses by manufacturer, as summarized in a 2024 perioperative review. The programmed magnet response is confirmed by interrogation before a procedure.

Surgery and procedures. Monopolar electrocautery is the main intraoperative source of electromagnetic interference; the device can count it as VF and deliver a shock, or, in a pacing-dependent patient, count it as intrinsic activity and withhold pacing. Bipolar cautery is used when possible, and monopolar cautery is applied in short bursts of less than 5 seconds with the dispersive electrode placed so that the current path does not cross the generator or leads. Reprogramming is generally unnecessary when no interference is expected, when only bipolar cautery is used at least 15 cm from the device, or when the procedure is below the umbilicus. For surgery above the umbilicus with monopolar cautery, tachyarrhythmia therapy is suspended by magnet or programming, with continuous ECG monitoring and external defibrillation pads available, and pacing-dependent patients are reprogrammed to asynchronous pacing. Therapies are restored and the device interrogated after the procedure. Radiofrequency ablation, external cardioversion, lithotripsy, transcutaneous electrical nerve stimulation, and radiation therapy are managed by the device team in advance.

Occupational sources. Arc welding, industrial induction equipment, and strong magnets held over the generator can cause oversensing or suspend therapy; patients whose work involves such equipment are assessed individually.

Magnetic resonance imaging in patients with an ICD

Most contemporary transvenous, subcutaneous, and extravascular ICD systems are labeled magnetic resonance (MR)-conditional, meaning that magnetic resonance imaging (MRI) is permitted under conditions specified for the complete system of generator and leads. Before scanning, tachyarrhythmia therapy is programmed off, pacing-dependent patients are programmed to an asynchronous mode, continuous ECG and oxygen saturation monitoring is used, an external defibrillator is available, and the device is reprogrammed and interrogated after the scan.

For devices that are not MR-conditional, the MagnaSafe registry performed 500 MRI examinations in patients with an ICD and 1,000 in patients with a pacemaker, all nonthoracic at 1.5 tesla (T), under a standardized protocol. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred. One ICD generator could not be interrogated after MRI and required replacement; it had not been programmed according to the protocol. Six partial electrical resets and six episodes of self-terminating atrial fibrillation or flutter occurred. Abandoned leads, epicardial leads, and lead fragments are higher-risk situations decided case by case.

Complications of transvenous and nonvascular systems

Procedural complications appear within days to weeks, and lead and generator complications accumulate over years.

ComplicationTiming and featuresRepresentative data
Pocket hematomaDays; swelling, pain; raises infection risk3.5% with continued warfarin vs 16.0% with heparin bridging (BRUISE CONTROL)
Pneumothorax and hemothoraxHours; subclavian punctureReduced by extrathoracic axillary or cephalic venous access
Lead dislodgementDays to weeks; loss of capture or sensing, change in impedanceHigher for atrial and coronary sinus leads
Cardiac perforation and tamponadeAcute or delayed; pericardial pain, rising threshold, diaphragmatic pacing, effusionRecognized with active-fixation leads; managed as on the pericardial effusion and tamponade page
Device infectionWeeks to years; pocket infection or lead-associated endocarditisMajor infection 1.2% at 12 months after high-risk procedures without an envelope (WRAP-IT); 4.9% of ICD recipients over a median 67.6 months in DANISH
Lead failureYears; noise, inappropriate shocks, loss of captureAdvisory leads (Sprint Fidelis, Riata) fail at rates that rise with time
Venous occlusionMonths to years; usually silent, sometimes arm swelling or superior vena cava syndromeComplicates lead addition or upgrade
Lead-related tricuspid regurgitationMonths; impingement or adhesion of a leafletManaged by lead repositioning, extraction, or valve intervention
Subcutaneous ICD complicationsInfection, erosion, lead or generator migration, inappropriate shocksComplications 31 patients vs 44 with transvenous ICD in PRAETORIAN (HR 0.69)
Extravascular ICD complicationsLead dislodgement, infection, substernal tunneling complications, inappropriate shocksMajor complications 7.3% at 6 months (pivotal study); 1-year freedom from major complications 89.1% (Enlighten)

Complications of ICD systems. The DANISH control group included patients with CRT pacemakers, in whom device infection occurred in 3.6%.

