CardioBoardExam.com

Library › Imaging › Transesophageal echocardiography

Transesophageal echocardiography

Transesophageal echocardiography in adults: indications, preparation, and standard views; appendage, septal, valve, aortic, and procedural imaging; complications; and training.

Transesophageal echocardiography (TEE) images the heart from the esophagus and stomach, where only the esophageal wall separates the transducer from the left atrium. It is the standard test for left atrial appendage thrombus, prosthetic valve and perivalvular complications, the interatrial septum, the thoracic aorta, and procedural guidance. Before elective cardioversion of atrial fibrillation lasting 48 hours or longer, the 2023 American College of Cardiology, American Heart Association, American College of Clinical Pharmacy, and Heart Rhythm Society (ACC/AHA/ACCP/HRS) guideline recommends 3 weeks of uninterrupted therapeutic anticoagulation or imaging to exclude thrombus (Class 1). Anticoagulation continues for at least 4 weeks afterward (Class 1). Serious complications are rare: mortality was 0.0098% in a survey of 10,419 diagnostic studies and morbidity 0.2% in 7,200 intraoperative studies.

Principles of transesophageal imaging and its advantages over transthoracic echocardiography

Transthoracic echocardiography (TTE) reaches the heart through the chest wall, ribs, and lung, and the posterior structures lie in its far field, where resolution is lowest and attenuation highest. The esophagus runs immediately behind the left atrium, so a transducer placed there has the left atrium, the left atrial appendage, the interatrial septum, the mitral valve, the pulmonary veins, and the descending aorta in its near field. The short acoustic path allows higher transmit frequencies, and higher frequency gives finer axial and lateral resolution. The same physics explains the weaknesses of TEE: structures far from the esophagus, such as the left ventricular apex, lie in its far field and are often foreshortened, and air in the trachea and main bronchus hides a short segment of the distal ascending aorta and proximal arch.

The transducer is a phased array mounted at the tip of a flexible gastroscope-like shaft. The imaging plane can be rotated electronically from 0° to 180° without moving the probe (multiplane imaging), and matrix-array probes acquire live three-dimensional volumes and two simultaneous orthogonal planes (biplane imaging). Piezoelectric elements lose their properties when heated near their Curie point, so TEE probes are disinfected chemically and are never heat-sterilized. A thermistor in the tip reports transducer temperature on the display and stops imaging when the tip overheats.

TEE is a semi-invasive procedure. It requires screening for esophageal disease, fasting, usually moderate sedation or anesthesia, and monitoring, and its complications range from sore throat to esophageal perforation. TTE therefore remains the first test for most questions, including ventricular size and function, valve gradients, and pulmonary pressure. TEE is chosen when a defined anatomic question requires esophageal resolution or concerns a structure that TTE images poorly. A nondiagnostic transthoracic study leaves the question open and is not reported as negative. Stress imaging is performed transthoracically, as described on the stress echocardiography page. Transthoracic chamber quantification, normal values, and the standard surface views are described on the transthoracic echocardiography and normal values page.

Indications for transesophageal echocardiography

The 2013 American Society of Echocardiography and Society of Cardiovascular Anesthesiologists (ASE/SCA) guideline groups indications into four categories. The first is evaluation of cardiac and aortic structure when the findings will alter management and TTE is nondiagnostic or likely to be. The other three are intraoperative TEE, guidance of transcatheter procedures, and critically ill patients in whom TTE cannot provide information expected to change management. Within the first category it lists structures in the transthoracic far field (the aorta and the left atrial appendage), prosthetic valves, paravalvular abscess, ventilated patients, chest wall injury, body habitus that prevents adequate imaging, and inability to turn to the left lateral position.

RatingClinical scenario
AppropriateHigh likelihood of a nondiagnostic TTE because of patient characteristics or inadequate visualization of the relevant structures
AppropriateRe-evaluation of a prior TEE finding for interval change when a change in therapy is anticipated (thrombus resolution after anticoagulation, vegetation after antibiotics)
AppropriateGuidance of percutaneous noncoronary interventions: closure devices, ablation, and percutaneous valve procedures
AppropriateSuspected acute aortic pathology, including dissection and transection
AppropriateValve structure and function to determine suitability for, and to plan, an intervention
AppropriateInfective endocarditis with moderate or high pretest probability (Staphylococcus bacteremia, fungemia, prosthetic valve, intracardiac device)
AppropriateCardiovascular source of embolus with no identified noncardiac source
AppropriateAtrial fibrillation or flutter: decisions on anticoagulation, cardioversion, or ablation
UncertainCardiovascular source of embolus with a previously identified noncardiac source
InappropriateRoutine TEE when a diagnostic TTE is reasonably anticipated to resolve all diagnostic and management concerns
InappropriateSurveillance of a prior TEE finding when no change in therapy is anticipated
InappropriateRoutine assessment of the pulmonary veins in an asymptomatic patient after pulmonary vein isolation
InappropriateInfective endocarditis with low pretest probability (transient fever, known alternative source, negative cultures, or an atypical organism)
InappropriateAtrial fibrillation or flutter when the decision has been made to anticoagulate and not to perform cardioversion

Appropriate use ratings for TEE as an initial or supplemental test, from the 2011 American College of Cardiology Foundation, American Society of Echocardiography, and American Heart Association appropriate use criteria as summarized in Table 5 of the 2013 ASE/SCA guideline.

Several class-based recommendations in ACC/AHA guidelines rest on TEE. In infective endocarditis, TEE is recommended when TTE is nondiagnostic, when complications are suspected, and when intracardiac leads are present (Class 1). In primary mitral regurgitation, TEE is recommended when TTE gives insufficient or discordant information on severity, mechanism, or ventricular function, and intraoperative TEE is recommended in patients undergoing intervention (Class 1). In atrial fibrillation, imaging to exclude intracardiac thrombus is the alternative to 3 weeks of anticoagulation before elective cardioversion (Class 1). During surgical exclusion of the left atrial appendage, intraoperative TEE confirms the absence of flow across the suture line and a stump shorter than 1 cm (Class 1). In suspected acute aortic syndrome, TEE is a reasonable alternative to CT angiography (Class 2a).

Choosing among transthoracic, transesophageal, and intracardiac echocardiography, CT, and MRI

The choice of test follows the structure in question and the decision it informs. An optimized transthoracic study, with an ultrasound-enhancing agent where appropriate, answers many questions for which TEE is requested by habit. CT supplies annular, coronary, access, and whole-aortic geometry; cardiac magnetic resonance supplies volumes, flow, and tissue characterization; intracardiac echocardiography (ICE) permits procedural imaging without general anesthesia.

QuestionPreferred testRole of TEELimitations of TEE
Left atrial appendage thrombusTEE; delayed-phase CT as an alternativeMultiplane and three-dimensional imaging, emptying velocity, spontaneous echo contrastPectinate muscles and the warfarin ridge mimic thrombus; dense smoke obscures the lumen
Left ventricular apical thrombusTTE with an enhancing agent; cardiac MRILimitedApex is in the far field and foreshortened
Native valve vegetation and abscessTEE after TTESmall vegetations, abscess, pseudoaneurysm, fistula, perforationEarly abscess may appear only as thickening; repeat imaging when suspicion remains
Prosthetic valve dysfunctionTTE for hemodynamics; TEE for mechanism; CT and fluoroscopy for leaflet motion and pannusMitral and tricuspid prostheses from the atrial side; paravalvular leak mappingThe anterior aspect of an aortic prosthesis is shadowed from the esophagus
Interatrial septum and PFOTTE with agitated saline to screen; TEE for anatomyTunnel length, rims, septal aneurysm, sinus venosus defectsSedation weakens the Valsalva maneuver
Acute aortic syndromeCT angiography (Class 1)Reasonable alternative (Class 2a); bedside and operating-room diagnosis; mechanism of aortic regurgitationTracheal blind spot; abdominal aorta not seen; reverberation can mimic a flap
Aortic arch atheromaTEE; CT or MRI for the blind spotPlaque thickness, ulceration, mobile componentsDistal ascending aorta and proximal arch partly hidden
Structural intervention guidanceTEE for most left-sided procedures; ICE for septal, appendage, and ablation procedures; CT before TAVRReal-time device guidance and complication detectionUsually requires deep sedation or general anesthesia; prolonged manipulation injures the esophagus

Modality selection for common TEE questions, after the 2016 ASE cardiac source of embolism guideline, the 2022 ACC/AHA aortic disease guideline, and the 2022 ASE standards for structural heart screening. PFO, patent foramen ovale; ICE, intracardiac echocardiography; TAVR, transcatheter aortic valve replacement.

Contraindications and gastrointestinal risk assessment

The principal hazard of TEE is mechanical injury to the oropharynx, esophagus, or stomach. Screening before the procedure covers dysphagia and odynophagia, esophageal or gastric disease and surgery, upper gastrointestinal bleeding, radiation to the neck or mediastinum, cervical spine disease that limits neck motion, liver disease with varices, and bleeding risk from coagulopathy, thrombocytopenia, or anticoagulant therapy. Absolute contraindications preclude passing the probe. Relative contraindications call for an individualized decision in which the expected benefit is weighed against the risk.

Absolute contraindicationsRelative contraindications
Perforated viscusHistory of radiation to the neck and mediastinum
Esophageal strictureHistory of gastrointestinal surgery
Esophageal tumorRecent upper gastrointestinal bleeding
Esophageal perforation or lacerationBarrett esophagus
Esophageal diverticulumHistory of dysphagia
Active upper gastrointestinal bleedingRestricted neck mobility (severe cervical arthritis, atlantoaxial joint disease)
Symptomatic hiatal hernia
Esophageal varices
Coagulopathy or thrombocytopenia
Active esophagitis
Active peptic ulcer disease

Contraindications to TEE, from Table 6 of the 2013 ASE/SCA guideline (modified from a 2010 ASE safety review). Radiation, scleroderma of the esophagus, and achalasia weaken esophageal tissue, and their severity determines whether they are absolute or relative contraindications.

When gastrointestinal symptoms or pathology raise doubt, the ASE/SCA guideline strongly recommends gastroenterology clearance before the study. For patients with relative contraindications it lists precautions such as not advancing the probe beyond the mid-esophagus and limiting manipulation. The 2010 American Society of Anesthesiologists (ASA) and SCA practice guideline allows TEE in patients with oral, esophageal, or gastric disease when the expected benefit outweighs the risk and appropriate precautions are applied. The precautions are consideration of other imaging (such as epicardial echocardiography), gastroenterology consultation, a smaller probe, a limited examination, avoidance of unnecessary manipulation, and the most experienced operator.

Esophageal varices. Varices are a relative contraindication. In a 2022 meta-analysis of 10 studies including 908 patients with gastroesophageal varices, gastrointestinal bleeding after TEE occurred in 0.84% (0.68% with intraoperative and 1.03% with diagnostic TEE), with no perforation or death, and the bleeding risk did not differ from that in patients without varices. Most included studies had moderate or high risk of bias, so the result supports a careful, limited examination when the indication is strong.

Anticoagulation. Therapeutic anticoagulation is not by itself a contraindication; most TEE studies before cardioversion are performed in patients receiving it. Coagulopathy and thrombocytopenia are relative contraindications, and correctable factors are addressed before an elective study.

Patient preparation, fasting, consent, and sedation

Fasting. The 2013 ASE/SCA guideline advises patients undergoing TEE with sedation to abstain from food and from beverages other than clear liquids for at least 6 hours and from all intake for 3 hours before the procedure. Patients with delayed gastric emptying or other aspiration risks may need a longer fast or a prokinetic agent such as metoclopramide, and routine medications are taken with a small sip of water. Glucagon-like peptide-1 receptor agonists slow gastric emptying. The 2024 multisociety guidance (American Gastroenterological Association, American Society for Metabolic and Bariatric Surgery, ASA, International Society of Perioperative Care of Patients with Obesity, and Society of American Gastrointestinal and Endoscopic Surgeons) allows most patients to continue these drugs. It recommends a liquid diet for 24 hours before the procedure in patients at higher risk of retained gastric contents; the risk categories are described on the GLP-1 receptor agonists page.

Assessment and consent. The preprocedural evaluation records the indication, the gastrointestinal screen, allergies, current medications including anticoagulants and sedatives, alcohol and drug use, and the ASA physical status class. Telemetry in addition to standard monitoring is advised for ASA class 3 or higher. The airway examination includes the Mallampati class, mouth opening, and neck mobility; a jaw-to-hyoid distance of less than three fingerbreadths in obese patients or those with limited neck mobility predicts difficult mask ventilation or intubation. Informed consent covers sore throat, dental and oral injury, bleeding, perforation, aspiration, arrhythmia, and sedation-related respiratory depression.

Intravenous access. A 20-gauge cannula is usually adequate. The left arm is recommended when an agitated saline study is anticipated, because a persistent left superior vena cava occurs in up to 4.3% of patients referred for cardiac evaluation and changes the path of the contrast.

