CardioBoardExam.com

Library › Imaging › Contrast, 3D and POCUS echo

Contrast, three-dimensional, and point-of-care echocardiography

Extensions of the standard echocardiogram in adults: ultrasound-enhancing agents and myocardial contrast, three-dimensional acquisition and quantification, and focused point-of-care cardiac and lung ultrasound.

An ultrasound-enhancing agent is indicated when two or more left ventricular segments are not adequately visualized for the assessment of function or wall motion (2018 American Society of Echocardiography guideline, Class I). Three-dimensional echocardiography measures chamber volumes without geometric assumptions and displays the mitral valve en face, as the surgeon sees it. Focused cardiac ultrasound answers a narrow bedside question with a qualitative scan by the treating clinician; it is documented as a focused examination and does not replace a diagnostic echocardiogram.

Extensions of the standard echocardiogram

Contrast, three-dimensional, and point-of-care echocardiography extend the standard transthoracic examination in different directions. Ultrasound-enhancing agents improve the signal from blood and the microcirculation, so that endocardial borders, apical anatomy, intracavitary masses, Doppler envelopes, and myocardial perfusion become measurable in patients whose unenhanced images are inadequate. Three-dimensional (3D) imaging adds a spatial dimension: chamber volumes are measured from the whole cavity, a valve or septal defect is displayed en face, and structural interventions are guided in real time. Point-of-care cardiac ultrasound (cardiac POCUS) moves imaging to the bedside and into the hands of the treating clinician, trading completeness and quantification for immediacy.

All three techniques change what the image shows, and all have artifacts and limitations that can be read as disease: dense bubbles shadow the basal segments, a stitched volume steps through the myocardium, and a normal pocket-device scan cannot exclude a valve lesion the device could not interrogate. Normal values for the standard examination are covered in the transthoracic echocardiography hub, and contrast during stress testing in the stress echocardiography hub. Three-dimensional transesophageal imaging is covered in the transesophageal echocardiography hub, Doppler principles in the Doppler hemodynamics hub, and ultrasound physics, the mechanical index, and bioeffects in the echocardiography physics and instrumentation hub.

Microbubble composition and the agents approved in the United States

An ultrasound-enhancing agent is a suspension of gas-filled microbubbles injected intravenously. A microbubble has a gas core inside a thin, flexible shell of phospholipid or denatured albumin. The gas, a perfluorocarbon or sulfur hexafluoride, has a high molecular weight and low solubility and diffuses out of the bubble slowly. Low solubility and the stabilizing shell let the bubbles survive passage through the pulmonary capillaries and persist in the circulation for minutes; an unshelled air bubble of the same size dissolves within seconds. Commercial microbubbles measure about 1.1 to 4.5 µm on average, similar to or smaller than a red blood cell, so they traverse pulmonary and systemic capillary beds and remain purely intravascular. They are blood-pool tracers: they do not cross the endothelium, and their distribution in the myocardium reflects microvascular blood volume.

Three agents are approved by the US Food and Drug Administration (FDA) for left ventricular opacification (LVO) and endocardial border delineation in patients with suboptimal echocardiograms. All three now carry adult and pediatric indications for this use. All other cardiovascular applications, including myocardial perfusion and stress imaging beyond border delineation, are off-label, although the 2018 American Society of Echocardiography (ASE) guideline recommends several such uses on the strength of the published evidence.

FeatureDefinity and Definity RTOptisonLumason
GasPerflutren (octafluoropropane)Perflutren (octafluoropropane)Sulfur hexafluoride
ShellPhospholipids, including a polyethylene glycol (PEG)–conjugated lipidHuman albumin (protein-type A microspheres)Phospholipids with palmitic acid; formulation contains PEG 4000
Mean diameter1.1 to 3.3 µm3.0 to 4.5 µm1.5 to 2.5 µm
PreparationMechanical activation of the vial (Definity refrigerated; Definity RT stored at room temperature)Resuspension by inversionReconstitution of lyophilized powder with saline
Labeled adult dose for LVOBolus 10 µL/kg over 30 to 60 s with a 10-mL saline flush, repeatable once after 30 min; or infusion of 1.3 mL in 50 mL saline at 4 mL/min, titrated to no more than 10 mL/min0.5 mL at no more than 1 mL/s, repeated as needed; maximum 5 mL in any 10 min and 8.7 mL per study2-mL bolus with a 5-mL saline flush; a second 2-mL dose may be given in the same examination
Labeled contraindicationHypersensitivity to perflutren lipid microspheres or their components, including PEGHypersensitivity to perflutren or albuminHypersensitivity to sulfur hexafluoride lipid microspheres or their components, including PEG
Maximum labeled mechanical index0.8 (0.3 in children)0.80.8