Twiddler syndrome, in which the patient rotates the generator in its pocket until the leads coil and retract, presents with loss of capture, noise, or inappropriate shocks, and the chest radiograph shows twisted leads. Generator erosion through thin skin is treated as an infected system. Phrenic nerve or intercostal stimulation suggests lead perforation or, with a coronary sinus lead, proximity to the phrenic nerve.

Device infection and lead extraction

Cardiac implantable electronic device infection ranges from superficial incisional infection to pocket infection, lead-associated endocarditis, and occult bacteremia. Staphylococci cause most infections. Pocket erythema, warmth, drainage, erosion, or pain prompts blood cultures and transesophageal echocardiography, as does any Staphylococcus aureus bacteremia in a patient with a device. Positron emission tomography or radiolabeled leukocyte imaging is added when the diagnosis remains uncertain.

Established device infection, including isolated pocket infection, is treated with complete removal of the generator and all leads in addition to antibiotics; antibiotic therapy alone is followed by relapse. Transvenous lead extraction is performed in centers with surgical backup. The 2026 expert consensus statement update on lead management and extraction addresses extraction facilities, technique, prevention of infection, and management of leads that cross a transcatheter tricuspid valve. Before reimplantation, the continued need for a device is reassessed, a new system is placed at a different site after bloodstream infection has cleared, and nonvascular systems are favored when pacing is not needed. A WCD is reasonable to bridge patients with prior cardiac arrest or sustained ventricular arrhythmia while the infection is treated (Class 2a). Diagnosis and antibiotic therapy are covered on the device infection and lead extraction section of the infective endocarditis page.

Follow-up and remote monitoring

After implantation, the wound and device function are checked in the first weeks, and follow-up continues by a combination of remote monitoring and in-person evaluation for the life of the system. Evaluations review battery voltage and charge time, pacing thresholds, sensed amplitudes, pacing and shock impedances, arrhythmia episodes with stored electrograms, therapies delivered and withheld, percentage of pacing (and of biventricular pacing in CRT-D), atrial arrhythmia burden, and heart failure diagnostics where available. Programming is reassessed when the clinical picture changes, for example, after new atrial fibrillation, new bundle branch block, or a change in antiarrhythmic drugs.

Remote monitoring transmits scheduled and alert-triggered data from a home communicator or smartphone application. The 2015 HRS expert consensus statement on remote interrogation and monitoring recommends remote monitoring combined with at least annual in-person evaluation over a calendar-based schedule of in-person visits alone (Class 1). The 2023 expert consensus statement on the remote device clinic states that starting remote monitoring before discharge or within 2 weeks of implantation can be beneficial (Class 2a). It also states that education about remote monitoring begins before implantation as part of shared decision-making on device selection (Class 1). In a pooled patient-level analysis of three randomized trials of daily remote monitoring (IN-TIME, ECOST, TRUST), the absolute risk of death at 1 year was 1.9 percentage points lower with monitoring (risk ratio 0.62). The composite of death or hospitalization for worsening heart failure was 5.6 percentage points lower (risk ratio 0.64). Remote alerts for lead integrity, therapies, atrial fibrillation, and battery status allow earlier action than scheduled visits.

Generator replacement and reassessment of the indication

Generator replacement at battery depletion is an elective procedure with higher infection risk than the initial implant, and it is a point at which the indication is reassessed. The 2017 guideline recommends that patients considering generator replacement be informed of their individual risk of sudden death and nonsudden death and of the efficacy, safety, and risks of the device (Class 1). Considerations include whether the ejection fraction has recovered, whether the patient has received appropriate therapy, the competing risk from heart failure and other illness, and the patient's goals. A patient with a recovered ejection fraction who has never had ventricular arrhythmia faces a different balance from one who has received appropriate shocks.

Replacement is also the point at which lead performance is reviewed, an upgrade to CRT is considered when right ventricular pacing burden or heart failure has increased, an antibiotic envelope is added in high-risk patients, and the patient's understanding of deactivation options is confirmed. Defibrillation testing at generator change is reasonable (Class 2a) but is often omitted when lead function is stable.