Monitoring and personnel. Minimum monitoring during moderate sedation is heart rate, noninvasive blood pressure, respiratory rate, and oxygen saturation. Capnography is strongly recommended for moderate sedation and in patients at risk of respiratory complications, and it is mandatory for deep sedation and general anesthesia. The room has oxygen, suction, airway equipment, and Advanced Cardiovascular Life Support drugs, with methylene blue, flumazenil, and naloxone available. A trained person other than the echocardiographer administers sedation and monitors the patient. A modified Aldrete score is recorded before sedation; a score of 9 or 10 is consistent with conscious sedation, and a score below 9 indicates oversedation.

AgentRolePharmacology and dosing pointsCautions
MidazolamAnxiolysis, sedation, amnesiaOnset 1–2 min; duration typically 15–30 min; small incremental doses in frail or older patientsDelirium in older adults; avoided in pregnancy, especially the first trimester; synergistic respiratory depression with opioids
FentanylAnalgesia for probe passageStarting dose halved in frail or older patients; tolerance in chronic opioid usersRespiratory depression, nausea; meperidine has fallen out of use because of respiratory depression and active metabolites in kidney failure
PropofolDeep sedationRapid onset and recovery without increased cardiopulmonary complications compared with benzodiazepine–opioid sedationApnea with titration; administered by or with anesthesia services; no reversal agent
FlumazenilBenzodiazepine reversal0.2 mg, repeated at 1-minute intervals until reversal; duration 10–15 min; monitoring for at least 30 min after useWithdrawal seizures in patients on chronic benzodiazepines; resedation
NaloxoneOpioid reversal0.4 mg for apnea; 0.1 mg increments at 1-minute intervals for partial reversal until respiratory rate exceeds 12 breaths/minShorter duration than some opioids

Sedation and reversal agents for TEE, from the 2013 ASE/SCA guideline. Local sedation policies, state regulations, and anesthesia department rules govern who administers these agents.

Anesthesia support. Anesthesia care is requested when deep sedation is expected to be needed, when hemodynamic instability is anticipated, when prior sedation failed, and in patients with a difficult airway or severe lung disease. Intubated patients in intensive care receive additional sedation as needed, and a respiratory therapist secures the endotracheal tube during insertion.

Recovery and discharge. Vital signs return to within 10% of baseline and the modified Aldrete score to baseline before discharge. Patients take nothing by mouth for at least 1 hour, until topical anesthesia and sedation have worn off, to reduce the risk of aspiration. Ambulatory patients do not drive and leave with a companion. Odynophagia or dysphagia lasting more than 1 day is reported, because it can signal esophageal injury.

Topical pharyngeal anesthesia and benzocaine-induced methemoglobinemia

Awake patients usually receive topical anesthesia of the tongue, palate, tonsillar pillars, and posterior pharynx with benzocaine, benzocaine-containing combinations, or lidocaine as spray, gargle, gel, or lozenge. Adequate topical anesthesia eases probe passage and reduces the sedative dose; some patients tolerate TEE with topical anesthesia alone. Spray reaches the posterior pharynx but can be inhaled and anesthetize the larynx, which raises the risk of aspiration. Anesthetized patients do not need topical anesthesia.

Benzocaine oxidizes hemoglobin iron to the ferric state. Methemoglobin cannot carry oxygen and shifts the dissociation curve of the remaining hemoglobin to the left. In a series of 28,478 TEE studies, methemoglobinemia occurred in 19 patients (1 in 1,499, 0.067%), with a mean methemoglobin level of 32%. All were cyanotic with low oxygen saturation, and affected patients were more often hospitalized, anemic, and septic than unaffected controls. Signs include dyspnea, tachycardia, cyanosis, a pulse oximetry reading that falls and does not respond to oxygen, and chocolate-brown arterial blood with a normal arterial partial pressure of oxygen. Co-oximetry establishes the diagnosis. Levels above 55% produce lethargy and stupor, and levels above 70% can cause arrhythmia, circulatory failure, and death.

Treatment is supplemental oxygen and intravenous methylene blue, 1 to 2 mg/kg of the 1% solution given slowly over 5 minutes. Methylene blue can precipitate serotonin syndrome in patients taking selective serotonin reuptake inhibitors and is avoided in them. Minimal benzocaine doses, lidocaine-based alternatives, and avoidance of benzocaine in hospitalized, anemic, or septic patients lower the risk.

Probe insertion technique

Before insertion, the probe is inspected for damage, its function is tested, and the controls are confirmed to be unlocked. In the awake patient, topical anesthesia is applied and a bite block is placed before sedation. The patient lies in the left lateral decubitus position with the echocardiographer facing the patient. The probe, lubricated and in a neutral position with slight anteflexion, is passed to the back of the pharynx in the midline, and the patient swallows as it advances into the esophagus. A midline course avoids the piriform fossa. With a bite block in place, one or two fingers can guide the probe toward the midline and depress the tongue. Resistance is never overcome with force; failure to pass the probe is accepted and the study abandoned or rescheduled with anesthesia support.

In the anesthetized, intubated patient lying supine, the probe is inserted from the head of the bed with slight anteflexion. Lifting the mandible anteriorly and caudally moves the tongue forward and opens the pharynx; a laryngoscope helps when passage is difficult. The bite block is placed after insertion in this setting, because a block placed first pushes the tongue backward and obstructs passage. A bite block is used even under anesthesia, because neuromuscular blockade is not always maintained throughout the procedure. Care is taken not to dislodge the endotracheal tube; endotracheal tube malposition occurred in 0.03% of intraoperative studies in a large series.

Probe manipulation, the multiplane angle, and image orientation

The ASE/SCA nomenclature describes four physical movements of the probe and one electronic rotation of the imaging plane.

MovementDefinitionWhat changesTypical use
Advance or withdrawPushing the tip distally or pulling it proximallyProbe depth and stationMoving between upper esophageal, mid-esophageal, transgastric, and deep transgastric stations; surveying the descending aorta
Turn right or leftRotating the shaft clockwise toward the patient's right or counterclockwise toward the leftDirection the transducer facesRight-sided structures (bicaval view, right pulmonary veins) or left-sided structures (appendage, descending aorta)
Anteflex or retroflexTilting the tip anteriorly or posteriorly with the large control wheelTilt of the plane up or down; gastric wall contactFive-chamber versus four-chamber view; transgastric contact
Flex right or leftTilting the tip toward the patient's right or left with the small control wheelLateral tiltFine alignment; transgastric right ventricular views
Rotate forward or backElectronic rotation of the imaging plane from 0° toward 180° and backPlane orientation at a fixed positionChanging view at one station without moving the probe

Probe manipulation terms from the 2013 ASE/SCA guideline. Flexion right or left keeps its name even when the shaft has been turned. Excessive flexion during advancement or withdrawal raises pressure on the esophageal wall and is avoided.

The display places the transducer at the top of the image, so posterior structures appear in the near field. At 0° (the transverse plane) the patient's right appears on the left of the display. Forward rotation to 90° (the longitudinal plane) moves the left side of the display inferiorly, so cephalad structures appear on the right. At 180° the image is the mirror image of 0°. The angles listed for the views are approximate, because heart position and the relation of the esophagus to the heart vary; in some patients the esophagus lies beside the atrioventricular groove and in others immediately behind the left atrium. A view is named for the structures it displays. The angle at which it was obtained does not define it. In simultaneous biplane imaging the secondary plane is oriented as if rotated forward from the primary plane, so the secondary image can appear reversed once the primary plane reaches 90°. Structures are therefore identified from landmarks such as the outflow tract, superior vena cava, appendage, and coronary origins.

The 28-view comprehensive examination

The 2013 ASE/SCA guideline expanded the 20 views of the 1999 intraoperative guideline to 28 views that include long- and short-axis images of all four valves, all four chambers, and the great vessels. The views are grouped by station: upper esophageal (at the aortic arch), mid-esophageal (behind the left atrium, about 30 cm from the incisors), transgastric (in the stomach, anteflexed beneath the left ventricle), and deep transgastric (advanced and fully anteflexed, looking up from the apex). The station is recognized from the near field: left atrium in mid-esophageal views, ventricular wall in transgastric views, and apex in the deep transgastric view. The clinical question determines which views are acquired first and in what detail, and additional off-axis images are often needed.

No.ViewAnglePrincipal structures and uses
1ME five-chamber0–10°Outflow tract, aortic valve, both atria and ventricles; mitral A1–A2 and P1–P2
2ME four-chamber0–10° (up to 20°)All chambers, both atrioventricular valves, both septa; global and regional function; mitral A2–P2
3ME mitral commissural50–70°Mitral valve along the coaptation line (P3, A2, P1); both papillary muscles
4ME two-chamber80–100°Left atrium, appendage, mitral valve, anterior and inferior walls; no right chambers
5ME long-axis120–140°Outflow tract, aortic valve, mitral A2–P2; systolic anterior motion; anteroseptal and inferolateral walls
6ME aortic valve long-axis120–140°Outflow tract, annulus, sinuses, sinotubular junction; root dimensions; aortic regurgitation
7ME ascending aortic long-axis90–110°Tubular ascending aorta, right pulmonary artery in cross section
8ME ascending aortic short-axis0–30°Circular ascending aorta, main and right pulmonary arteries, superior vena cava
9ME right pulmonary veinabout 0°Right upper and lower pulmonary veins, turned right from the ascending aortic short-axis view
10ME aortic valve short-axis25–45°Three cusps, commissures, coronary ostia, interatrial septum; planimetry; bicuspid morphology
11ME right ventricular inflow-outflow50–70°Right atrium, tricuspid valve, right ventricle, outflow tract, pulmonic valve around the central aortic valve
12ME modified bicaval tricuspid50–70°Tricuspid valve with the regurgitant jet often aligned with the beam
13ME bicaval90–110°Superior and inferior venae cavae, fossa ovalis, septum; PFO, sinus venosus defects, transseptal puncture
14ME right and left pulmonary veins90–110°Upper and lower veins on both sides, turned right or left
15ME left atrial appendage90–110°, then sweptAppendage body and lobes, ridge with the left upper pulmonary vein
16TG basal short-axis0–20°Mitral valve en face from below ("fish-mouth" view); posteromedial commissure in the near field
17TG midpapillary short-axis0–20°Six midventricular segments in all three coronary territories, both papillary muscles; monitoring of filling and function
18TG apical short-axis0–20°Apical segments without papillary muscles
19TG right ventricular basal0–20°Right ventricle and tricuspid valve in short axis
20TG right ventricular inflow-outflow0–20° with flexion, or 90–120°Tricuspid valve, outflow tract, pulmonic valve
21Deep TG five-chamber0–20°Outflow tract and aortic valve with flow parallel to the beam; aortic and outflow Doppler
22TG two-chamber90–110°Inferior and anterior walls, mitral subvalvular apparatus and chordae
23TG right ventricular inflow90–110°Right ventricle, tricuspid subvalvular apparatus
24TG long-axis120–140°Outflow tract and aortic valve; Doppler alignment for aortic flow
25Descending aortic short-axis0–10°Descending aorta in cross section; left pleural effusion
26Descending aortic long-axis90–100°Descending aorta along its length; atheroma, flap, intra-aortic balloon position
27UE aortic arch long-axis0–10°Mid and distal arch; left innominate vein
28UE aortic arch short-axis70–90°Arch in cross section, left subclavian origin, pulmonary artery and pulmonic valve

The 28 views of the comprehensive examination, with approximate multiplane angles, from the 2013 ASE/SCA guideline. ME, mid-esophageal; TG, transgastric; UE, upper esophageal; PFO, patent foramen ovale.

Several views share an angle and are separated by landmarks. At 0° to 10°, the five-chamber view contains the outflow tract and aortic valve and the four-chamber view does not. At 50° to 70°, the commissural view contains the left atrium, mitral valve, and left ventricle. The right ventricular inflow-outflow view and the modified bicaval tricuspid view contain the right atrium, tricuspid valve, and right ventricle. At 90° to 110°, the two-chamber view has no right-sided chamber, the bicaval view shows both venae cavae, and the ascending aortic long-axis view shows a tubular aorta without a ventricle. At 120° to 140°, the left atrium lies in the near field of the mid-esophageal long-axis view and the ventricular wall in the near field of the transgastric long-axis view. In the stomach, the basal short-axis view contains the mitral orifice, the midpapillary view both papillary muscles, and the apical view neither; a circular ventricle by itself does not identify the level.

The 11-view basic perioperative examination

The 2013 ASE/SCA consensus statement on basic perioperative TEE defines a focused examination of the 11 most relevant views for nondiagnostic intraoperative monitoring by anesthesiologists. The views are the mid-esophageal four-chamber, two-chamber, long-axis, ascending aortic long-axis, ascending aortic short-axis, aortic valve short-axis, right ventricular inflow-outflow, and bicaval views; the transgastric midpapillary short-axis view; and the descending aortic short- and long-axis views. The basic examination is directed at the causes of hemodynamic instability: ventricular size and function, regional wall motion, volume status, gross valve lesions, pericardial effusion, and pulmonary embolism or air. It does not replace a comprehensive diagnostic study; findings beyond its scope prompt a comprehensive examination by an echocardiographer with advanced training.

