Percutaneous coronary intervention: equipment, stents and technique
Percutaneous coronary intervention in adults from indications and access through guide catheters, guidewires, balloons, stent and drug-coated balloon evidence, procedural troubleshooting, antithrombotic therapy, complications, and same-day discharge.
Percutaneous coronary intervention (PCI) delivers a balloon, stent, or drug-coated balloon to a coronary stenosis over a coronary guidewire through a guide catheter seated in the coronary ostium. The 2021 ACC/AHA/SCAI revascularization guideline recommends radial access in acute coronary syndromes and stable ischemic heart disease, a contemporary drug-eluting stent over a bare-metal stent for all patients, and unfractionated heparin as the procedural anticoagulant unless it is contraindicated. Durable results depend on lesion preparation, correct sizing, full stent expansion, and recognition of mechanical problems before the procedure ends.
Definition and scope of percutaneous coronary intervention
Percutaneous coronary intervention is the catheter-based treatment of coronary obstruction. The term covers plain balloon angioplasty, implantation of bare-metal and drug-eluting stents, drug-coated balloon angioplasty, and the adjunctive devices that prepare a lesion, protect the distal bed, or support the circulation. All of these share one mechanical system: an arterial sheath, a guide catheter that engages the coronary ostium and transmits force from the operator's hands, a guidewire that crosses the lesion and serves as a rail, and the therapeutic devices that travel over the wire.
This page covers the equipment and technique of the procedure; the evidence behind stent and drug-coated balloon selection; troubleshooting of guide engagement, wire crossing, and device delivery; procedural antithrombotic therapy; complications; and discharge. The decision to revascularize is summarized here, with full coverage on the hubs for chronic coronary disease, ST-elevation myocardial infarction, and non-ST-elevation acute coronary syndrome. Diagnostic catheterization, angiographic grading, and invasive physiology belong to the coronary angiography and fractional flow reserve hub. Calcium modification, intravascular imaging guidance of stent implantation, and radiation safety in the catheterization laboratory are summarized in their own sections, with full coverage on the hubs for intravascular imaging, radiation safety in the catheterization laboratory, and complex PCI. Bifurcation, left main, and chronic total occlusion techniques appear here where they affect equipment choice.
Development of coronary angioplasty and stenting
Coronary balloon angioplasty was first reported in 1978 as transluminal dilatation of a coronary stenosis with a balloon catheter (Gruntzig 1978). Balloon angioplasty treated the fixed stenosis but left two failure modes: abrupt closure from elastic recoil and dissection in the first hours, and restenosis from negative remodeling and neointimal growth over the following months.
The first coronary stents were self-expanding stainless-steel mesh tubes. In the 1987 report, 24 coronary stents were implanted in 19 patients with restenosis, abrupt closure after angioplasty, or deteriorating bypass grafts. One stent occluded silently, and a second acute thrombosis was treated with thrombolysis (Sigwart 1987). Balloon-expandable stents then entered randomized comparison with angioplasty. In the Benestent trial, a primary clinical end point occurred in 20% of the stent group and 30% of the angioplasty group, and angiographic restenosis was 22% against 32%. Peripheral vascular complications requiring surgery, transfusion, or both were more frequent after stenting (13.5% against 3.1%), and hospital stay was longer (8.5 against 3.1 days) (Serruys 1994). In STRESS, stenting improved procedural success (96.1% against 89.6%) and reduced 6-month restenosis (31.6% against 42.1%) (Fischman 1994).
Bare-metal stents solved recoil and dissection but provoked neointimal hyperplasia, the smooth-muscle proliferation that produces in-stent restenosis. Drug-eluting stents added an antiproliferative drug carried on a polymer. The sirolimus-eluting stent reduced target-vessel failure from 21.0% to 8.6% in SIRIUS (Moses 2003), and the paclitaxel-eluting stent reduced 9-month ischemia-driven target-vessel revascularization from 12.0% to 4.7% in TAXUS IV (Stone 2004). These first-generation devices had thick struts and durable polymers. They were followed by thinner cobalt-chromium and platinum-chromium platforms eluting everolimus or zotarolimus, then by biodegradable-polymer and polymer-free designs, a bioresorbable scaffold that failed, and drug-coated balloons that leave no implant.
Indications for PCI in acute coronary syndromes and chronic coronary disease
Acute coronary syndromes. Primary PCI is the reperfusion strategy for ST-elevation myocardial infarction when it can be delivered within guideline time targets, and an invasive strategy with PCI of the culprit lesion is the default for most patients with non-ST-elevation acute coronary syndrome. In hemodynamically stable patients with ST-elevation myocardial infarction and multivessel disease, staged PCI of a significant nonculprit artery carries a Class 1 recommendation in the 2021 guideline. Routine PCI of nonculprit arteries at the time of primary PCI in cardiogenic shock is Class 3 (harm) because of the higher risk of death or renal failure (Lawton 2022). In CULPRIT-SHOCK, death or renal-replacement therapy at 30 days occurred in 45.9% of patients with culprit-lesion-only PCI and in 55.4% with immediate multivessel PCI (relative risk 0.83) (Thiele 2017). Time targets, fibrinolysis, and nonculprit strategy are covered on the ST-elevation myocardial infarction hub.
Chronic coronary disease. In stable patients, PCI relieves angina. Added to optimal medical therapy, it has not reduced death or myocardial infarction. In COURAGE, the 4.6-year rate of death or myocardial infarction was 19.0% with PCI and 18.5% with medical therapy alone (Boden 2007). ISCHEMIA randomized 5179 patients with moderate or severe ischemia; the primary event rate was higher with the invasive strategy at 6 months (5.3% against 3.4%) and lower at 5 years (16.4% against 18.2%), with no significant overall difference and 145 deaths against 144 (Maron 2020). ORBITA, with a placebo procedure, found no significant increase in exercise time with PCI (16.6 seconds, 95% confidence interval −8.9 to 42.0) (Al-Lamee 2018). The 2021 guideline downgraded revascularization for survival in multivessel disease with normal left ventricular function to Class 2b and gave revascularization a Class 2a recommendation to lower the risk of cardiovascular events such as spontaneous myocardial infarction, unplanned urgent revascularization, or cardiac death in multivessel disease appropriate for either CABG or PCI. It retained a Class 1 recommendation for revascularization to improve symptoms in refractory angina despite medical therapy with stenoses amenable to revascularization (Lawton 2022). These trials and the indications built on them are detailed on the chronic coronary disease hub.
Physiology before treatment. An angiographically intermediate stenosis without established ischemia is assessed with fractional flow reserve (FFR) or a nonhyperemic pressure ratio before PCI. In FAME, FFR guidance in multivessel disease reduced stents per patient (1.9 against 2.7) and 1-year death, myocardial infarction, or repeat revascularization (13.2% against 18.3%) (Tonino 2009). In DEFINE-FLAIR, guidance with the instantaneous wave-free ratio (iFR) was noninferior to FFR (6.8% against 7.0% at 1 year) with fewer procedural symptoms (3.1% against 30.8%) (Davies 2017). Thresholds, hyperemic agents, measurement technique, and discordance are on the FFR section of the angiography hub.
PCI or bypass surgery: anatomic complexity, the SYNTAX score, and the heart team
The 2021 guideline recommends a multidisciplinary heart team for patients in whom the best revascularization strategy is unclear, with patient-centered shared decision-making (Lawton 2022). Bypass surgery carries a Class 1 recommendation to improve survival in significant left main disease, and surgical revascularization is the recommended strategy in diabetes with three-vessel disease. PCI of the left main is a Class 2a alternative in selected patients with low to medium anatomic complexity. In complex multivessel disease with a SYNTAX score above 33, surgery has a Class 2a recommendation over PCI for survival. PCI may be considered in diabetes with three-vessel disease when the patient is a poor surgical candidate (Lawton 2022).
The SYNTAX score grades anatomic complexity from the angiogram: dominance, the number and location of lesions, and features such as total occlusion, bifurcation, trifurcation, aorto-ostial position, severe tortuosity, length, heavy calcification, thrombus, and diffuse disease. The guideline rates the usefulness of the score in treatment decisions as less well established than that of the Society of Thoracic Surgeons risk score, because the SYNTAX score varies between observers and contains no clinical variables (Lawton 2022). The comparative trials (SYNTAX, FREEDOM, EXCEL, NOBLE, FAME 3) are discussed on the chronic coronary disease hub, and the conduct of bypass surgery on the cardiac surgery and heart team hub.
Contraindications and procedural risk assessment
PCI has few absolute contraindications. The relative contraindications are those that make the procedure unsafe, unlikely to succeed, or unlikely to help:
- Inability to take dual antiplatelet therapy for the minimum period the chosen device requires, or active bleeding. Trials of stents with 1 month of dual antiplatelet therapy (high bleeding risk) and drug-coated balloons have narrowed this contraindication.
- Anatomy in which bypass surgery offers a clear survival or durability advantage and the patient is a reasonable surgical candidate.
- A stenosis without ischemic significance by physiology, or symptoms with a noncoronary explanation.
- Diffuse distal disease or a small vessel without a suitable landing zone.
- Uncorrected coagulopathy, a previous severe contrast reaction without a premedication plan, decompensated heart failure that prevents lying flat, or acute kidney injury when the procedure can safely wait.
- A patient who declines the procedure after informed discussion.
Risk assessment before elective PCI covers kidney function and contrast risk (contrast-associated kidney injury), bleeding risk (ARC high bleeding risk criteria), anticoagulant interruption (oral anticoagulants and PCI), access anatomy, left ventricular function, the myocardium at risk if the target vessel closes, and the need for mechanical support. Consent, documentation, and laboratory preparation are described on the angiography hub.
Appropriate use criteria and quality registries
The 2017 multisociety appropriate use criteria rate revascularization in stable ischemic heart disease by symptom severity, antianginal therapy, noninvasive risk findings, and coronary anatomy, and grade clinical scenarios for appropriateness (Patel 2017). In the National Cardiovascular Data Registry (NCDR) CathPCI Registry, 2.7 million PCIs from 766 hospitals showed a fall in nonacute PCI volume from 89,704 to 59,375 per year between 2010 and 2014. Among nonacute procedures, the share with Canadian Cardiovascular Society class III or IV angina rose from 15.8% to 38.4%, and the proportion rated inappropriate fell from 26.2% to 13.3% (Desai 2015).