Shared decision-making for ICD therapy

Shared decision-making is a structured conversation in which the clinician presents the options with their probable benefits and harms and the uncertainty around them, and the patient's values and preferences determine the choice. For the ICD, the decision is unusual because the device prevents one mode of death without improving symptoms and carries device-specific harms. It commits the patient to decades of follow-up and replacement and changes how the end of life unfolds.

Guideline and coverage requirements. The 2017 AHA/ACC/HRS guideline recommends shared decision-making for patients with ventricular arrhythmias or at increased risk of sudden cardiac death (Class 1). The same disclosure of sudden and nonsudden death risk and of device efficacy, safety, and risks applies to patients considering a new ICD (Class 1). The Medicare national coverage determination for ICDs, issued in 2018, requires a shared decision-making interaction using an evidence-based decision tool before initial implantation for primary prevention in patients with prior MI and reduced ejection fraction, ischemic or nonischemic dilated cardiomyopathy, or familial or genetic disorders with high arrhythmic risk. The interaction is held with a physician, physician assistant, nurse practitioner, or clinical nurse specialist and may occur at a separate visit. Patients with prior sustained VT or cardiac arrest from VF and patients receiving a replacement device are exempt.

ElementContent presented to the patient
Purpose of the devicePrevents sudden death from VT or VF; does not relieve heart failure symptoms or treat the underlying disease (unless combined with CRT)
Expected benefit in absolute termsFor example, mortality over about 20 months was 14.2% with an ICD and 19.8% without in MADIT-II; in nonischemic cardiomyopathy on contemporary therapy, all-cause death did not differ in DANISH while sudden death was halved
Individual modifiers of benefitAge, competing comorbidity, ejection fraction trajectory on therapy, scar burden, genotype, functional class, and expected survival
HarmsProcedural complications, infection, lead failure, inappropriate shocks (11.5% of MADIT-II recipients), anxiety after shocks, and the need for future generator replacement
AlternativesNo device with continued medical therapy; WCD during a waiting period; CRT pacemaker in place of CRT-D in selected patients; nonvascular ICD systems
Device choiceTransvenous, subcutaneous, or extravascular, with their trade-offs in pacing, ATP, lead risk, and inappropriate shocks
Living with the deviceDriving restrictions, occupational exposures, sports, remote monitoring, MRI conditions
DeactivationShocks can be turned off at any time consistent with the patient's goals, without surgery, while pacing continues

Elements of shared decision-making for a primary-prevention ICD, drawn from the 2017 AHA/ACC/HRS guideline, the 2018 Medicare coverage requirement, and the trials cited on this page.

Decision aids. Patient decision aids for primary-prevention ICDs present absolute risks in numerical and graphic form and help patients state what matters to them, such as avoiding sudden death, avoiding shocks or procedures, or living longer with progressive heart failure. Delivery of decision aids is uneven in practice, and the aids work best when they reach patients before the decision is made and present data that fit contemporary patients.

Decisions in specific groups. In older patients and those with significant comorbidity, the conversation gives particular weight to competing causes of death. In nonischemic cardiomyopathy, the DANISH result, including how benefit varies with age, is reviewed with the patient. A patient who declines an ICD after an informed discussion receives the same heart failure and arrhythmia care, and the decision is revisited if risk or preferences change.

Quality of life, anxiety, depression, and psychological care

Most ICD recipients adapt well, but psychological distress is common. In a systematic review of 45 studies including more than 5,000 patients, structured diagnostic interviews identified a depressive disorder in 11% to 28% and an anxiety disorder in 11% to 26% of ICD recipients; self-report rates ranged more widely. A working estimate of about 20% for both disorders is similar to other cardiac populations. Shocks, especially multiple shocks or storm, are associated with anxiety, avoidance behavior, and post-traumatic stress symptoms. A phantom shock is a shock the patient reports feeling when interrogation shows no therapy was delivered; it is identified by interrogation and is managed as a psychological symptom without reprogramming.