Mitral valve segmental anatomy by view

The Carpentier nomenclature divides the anterior leaflet into A1 (anterolateral), A2, and A3 (posteromedial) segments and the posterior leaflet into P1, P2, and P3 scallops. A1 and P1 lie toward the appendage and the anterolateral commissure; A3 and P3 lie toward the inferior vena cava and the posteromedial commissure. Different imaging planes cut the valve along different lines, so a single two-dimensional plane cannot localize all segments. A lesion is assigned to a segment only when two or more planes agree, and three-dimensional en face imaging settles residual doubt.

ViewAngleSegments usually shownComment
ME five-chamber0–10°A1–A2 with P1–P2Plane passes through the anterolateral valve near the outflow tract
ME four-chamber0–10°A2 and P2; A3 and P3 with advancement or retroflexionLevel changes with small probe movements
ME commissural50–70°P3, A2, P1 from left to rightBoth commissures; a central anterior segment flanked by two posterior scallops
ME two-chamber80–100°P3, then the anterior leaflet toward A1Appendage at the right of the display
ME long-axis120–140°P2 and A2Standard plane for vena contracta width and prolapse height
TG basal short-axis0–20°All six segments en facePosteromedial commissure in the near field
3D en face ("surgeon's view")VolumeAll segmentsAortic valve at 12 o'clock and appendage at 9 o'clock

Mitral segments by TEE view. Segment assignment requires agreement in at least two planes. ME, mid-esophageal; TG, transgastric; 3D, three-dimensional.

When a prolapsing segment appears at the right of the display in the commissural view while the central segment coapts normally, the lesion is in P1. A normal long-axis view (P2 and A2) together with a five-chamber view showing posterior prolapse near the outflow tract confirms isolated P1 disease. Grading of mitral regurgitation, the Carpentier classification, and repair planning are covered on the mitral regurgitation page.

Aortic valve, aortic root, and Doppler alignment

In the mid-esophageal aortic valve short-axis view at 25° to 45°, the noncoronary cusp lies beside the interatrial septum, the right coronary cusp lies anteriorly next to the right ventricular outflow tract, and the left coronary cusp lies toward the appendage next to the left main coronary origin. Cusps are named from these landmarks. A clock face on the display is not used. The view shows commissural fusion, raphe, and the systolic orifice shape of a bicuspid valve, and it allows planimetry of the anatomic orifice when calcification does not obscure the leaflet tips. The aortic valve long-axis view at 120° to 140° measures the annulus, sinuses, and sinotubular junction and shows cusp prolapse and the direction of aortic regurgitant jets.

Doppler accuracy depends on the angle between the beam and flow. Measured velocity equals true velocity multiplied by the cosine of the intercept angle. Because the pressure gradient varies with the square of velocity, a small percentage error in velocity produces roughly twice that percentage error in gradient, and larger angles produce disproportionately larger errors. Mid-esophageal views are nearly perpendicular to aortic flow. The deep transgastric five-chamber view and the transgastric long-axis view align the beam with the outflow tract and aortic valve. For a true peak aortic velocity of 4.6 m/s (gradient 85 mm Hg), a 50° intercept in the mid-esophageal long-axis view records 2.96 m/s (35 mm Hg), whereas a 15° intercept in the deep transgastric view records 4.44 m/s (79 mm Hg). Hemodynamic grading of aortic stenosis therefore rests on TTE, with TEE contributing morphology, planimetry, and root dimensions; grading thresholds are on the aortic stenosis page. For the other valves, mid-esophageal views align with mitral inflow and regurgitation, the modified bicaval view often aligns with tricuspid regurgitation, and the upper esophageal arch short-axis view aligns with pulmonary artery flow.

Right heart, venae cavae, coronary sinus, and pulmonary veins

The bicaval view at 90° to 110°, reached by turning right from the two-chamber view, places the left atrium in the near field, the right atrium beneath it, the superior vena cava at the right of the display, and the inferior vena cava at the left. It shows the fossa ovalis, the septum secundum, the Eustachian valve, the crista terminalis, and the right atrial appendage. A Chiari network, a filamentous remnant of the right sinus venosus valve, is seen in 1% to 4% of people and is associated with PFO and atrial septal aneurysm. The coronary sinus is imaged in deep esophageal views at 0° and, with counterclockwise turning, from the two-chamber view at 90° to 130°. It is examined when retrograde cardioplegia cannulation is planned and when a persistent left superior vena cava or unroofed coronary sinus is suspected.

The four pulmonary veins enter the posterior left atrium, with the left veins usually more superior than the right. The right veins are found by turning right from the ascending aortic short-axis view at about 0° or from the bicaval view at 90° to 110°; the left veins are found by turning left past the appendage. The left upper pulmonary vein is separated from the appendage by a ridge of infolded atrial wall. A vein is traced into the left atrium before it is sampled, because the superior vena cava and the appendage can be mistaken for a vein.

Pulmonary venous Doppler. Pulsed-wave Doppler 1 to 2 cm into the vein records a systolic (S) wave, a diastolic (D) wave, and an atrial reversal (Ar) wave. Systolic predominance is normal in middle-aged and older adults; diastolic predominance is normal in young adults with brisk early relaxation and abnormal in older patients, where it suggests elevated left atrial pressure. An atrial reversal that lasts at least 30 ms longer than the mitral A wave indicates elevated left ventricular end-diastolic pressure. Systolic flow reversal in the pulmonary veins is a specific sign of severe mitral regurgitation, while blunted systolic flow is nonspecific because atrial fibrillation and elevated left atrial pressure also blunt it. Accelerated, turbulent, continuous flow at a vein ostium after pulmonary vein isolation or other atrial procedures suggests stenosis.

Reference values and thresholds used in transesophageal echocardiography

Chamber and wall dimensions are measured on TTE, and TTE reference values do not transfer directly to TEE, where foreshortening and oblique planes are common. The values specific to TEE concern the appendage, the pulmonary veins, the septum, the aorta, and the device itself.

MeasurementValueInterpretationSource
Appendage peak emptying velocity, sinus rhythmMean 61 cm/sPatients with intermittent AF in sinus rhythm during TEESPAF-III TEE substudy
Appendage peak emptying velocity, AFMean 33 cm/sLower velocities reflect loss of organized contractionSPAF-III TEE substudy
Low appendage velocity<20 cm/sAssociated with dense spontaneous echo contrast, appendage thrombus (relative risk 2.6), and cardioembolic eventsSPAF-III TEE substudy
Appendage velocity quoted in the 2016 ASE guideline≤27 cm/sSecondary citation; the SPAF-III source it cites reports ≤20 cm/s (relative risk 1.7 for high clinical thromboembolic risk)2016 ASE; SPAF-III (Zabalgoitia 1998)
Appendage orifice diameter10–24 mmNormal range in anatomic series; device sizing uses multiple planes2013 ASE/SCA
Pulmonary vein Ar duration minus mitral A duration≥30 msElevated left ventricular end-diastolic pressure2016 ASE/EACVI diastolic function (updated by the 2025 ASE guideline)
Pulmonary vein systolic flowReversalSpecific for severe mitral regurgitation2017 ASE regurgitation
Atrial septal aneurysm>10 mm excursion, or 15 mm combinedExcursion into either atrium beyond 10 mm, or right plus left excursion totaling 15 mm2015 ASE/SCAI
Atrial septal defect rims<5 mm deficientA deficient aortic rim alone is usually acceptable for device closure2015 ASE/SCAI
Large right-to-left shunt>20 microbubblesCrossing into the left atrium; no grading scheme is universally accepted2016 ASE
Aortic plaque≥4 mmHigh-risk plaque in the ascending aorta or arch2016 ASE; case-control data
Intramural hematoma>5 mmCrescentic or circular wall thickening without flow2022 ACC/AHA aortic guideline
Valve strands≤2 mm wide, about 3–10 mm longFiliform, undulating, at the closure line; not associated with embolism in prospective data2016 ASE
Surgical appendage exclusionStump <1 cm; no flowAdequate exclusion on intraoperative TEE2023 ACC/AHA/ACCP/HRS (Class 1)
Probe tip temperatureWarning near 41 °C; imaging stops at 42.5 °CValues for one common platform; thresholds vary by manufacturerSingle-platform report (Choudhury 2016)

Reference values and decision thresholds specific to TEE. AF, atrial fibrillation; SPAF, Stroke Prevention in Atrial Fibrillation; EACVI, European Association of Cardiovascular Imaging; SCAI, Society for Cardiovascular Angiography and Interventions.

Left atrial appendage anatomy and the diagnosis of thrombus

The left atrial appendage is a blind-ending, trabeculated pouch whose orifice lies superiorly and laterally beside the left upper pulmonary vein. Its body varies in length, curvature, and number of lobes; autopsy studies describe one to five lobes. Because it contracts and changes shape through the cycle, a single static plane can miss a lobe or a thrombus. The appendage is imaged from the mid-esophageal station with reduced depth, a narrow sector, the focus at the appendage, and a high transmit frequency, and the plane is swept from 0° through 90° to about 135° or beyond. Simultaneous biplane imaging and three-dimensional zoom acquisition shorten the sweep and show lobes that a single plane foreshortens.

Pectinate muscles are parallel muscular ridges continuous with the appendage wall and moving with it. The ridge between the appendage orifice and the left upper pulmonary vein, variously called the warfarin ridge, the Q-tip sign, the posterolateral ridge, and the ligament of Marshall, can appear bulbous in cross section and is a frequent false-positive "mass." Thrombus is an echodense mass with a border distinct from the endocardium and pectinate muscles, often located in the apex or a lobe, sometimes mobile, and present in more than one plane. Reverberation from the ridge and side-lobe echoes from a calcified mitral annulus can project echoes into the appendage lumen. An echo that does not persist in an orthogonal plane, moves with the reflecting structure, or lies at a multiple of a reflector's depth is artifact.

Mid-esophageal transesophageal echocardiogram at a multiplane angle of 101 degrees in a patient with rheumatic mitral stenosis, showing dense spontaneous echo contrast in an enlarged left atrium and an arrow pointing to an echodense area in the left atrial appendage; the display reads a probe tip temperature of 39.8 degrees Celsius.
Echodense area in the left atrial appendage (arrow) on mid-esophageal TEE at 101° in a patient with permanent atrial fibrillation and severe rheumatic mitral stenosis, with dense spontaneous echo contrast filling the left atrium. Multiplane and three-dimensional imaging were interpreted as possible thrombus, but no thrombus was found at mitral valve surgery; the authors attributed the echodensity to blood stasis. The display also shows the probe tip temperature (TEE T). Credit: Libiran NBS, Becher E, Kataoka M, Cureus 2026;18:e109438, doi:10.7759/cureus.109438 (PMC13283740), Figure 1, CC BY 4.0.

When the appendage remains indeterminate, an ultrasound-enhancing agent fills the lumen and outlines a thrombus as a persistent filling defect, while pectinate muscles and artifact disappear or change with the contrast. The 2016 ASE guideline lists contrast and three-dimensional imaging as potentially useful for appendage assessment. Delayed-phase cardiac CT, in which slow flow fills in and thrombus persists, is the main nonechocardiographic alternative and is accepted by the 2023 ACC/AHA/ACCP/HRS guideline as an imaging method to exclude thrombus.

Anatomic illustration of the heart cut in the mid-esophageal two-chamber transesophageal plane, showing the left atrium at the top, the mitral valve and chordae, the left ventricle below, and the left atrial appendage on the right with its trabeculated pectinate muscle lining.
Anatomic section in the mid-esophageal two-chamber plane, oriented as on the TEE display with the transducer at the top. The left atrium lies in the near field and the appendage at the right of the display, where its pectinate muscles, ridges continuous with the wall, are the usual mimic of thrombus. Credit: Patrick J. Lynch, medical illustrator, via Wikimedia Commons, CC BY 2.5; resized.

Spontaneous echo contrast, sludge, and appendage emptying velocity

Spontaneous echo contrast is a swirling, smoke-like echogenicity produced by red-cell aggregation where flow is slow and shear is low. It accompanies atrial fibrillation, mitral stenosis, a large left atrium, and low appendage velocity, and it is diminished by mitral regurgitation, which increases atrial flow. Thrombus formation is a continuum: spontaneous echo contrast, then sludge (more viscid than smoke but less dense than thrombus), then thrombus. Sludge is reported as such, and thrombus is diagnosed only when a discrete mass with a defined border persists across planes. Excessive gain creates random speckle that flickers from frame to frame without moving with flow, and high reject settings can erase true smoke.

In a TEE study of 140 patients with atrial fibrillation, spontaneous echo contrast was present in 56% and was the only independent correlate of left atrial thrombus, found in 14 of 15 patients with thrombus; its grade rose as appendage velocity fell. In the SPAF trial's TEE substudy of 382 high-risk patients, dense spontaneous echo contrast (20% of patients) and appendage thrombus (10%) predicted ischemic stroke, and complex aortic plaque (35%) carried a fourfold stroke rate. Adjusted-dose warfarin reduced stroke in patients with dense contrast and complex plaque.