Vascular access for PCI: radial, femoral, and alternative routes
The 2021 guideline recommends radial access for PCI in patients with acute coronary syndromes or stable ischemic heart disease to reduce bleeding and vascular complications compared with femoral access, and notes a lower mortality rate with radial access in acute coronary syndromes (Lawton 2022). Diagnostic access technique and sedation are on the angiography hub.
| Trial | Population | Main result, radial compared with femoral |
|---|---|---|
| RIVAL (Jolly 2011) | 7021 patients with acute coronary syndrome | 30-day death, myocardial infarction, stroke, or non-CABG major bleeding 3.7% against 4.0% (HR 0.92, not significant); large hematoma 42 against 106 patients and pseudoaneurysm needing closure 7 against 23; benefit in ST-elevation myocardial infarction and high-volume radial centers |
| RIFLE-STEACS (Romagnoli 2012) | ST-elevation acute coronary syndrome | 30-day net adverse clinical events 13.6% against 21.0%; cardiac mortality 5.2% against 9.2%; bleeding 7.8% against 12.2% |
| MATRIX (Valgimigli 2015) | 8404 patients with acute coronary syndrome | Net adverse clinical events 9.8% against 11.7%; BARC major bleeding unrelated to CABG 1.6% against 2.3%; all-cause mortality 1.6% against 2.2% |
| SAFE-PCI for Women (Rao 2014) | 1787 women, 691 undergoing PCI | Bleeding or vascular complications in the PCI cohort 1.2% against 2.9% (not significant; stopped early); significant reduction across the whole randomized cohort (0.6% against 1.7%); crossover 6.1% against 1.7% |
| DISCO RADIAL (Aminian 2022) | Distal radial (snuffbox) against conventional radial, 1307 patients | Forearm radial occlusion 0.31% against 0.91% (not significant); crossover 7.4% against 3.5%; spasm 5.4% against 2.7%; shorter hemostasis |
Randomized comparisons of access routes for coronary angiography and PCI. HR, hazard ratio; BARC, Bleeding Academic Research Consortium; CABG, coronary artery bypass grafting.
Femoral access remains necessary for large-bore guides, mechanical circulatory support, some chronic total occlusion strategies, and patients whose radial and ulnar routes are unusable. Ultrasound or fluoroscopic guidance places the puncture over the femoral head below the inferior epigastric artery. A puncture above the inguinal ligament is the setting for retroperitoneal hemorrhage, and a puncture at or below the bifurcation is associated with pseudoaneurysm and arteriovenous fistula.
Alternative routes. Distal radial access in the anatomic snuffbox gave radial occlusion rates comparable to conventional radial access with a strict hemostasis protocol, with more crossover and spasm (Aminian 2022). Ulnar access serves operators trained in it when the radial artery is small, occluded, or spastic. Brachial access is used rarely because the brachial artery is an end artery for the forearm and hand and bleeding in the antecubital fossa can compress the median nerve.
Sheaths and French size. French size describes outer circumference in millimeters, so 1 French equals one third of a millimeter of outer diameter. A sheath is named by the largest catheter it accepts, and its outer diameter is larger than its label. In a randomized comparison of transradial PCI, 5 French and 6 French guides gave similar success and local complication rates, and 5 French caused less access-site pain (Gwon 2006). Thin-walled radial sheaths and sheathless guide systems reduce the outer diameter in the artery for a given guide lumen.
Hemostasis and closure. After radial procedures, a compression band applied with patent hemostasis preserves antegrade flow during compression. In PROPHET-II, which randomized 3000 patients after diagnostic transradial catheterization, adding prophylactic ipsilateral ulnar compression to patent hemostasis reduced 30-day radial artery occlusion from 3.0% to 0.9% (Pancholy 2016). After femoral procedures, vascular closure devices shorten time to hemostasis. In ISAR-CLOSURE, which randomized 4524 patients after diagnostic femoral angiography with a 6 French sheath, closure devices were noninferior to manual compression for the composite of access-site complications (6.9% against 7.9%), and median time to hemostasis was 1 minute against 10 minutes (Schulz-Schüpke 2014).
| Complication | Setting | Recognition | Management |
|---|---|---|---|
| Retroperitoneal hemorrhage | Femoral puncture above the inguinal ligament or through the back wall | Hypotension, flank or back pain, falling hemoglobin with an unremarkable groin; confirmed by CT | Volume and blood, anticoagulant reversal where appropriate; covered stent or surgery for continued bleeding |
| Groin hematoma | Femoral access, large sheath, potent antithrombotic therapy | Swelling and pain at the site | Manual compression, observation; exclusion of pseudoaneurysm |
| Femoral pseudoaneurysm | Low puncture, inadequate compression | Pulsatile mass, bruit; duplex ultrasound | Ultrasound-guided thrombin injection or compression; surgery for large or infected lesions |
| Arteriovenous fistula | Puncture traversing artery and vein | Continuous bruit; duplex ultrasound | Observation of small fistulas; repair for large or symptomatic fistulas |
| Radial artery spasm | Small radial artery, repeated catheter exchanges | Pain and resistance to catheter movement | Intra-arterial vasodilators, sedation and analgesia, smaller catheters |
| Radial artery occlusion | Occlusive compression, small artery-to-sheath ratio | Usually silent; absent radial pulse or flow on ultrasound | Prevention by patent hemostasis and procedural anticoagulation; ipsilateral ulnar compression can restore flow |
| Forearm hematoma and compartment syndrome | Radial or brachial wire perforation of a small branch | Tense, painful forearm | Compression bandage or blood pressure cuff; surgical decompression for compartment syndrome |
Access-site complications of PCI, their recognition, and treatment.
Guide catheters: shapes, sizes, and support
A guide catheter differs from a diagnostic catheter in its larger lumen, braided wall, and stiffer shaft, which transmit push and torque and hold position while devices move through it. Backup support is the resistance of the guide to being pushed out of the ostium when a device meets resistance. Two features of the guide set its backup force: the angle between the guide segment resting on the contralateral aortic wall and the coronary axis, which gives maximum support at 90 degrees, and the area of aortic wall against which the secondary curve rests (Lee Chuy 2024).
Passive support comes from the stiffness, size, and shape of the catheter, which lets the secondary curve rest firmly against the opposite sinus or aortic wall without manipulation. Extra-backup left shapes (EBU, XB, Voda) and Amplatz shapes provide passive support; their long tips require careful engagement to avoid dissection. Active support requires manipulation, such as deep intubation or seating a Judkins left or Ikari left guide in the "power position" with the secondary curve in the aortic sinus. This maneuver is reserved for arteries without ostial or proximal disease into which the guide advances coaxially (Lee Chuy 2024).
| Shape | Usual target | Support | Characteristic hazard |
|---|---|---|---|
| Judkins left | Left coronary artery; curve size chosen to aortic root width | Moderate, passive; higher in power position | Poor coaxial alignment in a wide or short root |
| Extra-backup left (EBU, XB, Voda) | Left coronary artery, including complex left-system PCI from radial or femoral access | High, passive, from a long segment resting on the contralateral aortic wall | Deep seating in a short left main |
| Judkins right | Right coronary artery with a horizontal origin | Low to moderate | Spasm or dissection with deep engagement; poor support for complex right coronary PCI |
| Amplatz left | Right coronary artery with a superior or shepherd's crook origin; anomalous origins; vein grafts; high-support left PCI | High, passive | Tip dives into the artery as devices are withdrawn; ostial dissection |
| Amplatz right and multipurpose | Inferiorly directed right coronary origins and some grafts | Moderate | Deep intubation |
| Ikari left | Left or right coronary artery from radial access | High in power position | Requires deliberate positioning |
Guide catheter shapes for PCI and their principal uses. Curve size is chosen from aortic root dimensions, ostial position, and access route; radial and femoral routes change the attachment point and therefore the support of a given shape.
Guide selection weighs aortic root size, ostial location, lesion location, and the tortuosity and calcification that devices must traverse (Lee Chuy 2024). Longer sheaths stiffen the system and bypass tortuosity in the brachiocephalic, subclavian, and iliofemoral arteries, improving torque transmission. Guide size follows the planned equipment: for rotational atherectomy, a 6 French guide accommodates burrs up to 1.75 mm, and larger burrs need a larger guide or a 7.5 French sheathless radial guide (Barbato 2015).
Pressure damping, ventricularization, and side holes
The pressure recorded at the guide tip is monitored continuously during PCI. Damping is a global fall in waveform amplitude with a rounded contour; it indicates that the catheter tip is obstructing flow, from ostial stenosis, spasm, a small vessel, or a tip pressed against the wall. Ventricularization is loss of the aortic diastolic contour, with a low diastolic pressure that resembles a ventricular tracing; it indicates that the catheter has wedged into the ostium and coronary inflow has stopped. Injection through a damped or ventricularized catheter risks ostial and aortic dissection and drives contrast into the arterial wall; sustained wedging causes ischemia and ventricular arrhythmia. The catheter is withdrawn until the waveform normalizes, and the ostium is reassessed with a nonselective injection.
Side holes near the tip of a guide catheter allow blood to enter the catheter when the tip is wedged, so the pressure tracing appears normal while the coronary artery is underperfused. Side holes therefore mask damping, reduce the contrast delivered to the coronary artery, and permit aortic reflux of contrast. They are used selectively for ostial right coronary or vein graft lesions in which engagement inevitably wedges the catheter. In pressure-wire studies, a damped or ventricularized aortic reference falsely lowers aortic pressure and falsely raises the measured FFR, so the guide is disengaged during measurement (wire technique).
Coronary guidewires
Coronary guidewires share a standard diameter of 0.014 inch, compatible with all rapid-exchange devices, and are built from a core (stainless steel or nitinol) that tapers toward the tip, a distal coil or polymer jacket, and a surface coating. The core and taper set support and torque response; the tip construction sets tip load (the force needed to deflect the tip, expressed in grams), steerability, and tactile feedback; the coating sets lubricity.
| Wire class | Construction | Use | Hazard |
|---|---|---|---|
| Workhorse | Soft coil tip, moderate support, hydrophilic or hydrophobic coating on the distal segment | Most native-vessel lesions; first wire in most procedures | Limited support in tortuous or calcified vessels |
| Polymer-jacketed | Polymer sleeve over the distal core, hydrophilic surface | Tortuous vessels, tight lesions, microchannels | Silent subintimal passage and distal perforation; reduced tactile feedback |
| Extra-support | Stiffer core extending closer to the tip | Straightening tortuous segments for stent delivery; buddy wire | Pseudostenosis (accordion effect) from vessel straightening; vessel trauma |
| Intermediate and high tip load | Heavier tip with tapered or untapered core | Fibrous lesions and occlusion caps, usually through a microcatheter | Perforation, especially with tapered stiff tips |
| Pressure and imaging-compatible wires | Pressure sensor near the tip; standard rail for imaging catheters | Physiology before and after PCI; imaging runs | Sensor drift; reduced handling compared with workhorse wires |
Classes of coronary guidewires by construction and use. Chronic total occlusion PCI adds escalation wires and retrograde collateral wires.
The wire tip is shaped to the vessel: a small distal bend for most lesions, a longer or second bend for wide-angle branches. Wire position is confirmed in two projections before devices are advanced, because a wire in a small side branch or in a subintimal track looks similar in one view. During long procedures the distal tip is kept in view; a tip that moves distally with device advancement can perforate a small terminal branch, a mechanism that underlies many distal perforations. A buddy wire, a second wire placed alongside the first or in a side branch, straightens the vessel, provides a second rail, anchors the guide, and protects branch access (device delivery).
Angioplasty balloons: compliance, nominal and rated burst pressure
Balloon catheters are built as rapid-exchange (monorail) devices, in which the wire exits the catheter a short distance behind the balloon, or over-the-wire devices, in which the wire runs the full catheter length. Over-the-wire balloons and microcatheters allow wire exchange and increase support but slow device handling.
Compliance describes how much a balloon diameter grows with pressure. Compliant balloons grow substantially beyond nominal diameter and are used for low-pressure inflations. Semicompliant balloons grow moderately and are used for predilation, crossing, and stent delivery. Noncompliant balloons grow little and are used for high-pressure predilation of fibrotic or calcified lesions and for postdilation of stents, where they concentrate force within the treated segment without overstretching the vessel beyond the balloon.