Psychological care includes preparation before implantation, explanation of what a shock means and what to do afterward, screening for depression and anxiety at follow-up, programming that minimizes avoidable shocks, cardiac rehabilitation, and referral for cognitive behavioral therapy or psychiatric care when needed. Patients with recurrent shocks receive psychological care with the same priority as arrhythmia control, as described on the psychological consequences of repeated shocks section.

Driving, occupation, sports, and daily activities

Driving. The AHA/HRS statement on driving restricts private driving for 6 months after an ICD implanted for secondary prevention, and a patient who receives appropriate therapy for VT or VF, especially with symptoms of cerebral hypoperfusion, becomes subject to the same 6-month restriction. After a primary-prevention ICD, driving is avoided for 1 week while the incision heals, and asymptomatic patients then have no restriction. For commercial drivers, the 2007 AHA/HRS addendum advises against driving with any ICD. The 2024 Federal Motor Carrier Safety Administration advisory criteria, which are guidance and not regulation, describe ICDs as likely to cause syncope or collapse, and medical examiners generally do not certify drivers with an ICD. Local laws, which vary, also apply. European guidance differs in its specific intervals.

Occupation and daily life. Arm movement on the implant side is limited for a few weeks after a transvenous implant to reduce lead dislodgement. Occupations with high-voltage, welding, or induction equipment, or in which sudden incapacitation endangers others, are assessed individually. Air travel is permitted, and the patient carries device identification for security screening.

Sports. Sports participation with an ICD is an individualized decision, as covered on the sports participation and sudden death in athletes section.

Deactivation of shocks at the end of life

As heart failure or another illness progresses, shocks can prolong dying and cause distress without adding meaningful life. Discussions have historically come late or not at all. Some patients receive shocks in their final minutes because deactivation was never discussed.

The 2017 AHA/ACC/HRS guideline recommends that patients be told at implantation, at generator replacement, and during advance care planning that shocks can be deactivated at any time consistent with their goals (Class 1). It also recommends that clinicians discuss deactivation with patients who have refractory heart failure or refractory ventricular arrhythmia or who are approaching death from another illness (Class 1). The HRS expert consensus on device management near the end of life holds that a patient, or the surrogate of a patient without decision-making capacity, may request withdrawal of a device therapy. Withdrawal is ethically and legally equivalent to withdrawal of any other life-sustaining treatment, and it is neither physician-assisted suicide nor euthanasia. A clinician who objects on conscientious grounds is not compelled to deactivate the device, does not abandon the patient, and involves a colleague who will carry out the deactivation.

Process. Deactivation of tachyarrhythmia therapy is performed with a programmer by the device team, in the hospital, in hospice, or at home, after the decision is documented; the record names the therapies turned off. Patients and families are told that turning off shocks does not stop the heart, cause pain, or hasten death; it removes a therapy that no longer serves the patient's goals. Bradycardia pacing is usually left active; turning off pacing in a pacing-dependent patient is a separate decision made on the same principles. In urgent situations, such as repeated shocks in a dying patient, a magnet taped over the generator suspends therapy in most devices until programming is available. The device is reactivated if the patient's goals change. Hospice and palliative care teams are informed of the device and its status. Goals of care during storm are discussed on the goals of care and ICD deactivation section.

ICDs in older adults, women, pregnancy, kidney failure, congenital heart disease, and mechanical support

Older adults. The trials enrolled few patients older than 75 years, and the share of deaths that are sudden falls with age and comorbidity. The 2017 guideline considers an ICD reasonable in older patients who meet a primary-prevention indication when meaningful survival is expected (Class 2a), and the decision weighs frailty, cognition, kidney function, and goals.

Women. Women were underrepresented in the landmark trials and make up a minority of contemporary cohorts (25.6% of UNTOUCHED and 28.7% of the Enlighten registry). Device selection and programming follow the same principles; breast tissue and implant position are considered for subcutaneous and extravascular systems.

Pregnancy. An ICD does not preclude pregnancy, and device follow-up, imaging, and delivery are planned with the device team. Arrhythmia management during pregnancy is covered on the ventricular arrhythmias in pregnancy section.