Pulsed-wave Doppler with the sample volume just inside the appendage orifice records emptying (outflow) and filling (inflow) waves. In sinus rhythm the emptying wave follows the P wave; in atrial fibrillation the signal is a sawtooth of irregular, lower-velocity waves; in atrial flutter the waves are rapid and regular. Among 721 patients with nonvalvular atrial fibrillation in SPAF-III, mean peak emptying velocity was 33 cm/s in patients in atrial fibrillation and 61 cm/s in those with intermittent atrial fibrillation who were in sinus rhythm at the time of TEE. A velocity below 20 cm/s was associated with dense spontaneous echo contrast, appendage thrombus (relative risk 2.6, independent of clinical risk factors), and subsequent cardioembolic events. The 2016 ASE guideline quotes a threshold of 27 cm/s or less with a relative risk of 1.7. The SPAF-III analysis it cites reports a threshold of 20 cm/s or less, with a relative risk of 1.7 for high clinical thromboembolic risk in a cross-sectional comparison, so 20 cm/s is the value supported by primary data and 27 cm/s is a secondary citation. Low velocity marks poor appendage function and stasis; it neither establishes nor excludes thrombus.

Atrial stunning. After cardioversion, atrial and appendage mechanical function can remain depressed for days to weeks despite organized P waves. Emptying velocity may fall below its value during fibrillation, and spontaneous echo contrast can appear or intensify within seconds of a successful shock. In a TEE study before and during cardioversion, thrombus was found in only 1 of 70 studies. Four patients had thromboembolic events without prior thrombus, and new or increased spontaneous echo contrast appeared after cardioversion in patients with impaired atrial contraction. This stunning, with partial recovery over 15 to 30 days and full recovery over 30 to 90 days in some series, is the mechanistic basis for anticoagulation after cardioversion even when TEE shows no thrombus.

TEE-guided cardioversion and the ACUTE trial

The conventional strategy before elective cardioversion of atrial fibrillation lasting 48 hours or more is 3 weeks of therapeutic anticoagulation, which allows existing thrombus to organize and adhere. The TEE-guided strategy shortens this interval: when TEE shows no thrombus in a patient already receiving therapeutic anticoagulation, cardioversion proceeds at once.

The Assessment of Cardioversion Using Transesophageal Echocardiography (ACUTE) trial randomized 1,222 patients with atrial fibrillation of more than 2 days' duration to TEE-guided early cardioversion after brief anticoagulation or to conventional treatment with 3 weeks of warfarin before cardioversion. Embolic events within 8 weeks did not differ significantly (5 of 619 patients, 0.8%, versus 3 of 603, 0.5%; P=0.50). Hemorrhagic events were less frequent with the TEE-guided strategy (2.9% versus 5.5%; P=0.03), time to cardioversion was shorter (mean 3.0 versus 30.6 days), and initial restoration of sinus rhythm was more frequent (71.1% versus 65.2%). Death, maintenance of sinus rhythm, and functional status at 8 weeks did not differ. The trial established TEE guidance as an alternative strategy that was not superior; patients in whom thrombus was found had cardioversion deferred and anticoagulation continued.

The direct oral anticoagulant trials of cardioversion (X-VeRT, ENSURE-AF, and EMANATE) allowed both strategies and showed low thromboembolic event rates similar to those with warfarin. Because there is no laboratory measure of adequate anticoagulation with a direct oral anticoagulant, adherence during the preceding 3 weeks is confirmed by history; uncertain adherence is an indication for TEE before cardioversion. Drug selection, dosing, and the cardioversion procedure itself are covered on the atrial fibrillation page.

Anticoagulation before and after cardioversion

SettingRecommendationClass (LOE)
AF of 48 hours or longer, elective cardioversion3 weeks of uninterrupted therapeutic anticoagulation or imaging to exclude intracardiac thrombus before cardioversion1 (B-R)
Any AF cardioversionTherapeutic anticoagulation established before cardioversion and continued without interruption for at least 4 weeks afterward1 (B-NR)
Appendage thrombus on precardioversion imagingCardioversion deferred; therapeutic anticoagulation for at least 3 to 6 weeks; imaging repeated before cardioversion1 (C-LD)
Previous LAAO, without anticoagulationImaging to assess occlusion and exclude device-related thrombus before cardioversion may be reasonable2b (B-NR)
AF under 48 hours, without anticoagulation, CHA₂DS₂-VASc ≥2Precardioversion imaging to exclude thrombus may be considered (not in the setting of cardiac surgery)2b (C-LD)

Recommendations from section 8.2.1 of the 2023 ACC/AHA/ACCP/HRS atrial fibrillation guideline that involve precardioversion imaging or follow a negative study; the same principles apply to pharmacologic and electrical cardioversion. The remaining recommendations in this section, including those for atrial fibrillation of under 12 hours and for residual leak after appendage occlusion, are on the atrial fibrillation page. LOE, level of evidence; LAAO, left atrial appendage occlusion; AF, atrial fibrillation; CHA₂DS₂-VASc, congestive heart failure, hypertension, age 75 years or older (2 points), diabetes, prior stroke or transient ischemic attack (2 points), vascular disease, age 65 to 74 years, sex category.

When thrombus is found in a patient already receiving anticoagulation, the 2023 guideline notes that the cause (nonadherence, inappropriate dosing, poor absorption, or drug interactions) is investigated. Switching agents can be considered, although the benefit is uncertain. In registry data, thrombus resolved on follow-up TEE in 61.2% of anticoagulant-naive patients after 3 to 12 weeks of a vitamin K antagonist and in 41.5% after 6 weeks of rivaroxaban. Long-term anticoagulation after the first 4 weeks follows stroke risk, whatever the rhythm. The 2024 European Society of Cardiology guideline uses 24 hours, not 48 hours, as the duration beyond which early cardioversion requires 3 weeks of anticoagulation or TEE.

Atrial flutter carries a similar pericardioversion risk. The guideline cites embolism after cardioversion in 0.72% of patients with flutter and 0.46% of those with fibrillation. It also cites left atrial thrombus in 0% to 38% and spontaneous echo contrast in 21% to 28% of patients with flutter, and thrombus or dense contrast in 3% of patients referred for flutter ablation without prior anticoagulation. Anticoagulation around cardioversion and ablation of flutter therefore follows the same rules.

Appendage imaging before ablation and after appendage occlusion

Before catheter ablation of atrial fibrillation, many centers image the appendage with TEE, CT, or intracardiac echocardiography to exclude thrombus, particularly in patients with persistent atrial fibrillation, high stroke risk, or interrupted anticoagulation. Routine TEE after pulmonary vein isolation in asymptomatic patients is rated inappropriate. After ablation, TEE is avoided when atrioesophageal fistula is suspected, because insufflation of air through an esophageal defect can cause massive air embolism; CT of the chest is the diagnostic test.

After percutaneous left atrial appendage occlusion, TEE or CT assesses device position, peridevice leak, and device-related thrombus. The 2025 Society for Cardiovascular Angiography and Interventions and Heart Rhythm Society (SCAI/HRS) guideline suggests postprocedural TEE or CT over no imaging (conditional recommendation, very low certainty) and makes no recommendation on the timing of repeat imaging. In a National Cardiovascular Data Registry analysis of 51,333 patients with the Watchman 2.5 device, a small residual leak (greater than 0 and up to 5 mm) at about 45 days was present in 25.8% and was associated with modestly more thromboembolic events at 1 year than no leak.

Cardioembolic source evaluation after stroke

An embolic source study searches the heart and thoracic aorta for material that could reach the systemic circulation, or for a pathway that lets venous thrombus cross to it. Potential sources fall into four groups: stasis within a chamber (the appendage and the left ventricle), material on a valve, atheroma of the thoracic aorta, and paradoxical embolism through a right-to-left shunt. Echocardiography identifies plausible sources and their high-risk features; it does not prove that a finding caused the event, and sources often coexist with large-artery or small-vessel disease. TEE for a cardiovascular source of embolus with no identified noncardiac source is rated appropriate, and with an identified noncardiac source it is rated uncertain.

Potential sourceTTETEEFeatures that raise embolic concern
Left atrial or appendage thrombus or spontaneous echo contrast−/+++++Dense contrast or sludge; emptying velocity <20 cm/s
Aortic atheroma−/+++++Thickness ≥4 mm; ulceration; mobile component
Prosthetic valve abnormality+++++Thrombus, pannus, vegetation; subtherapeutic anticoagulation
Native valve vegetation++++++Length >10 mm, mobility, mitral location
Atrial septal anomalies (PFO, aneurysm)++++++Large shunt; atrial septal aneurysm
Cardiac tumors+++++++Mobile, frond-like, or villous surface (fibroelastoma, friable myxoma)
Left ventricular thrombus++++++Protrusion and mobility; apical aneurysm

Relative yield of TTE and TEE by embolic source (symbols range from − for little or no yield to ++++ for the highest yield), from Table 3 of the 2016 ASE cardiac source of embolism guideline, with high-risk features added from the same document. TTE remains the first test and is better for left ventricular apical thrombus. PFO, patent foramen ovale.

In the 1991 comparison of transthoracic and transesophageal approaches in 93 patients with proven masses, TEE was superior for appendage thrombi, small flat left atrial thrombi, masses in the superior vena cava and right heart, and masses of the descending aorta, whereas TTE detected left ventricular apical thrombi more often. A cardioembolic workup therefore begins with TTE, which also screens for shunt with agitated saline in an awake patient who can perform a vigorous Valsalva maneuver. It proceeds to TEE when the answer would change management, most often in younger patients with cryptogenic stroke in whom PFO closure, anticoagulation, or surgery is under consideration.

Patent foramen ovale, atrial septal aneurysm, and the agitated saline study

A patent foramen ovale is a flap-valve channel between the septum primum and septum secundum with no deficient tissue; it opens only when right atrial pressure exceeds left atrial pressure. TEE defines its anatomy: tunnel length, septum secundum thickness, septal mobility, an associated atrial septal aneurysm, a prominent Eustachian valve or Chiari network, and the relation of the tunnel to the aorta. An atrial septal aneurysm is a redundant, mobile septum with excursion of more than 10 mm into either atrium, or right plus left excursion totaling 15 mm; mobility below these limits is described as a hypermobile septum.

Agitated saline is injected into an arm vein while the bicaval view is recorded, at rest and with provocation. Direct visualization of bubbles crossing the fossa ovalis is the strongest evidence of an intracardiac shunt. The preparation, the timing criteria that separate intracardiac from intrapulmonary shunts, and shunt grading are on the atrial septal defect and patent foramen ovale page.

Sedation is the main weakness of TEE for shunt detection. A sedated patient cannot strain effectively, so a negative TEE study may reflect an inadequate maneuver. Abdominal compression and release, or release of a sustained inspiratory hold under positive-pressure ventilation, substitute for the Valsalva maneuver. An effective maneuver is confirmed by septal bowing toward the left atrium. A prominent Eustachian valve can direct unopacified inferior caval blood at the fossa and produce a false-negative arm injection, which a femoral vein injection overcomes. Closure trials and candidacy for closure are on the atrial septal defect and patent foramen ovale page.

Atrial septal defects: secundum, sinus venosus, and rim assessment

TEE defines interatrial anatomy that TTE often cannot. The septum is swept from 0° in increments, and the bicaval view near 90° to 120° shows the superior and inferior rims and the junctions of both venae cavae. Secundum defects lie within the fossa ovalis. Superior sinus venosus defects lie at the junction of the superior vena cava with the atria, above the fossa, and are usually accompanied by anomalous drainage of the right upper pulmonary vein; inferior sinus venosus defects lie at the inferior caval junction. Both are frequently missed on TTE in adults and are best shown in the bicaval view with low-velocity color Doppler. Primum defects lie at the crux next to the atrioventricular valves, and an unroofed coronary sinus is shown by flow from the left atrium into a dilated coronary sinus.

Before device closure of a secundum defect, the report gives the number of defects, maximal and minimal diameters (ideally from three-dimensional data), dynamic change through the cycle, flow direction, septal aneurysm, and all six rims. The rims are aortic (anterosuperior), superior vena caval, right upper pulmonary vein (posterosuperior), posterior, inferior vena caval, and atrioventricular valve (anteroinferior). A rim shorter than 5 mm is deficient. A deficient aortic rim alone is common and usually acceptable, whereas deficient inferior caval or posterior rims give less support to the device, and a defect larger than 38 mm is generally unsuitable for a device. During closure, TEE or ICE confirms the defect, guides the wire and balloon sizing, shows both discs straddling the septum without impinging on the aortic root, atrioventricular valves, coronary sinus, or venous inflows, and assesses residual shunt and the pericardium. Indications for closure are on the atrial septal defect and patent foramen ovale page.

Aortic atheroma as an embolic source

TEE shows the descending thoracic aorta from the subclavian origin to about the level of the superior mesenteric artery and most of the ascending aorta and arch, and it is the preferred echocardiographic method for aortic sources of embolism. Plaque is described by maximal thickness, surface (smooth or ulcerated), calcification, and mobile components. In a case-control study of 250 consecutive patients with ischemic stroke and 250 controls older than 60 years, plaque 4 mm or thicker in the ascending aorta or proximal arch was present in 14.4% of patients and 2% of controls, with an adjusted odds ratio for stroke of 9.1. Among patients with strokes of no obvious cause, 28.2% had such plaque. The 2016 ASE guideline adopts 4 mm in the ascending aorta or arch as the high-risk threshold. Ulceration and mobile components add risk, and a mobile component is reported separately because it can occur on plaque of any thickness. Plaque distal to the left subclavian artery reaches the cerebral circulation only with retrograde diastolic flow, so ascending and arch plaque carry most weight in stroke evaluation.