Nominal pressure and rated burst pressure. The manufacturer's compliance chart lists balloon diameter at a series of pressures and identifies the nominal pressure, at which the balloon reaches its labeled diameter, and the rated burst pressure. The US Food and Drug Administration (FDA) defines the rated burst pressure as the pressure at which 99.9% of balloons can survive with 95% confidence; a balloon that fails at that pressure can cause device malfunction or vessel damage (FDA 2010). The FDA guidance permits a compliance chart to list pressures up to 25% above the rated burst pressure only when data show that 99% of balloons will not fail at those pressures with 95% confidence (FDA 2010). Inflation beyond the rated burst pressure increases the risk of balloon rupture, which can dissect or perforate the artery or embolize balloon material; a pinhole rupture in a calcified lesion is the usual mechanism. Stent delivery balloons carry their own compliance chart, and the deployed stent diameter follows that chart.
| Balloon type | Diameter growth with pressure | Typical role | Limitation |
|---|---|---|---|
| Compliant | Large | Low-pressure dilation, anchoring, occlusion of a perforation | Dog-bone expansion into normal segments at higher pressure |
| Semicompliant | Moderate | Predilation, crossing, most stent delivery systems | Uneven expansion in resistant lesions |
| Noncompliant | Small | High-pressure predilation, stent postdilation, proximal optimization | Stiffer and bulkier to deliver |
| Super high-pressure noncompliant | Very small at very high pressure | Resistant calcified lesions and stent underexpansion | Perforation and dissection at extreme pressure |
| Cutting | Noncompliant with mounted atherotomes | Fibrotic, ostial, and in-stent restenotic lesions | Bulky profile; entrapment in tortuous vessels |
| Scoring | Semicompliant or noncompliant with external wire or nitinol element | Lesion preparation; prevents slippage (watermelon seeding) | Less effective in deep circumferential calcium |
| Drug-coated | Semicompliant carrier coated with antiproliferative drug and excipient | In-stent restenosis; selected de novo lesions | Requires prior lesion preparation; no scaffold for recoil or dissection |
Coronary balloon types by compliance and role. Exact nominal and rated burst pressures are device specific and are read from the compliance chart of the balloon in use.
Cutting, scoring, and other lesion-preparation balloons
Cutting balloons carry longitudinal microblades that incise the plaque during inflation, and scoring balloons carry a wire or nitinol element that concentrates force along lines. Both direct plaque fracture at lower pressures, reduce balloon slippage, and are used to prepare fibrotic or fibrocalcific lesions, ostial lesions, and in-stent restenosis before a stent or drug-coated balloon. In a network meta-analysis of in-stent restenosis treatments, plain cutting balloon angioplasty did not differ from plain balloon angioplasty for target-vessel revascularization, while everolimus-eluting stents, paclitaxel-eluting balloons, paclitaxel-eluting stents, and sirolimus-eluting stents were associated with lower rates (Sethi 2015). Cutting and scoring balloons are therefore preparation tools that precede a definitive antiproliferative treatment.
Deep or circumferential calcium resistant to balloon preparation requires complex PCI techniques of calcium modification: rotational or orbital atherectomy, intravascular lithotripsy, or laser atherectomy, selected by calcium morphology on intravascular imaging.
Bare-metal stents
A bare-metal stent is a balloon-expandable metal scaffold without drug or polymer. It eliminates elastic recoil and tacks dissection flaps, which ended most abrupt closures after balloon angioplasty. The vessel responds to the struts with neointimal hyperplasia that peaks in the first months, and in-stent restenosis from this tissue was the dominant failure mode of bare-metal stents.
Strut thickness itself affects restenosis. ISAR-STEREO randomized 651 patients to stents of comparable design with 50 µm or 140 µm struts; angiographic restenosis was 15.0% against 25.8% and reintervention 8.6% against 13.8% (Kastrati 2001). Thin struts, which disturb flow less and injure the wall less, became a design goal of later drug-eluting platforms.
The traditional reason to choose a bare-metal stent was a short dual antiplatelet period in patients at high bleeding risk or facing early surgery. Randomized trials with 1 month of dual antiplatelet therapy have removed that rationale (high bleeding risk), and the 2021 guideline recommends contemporary drug-eluting stents over bare-metal stents for all patients (Lawton 2022).
Drug-eluting stents: drugs, polymers, and platforms
A drug-eluting stent combines a metal platform, an antiproliferative drug, and in most designs a polymer that controls drug release. The limus drugs (sirolimus and its analogs everolimus, zotarolimus, biolimus, and umirolimus) inhibit the mammalian target of rapamycin and arrest smooth-muscle proliferation; paclitaxel stabilizes microtubules. The drug suppresses neointimal hyperplasia and also delays endothelial coverage of the struts, which is why antiplatelet therapy is needed while the stent heals.
| Generation | Platform and struts | Drug and carrier | Advance | Remaining limitation |
|---|---|---|---|---|
| Balloon angioplasty | No implant | None | Dilated the fixed stenosis | Recoil, abrupt closure from dissection, restenosis |
| Bare-metal stent | Stainless steel, later cobalt-chromium; thick struts in early designs | None | Prevented recoil and abrupt closure | Neointimal hyperplasia and in-stent restenosis |
| First-generation drug-eluting stent | Stainless steel, thick struts | Sirolimus or paclitaxel on durable polymer | Large reduction in restenosis (SIRIUS, TAXUS IV) | Delayed healing and late and very late stent thrombosis |
| Second-generation drug-eluting stent | Cobalt-chromium or platinum-chromium, thin struts | Everolimus or zotarolimus on biocompatible durable polymer | Lower stent thrombosis, myocardial infarction, and repeat revascularization than paclitaxel stents (SPIRIT IV, COMPARE) | Persistent polymer; low rate of very late events |
| Biodegradable-polymer drug-eluting stent | Thin or ultrathin struts | Limus drug on polymer that degrades over months | Noninferior or better outcomes (BIOFLOW V) | Clinical advantage over durable-polymer designs inconsistent |
| Polymer-free drug-coated stent | Textured or microporous metal surface | Limus drug without polymer | Superior to bare-metal stent with 1-month DAPT (LEADERS FREE) | Durable-polymer stents performed equally (ONYX ONE) |
| Bioresorbable scaffold | Poly-L-lactic acid, thick struts | Everolimus | Temporary scaffold that resorbs | Excess scaffold thrombosis and target-vessel myocardial infarction (ABSORB III, AIDA) |
Generations of coronary stent technology. DAPT, dual antiplatelet therapy.
From paclitaxel to limus stents. SPIRIT IV compared everolimus-eluting with paclitaxel-eluting stents: 1-year target-lesion failure was 4.2% against 6.8%, myocardial infarction 1.9% against 3.1%, and stent thrombosis 0.17% against 0.85%; the benefit was absent in diabetes (6.4% against 6.9%) (Stone 2010). In the all-comer COMPARE trial, the everolimus stent reduced the primary composite (6% against 9%), with lower stent thrombosis (6 against 23 patients), myocardial infarction, and target-vessel revascularization (Kedhi 2010).
Polymers. Durable polymers persist after drug release; biodegradable polymers dissolve over months and leave a bare platform; polymer-free designs carry the drug on a modified metal surface. In BIOFLOW V, an ultrathin-strut biodegradable-polymer sirolimus-eluting stent had a 12-month target-lesion failure rate of 6% against 10% with a thin-strut durable-polymer everolimus stent, driven by target-vessel myocardial infarction (Kandzari 2017). Polymer type has not determined dual antiplatelet duration in US guidance; duration follows clinical presentation and the balance of ischemic and bleeding risk (DAPT duration).
Evidence comparing drug-eluting and bare-metal stents
In randomized comparisons including NORSTENT, ZEUS, and LEADERS FREE, drug-eluting stents reduced repeat revascularization compared with bare-metal stents. NORSTENT randomized 9013 patients with stable or unstable coronary disease; at 6 years death or nonfatal spontaneous myocardial infarction did not differ (16.6% against 17.1%), while repeat revascularization (16.5% against 19.8%) and definite stent thrombosis (0.8% against 1.2%) were lower with drug-eluting stents (Bønaa 2016). The absence of a hard-outcome difference in NORSTENT and the reductions in repeat revascularization and stent thrombosis together underpin the guideline preference.
The remaining argument for bare-metal stents, a shorter antiplatelet period, was tested in patients with high bleeding risk or advanced age.
Drug-eluting stents in patients at high bleeding risk
| Trial | Population | Comparison and DAPT | Result |
|---|---|---|---|
| ZEUS (Valgimigli 2015) | Uncertain drug-eluting stent candidates, including high bleeding risk | Zotarolimus-eluting stent against bare-metal stent; median DAPT 32 days | 1-year major adverse cardiac events 17.5% against 22.1%; myocardial infarction 2.9% against 8.1%; definite or probable stent thrombosis 2.0% against 4.1% |
| LEADERS FREE (Urban 2015) | 2466 patients at high bleeding risk | Polymer-free umirolimus-coated stent against bare-metal stent; 1-month DAPT | Primary safety end point 9.4% against 12.9% (superior); clinically driven target-lesion revascularization 5.1% against 9.8% |
| SENIOR (Varenne 2018) | 1200 patients aged 75 years or older | Drug-eluting against bare-metal stent with identical short DAPT | Primary composite 12% against 16% (RR 0.71); bleeding 5% in both groups |
| ONYX ONE (Windecker 2020) | 1996 patients at high bleeding risk | Durable-polymer zotarolimus-eluting stent against polymer-free stent; 1-month DAPT | 1-year safety composite 17.1% against 16.9% (noninferior) |
| XIENCE 90 and XIENCE 28 (Valgimigli 2021) | High bleeding risk after everolimus-eluting stent | 1-month against 3-month DAPT, compared across single-arm studies | Death or myocardial infarction 7.3% against 7.5%; BARC type 2–5 bleeding 7.6% against 10.0% |
Stent and short dual antiplatelet therapy trials in patients at high bleeding risk or of advanced age. RR, relative risk; BARC, Bleeding Academic Research Consortium; DAPT, dual antiplatelet therapy.
These trials established that a contemporary drug-eluting stent with 1 month of dual antiplatelet therapy outperforms a bare-metal stent with the same regimen, and ONYX ONE showed that the advantage does not depend on a polymer-free design. In stable ischemic heart disease, the 2021 guideline considers a short course of dual antiplatelet therapy after PCI reasonable to reduce bleeding, with transition to P2Y12 inhibitor monotherapy after 1 to 3 months in selected patients (Class 2a) (Lawton 2022). Definitions of high bleeding risk, the trials of P2Y12 inhibitor monotherapy, and duration tables are on the dual antiplatelet therapy hub.
Bioresorbable scaffolds
The everolimus-eluting poly-L-lactic acid scaffold was designed to support the vessel for months and then resorb, restoring vasomotion and leaving no permanent implant. In ABSORB III, 1-year target-lesion failure met noninferiority against a cobalt-chromium everolimus stent (7.8% against 6.1%), with device thrombosis of 1.5% against 0.7% (Ellis 2015). Over 5 years, target-lesion failure, target-vessel myocardial infarction, and scaffold thrombosis were higher with the scaffold; the excess hazard ended at 3 years, coincident with complete resorption (Kereiakes 2019). In the all-comer AIDA trial, target-vessel failure did not differ at 2 years (11.7% against 10.7%), but definite or probable device thrombosis was 3.5% against 0.9% (HR 3.87) (Wykrzykowska 2017).