Kidney failure and dialysis. Patients on hemodialysis have higher infection risk and venous access that must be preserved for fistulas and catheters, which favors subcutaneous or extravascular systems when pacing is not needed. Competing mortality is high in dialysis patients, and primary-prevention decisions are individualized.

Adult congenital heart disease. Venous anomalies, intracardiac shunts that allow paradoxical embolism from transvenous leads, prior surgery, and absent venous access to the ventricle make nonvascular or epicardial systems common choices. Subcutaneous ICD screening is affected by abnormal QRS and T-wave morphology.

Left ventricular assist devices and transplantation. An ICD is reasonable in patients awaiting transplantation who are discharged home and do not otherwise qualify (Class 2a), and in patients with a left ventricular assist device and sustained ventricular arrhythmia (Class 2a). These are covered on the ICD for primary prevention section and the advanced heart failure section.

Cost-effectiveness of the ICD

A Markov model of the primary-prevention trial populations found that the ICD increased lifetime costs in all trials. In the two trials without a mortality benefit, the Coronary Artery Bypass Graft (CABG) Patch trial and DINAMIT, the ICD was more costly and less effective than control therapy. For the six trials that showed benefit (MADIT, MADIT-II, MUSTT, DEFINITE, COMPANION, and SCD-HeFT), the ICD added 1.01 to 2.99 quality-adjusted life-years (QALYs) at an added cost of $68,300 to $101,500, giving $34,000 to $70,200 per QALY gained. The ratio remained below $100,000 per QALY as long as the ICD reduced mortality for 7 years or more. Cost-effectiveness is therefore greatest where the absolute reduction in sudden death is largest and competing mortality is lowest, and it falls with frequent generator replacement, complications, and older age. Programming that avoids shocks and remote monitoring that reduces visits and hospitalizations improve the value of the therapy.