A short segment of the distal ascending aorta just proximal to the innominate artery is hidden by air in the trachea and main bronchus (the tracheal blind spot). It is reported as not visualized, and CT, MRI, or intraoperative epiaortic scanning covers it when it matters. Before cardiac surgery, intraoperative identification of severe ascending aortic atheroma guides the sites of cannulation and cross-clamping or leads to an alternative strategy.

Aortic dissection, intramural hematoma, and reverberation that mimics a flap

CT angiography is the Class 1 initial test for suspected acute aortic syndrome in the 2022 ACC/AHA aortic disease guideline, and TEE and MRI are reasonable alternatives (Class 2a). The guideline notes that the sensitivity and specificity of CT, TEE, and MRI are all high. TEE is chosen when the patient is too unstable to leave the emergency department or intensive care unit, when iodinated contrast is contraindicated, and in the operating room before repair of type A dissection.

The diagnostic finding is an intimal flap separating a true and a false lumen in more than one plane, with different color Doppler flow in the two lumens. The true lumen expands in systole; the false lumen is usually larger, may contain spontaneous echo contrast or thrombus, and may show entry and re-entry tears as color jets across the flap. TEE identifies the features that change the operation. These are the proximal extent of the flap, involvement of the coronary ostia, the mechanism and severity of aortic regurgitation (root dilation, commissural detachment with cusp prolapse, or flap prolapse through the valve), pericardial effusion and tamponade, and regional wall motion abnormalities suggesting coronary involvement. Intramural hematoma appears as crescentic or circular wall thickening of more than 5 mm without detectable flow, and penetrating atherosclerotic ulcer as a crater through plaque into the media, most often in the descending aorta.

The ascending aorta is a common site of reverberation artifact. A linear echo produced by reverberation from the posterior aortic wall or a catheter lies at twice the depth of its source, moves with the source, crosses anatomic boundaries, and does not separate two streams of color flow; M-mode shows motion parallel to the reflecting wall. A true flap moves independently and is seen in orthogonal planes. Classification, medical therapy, and surgical timing are on the acute aortic syndromes page.

Infective endocarditis: vegetations and perivalvular extension

TEE has sensitivity and specificity above 90% for vegetations, whereas transthoracic sensitivity in meta-analyses ranges from 62% to 79%, and vegetations smaller than 2 to 3 mm may be missed on TTE. The gain is largest for perivalvular extension. In a prospective study of 118 patients with infective endocarditis on 137 native or prosthetic valves, confirmed at surgery or autopsy, abscesses were present in 37.3%; TTE identified 13 of 46 abscess regions and TEE identified 40, with sensitivity of 28.3% versus 87.0% and specificity of 98.6% versus 94.6%. Abscesses were more frequent with aortic valve infection and with staphylococci, and in-hospital mortality was 22.7% with abscess and 13.5% without.

LesionTEE appearanceBest views
VegetationIrregular, oscillating mass on the upstream (low-pressure) side of the valve near the coaptation line, with motion independent of the leafletMitral: ME four-chamber, commissural, long-axis. Aortic: ME aortic valve short- and long-axis. Tricuspid: ME right ventricular inflow-outflow, modified bicaval, TG right ventricular inflow
AbscessPerivalvular thickening or an echolucent or echodense cavity in the annulus, root, or septum without flowME aortic valve short-axis (root circumference); long-axis (aortomitral continuity)
PseudoaneurysmPulsatile echo-free space with systolic expansion and flow through a neck; the mitral–aortic intervalvular fibrosa is a classic siteME long-axis and aortic valve long-axis
FistulaColor jet between the aorta and a chamber or between chambersME aortic valve short-axis; ME right ventricular inflow-outflow
PerforationLeaflet defect with a color jet through the body of the leafletThree-dimensional en face view
Prosthetic dehiscenceRocking motion of the sewing ring with paravalvular regurgitationMitral: ME views from the atrial side. Aortic: ME short-axis and TG views

TEE findings in infective endocarditis and the views that show them best. ME, mid-esophageal; TG, transgastric.

In addition to the Class 1 indications for TEE in endocarditis, the 2020 ACC/AHA valve guideline considers TEE reasonable in Staphylococcus aureus bacteremia without a known source and in prosthetic valve patients with persistent fever without bacteremia (Class 2a). The 2023 European Society of Cardiology guideline recommends TEE in all patients with suspected endocarditis, even when TTE is positive, except for isolated right-sided native valve infection with a good-quality, unequivocal transthoracic study. An early TEE can miss a small vegetation or an abscess that has not yet cavitated, so a negative study is repeated within about a week when suspicion remains high. Duke-ISCVID criteria, surgical indications, and antibiotic regimens are on the infective endocarditis page.

Lead and device-related endocarditis

Pacemaker and defibrillator leads cross the tricuspid valve in the transthoracic far field and cast reverberations and shadows, so TTE is insensitive for lead infection. In 23 patients with definite pacemaker lead infection, TTE showed vegetations in 7 (30%) and TEE in 21 (91%); among 17 patients with uninfected leads, TEE showed strands in 5 and vegetations in none. TEE images the lead course from the superior vena cava through the right atrium and tricuspid valve, often in the bicaval, modified bicaval, and right ventricular inflow views and in deep esophageal views, and it examines left-sided valves for concomitant endocarditis.

Fibrin strands and small thrombi on leads are common in uninfected patients, so a lead mass by itself does not establish infection. Bacteremia, pocket findings, septic pulmonary emboli, positron emission tomography, and comparison with prior studies establish the diagnosis. Vegetation size is recorded because very large lead vegetations raise the risk of pulmonary embolization during transvenous extraction; in the series above, percutaneous extraction of leads with vegetations of 10 mm or more was attempted in 12 patients and succeeded in 9 (75%) without clinical pulmonary embolism. TEE during extraction detects tricuspid injury and pericardial effusion. Intracardiac echocardiography is an option when transthoracic and transesophageal studies are negative and bacteremia persists. Extraction and reimplantation are covered on the infective endocarditis page.

Prosthetic valves: obstruction, paravalvular regurgitation, and leaflet thrombosis

Prosthetic material reflects strongly and shadows what lies behind it, so the direction of insonation determines what is seen. From the esophagus, the atrial side of a mitral or tricuspid prosthesis lies in the near field, and mitral regurgitation, paravalvular leaks, atrial thrombus, and vegetations on the atrial side are well seen; the same structures are shadowed from the apical transthoracic window. The anterior aspect of an aortic prosthesis is shadowed from the esophagus and is examined from transgastric and deep transgastric views, so TTE and TEE are complementary for aortic prostheses. A missing jet behind a prosthesis is nondiagnostic.

Small closing and leakage ("washing") jets within a mechanical prosthesis are normal and are distinguished by their location within the sewing ring, short length, and uniform color. Paravalvular jets arise outside the sewing ring. For mitral prostheses, defects are located on the three-dimensional en face view with the aortic valve at 12 o'clock and the appendage at 9 o'clock, which gives the defects clock-face positions for transcatheter closure. For aortic prostheses, the circumferential extent of paravalvular regurgitation is measured in the short-axis view just below the sewing ring or stent.

In obstruction, TEE distinguishes thrombus from pannus. Thrombus is a larger, soft, often mobile mass on the occluder or leaflets, extending into the atrium with mitral valves, usually early or after a lapse in anticoagulation. Pannus is small, dense, fixed tissue growing inward from the sewing ring, usually late and often in the aortic position. Fluoroscopy and CT measure leaflet opening angles and attenuation. TEE before and during fibrinolysis measures thrombus size and response. Hypoattenuated leaflet thickening after TAVR or surgical bioprosthetic replacement is detected mainly by CT, and three-dimensional TEE or four-dimensional CT confirms leaflet thrombosis when the gradient rises. Normal Doppler values, patient–prosthesis mismatch, and treatment of obstruction are on the prosthetic heart valves page.

Mitral regurgitation, mitral stenosis, and tricuspid disease

Mitral regurgitation. TEE defines the mechanism (Carpentier type), the diseased segments, chordal rupture, clefts, annular dimensions and calcification, and the predictors of systolic anterior motion after repair, and it refines severity when transthoracic measures disagree. Vena contracta width is measured in the long-axis view, and the flow convergence method, also called the proximal isovelocity surface area (PISA) method, uses a view that aligns the beam with the proximal flow field. Three-dimensional color Doppler measures vena contracta area without the hemispheric assumption of two-dimensional PISA, which matters for the crescentic orifice of secondary regurgitation. Systolic reversal of pulmonary venous flow supports severe regurgitation. Under sedation or anesthesia, afterload falls and regurgitation appears smaller, so loading conditions are recorded with all grades. Severity criteria, stages, and repair and edge-to-edge criteria are on the mitral regurgitation page.

Mitral stenosis. Before percutaneous mitral balloon commissurotomy, TEE excludes left atrial and appendage thrombus, which contraindicates the procedure; the 2020 ACC/AHA Class 1 recommendation for commissurotomy applies to patients with favorable morphology, less than moderate regurgitation, and no left atrial thrombus. TEE also grades commissural fusion and calcification, subvalvular thickening, and regurgitation, and three-dimensional en face imaging planimeters the orifice at the leaflet tips. Rheumatic mitral stenosis with atrial fibrillation is often accompanied by dense spontaneous echo contrast, which can be difficult to separate from appendage thrombus, as in the figure above.

Tricuspid valve. The tricuspid valve lies in the far field of mid-esophageal views and its leaflets are thin, so leaflet identification uses the mid-esophageal four-chamber, right ventricular inflow-outflow, and modified bicaval views, transgastric short-axis views of all three leaflets en face, and deep esophageal views. Leads shadow the valve. Before transcatheter tricuspid repair or replacement, TEE is combined with TTE, three-dimensional imaging, and CT; ICE is increasingly used during these procedures when esophageal windows are poor.

Cardiac masses, tumors, and normal structures that mimic them

Primary cardiac tumors are rare (about 0.05% in autopsy series), and metastatic tumors are about 20 times more common. In adults the two most frequent primary tumors are myxoma and papillary fibroelastoma, both of which present with embolism. Myxomas arise in the left atrium in more than 75% of cases, attached by a stalk to the region of the fossa ovalis in more than 90%; they are gelatinous, pedunculated, and smooth or villous, may prolapse through the mitral valve, and are multiple in up to 5%. About 7% are associated with the Carney complex. Papillary fibroelastomas arise mainly on valve surfaces, are usually smaller than 20 mm (mean about 8 to 9 mm), are solitary in more than 90% of cases, and have a central stalk with frond-like projections. Reported sensitivities for detection are about 62% for TTE and 77% for TEE. TEE defines the attachment site and stalk, mobility, and relation to valves, and contrast perfusion helps separate a vascular tumor (enhancing) from thrombus or vegetation (nonenhancing).

StructureLocationDistinguishing features
Warfarin ridge (left lateral ridge)Between the appendage orifice and the left upper pulmonary veinSmooth, continuous with the atrial wall; bulbous "Q-tip" tip in cross section; flow on both sides
Pectinate musclesLeft and right atrial appendagesParallel ridges continuous with the wall and moving with it; seen in all planes through the appendage
Lipomatous hypertrophy of the septumInteratrial septumThick, bright septum sparing the fossa ovalis ("dumbbell" shape); fat attenuation on CT
Eustachian valveInferior vena cava–right atrial junctionLinear, mobile, attached at the caval orifice; directs caval flow toward the fossa
Chiari networkRight atriumThin, fenestrated, highly mobile filaments from the region of the inferior vena cava and coronary sinus
Crista terminalisPosterolateral right atrium at the superior caval junctionC-shaped muscular ridge separating smooth venous right atrium from the trabeculated appendage
Valve strands (Lambl excrescences)Closure lines; atrial side of mitral, ventricular side of aorticFiliform, ≤2 mm wide, about 3–10 mm long, undulating, often multiple
Mitral annular calcificationPosterior annulusDense, shadowing, fixed; caseous form is round with an echolucent center
Transverse sinus and pericardial fatBetween the left atrium and the great arteries; atrioventricular groovesExtracardiac location; fluid-filled sinus can resemble a cyst or abscess beside the appendage
Catheters, leads, and cannulasVenae cavae, right heartBright, parallel reverberations; continuity with a known device

Normal structures and variants that mimic masses or thrombus on TEE. Mimics are identified by location, continuity with normal anatomy, behavior in orthogonal planes, and color flow. Details for strands, the Eustachian valve, the Chiari network, and the crista terminalis follow the 2013 ASE/SCA and 2016 ASE guidelines.

Prospective TEE data found valve strands in 40% to 50% of patients regardless of age, sex, or prior embolic events, without change over time and without association with systemic embolism. Nonbacterial thrombotic endocarditis in cancer and autoimmune disease produces small sterile vegetations along closure lines and belongs in the differential diagnosis of embolic stroke with a valve mass and negative cultures. Right heart thrombus in transit, a serpiginous mobile mass in the right atrium or ventricle, is seen in some patients with pulmonary embolism and is usually shown on TTE.