The mechanisms were thick struts that disturbed flow and protruded into the lumen, sensitivity to underexpansion and incomplete lesion preparation, and late structural discontinuity during resorption. Magnesium and thinner polymer scaffolds remain investigational. In the ABSORB experience, mechanical failure modes that appeared only with longer follow-up outweighed the theoretical advantage of resorption.
Stent thrombosis: definitions, timing, and mechanisms
The Academic Research Consortium (ARC) standardized stent thrombosis definitions in 2007 and revised them in ARC-2 (Cutlip 2007; Garcia-Garcia 2018). Certainty is graded as definite, probable, or possible, and timing as early or late relative to implantation.
| Category | Definition |
|---|---|
| Definite | Angiographic confirmation of thrombus originating in the stent or within 5 mm of it with a clinical ischemic event, or pathologic confirmation of recent thrombus within the stent at autopsy or in retrieved tissue |
| Probable | Unexplained death within 30 days of implantation, or myocardial infarction related to acute ischemia in the territory of the stent without angiographic confirmation and without another obvious cause |
| Possible | Unexplained death from 30 days after implantation onward |
| Acute | 0 to 24 hours after implantation |
| Subacute | More than 24 hours to 30 days |
| Late | More than 30 days to 1 year |
| Very late | More than 1 year |
Academic Research Consortium definitions of stent thrombosis by certainty and timing. Acute and subacute thrombosis together form early stent thrombosis.
Stent thrombosis usually presents as ST-elevation myocardial infarction from abrupt occlusion of a previously treated artery. Early thrombosis is dominated by mechanical and pharmacologic causes: underexpansion, malapposition, edge dissection, residual inflow or outflow disease, geographic miss, thrombus prolapse, and interruption or inadequate effect of antiplatelet therapy. Late and very late thrombosis add biologic causes: delayed strut coverage, hypersensitivity to polymer, late malapposition from positive remodeling, and rupture of neoatherosclerotic plaque within the stent. Treatment is emergency PCI with restoration of flow, intravascular imaging to define and correct the mechanism, and review of antiplatelet adherence and response.
In-stent restenosis and neoatherosclerosis
In-stent restenosis is renarrowing within a stent or at its edges. Despite contemporary drug-eluting stents, in-stent restenosis and the resulting repeat revascularization occur at about 1% to 2% per year (Giustino 2022). Presentation ranges from recurrent exertional angina to an acute coronary syndrome or total occlusion of the stented segment.
Mechanisms. Neointimal hyperplasia is homogeneous smooth-muscle tissue growing through the struts and dominates early restenosis. Neoatherosclerosis is new lipid-laden plaque, often with a thin fibrous cap, calcification, or rupture, forming within the neointima years after implantation; it is best characterized by optical coherence tomography. Mechanical causes include stent underexpansion (often from unprepared calcium), stent fracture, edge injury, and geographic miss, in which the stent fails to cover the full diseased or balloon-injured segment. The mechanism directs treatment: underexpansion requires high-pressure noncompliant dilation or calcium modification, and an additional stent seldom corrects it; neointimal hyperplasia requires an antiproliferative treatment; uncovered disease at an edge requires coverage.
| Mehran pattern | Definition | Frequency in original series | Target-lesion revascularization |
|---|---|---|---|
| I, focal | Lesion 10 mm or shorter | 42% | 19% |
| II, diffuse intrastent | Longer than 10 mm, confined to the stent | 21% | 35% |
| III, diffuse proliferative | Longer than 10 mm, extending beyond the stent edges | 30% | 50% |
| IV, total occlusion | Restenosis with Thrombolysis in Myocardial Infarction (TIMI) grade 0 flow | 7% | 83% |
Angiographic classification of in-stent restenosis after bare-metal stenting (Mehran 1999). Diabetes and previously recurrent restenosis were more frequent in higher classes, and a higher class predicted repeat revascularization.
Intravascular imaging defines the mechanism before reintervention, and pressure-wire physiology confirms significance when the angiogram is equivocal. Lesion preparation with noncompliant, cutting, or scoring balloons precedes treatment with a drug-eluting stent or a drug-coated balloon.
Drug-coated balloons for in-stent restenosis
A drug-coated balloon transfers an antiproliferative drug, carried in an excipient matrix, to the vessel wall during a single inflation and leaves no new metal layer. The International DCB Consensus Group describes drug-coated balloons as established for in-stent restenosis of bare-metal and drug-eluting stents and notes that differences in balloon design, dose, formulation, and release kinetics mean there is no class effect (Jeger 2020).
| Trial | Population | Comparison | Result |
|---|---|---|---|
| ISAR-DESIRE 3 (Byrne 2013) | 402 patients with drug-eluting stent restenosis | Paclitaxel balloon against paclitaxel stent against plain balloon | Diameter stenosis at 6–8 months 38.0% against 37.4% (noninferior); both superior to plain balloon (54.1%) |
| ISAR-DESIRE 3 at 10 years (Giacoppo 2023) | Same cohort | Same | Primary composite 55.9% with paclitaxel balloon, 62.4% with paclitaxel stent, 72.0% with plain balloon; no significant difference between balloon and stent |
| DAEDALUS (Giacoppo 2020) | Pooled patient-level data from 10 trials: 710 with bare-metal and 1248 with drug-eluting stent restenosis | Drug-coated balloon against repeat drug-eluting stent | Bare-metal stent restenosis: no efficacy difference. Drug-eluting stent restenosis: 3-year target-lesion revascularization higher with balloon (20.3% against 13.4%), safety end point numerically lower (9.5% against 13.3%) |
| AGENT IDE (Yeh 2024) | 600 US patients with in-stent restenosis | Paclitaxel-coated balloon against uncoated balloon | 1-year target-lesion failure 17.9% against 28.6%; target-lesion revascularization 13.0% against 24.7%; target-vessel myocardial infarction 5.8% against 11.1% |
Randomized and patient-level evidence for drug-coated balloons in coronary in-stent restenosis.
AGENT IDE was the basis for FDA premarket approval of the AGENT paclitaxel-coated balloon, indicated after appropriate vessel preparation in adults undergoing PCI for in-stent restenosis in coronary arteries 2.0 to 4.0 mm in diameter with lesions up to 26 mm long (FDA 2024). Its contraindications include known hypersensitivity to paclitaxel, unprotected native left main disease, coronary spasm without a significant stenosis, and inability to take the recommended antiplatelet or anticoagulant therapy (FDA 2024).
The choice between a drug-coated balloon and a repeat drug-eluting stent follows the mechanism and the number of stent layers. A balloon avoids a further metal layer in multilayer restenosis and allows short antiplatelet therapy, while DAEDALUS favored a repeat drug-eluting stent for efficacy in drug-eluting stent restenosis. Underexpansion is corrected before either treatment.
Drug-coated balloons for de novo and small-vessel disease
For de novo lesions, drug-coated balloon angioplasty follows a "leave nothing behind" strategy: the lesion is prepared, and a stent is placed only for recoil, a flow-limiting dissection, or a residual stenosis. BASKET-SMALL 2 randomized 758 patients with vessels under 3 mm; 12-month major adverse cardiac events were 7.5% with the drug-coated balloon and 7.3% with a drug-eluting stent, meeting noninferiority (Jeger 2018). REC-CAGEFREE I randomized 2272 Chinese patients with de novo noncomplex lesions. Rescue stenting was needed in 9.4% of the balloon group, and the 24-month device-oriented composite was 6.4% with the balloon strategy against 3.4% with intended stenting, so noninferiority was not met (Gao 2024).
The de novo evidence therefore supports drug-coated balloons in selected small vessels, in side branches, and in patients for whom a permanent implant or prolonged antiplatelet therapy is undesirable. In larger de novo coronary arteries, the evidence favors stenting. In the United States, the labeled coronary indication of the approved paclitaxel-coated balloon is in-stent restenosis (FDA 2024).
Physiology before PCI and intravascular imaging during PCI
Pressure-wire physiology establishes whether a stenosis causes ischemia; intravascular imaging defines how to treat it. The physiologic evidence base (FAME, DEFINE-FLAIR, iFR-SWEDEHEART, FAME 3) and post-PCI physiology are on the angiography hub, and the guideline recommendations for intravascular imaging in stable disease are on the chronic coronary disease hub. The 2025 ACC/AHA acute coronary syndromes guideline gives intravascular imaging guidance of PCI a Class 1 recommendation (Level of Evidence A) (Rao 2025).
Intravascular ultrasound (IVUS) and optical coherence tomography (OCT) measure reference vessel size and lesion length for stent selection, characterize calcium, and after stenting identify underexpansion, malapposition, edge dissection, tissue protrusion, and geographic miss. IVUS needs no blood clearance and images the aorto-ostial left main; OCT has higher resolution, distinguishes calcium thickness, thrombus, and neoatherosclerosis, and requires a contrast or flush clearance.
| Trial | Population | Result, imaging compared with angiography guidance |
|---|---|---|
| IVUS-XPL (Hong 2015) | 1400 patients with long lesions | 1-year major adverse cardiac events 2.9% against 5.8%, driven by target-lesion revascularization |
| RENOVATE-COMPLEX-PCI (Lee 2023) | 1639 patients with complex lesions; IVUS or OCT | Target-vessel failure 7.7% against 12.3% at median 2.1 years (HR 0.64) |
| OCTOBER (Holm 2023) | 1201 patients with complex bifurcation lesions | 2-year major adverse cardiac events 10.1% against 14.1% (HR 0.70) |
| ILUMIEN IV (Ali 2023) | 2487 patients with medication-treated diabetes or complex lesions | Larger minimum stent area (5.72 against 5.36 mm²); 2-year target-vessel failure 7.4% against 8.2% (not significant); stent thrombosis 0.5% against 1.4% |
| Network meta-analysis (Stone 2024) | 22 trials, 15,964 patients | Target-lesion failure RR 0.71; cardiac death RR 0.55; stent thrombosis RR 0.52; all-cause death RR 0.75; OCT and IVUS similar |
Randomized evidence for intravascular imaging guidance of drug-eluting stent implantation. HR, hazard ratio; RR, relative risk.
Stent sizing, deployment, and optimization
Lesion preparation. Predilation with a semicompliant or noncompliant balloon tests lesion compliance, opens a channel for the stent, and reveals calcium that will resist expansion. A balloon that does not expand fully at high pressure, leaving a persistent waist, predicts stent underexpansion and calls for specialty balloons or calcium modification before the stent is placed. Direct stenting without predilation suits soft, noncalcified lesions, particularly thrombotic lesions in acute coronary syndromes where repeated balloon inflation may embolize debris.
Sizing. Stent diameter is chosen from the distal reference diameter, measured by quantitative angiography or, more reliably, by intravascular imaging. Imaging sizing uses the external elastic lamina or lumen diameter at the reference segments according to the protocol in use. Stent length covers the lesion from normal segment to normal segment so that both edges land in segments with limited plaque; landing an edge in a plaque-heavy segment increases the risk of edge dissection and edge restenosis.