Sources of error in ICD selection, programming, and follow-up

References

  1. Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2018;138:e272–e391. doi:10.1161/CIR.0000000000000549
  2. Heidenreich PA, Bozkurt B, Aguilar D, et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022;145:e895–e1032. doi:10.1161/CIR.0000000000001063
  3. Zeppenfeld K, Tfelt-Hansen J, de Riva M, et al. 2022 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur Heart J. 2022;43:3997–4126. doi:10.1093/eurheartj/ehac262
  4. Wilkoff BL, Fauchier L, Stiles MK, et al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Heart Rhythm. 2016;13:e50–e86. doi:10.1016/j.hrthm.2015.11.018 Also published in J Arrhythm. 2016;32:1–28 (erratum in J Arrhythm. 2016;32:441–442, correcting the statement on dual-chamber discrimination)
  5. Ferrick AM, Raj SR, Deneke T, et al. 2023 HRS/EHRA/APHRS/LAHRS expert consensus statement on practical management of the remote device clinic. J Arrhythm. 2023;39:250–302. doi:10.1002/joa3.12851
  6. Cha YM, El-Chami MF, Liu CF, et al. 2026 HRS/AHA/APHRS/EHRA/IDSA/LAHRS/PACES/STS expert consensus statement update on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2026;23:e1716–e1792. doi:10.1016/j.hrthm.2026.04.015
  7. Epstein AE, Baessler CA, Curtis AB, et al. Addendum to “Personal and public safety issues related to arrhythmias that may affect consciousness: implications for regulation and physician recommendations.” Circulation. 2007;115:1170–1176. doi:10.1161/CIRCULATIONAHA.106.180203
  8. Centers for Medicare & Medicaid Services. Decision memo for implantable cardioverter defibrillators (CAG-00157R4), national coverage determination 20.4. February 15, 2018. cms.gov
  9. Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Eur Heart J. 2000;21:2071–2078. doi:10.1053/euhj.2000.2476
  10. Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–883. doi:10.1056/NEJMoa013474
  11. Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure (SCD-HeFT). N Engl J Med. 2005;352:225–237. doi:10.1056/NEJMoa043399
  12. Køber L, Thune JJ, Nielsen JC, et al. Defibrillator implantation in patients with nonischemic systolic heart failure. N Engl J Med. 2016;375:1221–1230. doi:10.1056/NEJMoa1608029
  13. Hohnloser SH, Kuck KH, Dorian P, et al. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction (DINAMIT). N Engl J Med. 2004;351:2481–2488. doi:10.1056/NEJMoa041489
  14. Weiss R, Knight BP, Gold MR, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–953. doi:10.1161/CIRCULATIONAHA.113.003042
  15. Boersma L, Barr C, Knops R, et al. Implant and midterm outcomes of the subcutaneous implantable cardioverter-defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol. 2017;70:830–841. doi:10.1016/j.jacc.2017.06.040
  16. Gold MR, Lambiase PD, El-Chami MF, et al. Primary results from the Understanding Outcomes With the S-ICD in Primary Prevention Patients With Low Ejection Fraction (UNTOUCHED) trial. Circulation. 2021;143:7–17. doi:10.1161/CIRCULATIONAHA.120.048728
  17. Knops RE, Olde Nordkamp LRA, Delnoy PHM, et al. Subcutaneous or transvenous defibrillator therapy (PRAETORIAN). N Engl J Med. 2020;383:526–536. doi:10.1056/NEJMoa1915932
  18. Healey JS, Krahn AD, Bashir J, et al. Perioperative safety and early patient and device outcomes among subcutaneous versus transvenous implantable cardioverter defibrillator implantations: a randomized, multicenter trial (ATLAS). Ann Intern Med. 2022;175:1658–1665. doi:10.7326/M22-1566
  19. Knops RE, Lloyd MS, Roberts PR, et al. A modular communicative leadless pacing-defibrillator system. N Engl J Med. 2024;391:1402–1412. doi:10.1056/NEJMoa2401807
  20. Friedman P, Murgatroyd F, Boersma LVA, et al. Efficacy and safety of an extravascular implantable cardioverter-defibrillator. N Engl J Med. 2022;387:1292–1302. doi:10.1056/NEJMoa2206485
  21. Boersma LVA, Amin A, Murgatroyd F, et al. Real-world experience with the extravascular implantable cardioverter defibrillator through 1 year: the Enlighten post approval registry study. Circulation. 2026;154:1925–1938. doi:10.1161/CIRCULATIONAHA.126.080942
  22. Olgin JE, Pletcher MJ, Vittinghoff E, et al. Wearable cardioverter-defibrillator after myocardial infarction (VEST). N Engl J Med. 2018;379:1205–1215. doi:10.1056/NEJMoa1800781
  23. Kutyifa V, Moss AJ, Klein H, et al. Use of the wearable cardioverter defibrillator in high-risk cardiac patients: data from the Prospective Registry of Patients Using the Wearable Cardioverter Defibrillator (WEARIT-II Registry). Circulation. 2015;132:1613–1619. doi:10.1161/CIRCULATIONAHA.115.015677
  24. Duncker D, Garcia R, Bonderman D, et al. A review on wearable cardioverter-defibrillators in clinical practice for sudden cardiac death prevention and heart failure management. Europace. 2025;27:euaf257. doi:10.1093/europace/euaf257