3D transesophageal echocardiography

Matrix-array TEE probes contain thousands of elements in a two-dimensional grid, timed independently to steer and focus the beam in both azimuth and elevation. They acquire live three-dimensional volumes, zoomed volumes of a target, full volumes stitched from several beats with electrocardiographic gating, three-dimensional color Doppler, and simultaneous biplane or multiplane displays. Three-dimensional data have lower spatial and temporal resolution than two-dimensional images, and stitched acquisitions show artifacts with arrhythmia and respiratory motion, so findings on rendered images are confirmed in the source planes.

The principal clinical uses are en face display of the mitral valve from the left atrium ("surgeon's view," aortic valve at 12 o'clock) to localize prolapse, flail segments, clefts, and perforations, and planimetry of mitral and aortic orifices. Other uses are measurement of the mitral annulus and the appendage orifice, vena contracta area of regurgitant jets, mapping of paravalvular leaks, assessment of septal defect size and shape, and real-time guidance of transseptal puncture and device deployment. The 2012 European Association of Echocardiography and ASE recommendations standardize acquisition and display. Thin leaflet or septal tissue drops out at low gain and can mimic a defect, so an apparent cleft, perforation, or fenestration is confirmed by a color jet through it in orthogonal two-dimensional planes.

Guidance of structural heart interventions

Procedural echocardiography defines the target before a device enters the heart, guides catheters and devices in real time, recognizes complications, and judges the result. The 2022 ASE standards for TEE screening before structural intervention describe procedure-specific screening: mechanism and severity, target dimensions, adjacent structures at risk, access, and exclusionary anatomy. During the procedure the imager identifies the view, the structure, and the finding in anatomic terms, gives measurements with heart rate, blood pressure, and inotrope use, and states whether the image is adequate for the decision.

ProcedureIntraprocedural guidanceOther imaging rolesPrincipal echocardiographic end points
Mitral edge-to-edge repairTwo- and three-dimensional TEETTE follow-upTransseptal height, coaxial alignment, leaflet insertion, residual MR, mean gradient, iatrogenic ASD
TAVRFluoroscopy with TTE or TEECT: annulus, coronary heights, accessValve position, paravalvular regurgitation, new effusion, wall motion, mitral apparatus
Transcatheter mitral replacement or valve-in-valveTwo- and three-dimensional TEECT: annulus, predicted neo-LVOTOutflow and transmitral gradients, paravalvular leak, stability
Tricuspid repair or replacementTEE with ICE or TTETTE and CT: right ventricle, annulus, leadsLeaflet capture, residual TR, diastolic gradient, lead position
Left atrial appendage occlusionTEE or ICE (SCAI/HRS 2025)TEE or CT before and afterThrombus exclusion, landing zone, compression, peridevice leak, effusion
ASD or PFO closureTEE or ICETTE afterRims, disc position, residual shunt, impingement, effusion
Paravalvular leak closureTwo- and three-dimensional TEECT for complex defectsDefect map, wire crossing, residual leak, occluder interference with leaflets
Alcohol septal ablationTTE with myocardial contrastTTE afterPerforator territory limited to the basal septum at the point of mitral–septal contact
Cardiac surgeryTEEEpiaortic scanningPre-bypass diagnosis, post-bypass result, air, ventricular function

Imaging by structural procedure. MR, mitral regurgitation; ASD, atrial septal defect; PFO, patent foramen ovale; TR, tricuspid regurgitation; LVOT, left ventricular outflow tract; ICE, intracardiac echocardiography.

Transseptal puncture. Imaging shows the "tent" that the needle raises in the fossa ovalis before it advances. The bicaval view shows superior–inferior position, the aortic valve short-axis view shows anterior–posterior position, and the four-chamber view shows the height of the tent above the target. A tent directed toward the aortic root risks aortic puncture, and one against the posterior atrial wall risks exit from the heart. The site is chosen for the procedure. Mitral edge-to-edge repair needs a superior and posterior site with adequate height. The 2026 American Society of Echocardiography guideline gives typically 4.5 to 5.0 cm above the anticipated point of device deployment. Medial targets and prolapsing or flail leaflets need the higher end of this range, and lateral targets and the restricted leaflets of secondary regurgitation tolerate less. Appendage occlusion usually uses an inferior and posterior site aligned with the appendage axis. PFO closure crosses the tunnel. A new pericardial effusion is excluded after crossing.

Mitral edge-to-edge repair. The commissural view at about 60° shows medial–lateral position and the long-axis view at about 120° shows anterior–posterior position; biplane imaging displays both. The three-dimensional en face view shows the device arms perpendicular to the line of coaptation. Leaflet capture is proved anatomically by both leaflets inserted to adequate depth in more than one view and a tissue bridge forming a double orifice in the transgastric short axis; a smaller color jet in one plane does not prove capture. Before release, imaging assesses residual regurgitation by integrating several parameters. It also assesses the mean gradient at a stated heart rate (a result at or below about 5 mm Hg is the usual target), pulmonary venous flow, stability, the iatrogenic atrial septal defect, and the pericardium. Anatomic selection and outcomes are covered on the mitral regurgitation page.

Transcatheter aortic valve replacement. Gated CT provides annular sizing, coronary heights, and access; two-dimensional TEE annular diameter is not the modern sizing method. Many transfemoral procedures are performed under moderate sedation with TTE and fluoroscopy, and TEE is used under general anesthesia or when complications arise. After deployment, echocardiography confirms valve position and function, records the baseline gradient, and grades paravalvular regurgitation in short axis at and below the valve frame. Grading uses the combined circumferential extent of the jets (less than 10% mild, 10% to 29% moderate, 30% or more severe in the 2024 ASE guideline), integrated with other parameters. Sudden hypotension prompts a search for tamponade from annular rupture or wire perforation of the ventricle, severe regurgitation, coronary obstruction (new regional wall motion abnormality), mitral apparatus injury, aortic dissection, and valve embolization.

Transcatheter mitral replacement. The valve frame displaces the anterior mitral leaflet, or the leaflets of a surgical bioprosthesis, toward the septum and creates a neo-left ventricular outflow tract. Virtual implantation on CT predicts the risk of obstruction before the procedure; during it, TEE guides access and deployment and then measures outflow and transmitral gradients, paravalvular regurgitation, stability, and the pericardium.

Left atrial appendage occlusion. Before implantation, TEE or CT excludes thrombus and defines the ostium, landing zone, depth, and lobes; the 2025 SCAI/HRS guideline suggests TEE or CT before the day of the procedure over omitting it (conditional recommendation, very low certainty). During implantation the guideline suggests either ICE or TEE (conditional recommendation, low certainty), chosen by expertise, anesthesia plan, and visualization of the septum, device, and pericardium. Imaging confirms device position, stability, compression, and peridevice flow in several planes using device-specific criteria. Before discharge, TTE detects effusion and device embolization; it cannot exclude device-related thrombus or grade peridevice leak, which TEE or CT assesses at follow-up. Patient selection and antithrombotic therapy are on the atrial fibrillation page.

Septal ablation, pericardiocentesis, and biopsy. For alcohol septal ablation, an ultrasound-enhancing agent injected into the candidate septal perforator shows its territory before ethanol is given; opacification of a papillary muscle, the right ventricle, or a distant wall excludes the branch. During pericardiocentesis, agitated saline through the needle that opacifies the right ventricle indicates an intracardiac needle position. During endomyocardial biopsy and electrophysiology procedures, a new effusion is treated as perforation until shown otherwise.

Intracardiac echocardiography as an alternative to TEE

Intracardiac echocardiography uses a steerable phased-array catheter, usually advanced from the femoral vein into the right atrium, and is manipulated by the interventional operator. It avoids esophageal intubation and general anesthesia, and it is standard during many electrophysiology procedures and increasingly used for septal closure, appendage occlusion, and tricuspid interventions. Its planes depend on catheter position, rotation, and flexion, so orientation comes from chamber landmarks. From the mid right atrium the "home" view shows the tricuspid valve, right ventricle, and outflow tract, and clockwise rotation brings the aortic valve, the interatrial septum with the left atrium and mitral valve, the appendage, and the left pulmonary veins into view. Near-field saturation, reverberations from catheters and wires, and shadowing from devices are its common artifacts, and its catheter can perforate the atrium.

In a 2026 meta-analysis of 27 observational studies of appendage occlusion, procedural success with ICE was comparable to TEE, procedure time was shorter with ICE in adjusted analyses (mean difference about 21 minutes), and ICE was associated with more in-hospital pericardial effusions requiring intervention (adjusted odds ratio 1.74) and more residual iatrogenic atrial septal defects at less than 4 months (unadjusted odds ratio 1.61). The data are observational and reflect operator learning curves; the choice is made by program experience and the need to avoid general anesthesia.

Intraoperative transesophageal echocardiography in cardiac surgery

The 2010 ASA/SCA practice guideline states that, for adults without contraindications, TEE should be used in all open heart (for example, valvular) and thoracic aortic surgical procedures and should be considered in coronary artery bypass surgery. Its purposes are to confirm and refine the preoperative diagnosis, detect new or unsuspected pathology, adjust the anesthetic and surgical plan, and assess the result of the operation. The 2020 ASE/SCA/Society of Thoracic Surgeons guideline organizes intraoperative TEE around the surgical decisions specific to the operation.

Before bypass. The study confirms the diagnosis and severity and surveys ventricular function, all valves, the aorta, the appendage, and the septum even when the lesion is known, because unexpected tricuspid regurgitation, a PFO, aortic atheroma, or appendage thrombus can change the operation. It guides placement of cannulas, the coronary sinus catheter for retrograde cardioplegia, and intra-aortic balloon pumps, and disagreement with preoperative imaging is communicated before bypass.

Loading conditions. General anesthesia, positive-pressure ventilation, vasodilation, volume shifts, and inotropes alter regurgitant lesions, dynamic obstruction, and gradients. Lower afterload makes mitral regurgitation, especially secondary regurgitation, appear smaller; lower preload and inotropes worsen dynamic outflow obstruction; positive-pressure ventilation raises right ventricular afterload. When intraoperative and preoperative grades conflict, loading is restored toward awake values with a vasopressor such as phenylephrine and volume before the lesion is regraded. Diastolic function algorithms derived from awake patients are not intended for intraoperative use.

After bypass. Transient global dysfunction after cardioplegia is common and is interpreted with pacing, loading, and inotropes. A new regional abnormality in a coronary distribution suggests coronary compromise: the circumflex artery runs near the posterior mitral annulus, and the coronary ostia are at risk in aortic valve and root surgery. Air collects in the uppermost structures of the supine patient (left ventricular apex, left atrial dome, right upper pulmonary vein, and right coronary sinus) and appears as mobile bright echoes that change with de-airing. Air entering the right coronary artery, whose ostium is uppermost, produces transient inferior and right ventricular dysfunction.

After mitral repair, TEE assesses residual regurgitation and its mechanism, the transmitral gradient, and systolic anterior motion with blood pressure restored toward awake values. Residual regurgitation more than mild, a significant gradient, or persistent obstruction can justify a return to bypass. Systolic anterior motion with outflow obstruction after repair often resolves when inotropes are stopped, volume is given, afterload is raised, and heart rate is slowed. After valve replacement, TEE confirms leaflet or occluder motion, seating, gradients, and the absence of significant paravalvular regurgitation before chest closure; small washing jets within a mechanical valve are normal. A high early gradient is interpreted with flow (anemia, inotropes, tachycardia), using the dimensionless index and acceleration time.

Surgical appendage exclusion. In the Left Atrial Appendage Occlusion Study III (LAAOS III), 4,811 patients with atrial fibrillation and a CHA₂DS₂-VASc score of 2 or more undergoing cardiac surgery were randomized to occlusion or no occlusion, with intraoperative TEE recommended to confirm closure. In the primary analysis population of 4,770, stroke or systemic embolism over a mean of 3.8 years was 4.8% versus 7.0% (hazard ratio 0.67). The 2023 guideline recommends a technique that leaves no flow across the suture line and a stump shorter than 1 cm on intraoperative TEE (Class 1, LOE A).

Aortic surgery. TEE evaluates the visible aorta, aortic regurgitation and its mechanism, ventricular function, and the pericardium and, in dissection, the flap, entry site, and coronary involvement. It cannot establish that the whole thoracic aorta is normal, because of the tracheal blind spot; epiaortic scanning with a sheathed high-frequency transducer placed on the ascending aorta covers that segment and identifies atheroma at cannulation and cross-clamp sites.

TEE in noncardiac surgery, critical care, and cardiac arrest

Noncardiac surgery. The 2010 ASA/SCA guideline states that TEE may be used when the nature of the planned surgery or the patient's known or suspected cardiovascular disease might result in severe hemodynamic, pulmonary, or neurologic compromise, and that, if equipment and expertise are available, TEE should be used when unexplained, life-threatening circulatory instability persists despite corrective therapy. The consultant and member surveys underlying the guideline supported TEE during lung transplantation and major abdominal or thoracic trauma and opposed its routine use in orthopedic surgery. The basic 11-view examination answers most monitoring questions: ventricular filling and function, regional wall motion, pericardial effusion, valve failure, and right ventricular failure from pulmonary embolism, gas, or fat embolism.