Deployment. The stent is positioned in two projections, using radiopaque balloon markers and fixed landmarks such as side-branch origins. It is deployed at or above nominal pressure according to the compliance chart, with inflation held long enough for full expansion. In tapered vessels, the stent is sized to the distal reference and the proximal segment is expanded afterward with a larger noncompliant balloon; this proximal optimization technique is standard in bifurcation and left main stenting.
Postdilation and optimization. High-pressure postdilation with a noncompliant balloon shorter than the stent expands the stent without injuring the edges. Intravascular imaging then confirms that expansion is adequate relative to the reference segments, that struts are apposed in large vessels, and that no edge dissection or geographic miss requires treatment. Angiographic signs of a good result include a smooth contour without step-up or step-down, TIMI 3 flow, no dissection beyond the stent edges, and patent side branches. A final angiogram is taken in two projections. In a tortuous vessel the guidewire is removed first, because a stiff wire straightens the artery and creates pseudostenosis (accordion effect) that disappears when the wire is withdrawn.
Troubleshooting guide engagement and support
Failure to engage a coronary ostium coaxially usually reflects a mismatch between catheter curve and aortic root, an unusual ostial origin, or access-route tortuosity. Wider roots need larger curves; elongated or unfolded aortas in older adults change the angle of approach; radial access from the right reverses the geometry compared with femoral access. A standard 0.035-inch wire left in the ascending aorta straightens the subclavian route and transmits torque when engagement fails from radial access. Anomalous coronary origins and grafts often need Amplatz, multipurpose, or dedicated shapes (coronary anomalies).
When the guide backs out as devices advance, support is increased with a more supportive passive shape (extra-backup left, Amplatz), a larger guide, a longer sheath, deep intubation over a balloon shaft or in the power position, a buddy wire, an anchor balloon, or a guide extension catheter (Lee Chuy 2024). Deep intubation is temporary and is performed only in an artery without proximal disease, with the pressure waveform watched for damping. Ostial dissection from an aggressive or deeply seated guide can propagate retrograde into the aortic root; recognition requires a nonselective injection, and treatment is prompt stenting of the ostium, with surgery for extension into the ascending aorta.
Troubleshooting wire crossing
Wire crossing fails for three reasons: the tip shape does not match the vessel angle, the vessel is too tortuous for the wire to transmit torque, or the lesion is too tight or fibrotic for the tip. A smaller or second bend, a polymer-jacketed wire, or a microcatheter or over-the-wire balloon placed close to the lesion addresses these in turn. The microcatheter supports the wire tip, allows reshaping without losing position, and permits escalation to wires with higher tip load. In angulated side branches, dual-lumen microcatheters and deflectable-tip catheters direct the wire into the branch.
A wire that crosses with no resistance but fails to reach a distal branch, that moves freely without following the vessel contour, or that produces a lucent channel on injection may lie in a subintimal space. Position is verified in orthogonal projections, by contralateral injection when collateral filling is the only view of the distal vessel, or by intravascular imaging, before any balloon is advanced. Chronic total occlusions are crossed with dedicated techniques, including antegrade dissection and reentry and retrograde approaches.
Troubleshooting device delivery: buddy wire, anchor balloon, and guide extensions
A balloon or stent that will not reach or cross a lesion meets resistance from proximal tortuosity, calcium, a previously implanted stent edge, or inadequate guide support. The causes are separated by watching the guide: a guide that backs out signals insufficient support, while a device that stalls with a stable guide signals resistance in the vessel.
Lesion preparation first. Predilation with a small balloon and then a noncompliant balloon at higher pressure opens the lesion; calcium that prevents balloon expansion requires calcium modification before a stent is forced through it. A forced stent is the most common setting for stent stripping and loss.
Buddy wire. A second wire, parallel in the target vessel or in a side branch, straightens the vessel and anchors the guide. Parallel wiring provides a rail; side-branch wiring anchors the system for tight main-branch lesions, prevents deep guide intubation as balloons are withdrawn, and helps re-engagement for ostial lesions (Lee Chuy 2024).
Anchor balloon. A balloon inflated at low pressure in a side branch or distal segment fixes the guide position while the target device advances. Alternatively, a balloon inflated beside the device within the guide locks it against the guide wall so that the system advances as a unit. Anchoring is used to deliver equipment through tortuous or diseased proximal segments.
Guide extension catheters. A rapid-exchange guide extension is a short catheter segment on a pushrod that advances through the guide into the coronary artery, extending the effective length of the guide and providing coaxial support close to the lesion. Extensions are 1 French smaller than the corresponding guide, and a 6 French extension generally accepts the equipment used with a 5 French guide, typically one wire and one balloon or stent; thin-walled designs accept more (Lee Chuy 2024; Frizzell 2026). The extension is delivered over a balloon shaft or an inflated anchor balloon to keep its leading edge off the vessel wall, because pushing it over a bare wire biases it to the outer curve and causes dissection. Repeated balloon-assisted advancement (the inchworm technique) delivers it further. Contrast injection through a deeply placed extension risks barotrauma, dissection, and perforation, so position is checked with the extension withdrawn or against earlier reference images. Pushrod rotation during advancement wraps the pushrod around the wire and can block stent passage; clamping a hemostat to the end tab during insertion prevents wire wrap (Frizzell 2026).
| Problem | Likely cause | Escalation in order |
|---|---|---|
| Guide will not engage coaxially | Curve mismatch with aortic root; anomalous origin; access tortuosity | Different curve size or shape; stiff 0.035-inch wire to straighten the route; alternative access |
| Guide backs out as devices advance | Insufficient support | Supportive shape; larger guide; long sheath; buddy wire; anchor balloon; guide extension; deep intubation |
| Wire will not cross | Tip shape; tortuosity; tight or fibrotic lesion | Tip reshaping; polymer-jacketed wire; microcatheter or over-the-wire balloon support; higher tip load |
| Balloon will not cross | Tight, fibrotic, or calcified lesion; poor support | Small low-profile balloon; increased support; guide extension; atherectomy for calcium |
| Balloon will not expand | Calcium; fibrosis | Noncompliant or super high-pressure balloon; cutting or scoring balloon; calcium modification guided by imaging |
| Stent will not reach lesion | Proximal tortuosity or calcium; prior stent edge; inadequate preparation | Further preparation; buddy wire; anchor balloon; guide extension; shorter stent |
| Stent will not cross into prior stent | Strut catching at an edge | Guide extension across the edge; buddy wire; rotating the delivery system; balloon-assisted tracking |
Troubleshooting engagement, wire crossing, and device delivery during PCI.
Stent loss, device entrapment, and longitudinal deformation
Stent loss. An undeployed stent can strip from its delivery balloon when forced through calcium, tortuosity, or a prior stent edge, or when withdrawn into a guide that is not coaxial. If the wire remains through the stent, a small balloon advanced through the lost stent and inflated at low pressure can pull it back into the guide or deploy it in a safe segment; alternatively, the stent is crushed against the wall with a second stent. Without wire position, retrieval uses snares, double-wire twisting, or small-balloon techniques. A stent lost in the systemic circulation is usually retrieved with a snare or left when embolization is to a peripheral site without consequence.
Device entrapment. Balloons, wires, atherectomy burrs, and jailed side-branch wires can become entrapped. Jailed wires are best withdrawn before high-pressure postdilation, and polymer-jacketed wires are less likely to fracture when withdrawn from under struts. Fractured wire fragments are snared when accessible; fragments embedded under deployed struts without a mechanical consequence are sometimes left. Surgical retrieval is reserved for failure of percutaneous methods with ischemia.
Longitudinal stent deformation. After deployment, a guide, guide extension, imaging catheter, or balloon that strikes the proximal stent edge can compress or elongate the stent along its axis. Deformation is identified on fluoroscopy by shortening or overlap of the stent markers or on intravascular imaging and is treated with balloon dilation of the compressed segment, with an additional stent when struts protrude into the lumen.
Procedural antithrombotic therapy
Unfractionated heparin is the only Class 1 anticoagulant for PCI in the 2021 guideline unless it is contraindicated (Lawton 2022). It is given as a weight-based bolus, adjusted to the activated clotting time and to the use of a glycoprotein IIb/IIIa inhibitor, and can be reversed with protamine if perforation occurs. Weight-based doses, activated clotting time targets, and the use of enoxaparin and fondaparinux at PCI are on the parenteral anticoagulant hub.
Bivalirudin, a direct thrombin inhibitor, is the anticoagulant for patients with heparin-induced thrombocytopenia and an alternative in others. In VALIDATE-SWEDEHEART, 6006 patients with myocardial infarction undergoing PCI had a 180-day rate of death, myocardial infarction, or major bleeding of 12.3% with bivalirudin and 12.8% with heparin (Erlinge 2017).
| Drug | Role at PCI | Labeled dosing | Notes |
|---|---|---|---|
| Aspirin | Background antiplatelet therapy | Loading before PCI, then low-dose maintenance | Loading and maintenance doses on the DAPT hub |
| Clopidogrel | P2Y12 inhibitor for elective PCI and with oral anticoagulation | 300 mg loading dose, then 75 mg once daily (FDA label) | A 600 mg dose is labeled for the transition from cangrelor |
| Prasugrel | P2Y12 inhibitor in acute coronary syndrome with known anatomy | 60 mg loading dose, then 10 mg once daily; 5 mg once daily considered below 60 kg | In the pivotal trial, loading in non-ST-elevation acute coronary syndrome followed definition of the coronary anatomy |
| Ticagrelor | P2Y12 inhibitor in acute coronary syndrome | 180 mg loading dose, then 90 mg twice daily for the first year | Aspirin maintenance 75–100 mg daily with ticagrelor |
| Cangrelor | Intravenous P2Y12 inhibitor in P2Y12-naive patients | 30 mcg/kg bolus before PCI, then 4 mcg/kg/min for at least 2 hours or the duration of PCI, whichever is longer | Transition: ticagrelor 180 mg during or after the infusion; prasugrel 60 mg or clopidogrel 600 mg immediately after the infusion stops |
| Bivalirudin | Procedural anticoagulant; heparin-induced thrombocytopenia | 0.75 mg/kg bolus, then 1.75 mg/kg/h for the procedure; activated clotting time 5 minutes after the bolus, extra 0.3 mg/kg bolus if needed | Infusion up to 4 hours after PCI considered in ST-elevation myocardial infarction; reduced infusion below creatinine clearance 30 mL/min and on dialysis |
| Eptifibatide | Glycoprotein IIb/IIIa inhibitor, bailout for thrombus or no-reflow | 180 mcg/kg bolus, second 180 mcg/kg bolus at 10 minutes, then 2 mcg/kg/min | Infusion 1 mcg/kg/min below creatinine clearance 50 mL/min |
| Tirofiban | Glycoprotein IIb/IIIa inhibitor | 25 mcg/kg within 5 minutes, then 0.15 mcg/kg/min for up to 18 hours | Infusion 0.075 mcg/kg/min at creatinine clearance 60 mL/min or less |
Antiplatelet and anticoagulant drugs used at PCI with FDA-labeled dosing. Choice of P2Y12 inhibitor, pretreatment, and dual antiplatelet duration are covered on the dual antiplatelet therapy hub; heparin dosing is on the parenteral anticoagulant hub.