  25. Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming (MADIT-RIT). N Engl J Med. 2012;367:2275–2283. doi:10.1056/NEJMoa1211107
  26. Gasparini M, Proclemer A, Klersy C, et al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA. 2013;309:1903–1911. doi:10.1001/jama.2013.4598 (erratum in JAMA. 2013;309:2552)
  27. Saeed M, Hanna I, Robotis D, et al. Programming implantable cardioverter-defibrillators in patients with primary prevention indication to prolong time to first shock: results from the PROVIDE study. J Cardiovasc Electrophysiol. 2014;25:52–59. doi:10.1111/jce.12273
  28. Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation. 2004;110:2591–2596. doi:10.1161/01.CIR.0000145610.64014.E4
  29. Healey JS, Hohnloser SH, Glikson M, et al. Cardioverter defibrillator implantation without induction of ventricular fibrillation: a single-blind, non-inferiority, randomised controlled trial (SIMPLE). Lancet. 2015;385:785–791. doi:10.1016/S0140-6736(14)61903-6
  30. Bänsch D, Bonnemeier H, Brandt J, et al. Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial. Eur Heart J. 2015;36:2500–2507. doi:10.1093/eurheartj/ehv292
  31. Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) trial. JAMA. 2002;288:3115–3123. doi:10.1001/jama.288.24.3115
  32. Peterson PN, Varosy PD, Heidenreich PA, et al. Association of single- vs dual-chamber ICDs with mortality, readmissions, and complications among patients receiving an ICD for primary prevention. JAMA. 2013;309:2025–2034. doi:10.1001/jama.2013.4982
  33. Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–1017. doi:10.1056/NEJMoa071098
  34. Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–1365. doi:10.1016/j.jacc.2007.09.073
  35. Birnie DH, Healey JS, Wells GA, et al. Pacemaker or defibrillator surgery without interruption of anticoagulation (BRUISE CONTROL). N Engl J Med. 2013;368:2084–2093. doi:10.1056/NEJMoa1302946
  36. Birnie DH, Healey JS, Wells GA, et al. Continued vs. interrupted direct oral anticoagulants at the time of device surgery, in patients with moderate to high risk of arterial thrombo-embolic events (BRUISE CONTROL-2). Eur Heart J. 2018;39:3973–3979. doi:10.1093/eurheartj/ehy413
  37. Tarakji KG, Mittal S, Kennergren C, et al. Antibacterial envelope to prevent cardiac implantable device infection (WRAP-IT). N Engl J Med. 2019;380:1895–1905. doi:10.1056/NEJMoa1901111
  38. Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator (MagnaSafe). N Engl J Med. 2017;376:755–764. doi:10.1056/NEJMoa1603265
  39. Hindricks G, Varma N, Kacet S, et al. Daily remote monitoring of implantable cardioverter-defibrillators: insights from the pooled patient-level data from three randomized controlled trials (IN-TIME, ECOST, TRUST). Eur Heart J. 2017;38:1749–1755. doi:10.1093/eurheartj/ehx015
  40. Magyar-Russell G, Thombs BD, Cai JX, et al. The prevalence of anxiety and depression in adults with implantable cardioverter defibrillators: a systematic review. J Psychosom Res. 2011;71:223–231. doi:10.1016/j.jpsychores.2011.02.014
  41. Sanders GD, Hlatky MA, Owens DK. Cost-effectiveness of implantable cardioverter-defibrillators. N Engl J Med. 2005;353:1471–1480. doi:10.1056/NEJMsa051989
  42. Kim M, Kwon CH. Perioperative management of patients with cardiac implantable electronic devices. Korean J Anesthesiol. 2024;77:306–315. doi:10.4097/kja.23826
  43. Schiavone M, Gasperetti A, Tilz R, et al. Integration of PRAETORIAN score, defibrillation testing and follow-up outcomes for the assessment of conversion failure in patients with subcutaneous implantable cardioverter defibrillator: results from the iSUSI registry. Heart Rhythm. 2026 (published online June 26, 2026). doi:10.1016/j.hrthm.2026.06.024
  44. Lampert R, Hayes DL, Annas GJ, et al. HRS expert consensus statement on the management of cardiovascular implantable electronic devices (CIEDs) in patients nearing end of life or requesting withdrawal of therapy. Heart Rhythm. 2010;7:1008–1026. doi:10.1016/j.hrthm.2010.04.033
  45. Krahn AD, Longtin Y, Philippon F, et al. Prevention of Arrhythmia Device Infection Trial: the PADIT trial. J Am Coll Cardiol. 2018;72:3098–3109. doi:10.1016/j.jacc.2018.09.068
  46. Slotwiner D, Varma N, Akar JG, et al. HRS expert consensus statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Heart Rhythm. 2015;12:e69–e100. doi:10.1016/j.hrthm.2015.05.008
  47. Federal Motor Carrier Safety Administration. Medical advisory criteria for evaluation under 49 CFR part 391, subpart E. 49 CFR part 391, appendix A. Fed Regist. 2024;89:3577. govinfo.gov/content/pkg/FR-2024-01-19/html/2024-00980.htm