Critical care. The same guideline states that TEE should be used in critically ill patients when diagnostic information expected to alter management cannot be obtained by TTE or other modalities in a timely manner. Mechanical ventilation, dressings, chest tubes, obesity, and positioning limit transthoracic windows, and TEE then answers questions such as tamponade after cardiac surgery (including loculated posterior collections compressing the left atrium), acute valve dysfunction, hypoxemia from a PFO under positive-pressure ventilation, endocarditis, and aortic dissection. The 2024 Society of Critical Care Medicine focused update suggests critical care ultrasonography to guide management of septic shock, acute respiratory failure, and cardiogenic shock, and found insufficient data to recommend it over standard care in cardiac arrest. Postoperative tamponade is covered on the pericardial effusion and tamponade page.

Cardiac arrest. TEE images the heart continuously during chest compressions without requiring pauses, identifies the area of maximal compression, and can reveal causes such as tamponade, massive pulmonary embolism, hypovolemia, aortic dissection, and ventricular fibrillation that is hard to see on the monitor. A 2026 International Liaison Committee on Resuscitation scoping review of 22 studies including 1,254 patients, 84% with out-of-hospital arrest, found successful insertion in 80% to 100%, changes in compression position in 19% to 76%, changes in management in 31% to 48%, and identification of a possible cause in 20% to 86%. Two randomized trials found no difference in return of spontaneous circulation, survival, or neurologic outcome with TEE-guided compared with conventional resuscitation. In observational data, compression over the outflow tract or aortic valve was associated with lower rates of return of spontaneous circulation. Resuscitative TEE is therefore feasible and changes management, but it has not been shown to improve patient outcomes. Point-of-care ultrasound during arrest is discussed on the adult cardiac arrest page.

Complications of transesophageal echocardiography and their rates

The largest early safety survey, from 15 European centers, analyzed 10,419 attempted examinations, most in conscious patients without intravenous sedation. Insertion failed in 1.9%, mostly because of poor cooperation or operator inexperience. Among 10,218 successful insertions, 0.88% of studies were interrupted, most often because of intolerance (65 cases) and less often because of pulmonary (8), cardiac (8), or bleeding (2) complications. One death occurred, from bleeding of a malignant lung tumor that infiltrated the esophagus (mortality 0.0098%). In 7,200 anesthetized adult cardiac surgical patients, TEE-associated morbidity was 0.2% and mortality 0%. Severe odynophagia occurred in 0.1%, dental injury, endotracheal tube malposition, and upper gastrointestinal hemorrhage at the same rate of 0.03%, and esophageal perforation in 0.01%, while insertion was unsuccessful in 0.18% and contraindicated in 0.5%.

ComplicationDiagnostic TEEIntraoperative TEE
Overall complication rate0.18–2.8%0.2%
Mortality<0.01–0.02%0%
Major morbidity0.2%0–1.2%
Major bleeding<0.01%0.03–0.8%
Esophageal perforation<0.01%0–0.3%
Heart failure0.05%—
Arrhythmia0.06–0.3%—
Laryngospasm0.14%—
Bronchospasm0.06–0.07%—
Unplanned tracheal intubation0.02%—
Endotracheal tube malposition—0.03%
Minor pharyngeal bleeding0.01–0.2%0.01%
Severe odynophagia—0.1%
Dysphagia—1.8%
Hoarseness—12%
Lip injury—13%
Dental injury—0.03–0.1%

Reported complication rates, from Table 7 of the 2013 ASE/SCA guideline, which compiles diagnostic and intraoperative series. Rates of minor injury depend on how carefully they are sought; the hoarseness, lip injury, and dysphagia figures come from a series that questioned patients systematically.

Oropharyngeal and dental injury. Sore throat is the most common complaint after TEE. Lip, tongue, and dental injuries occur during insertion and biting; loose teeth and dental work are documented beforehand, and dentures are removed. Pharyngeal hematoma and vocal cord or arytenoid injury are rare.

Esophageal and gastric injury. Perforation is the most serious complication. It results from forceful insertion, excessive flexion, or prolonged manipulation, and esophageal disease and long procedures increase the risk. Presentation with chest or neck pain, fever, subcutaneous emphysema, or sepsis can be delayed after an apparently uneventful study, so new odynophagia, dysphagia, or chest pain after TEE leads to CT with oral contrast or endoscopy. Structural interventions carry particular risk because the probe remains in position, often flexed, for long periods. In a prospective study of 50 patients undergoing TEE-guided structural interventions with endoscopy before and after the procedure, 86% had a new esophageal or gastric injury and 40% had complex lesions (intramural hematoma or mucosal laceration). Longer TEE manipulation time (odds ratio 1.27 per 10 minutes) and poor image quality predicted complex lesions, and these patients more often had dysphagia or odynophagia. Minimizing time with the probe flexed, releasing flexion during pauses, and avoiding unnecessary manipulation reduce this risk.

Bleeding. Major bleeding is reported in fewer than 0.01% of diagnostic studies and 0.03% to 0.8% of intraoperative studies; the one death in the European survey followed bleeding from a tumor infiltrating the esophagus. The risk with varices is described under contraindications.

Cardiorespiratory complications. Hypoxemia from sedation or aspiration, laryngospasm, bronchospasm, transient arrhythmias (supraventricular tachycardia, atrial fibrillation, bradycardia, and rarely ventricular tachycardia), hypotension, and hypertension occur. Benzocaine-induced methemoglobinemia is described under topical pharyngeal anesthesia. A 2024 review groups complications by mechanism into complications of insertion (oropharyngeal and esophageal injury), complications of manipulation (gastrointestinal bleeding and swallowing dysfunction), and systemic complications (cardiovascular and respiratory).

Probe care: inspection, disinfection, electrical safety, and tip temperature

Inspection. Before use, the shaft, tip, and handle are inspected for cracks, punctures, bite marks, and seal failure, and the controls are tested through their full range. A damaged insertion tube is withdrawn from use, because cracks harbor organisms and can expose the patient to electrical leakage current. Electrical safety and leakage current testing follow the manufacturer's schedule, and competence in TEE includes knowledge of infection control and electrical safety.

Disinfection. TEE probes contact mucous membranes and are semicritical devices in the Spaulding classification, which requires high-level disinfection between patients. The probe is precleaned immediately after use, washed with an enzymatic detergent, disinfected with an agent approved for the probe (such as ortho-phthalaldehyde, glutaraldehyde, hydrogen peroxide, or peracetic acid, manually or in an automated reprocessor), rinsed, dried, and stored hanging in a clean, ventilated cabinet. Reprocessing is documented so that a probe can be traced to the patients it was used on.

Tip temperature. Ultrasound energy and probe electronics heat the transducer, and a thermistor in the tip measures its temperature. On one common platform the system warns when the tip reaches about 41 °C and freezes imaging at 42.5 °C until the tip cools below 42 °C, with a lower limit of 17.5 °C; thermal limits on other platforms are set by the manufacturer. Fever and long periods of continuous imaging raise tip temperature, and the system can freeze unexpectedly in a febrile patient. Increasing depth, reducing pulse repetition frequency, and freezing the image between acquisitions lower it. In 2019 one manufacturer recalled certain S7-3t and S8-3t TEE transducers that had been programmed incorrectly and could fail to initiate automatic cooling (US Food and Drug Administration Class 2 recall), illustrating that the safeguard depends on correct device programming. During long monitoring in the operating room or intensive care unit, the probe is left in a neutral position and imaging is unfrozen only when images are needed.

Artifacts on transesophageal imaging

All transthoracic artifacts also occur on TEE, and strong reflectors close to the transducer make several of them prominent.

ArtifactTypical TEE settingRecognitionResolution
Linear reverberationAscending aorta (mimics a dissection flap); pulmonary artery catheter; posterior aortic wallLies at twice the depth of the reflector, moves with it, crosses anatomic boundaries, no color flow difference across itDifferent plane and depth; M-mode; CT when doubt persists
Mirror imageDescending aorta against aerated lung ("double aorta")Second image deep to a strong interface, identical in shapeRecognition of the anatomy; a change in angle
Side lobe and grating lobeCalcified mitral annulus, prosthetic rings, devices projecting into the atrium or appendageFaint, curved echo at the reflector's depth that sweeps across boundariesNarrower sector, different plane, lower gain
Acoustic shadowingProsthetic valves, annular calcification, devicesAbsence of echoes and color beyond a strong reflectorTransgastric or deep transgastric views; TTE for the opposite side
Near-field clutterPosterior left atrial wall, septumHaze close to the transducerHarmonic imaging, gain and focus adjustment
Septal dropoutThin fossa ovalis parallel to the beamApparent defect without flow through itPerpendicular plane; low-velocity color Doppler; saline contrast
Gain-related speckleLeft atrium and appendageRandom flicker that does not move with flow (distinct from spontaneous echo contrast)Lower gain; confirmation that the echoes do not move with flow
Beam-width and slice-thicknessAppendage, aortaOut-of-plane structure projected into a cavity; disappears with slight angulationOrthogonal plane; biplane imaging

Common artifacts on TEE and how they are recognized. A suspected mass or flap is confirmed in an orthogonal plane, checked against multiples of reflector depths, and tested with color Doppler and changes in depth or frequency before it is reported.

Pregnancy, older adults, and other special populations

Pregnancy. TEE is performed in pregnancy when the answer will change management, most often for endocarditis, prosthetic valve thrombosis, aortic dissection, or mitral stenosis before balloon commissurotomy. Aspiration risk rises in later pregnancy because of delayed gastric emptying and lower esophageal sphincter tone, so anesthesia support and aspiration precautions are used, and the left lateral position avoids aortocaval compression. The 2013 ASE/SCA guideline advises avoiding benzodiazepines in pregnancy, particularly in the first trimester. Mechanical valve thrombosis in pregnancy is covered on the prosthetic heart valves page.

Older adults. Older patients are more sensitive to benzodiazepines and opioids, so starting doses are reduced (opioid doses by half in the frail), and delirium after sedation is more frequent. Cervical spine disease limits neck flexion, and Zenker diverticulum, which is more common with age, is a reason to stop when resistance is felt. Osteophytes and kyphosis make insertion more difficult.

Patients with obesity, sleep apnea, or a difficult airway. Moderate sedation carries a higher risk of airway obstruction and hypoxemia, so capnography and anesthesia support are used at a low threshold.

Patients receiving anticoagulation or with thrombocytopenia. Most precardioversion TEE studies are performed on therapeutic anticoagulation. Severe thrombocytopenia or coagulopathy is a relative contraindication, and correctable factors are addressed when the study is not urgent.

Patients with ventricular assist devices, extracorporeal support, or transplantation. TEE guides cannula position, septal position, and aortic valve opening during device optimization and detects inflow cannula obstruction, thrombus, and tamponade. Rotary pumps and cannulas produce characteristic artifacts, and Doppler of the inflow and outflow cannulas follows device-specific criteria.

Training, competence, and certification

The Core Cardiovascular Training Statement 4 (COCATS 4, 2015) places competence in nonintraoperative TEE within Level II echocardiography training. Minimum training for independent performance and interpretation is 50 complete TEE examinations, including 25 esophageal intubations, in addition to Level II transthoracic training. The ASE also endorses 50 additional TEE studies with partial performance or bedside instruction, and training in an institution that performs at least 500 TEE studies a year. Simulation increasingly supplements clinical experience. Competence includes knowledge of indications, contraindications, and complications and experience with conscious sedation. For cardiovascular fellows, competence in intraoperative TEE requires additional supervised intraoperative experience and an understanding of cardiac surgery and cardiopulmonary bypass. The 2019 ACC/AHA/ASE advanced training statement defines Level III competencies, including advanced TEE, three-dimensional imaging, and procedural guidance.

The National Board of Echocardiography administers the Adult Comprehensive Echocardiography examination (ASCeXAM), which includes TEE, and the Advanced Perioperative Transesophageal Echocardiography examination (Advanced PTEeXAM); certification combines a passing examination with documented clinical experience. Laboratory accreditation by the Intersocietal Accreditation Commission includes standards for adult TEE.

Reporting a transesophageal study

The report states the indication and the clinical question, the type of sedation or anesthesia and who administered it, topical anesthesia, the number of insertion attempts, any difficulty or complication, and the patient's condition at the end. It distinguishes a comprehensive examination from a targeted one and lists the structures not examined.

Findings are given by structure with the views used. For the appendage, the report states whether thrombus is present, absent, or indeterminate, the planes examined, the spontaneous echo contrast and its density, and the emptying velocity. For the septum, it gives PFO, aneurysm, defect, shunt direction, and the results of saline injections with and without provocation. For the valves, it gives the mechanism, segments, and severity with the loading conditions at the time. For the aorta, it gives plaque thickness and morphology and the segments not visualized because of the tracheal blind spot. Any masses are described by attachment, size, and mobility. Measurements made under sedation or anesthesia are reported with heart rate and blood pressure, because both alter regurgitant severity and gradients. Findings that require immediate action, such as thrombus before a planned cardioversion, a new abscess, or an aortic flap, are communicated to the treating team.