Cangrelor reduced periprocedural ischemic events in CHAMPION PHOENIX (4.7% against 5.9% with clopidogrel), including stent thrombosis (0.8% against 1.4%), without a significant increase in severe bleeding (Bhatt 2013). It provides rapid platelet inhibition in patients who have not received an oral P2Y12 inhibitor or cannot absorb one, such as intubated patients after cardiac arrest. Glycoprotein IIb/IIIa inhibitors are reserved for large thrombus burden, no-reflow, or thrombotic complications in patients already receiving dual antiplatelet therapy. Antithrombotic management in patients taking oral anticoagulants is on the atrial fibrillation and PCI hub.
Thrombus aspiration
Manual aspiration thrombectomy was tested as a routine step in primary PCI. TASTE randomized 7244 patients with ST-elevation myocardial infarction; 30-day mortality was 2.8% with aspiration and 3.0% without (Fröbert 2013). TOTAL randomized 10,732 patients, and the primary analysis included 10,063; the 180-day composite of cardiovascular death, recurrent myocardial infarction, cardiogenic shock, or class IV heart failure was 6.9% against 7.0%, and 30-day stroke was higher with thrombectomy (0.7% against 0.3%) (Jolly 2015). Aspiration is therefore limited to selective bailout use for large residual thrombus that prevents flow or visualization after initial balloon dilation. The guideline position is summarized on the ST-elevation myocardial infarction hub.
Complications of PCI
| Complication | Mechanism | Recognition | Immediate management |
|---|---|---|---|
| Abrupt closure | Dissection, thrombus, spasm, side-branch or plaque shift | Loss of flow, chest pain, ST elevation, hypotension | Wire position maintained; nitroglycerin for spasm; stenting of dissection; antithrombotic escalation for thrombus |
| Coronary dissection | Balloon or stent edge injury, guide or wire trauma | Linear lucency, contrast staining, or spiral filling defect (NHLBI types A to F) | Stenting of flow-limiting or extensive dissection; observation of small non-flow-limiting edge tears |
| Coronary perforation | Wire exit, oversized balloon, rupture at high pressure, atherectomy | Contrast extravasation, pericardial staining, chest pain, hypotension | Balloon tamponade; echocardiography; pericardiocentesis for tamponade; covered stent or embolization; reversal of heparin; surgery |
| No-reflow | Distal embolization, microvascular spasm and obstruction | Reduced TIMI flow with a patent epicardial artery | Exclusion of mechanical cause; intracoronary vasodilators delivered distally; hemodynamic support |
| Side-branch occlusion | Plaque or carina shift | Loss of branch filling | Rewiring through struts and balloon dilation; protection wire in branches at risk |
| Stent thrombosis | Mechanical stent problems, antiplatelet failure | Abrupt occlusion with thrombus | Emergency PCI; imaging to correct mechanism; review of antiplatelet therapy |
| Stent loss and device entrapment | Forced delivery, poor coaxial alignment | Stent visible off its balloon; device will not withdraw | Small-balloon or snare retrieval; crushing or deployment in place |
| Air embolism | Air in the manifold, syringe, or device | Sudden ischemia with a contrast-free segment | Oxygen, aspiration, forceful saline flush, support |
| Aortocoronary dissection | Guide trauma to the ostium | Contrast staining of the sinus or ascending aorta | Ostial stenting; surgery for extension |
| Periprocedural myocardial infarction | Side-branch loss, distal embolization, dissection | Troponin elevation with ischemic evidence | Treatment directed at the cause; prognostic significance depends on the biomarker threshold |
| Contrast-associated kidney injury | Contrast, hypotension, atheroembolism | Creatinine rise at 48–72 hours | Contrast limitation and volume management |
| Stroke | Embolism from aortic arch or catheters, hemorrhage on antithrombotic therapy | New neurologic deficit | Immediate imaging and neurologic consultation; thrombectomy for large-vessel occlusion |
| Cholesterol embolization | Catheter disruption of aortic plaque | Livedo, blue toes, eosinophilia, kidney injury days to weeks later | Supportive care |
Coronary and systemic complications of PCI by mechanism. Access-site complications are listed in the access section. NHLBI, National Heart, Lung, and Blood Institute.
Periprocedural myocardial infarction
Troponin rises after many uncomplicated elective PCIs. By the Fourth Universal Definition of Myocardial Infarction, PCI-related (type 4a) myocardial infarction in patients with chronic coronary syndrome requires a rise in cardiac troponin of more than 5 times the 99th percentile upper reference limit. The rise must be accompanied by new ischemic evidence on ECG, imaging, or angiography showing a flow-limiting complication (Mair 2022). The SCAI expert consensus proposed a separate "clinically relevant" definition for use in trials, patient care, and quality assessment, built on a biomarker threshold strongly linked to subsequent adverse events in clinical studies (Moussa 2013). As an isolated biomarker criterion, that threshold is a rise in creatine kinase MB isoenzyme (CK-MB) above 10 times the upper limit of normal or a rise in troponin above 70 times that limit. In ISCHEMIA, the incidence of the primary outcome was sensitive to the definition of myocardial infarction, and a secondary definition yielded more procedural infarctions of uncertain clinical importance (Maron 2020).
No-reflow and slow flow
No-reflow is reduced antegrade flow (TIMI 0 to 2) after the epicardial obstruction has been relieved, caused by distal embolization of thrombus and plaque, microvascular spasm, and reperfusion injury. It is most frequent in primary PCI with heavy thrombus, degenerated saphenous vein grafts, and atherectomy. Before microvascular treatment, mechanical causes are excluded: guide damping, dissection, thrombus, spasm, air, and residual stenosis. Intracoronary vasodilators (adenosine, nitroprusside, verapamil or another calcium channel blocker, and epinephrine in hypotension) are delivered distally through a microcatheter or aspiration catheter, and blood pressure is supported to maintain microvascular perfusion pressure. Embolic protection in saphenous vein graft PCI has a Class 2a recommendation in the 2021 guideline (Lawton 2022).
Coronary perforation
The Ellis classification grades perforation as type I, an extraluminal crater without extravasation; type II, pericardial or myocardial blushing; type III, a perforation of 1 mm or more with contrast streaming; and type III cavity spilling (Ellis 1994), extravasation into a cardiac chamber, an anatomic cavity, or the coronary sinus or a coronary vein, most often the left ventricle (Majeed 2023). In the original series, perforation occurred in 0.5% of procedures and was more common with atherectomy and laser devices than with balloon angioplasty. In the NCDR CathPCI Registry, perforation complicated 0.37% of 3,759,268 PCIs. Chronic total occlusion PCI, atherectomy, saphenous vein graft PCI, ACC type C lesions, female sex, and cardiogenic shock predicted it, and patients with perforation had higher rates of tamponade (9.6% against 0.05%) and in-hospital death (4.87% against 1.14%) (Nairooz 2020). Angiographic grading and images are on the angiography hub.
Management begins with a balloon inflated at low pressure proximal to or across the perforation to stop bleeding, urgent echocardiography, and pericardiocentesis for tamponade. Large-vessel perforation is sealed with a covered stent delivered through a guide large enough for it, often with a second guide (the dual-guide or "ping-pong" technique) so that the occluding balloon remains inflated while the covered stent is positioned. Distal wire perforation is sealed by embolization with coils, fat, thrombin, or other material through a microcatheter. Heparin is reversed with protamine once the need for further equipment has been decided, and glycoprotein IIb/IIIa infusions are stopped. Surgery is required when percutaneous sealing fails. Tamponade can appear hours after an apparently sealed perforation, so echocardiography is repeated.
Dissection and abrupt closure
Dissection after balloon dilation is graded by the National Heart, Lung, and Blood Institute classification, from type A intimal lucency to type F total occlusion (dissection grades). Flow-limiting or extensive dissections are stented while wire position is maintained. Edge dissections seen only on intravascular imaging are treated when they are large, extend into the media, or reduce lumen area at the edge. Abrupt closure from dissection, thrombus, or spasm is treated by the mechanism; loss of wire position across a dissection makes rewiring the true lumen difficult, which is why the wire is never withdrawn from a dissected vessel until the result is secure.
Contrast-associated kidney injury, radiation, stroke, and emergency surgery
Kidney injury after PCI is prevented by contrast limitation, volume management guided by hemodynamics, and avoidance of hypotension; risk scores, contrast volume limits, and the evidence for hydration are on the contrast hub. Prolonged complex procedures can deliver radiation doses that cause skin injury. Dose monitoring, collimation, low frame rates, table and detector positioning, and changing projections limit patient and operator exposure. The principles of radiation protection are described on the radiation safety hub. Stroke after PCI arises from catheter-related embolism or from hemorrhage on antithrombotic therapy and is evaluated immediately with brain imaging. Emergency bypass surgery after PCI has become rare with stenting and is reserved for failed treatment of perforation, extensive dissection with ischemia, or entrapped equipment.
Mechanical circulatory support in high-risk PCI
High-risk PCI refers to intervention in patients with severely reduced ventricular function, complex multivessel or unprotected left main disease, or a last remaining vessel, in whom transient ischemia during balloon inflation or a complication could cause hemodynamic collapse. PROTECT II randomized 452 such patients to Impella 2.5 or intra-aortic balloon pump support. The Impella provided greater hemodynamic support, but 30-day major adverse events did not differ significantly (35.1% against 40.1%) (O'Neill 2012). Routine prophylactic support has not been shown to improve outcomes, and device choice weighs the jeopardized myocardium, ventricular function, the procedure plan, and access risk. Device physiology, the evidence for protected PCI, and complications of large-bore access are on the temporary mechanical support hub and the Impella hub.
PCI at centers without on-site cardiac surgery
The need for emergency surgery after PCI fell sharply with stenting, which led to trials of PCI at hospitals without cardiac surgery. CPORT-E randomized 18,867 patients in a 3:1 ratio to PCI at hospitals with or without on-site surgery; 6-week mortality was 0.9% without and 1.0% with on-site surgery, and 9-month major adverse cardiac events were 12.1% without and 11.2% with on-site surgery. Both end points met noninferiority, although target-vessel revascularization was higher at hospitals without on-site surgery (6.5% against 5.4%) (Aversano 2012). Program standards for PCI without on-site surgical backup are set out in the SCAI/ACC/AHA expert consensus document (Dehmer 2014).
Same-day discharge after PCI
In a meta-analysis of 37 studies with 12,803 patients, mostly undergoing elective PCI, randomized trials showed no difference between same-day discharge and overnight observation in death, myocardial infarction, or target-lesion revascularization (odds ratio [OR] 0.90) or in major bleeding or vascular complications (OR 1.69, not significant). In observational studies of same-day discharge, the pooled rate of death, myocardial infarction, or target-lesion revascularization was 1.00%, and that of major bleeding or vascular complications was 0.68% (Brayton 2013).
The 2021 ACC expert consensus decision pathway provides a checklist of clinical, social, and facility or system factors to determine whether a patient can be considered for same-day discharge (Rao 2021). The pathway notes that same-day discharge in selected adults has not been accompanied by increases in complications, mortality, or readmissions. Pre-discharge processes include confirmation that the patient has received a P2Y12 inhibitor prescription, instructions on monitoring the access site, and confirmed outpatient follow-up. The checklist can begin before the procedure or be completed afterward, is adapted to institutional protocols, and depends on shared decision-making with the patient.
Care after PCI: medications, follow-up, and cardiac rehabilitation
After PCI, dual antiplatelet therapy continues for the period set by presentation and bleeding risk, with a single antiplatelet drug long term (DAPT duration). Patients who require oral anticoagulation follow the regimens on the atrial fibrillation and PCI hub. Premature discontinuation of antiplatelet therapy was the strongest independent predictor of stent thrombosis after drug-eluting stent implantation in a 9-month registry (hazard ratio 89.78) (Iakovou 2005); elective noncardiac surgery is timed accordingly (surgery on DAPT). High-intensity statin therapy and risk-factor control for secondary prevention are covered on the chronic coronary disease and dyslipidemia hubs.
Cardiac rehabilitation is indicated after PCI, and the content of rehabilitation after PCI and surgery is on the cardiac rehabilitation hub. Recurrent angina after PCI is evaluated for restenosis, progression of disease outside the treated segment, residual untreated lesions, and vasomotor or microvascular causes (recurrent angina).
Sources of error in PCI equipment and technique
- Treating an angiographically intermediate lesion without physiology. Stenting a lesion without ischemic significance adds the risks of an implant with no benefit, and FFR guidance reduced stents and events in FAME.
- Injecting through a damped or ventricularized guide. A wedged catheter tip turns an injection into a dissection of the ostium or aortic root and causes ischemia; the guide is withdrawn until the waveform normalizes.
- Reading a normal pressure tracing through a side-hole guide as proof of unobstructed engagement. Side holes mask wedging.
- Measuring FFR with the guide engaged and damped. A falsely low aortic reference raises the ratio and hides a significant lesion.
- Advancing a balloon over a wire whose position was checked in one projection. A wire in a side branch or subintimal track looks similar in a single view.
- Losing sight of the distal wire tip during device advancement, which permits distal perforation.
- Inflating beyond the rated burst pressure in a calcified lesion. Balloon rupture can dissect or perforate the artery.
- Deploying a stent across a lesion that a noncompliant balloon could not expand. Unprepared calcium causes underexpansion, which is the leading correctable cause of restenosis and stent thrombosis.
- Sizing a stent by eye in a diffusely diseased or tapered vessel. Imaging or quantitative angiography of the reference segments avoids undersizing and edge injury.
- Landing a stent edge in a plaque-heavy segment or leaving disease uncovered (geographic miss), which predisposes to edge dissection and edge restenosis.
- Forcing a stent through tortuosity, calcium, or a prior stent edge. Stent stripping and loss follow; preparation, buddy wires, anchors, and guide extensions are used first.
- Pushing a guide extension over a bare wire into a diseased vessel or injecting contrast through a deeply placed extension. Both cause dissection or perforation.
- Withdrawing the wire from a dissected vessel before the result is secure, which may make true-lumen rewiring impossible.
- Reading accordion pseudostenosis from a stiff wire as a dissection or residual lesion. The final angiogram is taken with the wire withdrawn from a tortuous vessel.
- Treating in-stent restenosis without defining its mechanism. Underexpansion requires high-pressure dilation or calcium modification, and an additional stent seldom corrects it.
- Choosing a bare-metal stent because the patient can take only 1 month of dual antiplatelet therapy. Drug-eluting stents outperform bare-metal stents with that regimen.
- Extrapolating drug-coated balloon results in small vessels or in-stent restenosis to larger de novo arteries, where REC-CAGEFREE I did not show noninferiority to stenting.
- Routine aspiration thrombectomy in primary PCI. TASTE and TOTAL showed no mortality benefit, and TOTAL showed more strokes.
- Assuming that a sealed perforation is safe. Tamponade can develop hours later, and echocardiography is repeated.
- Giving protamine before deciding whether more coronary equipment will be needed, which risks thrombosis of wires and guides in the coronary artery.
- Treating no-reflow with vasodilators before excluding a mechanical cause such as dissection, thrombus, or guide damping.
- Labeling any post-PCI troponin rise as myocardial infarction. Type 4a infarction requires a rise above 5 times the upper reference limit with ischemic evidence.
- Femoral puncture above the inguinal ligament, which sets up retroperitoneal hemorrhage, or at or below the bifurcation, which sets up pseudoaneurysm.
- Occlusive radial compression after the procedure. Patent hemostasis preserves flow and reduces radial artery occlusion.
- Discharging a patient the same day without confirmation of the P2Y12 inhibitor prescription, access-site instructions, and follow-up.
References
- Lawton JS, Tamis-Holland JE, Bangalore S, et al. 2021 ACC/AHA/SCAI guideline for coronary artery revascularization: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022;145:e18–e114. doi:10.1161/CIR.0000000000001038
- Rao SV, Vidovich MI, Gilchrist IC, et al. 2021 ACC expert consensus decision pathway on same-day discharge after percutaneous coronary intervention: a report of the American College of Cardiology Solution Set Oversight Committee. J Am Coll Cardiol. 2021;77:811–825. doi:10.1016/j.jacc.2020.11.013
- Gruntzig A. Transluminal dilatation of coronary-artery stenosis. Lancet. 1978;1:263. doi:10.1016/s0140-6736(78)90500-7
- Sigwart U, Puel J, Mirkovitch V, Joffre F, Kappenberger L. Intravascular stents to prevent occlusion and restenosis after transluminal angioplasty. N Engl J Med. 1987;316:701–706. doi:10.1056/NEJM198703193161201
- Serruys PW, de Jaegere P, Kiemeneij F, et al. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease (Benestent). N Engl J Med. 1994;331:489–495. doi:10.1056/NEJM199408253310801
- Fischman DL, Leon MB, Baim DS, et al. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease (STRESS). N Engl J Med. 1994;331:496–501. doi:10.1056/NEJM199408253310802
- Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery (SIRIUS). N Engl J Med. 2003;349:1315–1323. doi:10.1056/NEJMoa035071
- Stone GW, Ellis SG, Cox DA, et al. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease (TAXUS IV). N Engl J Med. 2004;350:221–231. doi:10.1056/NEJMoa032441
- Stone GW, Rizvi A, Newman W, et al. Everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease (SPIRIT IV). N Engl J Med. 2010;362:1663–1674. doi:10.1056/NEJMoa0910496
- Kedhi E, Joesoef KS, McFadden E, et al. Second-generation everolimus-eluting and paclitaxel-eluting stents in real-life practice (COMPARE): a randomised trial. Lancet. 2010;375:201–209. doi:10.1016/S0140-6736(09)62127-9
- Kastrati A, Mehilli J, Dirschinger J, et al. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO) trial. Circulation. 2001;103:2816–2821. doi:10.1161/01.cir.103.23.2816
- Kandzari DE, Mauri L, Koolen JJ, et al. Ultrathin, bioresorbable polymer sirolimus-eluting stents versus thin, durable polymer everolimus-eluting stents in patients undergoing coronary revascularisation (BIOFLOW V): a randomised trial. Lancet. 2017;390:1843–1852. doi:10.1016/S0140-6736(17)32249-3
- Bønaa KH, Mannsverk J, Wiseth R, et al. Drug-eluting or bare-metal stents for coronary artery disease (NORSTENT). N Engl J Med. 2016;375:1242–1252. doi:10.1056/NEJMoa1607991
- Varenne O, Cook S, Sideris G, et al. Drug-eluting stents in elderly patients with coronary artery disease (SENIOR): a randomised single-blind trial. Lancet. 2018;391:41–50. doi:10.1016/S0140-6736(17)32713-7
- Valgimigli M, Patialiakas A, Thury A, et al. Zotarolimus-eluting versus bare-metal stents in uncertain drug-eluting stent candidates (ZEUS). J Am Coll Cardiol. 2015;65:805–815. doi:10.1016/j.jacc.2014.11.053
- Urban P, Meredith IT, Abizaid A, et al. Polymer-free drug-coated coronary stents in patients at high bleeding risk (LEADERS FREE). N Engl J Med. 2015;373:2038–2047. doi:10.1056/NEJMoa1503943
- Windecker S, Latib A, Kedhi E, et al. Polymer-based or polymer-free stents in patients at high bleeding risk (ONYX ONE). N Engl J Med. 2020;382:1208–1218. doi:10.1056/NEJMoa1910021
- Valgimigli M, Cao D, Angiolillo DJ, et al. Duration of dual antiplatelet therapy for patients at high bleeding risk undergoing PCI (XIENCE 90 and XIENCE 28). J Am Coll Cardiol. 2021;78:2060–2072. doi:10.1016/j.jacc.2021.08.074
- Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease (ABSORB III). N Engl J Med. 2015;373:1905–1915. doi:10.1056/NEJMoa1509038
- Kereiakes DJ, Ellis SG, Metzger DC, et al. Clinical outcomes before and after complete everolimus-eluting bioresorbable scaffold resorption: five-year follow-up from the ABSORB III trial. Circulation. 2019;140:1895–1903. doi:10.1161/CIRCULATIONAHA.119.042584
- Wykrzykowska JJ, Kraak RP, Hofma SH, et al. Bioresorbable scaffolds versus metallic stents in routine PCI (AIDA). N Engl J Med. 2017;376:2319–2328. doi:10.1056/NEJMoa1614954
- Cutlip DE, Windecker S, Mehran R, et al. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007;115:2344–2351. doi:10.1161/CIRCULATIONAHA.106.685313
- Garcia-Garcia HM, McFadden EP, Farb A, et al. Standardized end point definitions for coronary intervention trials: the Academic Research Consortium-2 consensus document. Circulation. 2018;137:2635–2650. doi:10.1161/CIRCULATIONAHA.117.029289
- Mehran R, Dangas G, Abizaid AS, et al. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome. Circulation. 1999;100:1872–1878. doi:10.1161/01.cir.100.18.1872
- Giustino G, Colombo A, Camaj A, et al. Coronary in-stent restenosis: JACC state-of-the-art review. J Am Coll Cardiol. 2022;80:348–372. doi:10.1016/j.jacc.2022.05.017
- Yeh RW, Shlofmitz R, Moses J, et al. Paclitaxel-coated balloon vs uncoated balloon for coronary in-stent restenosis: the AGENT IDE randomized clinical trial. JAMA. 2024;331:1015–1024. doi:10.1001/jama.2024.1361
- Giacoppo D, Alfonso F, Xu B, et al. Drug-coated balloon angioplasty versus drug-eluting stent implantation in patients with coronary stent restenosis (DAEDALUS). J Am Coll Cardiol. 2020;75:2664–2678. doi:10.1016/j.jacc.2020.04.006