Sources of error in transesophageal echocardiography

References

  1. Hahn RT, Abraham T, Adams MS, et al. Guidelines for performing a comprehensive transesophageal echocardiographic examination: recommendations from the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists. J Am Soc Echocardiogr. 2013;26:921–964. doi:10.1016/j.echo.2013.07.009
  2. Reeves ST, Finley AC, Skubas NJ, et al. Basic perioperative transesophageal echocardiography examination: a consensus statement of the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists. J Am Soc Echocardiogr. 2013;26:443–456. doi:10.1016/j.echo.2013.02.015
  3. Nicoara A, Skubas N, Ad N, et al. Guidelines for the use of transesophageal echocardiography to assist with surgical decision-making in the operating room: a surgery-based approach. From the American Society of Echocardiography in collaboration with the Society of Cardiovascular Anesthesiologists and the Society of Thoracic Surgeons. J Am Soc Echocardiogr. 2020;33:692–734. doi:10.1016/j.echo.2020.03.002
  4. American Society of Anesthesiologists and Society of Cardiovascular Anesthesiologists Task Force on Transesophageal Echocardiography. Practice guidelines for perioperative transesophageal echocardiography: an updated report. Anesthesiology. 2010;112:1084–1096. doi:10.1097/ALN.0b013e3181c51e90
  5. Douglas PS, Garcia MJ, Haines DE, et al. ACCF/ASE/AHA/ASNC/HFSA/HRS/SCAI/SCCM/SCCT/SCMR 2011 appropriate use criteria for echocardiography. J Am Coll Cardiol. 2011;57:1126–1166. doi:10.1016/j.jacc.2010.11.002
  6. Joglar JA, Chung MK, Armbruster AL, et al. 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2024;83:109–279. doi:10.1016/j.jacc.2023.08.017
  7. Van Gelder IC, Rienstra M, Bunting KV, et al. 2024 ESC guidelines for the management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. 2024;45:3314–3414. doi:10.1093/eurheartj/ehae176
  8. Klein AL, Grimm RA, Murray RD, et al. Use of transesophageal echocardiography to guide cardioversion in patients with atrial fibrillation (ACUTE). N Engl J Med. 2001;344:1411–1420. doi:10.1056/NEJM200105103441901
  9. Fatkin D, Kuchar DL, Thorburn CW, Feneley MP. Transesophageal echocardiography before and during direct current cardioversion of atrial fibrillation: evidence for “atrial stunning” as a mechanism of thromboembolic complications. J Am Coll Cardiol. 1994;23:307–316. doi:10.1016/0735-1097(94)90412-x
  10. Fatkin D, Kelly RP, Feneley MP. Relations between left atrial appendage blood flow velocity, spontaneous echocardiographic contrast and thromboembolic risk in vivo. J Am Coll Cardiol. 1994;23:961–969. doi:10.1016/0735-1097(94)90644-0
  11. Goldman ME, Pearce LA, Hart RG, et al. Pathophysiologic correlates of thromboembolism in nonvalvular atrial fibrillation: I. Reduced flow velocity in the left atrial appendage (the Stroke Prevention in Atrial Fibrillation [SPAF-III] study). J Am Soc Echocardiogr. 1999;12:1080–1087. doi:10.1016/s0894-7317(99)70105-7
  12. Stroke Prevention in Atrial Fibrillation Investigators Committee on Echocardiography. Transesophageal echocardiographic correlates of thromboembolism in high-risk patients with nonvalvular atrial fibrillation. Ann Intern Med. 1998;128:639–647. doi:10.7326/0003-4819-128-8-199804150-00005
  13. Agmon Y, Khandheria BK, Gentile F, Seward JB. Echocardiographic assessment of the left atrial appendage. J Am Coll Cardiol. 1999;34:1867–1877. doi:10.1016/s0735-1097(99)00472-6
  14. Saric M, Armour AC, Arnaout MS, et al. Guidelines for the use of echocardiography in the evaluation of a cardiac source of embolism. J Am Soc Echocardiogr. 2016;29:1–42. doi:10.1016/j.echo.2015.09.011
  15. Mügge A, Daniel WG, Haverich A, Lichtlen PR. Diagnosis of noninfective cardiac mass lesions by two-dimensional echocardiography: comparison of the transthoracic and transesophageal approaches. Circulation. 1991;83:70–78. doi:10.1161/01.cir.83.1.70
  16. Silvestry FE, Cohen MS, Armsby LB, et al. Guidelines for the echocardiographic assessment of atrial septal defect and patent foramen ovale: from the American Society of Echocardiography and Society for Cardiac Angiography and Interventions. J Am Soc Echocardiogr. 2015;28:910–958. doi:10.1016/j.echo.2015.05.015
  17. Amarenco P, Cohen A, Tzourio C, et al. Atherosclerotic disease of the aortic arch and the risk of ischemic stroke. N Engl J Med. 1994;331:1474–1479. doi:10.1056/NEJM199412013312202
  18. Isselbacher EM, Preventza O, Hamilton Black J 3rd, et al. 2022 ACC/AHA guideline for the diagnosis and management of aortic disease: a report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines. Circulation. 2022;146:e334–e482. doi:10.1161/CIR.0000000000001106
  19. Daniel WG, Mügge A, Martin RP, et al. Improvement in the diagnosis of abscesses associated with endocarditis by transesophageal echocardiography. N Engl J Med. 1991;324:795–800. doi:10.1056/NEJM199103213241203
  20. Otto CM, Nishimura RA, Bonow RO, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2021;143:e72–e227. doi:10.1161/CIR.0000000000000923
  21. Delgado V, Ajmone Marsan N, de Waha S, et al. 2023 ESC guidelines for the management of endocarditis. Eur Heart J. 2023;44:3948–4042. doi:10.1093/eurheartj/ehad193
  22. Victor F, De Place C, Camus C, et al. Pacemaker lead infection: echocardiographic features, management, and outcome. Heart. 1999;81:82–87. doi:10.1136/hrt.81.1.82
  23. Zoghbi WA, Jone PN, Chamsi-Pasha MA, et al. Guidelines for the evaluation of prosthetic valve function with cardiovascular imaging: a report from the American Society of Echocardiography developed in collaboration with the Society for Cardiovascular Magnetic Resonance and the Society of Cardiovascular Computed Tomography. J Am Soc Echocardiogr. 2024;37:2–63. doi:10.1016/j.echo.2023.10.004
  24. Zoghbi WA, Adams D, Bonow RO, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American Society of Echocardiography developed in collaboration with the Society for Cardiovascular Magnetic Resonance. J Am Soc Echocardiogr. 2017;30:303–371. doi:10.1016/j.echo.2017.01.007
  25. Nagueh SF, Smiseth OA, Appleton CP, et al. Recommendations for the evaluation of left ventricular diastolic function by echocardiography: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2016;29:277–314. doi:10.1016/j.echo.2016.01.011
  26. Lang RM, Badano LP, Tsang W, et al. EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography. J Am Soc Echocardiogr. 2012;25:3–46. doi:10.1016/j.echo.2011.11.010
  27. Hahn RT, Saric M, Faletra FF, et al. Recommended standards for the performance of transesophageal echocardiographic screening for structural heart intervention: from the American Society of Echocardiography. J Am Soc Echocardiogr. 2022;35:1–76. doi:10.1016/j.echo.2021.07.006
  28. Goldsweig AM, Glikson M, Joza J, et al. 2025 SCAI/HRS clinical practice guidelines on transcatheter left atrial appendage occlusion. J Soc Cardiovasc Angiogr Interv. 2025;4:103783. Corrigendum: J Soc Cardiovasc Angiogr Interv. 2026;5:104164. doi:10.1016/j.jscai.2025.103783
  29. Yaser H, Yaser S, Mohammed HE, et al. Intracardiac versus transesophageal echocardiography for left atrial appendage occlusion: a meta-analysis. JACC Adv. 2026;5:102815. doi:10.1016/j.jacadv.2026.102815
  30. Whitlock RP, Belley-Cote EP, Paparella D, et al. Left atrial appendage occlusion during cardiac surgery to prevent stroke (LAAOS III). N Engl J Med. 2021;384:2081–2091. doi:10.1056/NEJMoa2101897
  31. Daniel WG, Erbel R, Kasper W, et al. Safety of transesophageal echocardiography: a multicenter survey of 10,419 examinations. Circulation. 1991;83:817–821. doi:10.1161/01.cir.83.3.817
  32. Kallmeyer IJ, Collard CD, Fox JA, Body SC, Shernan SK. The safety of intraoperative transesophageal echocardiography: a case series of 7200 cardiac surgical patients. Anesth Analg. 2001;92:1126–1130. doi:10.1097/00000539-200105000-00009
  33. Hilberath JN, Oakes DA, Shernan SK, et al. Safety of transesophageal echocardiography. J Am Soc Echocardiogr. 2010;23:1115–1127. doi:10.1016/j.echo.2010.08.013
  34. Freitas-Ferraz AB, Bernier M, Vaillancourt R, et al. Safety of transesophageal echocardiography to guide structural cardiac interventions. J Am Coll Cardiol. 2020;75:3164–3173. doi:10.1016/j.jacc.2020.04.069
  35. Zhang L, Xie Y, Ren Z, Xie M. Transesophageal echocardiography related complications. Front Cardiovasc Med. 2024;11:1410594. doi:10.3389/fcvm.2024.1410594
  36. Hui RW, Leung CM. Incidence of gastrointestinal bleeding after transesophageal echocardiography in patients with gastroesophageal varices: a systematic review and meta-analysis. J Am Soc Echocardiogr. 2022;35:387–394. doi:10.1016/j.echo.2021.11.014
  37. Kane GC, Hoehn SM, Behrenbeck TR, Mulvagh SL. Benzocaine-induced methemoglobinemia based on the Mayo Clinic experience from 28 478 transesophageal echocardiograms: incidence, outcomes, and predisposing factors. Arch Intern Med. 2007;167:1977–1982. doi:10.1001/archinte.167.18.1977
  38. Kindel TL, Wang AY, Wadhwa A, et al. Multisociety clinical practice guidance for the safe use of glucagon-like peptide-1 receptor agonists in the perioperative period. Surg Obes Relat Dis. 2024;20:1183–1186. doi:10.1016/j.soard.2024.08.033
  39. Choudhury M, Chauhan S. Hyperthermia and malfunction of transesophageal echocardiographic probe. Indian J Crit Care Med. 2016;20:380–381. doi:10.4103/0972-5229.183902
  40. Kanagala P, Bradley C, Hoffman P, Steeds RP. Guidelines for transoesophageal echocardiographic probe cleaning and disinfection from the British Society of Echocardiography. Eur J Echocardiogr. 2011;12:i17–i23. doi:10.1093/ejechocard/jer095
  41. Scquizzato T, Zelop CM, Purkarthofer D, et al. Transoesophageal echocardiography during cardiopulmonary resuscitation for cardiac arrest: a scoping review. Resuscitation. 2026;227:111261. doi:10.1016/j.resuscitation.2026.111261
  42. Díaz-Gómez JL, Sharif S, Ablordeppey E, et al. Society of Critical Care Medicine guidelines on adult critical care ultrasonography: focused update 2024. Crit Care Med. 2025;53:e447–e458. doi:10.1097/CCM.0000000000006530
  43. Ryan T, Berlacher K, Lindner JR, et al. COCATS 4 Task Force 5: training in echocardiography. J Am Coll Cardiol. 2015;65:1786–1799. doi:10.1016/j.jacc.2015.03.035
  44. Wiegers SE, Ryan T, Arrighi JA, et al. 2019 ACC/AHA/ASE advanced training statement on echocardiography (revision of the 2003 ACC/AHA clinical competence statement on echocardiography). J Am Coll Cardiol. 2019;74:377–402. doi:10.1016/j.jacc.2019.02.003
  45. Zabalgoitia M, Halperin JL, Pearce LA, et al. Transesophageal echocardiographic correlates of clinical risk of thromboembolism in nonvalvular atrial fibrillation. J Am Coll Cardiol. 1998;31:1622–1626. doi:10.1016/s0735-1097(98)00146-6
  46. American Society of Echocardiography. Scope of practice for transesophageal echocardiography. Durham, NC: American Society of Echocardiography; 2018. asecho.org
  47. Alkhouli M, Du C, Killu A, et al. Clinical impact of residual leaks following left atrial appendage occlusion: insights from the NCDR LAAO Registry. JACC Clin Electrophysiol. 2022;8:766–778. doi:10.1016/j.jacep.2022.03.001
  48. US Food and Drug Administration. Class 2 device recall: S7-3t and S8-3t transesophageal transducers (recall Z-1843-2020; initiated August 30, 2019). accessdata.fda.gov
  49. Nagueh SF, Sanborn DY, Oh JK, et al. Recommendations for the evaluation of left ventricular diastolic function by echocardiography and for heart failure with preserved ejection fraction diagnosis: an update from the American Society of Echocardiography. J Am Soc Echocardiogr. 2025;38:537–569. doi:10.1016/j.echo.2025.03.011
  50. Little SH, Quader N, Brady M, et al. Guidelines for the intraprocedural imaging for mitral valve transcatheter edge-to-edge repair (M-TEER): recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2026;39:529–550. doi:10.1016/j.echo.2026.03.003