- Byrne RA, Neumann FJ, Mehilli J, et al. Paclitaxel-eluting balloons, paclitaxel-eluting stents, and balloon angioplasty in patients with restenosis after implantation of a drug-eluting stent (ISAR-DESIRE 3): a randomised, open-label trial. Lancet. 2013;381:461–467. doi:10.1016/S0140-6736(12)61964-3
- Giacoppo D, Alvarez-Covarrubias HA, Koch T, et al. Coronary artery restenosis treatment with plain balloon, drug-coated balloon, or drug-eluting stent: 10-year outcomes of the ISAR-DESIRE 3 trial. Eur Heart J. 2023;44:1343–1357. doi:10.1093/eurheartj/ehad026
- Sethi A, Malhotra G, Singh S, et al. Efficacy of various percutaneous interventions for in-stent restenosis: comprehensive network meta-analysis of randomized controlled trials. Circ Cardiovasc Interv. 2015;8:e002778. doi:10.1161/CIRCINTERVENTIONS.115.002778
- Jeger RV, Farah A, Ohlow MA, et al. Drug-coated balloons for small coronary artery disease (BASKET-SMALL 2): an open-label randomised non-inferiority trial. Lancet. 2018;392:849–856. doi:10.1016/S0140-6736(18)31719-7
- Gao C, He X, Ouyang F, et al. Drug-coated balloon angioplasty with rescue stenting versus intended stenting for the treatment of patients with de novo coronary artery lesions (REC-CAGEFREE I): an open-label, randomised, non-inferiority trial. Lancet. 2024;404:1040–1050. doi:10.1016/S0140-6736(24)01594-0
- Jeger RV, Eccleshall S, Wan Ahmad WA, et al. Drug-coated balloons for coronary artery disease: third report of the International DCB Consensus Group. JACC Cardiovasc Interv. 2020;13:1391–1402. doi:10.1016/j.jcin.2020.02.043
- US Food and Drug Administration. AGENT paclitaxel-coated balloon catheter, premarket approval P230035. 2024. accessdata.fda.gov
- US Food and Drug Administration. Class II special controls guidance document: certain percutaneous transluminal coronary angioplasty (PTCA) catheters. 2010. fda.gov
- Lee Chuy K, Williams SO, Prasad VS, Latif F. Maximizing guide catheter support in percutaneous coronary interventions. Society for Cardiovascular Angiography & Interventions quality improvement tip. 2024. scai.org
- Frizzell JD. Guide catheter extensions: a brief, practical user's guide. Cardiac Interventions Today. 2026;May–June. citoday.com
- Ali ZA, Landmesser U, Maehara A, et al. Optical coherence tomography-guided versus angiography-guided PCI (ILUMIEN IV). N Engl J Med. 2023;389:1466–1476. doi:10.1056/NEJMoa2305861
- Lee JM, Choi KH, Song YB, et al. Intravascular imaging-guided or angiography-guided complex PCI (RENOVATE-COMPLEX-PCI). N Engl J Med. 2023;388:1668–1679. doi:10.1056/NEJMoa2216607
- Holm NR, Andreasen LD, Neghabat O, et al. OCT or angiography guidance for PCI in complex bifurcation lesions (OCTOBER). N Engl J Med. 2023;389:1477–1487. doi:10.1056/NEJMoa2307770
- Stone GW, Christiansen EH, Ali ZA, et al. Intravascular imaging-guided coronary drug-eluting stent implantation: an updated network meta-analysis. Lancet. 2024;403:824–837. doi:10.1016/S0140-6736(23)02454-6
- Hong SJ, Kim BK, Shin DH, et al. Effect of intravascular ultrasound-guided vs angiography-guided everolimus-eluting stent implantation: the IVUS-XPL randomized clinical trial. JAMA. 2015;314:2155–2163. doi:10.1001/jama.2015.15454
- Tonino PA, De Bruyne B, Pijls NH, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention (FAME). N Engl J Med. 2009;360:213–224. doi:10.1056/NEJMoa0807611
- Davies JE, Sen S, Dehbi HM, et al. Use of the instantaneous wave-free ratio or fractional flow reserve in PCI (DEFINE-FLAIR). N Engl J Med. 2017;376:1824–1834. doi:10.1056/NEJMoa1700445
- Boden WE, O'Rourke RA, Teo KK, et al. Optimal medical therapy with or without PCI for stable coronary disease (COURAGE). N Engl J Med. 2007;356:1503–1516. doi:10.1056/NEJMoa070829
- Maron DJ, Hochman JS, Reynolds HR, et al. Initial invasive or conservative strategy for stable coronary disease (ISCHEMIA). N Engl J Med. 2020;382:1395–1407. doi:10.1056/NEJMoa1915922
- Al-Lamee R, Thompson D, Dehbi HM, et al. Percutaneous coronary intervention in stable angina (ORBITA): a double-blind, randomised controlled trial. Lancet. 2018;391:31–40. doi:10.1016/S0140-6736(17)32714-9
- Patel MR, Calhoon JH, Dehmer GJ, et al. ACC/AATS/AHA/ASE/ASNC/SCAI/SCCT/STS 2017 appropriate use criteria for coronary revascularization in patients with stable ischemic heart disease. J Nucl Cardiol. 2017;24:1759–1792. doi:10.1007/s12350-017-0917-9
- Desai NR, Bradley SM, Parzynski CS, et al. Appropriate use criteria for coronary revascularization and trends in utilization, patient selection, and appropriateness of percutaneous coronary intervention. JAMA. 2015;314:2045–2053. doi:10.1001/jama.2015.13764
- Jolly SS, Yusuf S, Cairns J, et al. Radial versus femoral access for coronary angiography and intervention in patients with acute coronary syndromes (RIVAL): a randomised, parallel group, multicentre trial. Lancet. 2011;377:1409–1420. doi:10.1016/S0140-6736(11)60404-2
- Valgimigli M, Gagnor A, Calabró P, et al. Radial versus femoral access in patients with acute coronary syndromes undergoing invasive management (MATRIX): a randomised multicentre trial. Lancet. 2015;385:2465–2476. doi:10.1016/S0140-6736(15)60292-6
- Romagnoli E, Biondi-Zoccai G, Sciahbasi A, et al. Radial versus femoral randomized investigation in ST-segment elevation acute coronary syndrome: the RIFLE-STEACS study. J Am Coll Cardiol. 2012;60:2481–2489. doi:10.1016/j.jacc.2012.06.017
- Rao SV, Hess CN, Barham B, et al. A registry-based randomized trial comparing radial and femoral approaches in women undergoing percutaneous coronary intervention: the SAFE-PCI for Women trial. JACC Cardiovasc Interv. 2014;7:857–867. doi:10.1016/j.jcin.2014.04.007
- Aminian A, Sgueglia GA, Wiemer M, et al. Distal versus conventional radial access for coronary angiography and intervention: the DISCO RADIAL trial. JACC Cardiovasc Interv. 2022;15:1191–1201. doi:10.1016/j.jcin.2022.04.032
- Pancholy SB, Bernat I, Bertrand OF, Patel TM. Prevention of radial artery occlusion after transradial catheterization: the PROPHET-II randomized trial. JACC Cardiovasc Interv. 2016;9:1992–1999. doi:10.1016/j.jcin.2016.07.020
- Schulz-Schüpke S, Helde S, Gewalt S, et al. Comparison of vascular closure devices vs manual compression after femoral artery puncture: the ISAR-CLOSURE randomized clinical trial. JAMA. 2014;312:1981–1987. doi:10.1001/jama.2014.15305
- Bhatt DL, Stone GW, Mahaffey KW, et al. Effect of platelet inhibition with cangrelor during PCI on ischemic events (CHAMPION PHOENIX). N Engl J Med. 2013;368:1303–1313. doi:10.1056/NEJMoa1300815
- Erlinge D, Omerovic E, Fröbert O, et al. Bivalirudin versus heparin monotherapy in myocardial infarction (VALIDATE-SWEDEHEART). N Engl J Med. 2017;377:1132–1142. doi:10.1056/NEJMoa1706443
- Fröbert O, Lagerqvist B, Olivecrona GK, et al. Thrombus aspiration during ST-segment elevation myocardial infarction (TASTE). N Engl J Med. 2013;369:1587–1597. doi:10.1056/NEJMoa1308789
- Jolly SS, Cairns JA, Yusuf S, et al. Randomized trial of primary PCI with or without routine manual thrombectomy (TOTAL). N Engl J Med. 2015;372:1389–1398. doi:10.1056/NEJMoa1415098
- Mair J, Jaffe A, Lindahl B, et al. The clinical approach to diagnosing peri-procedural myocardial infarction after percutaneous coronary interventions according to the fourth universal definition of myocardial infarction. Biomarkers. 2022;27:407–417. doi:10.1080/1354750X.2022.2055792
- Moussa ID, Klein LW, Shah B, et al. Consideration of a new definition of clinically relevant myocardial infarction after coronary revascularization: an expert consensus document from the Society for Cardiovascular Angiography and Interventions (SCAI). J Am Coll Cardiol. 2013;62:1563–1570. doi:10.1016/j.jacc.2013.08.720
- Ellis SG, Ajluni S, Arnold AZ, et al. Increased coronary perforation in the new device era: incidence, classification, management, and outcome. Circulation. 1994;90:2725–2730. doi:10.1161/01.cir.90.6.2725
- Nairooz R, Parzynski CS, Curtis JP, et al. Contemporary trends, predictors and outcomes of perforation during percutaneous coronary intervention (from the NCDR CathPCI Registry). Am J Cardiol. 2020;130:37–45. doi:10.1016/j.amjcard.2020.06.014
- O'Neill WW, Kleiman NS, Moses J, et al. A prospective, randomized clinical trial of hemodynamic support with Impella 2.5 versus intra-aortic balloon pump in patients undergoing high-risk percutaneous coronary intervention: the PROTECT II study. Circulation. 2012;126:1717–1727. doi:10.1161/CIRCULATIONAHA.112.098194
- Aversano T, Lemmon CC, Liu L. Outcomes of PCI at hospitals with or without on-site cardiac surgery (CPORT-E). N Engl J Med. 2012;366:1792–1802. doi:10.1056/NEJMoa1114540
- Dehmer GJ, Blankenship JC, Cilingiroglu M, et al. SCAI/ACC/AHA expert consensus document: 2014 update on percutaneous coronary intervention without on-site surgical backup. J Am Coll Cardiol. 2014;63:2624–2641. doi:10.1016/j.jacc.2014.03.002
- Brayton KM, Patel VG, Stave C, de Lemos JA, Kumbhani DJ. Same-day discharge after percutaneous coronary intervention: a meta-analysis. J Am Coll Cardiol. 2013;62:275–285. doi:10.1016/j.jacc.2013.03.051
- Rao SV, O’Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2025;151:e771–e862. doi:10.1161/CIR.0000000000001309
- Thiele H, Akin I, Sandri M, et al. PCI strategies in patients with acute myocardial infarction and cardiogenic shock. N Engl J Med. 2017;377:2419–2432. doi:10.1056/NEJMoa1710261
- Majeed H, Blankenship JC. Ellis type 4 coronary artery perforation during percutaneous coronary intervention: case series and review. Catheter Cardiovasc Interv. 2023;102:1252–1258. doi:10.1002/ccd.30873
- Gwon HC, Doh JH, Choi JH, et al. A 5Fr catheter approach reduces patient discomfort during transradial coronary intervention compared with a 6Fr approach: a prospective randomized study. J Interv Cardiol. 2006;19:141–147. doi:10.1111/j.1540-8183.2006.00121.x
- Barbato E, Carrié D, Dardas P, et al. European expert consensus on rotational atherectomy. EuroIntervention. 2015;11:30–6. doi:10.4244/EIJV11I1A6
- Iakovou I, Schmidt T, Bonizzoni E, et al. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA. 2005;293:2126–2130. doi:10.1001/jama.293.17.2126