Ultrasound-enhancing agents approved in the United States, from the current FDA prescribing information (Definity revised 2024, Optison revised 2025, Lumason revised 2024). Outside the United States, Definity is marketed as Luminity and Lumason as SonoVue. The labels advise against imaging above a mechanical index of 0.8 because higher acoustic pressure can rupture the microspheres and has been associated with ventricular arrhythmias.

Doses used in practice are often smaller than the labeled bolus and are titrated to the image. The 2018 ASE guideline describes small boluses of 0.1 to 0.2 mL of Definity or 0.3 to 0.5 mL of Lumason or Optison, followed by a slow 5- to 10-mL saline flush over about 10 s, or continuous infusions at 3 to 5 mL/min of dilute agent (3% to 5% Definity or 10% Optison). A slow flush keeps a dense bolus from reaching the ventricle at once and shadowing the basal segments.

Microbubble acoustics: nonlinear oscillation and the mechanical index

Red blood cells are weak scatterers of ultrasound. A microbubble is far more compressible than blood or tissue, and it contracts during the compressional half of the acoustic cycle and expands during the rarefactional half. An oscillating bubble scatters sound many times more strongly than a red cell, most strongly near its resonant frequency, and bubbles of the clinical size resonate within the frequency range of transthoracic imaging.

The behavior of a bubble depends on the acoustic pressure it experiences, which the system displays as the mechanical index (MI). The MI is the peak rarefactional pressure (in megapascals) divided by the square root of the center frequency (in megahertz). At a very low MI, below 0.2, bubbles oscillate asymmetrically: the expansion phase is larger than the compression phase. The asymmetric oscillation returns energy at harmonics of the transmitted frequency, chiefly the second harmonic, and at other nonlinear frequencies. Tissue produces little harmonic signal at such low pressures, so at a very low MI the nonlinear signal comes almost entirely from bubbles, and bubbles survive for continuous imaging. As the MI rises, the shell ruptures and the gas dissolves or the bubble collapses. A brief burst at high MI is used deliberately to destroy bubbles in the scan plane.

SettingMechanical indexBubble behaviorUse
Very low MI, contrast-specific multipulse imagingBelow 0.2Stable nonlinear oscillation; bubbles preservedReal-time opacification, wall motion, myocardial perfusion
Low-MI harmonic imagingAbout 0.2 to 0.3Moderate destruction, most in the near fieldOpacification on systems without multipulse sequences
Flash (destruction) impulse0.8 to 1.2 for 5 to 15 framesDeliberate destruction in the scan planeClears myocardial contrast; starts the replenishment clock; sharpens the border
Conventional tissue harmonic imagingHigher than contrast presetsContinuous destruction, greatest at the apexNot used for contrast imaging

Mechanical index ranges for imaging with ultrasound-enhancing agents, from the 2018 ASE guideline. For left ventricular opacification at rest, the guideline describes a 5- to 10-frame flash at an MI of 0.8 to 1.2; for stress imaging, 5 to 15 frames at 0.8 to 1.0. Receiver gain changes neither the MI nor bubble behavior.

Because the MI is pressure divided by the square root of frequency, the same output power produces a higher MI at a lower transmit frequency. At a center frequency of 2.0 MHz, an MI of 0.15 corresponds to a peak rarefactional pressure of 0.15 × √2.0 = 0.21 MPa, and a flash at an MI of 1.0 corresponds to 1.41 MPa, about 6.7 times higher. If the frequency is lowered to 1.5 MHz at the same pressure, the MI rises to about 0.17. The displayed MI is therefore rechecked after any change in frequency or preset.

Contrast-specific pulse sequences and machine settings

Contrast presets separate the nonlinear bubble signal from the largely linear tissue echo by transmitting more than one pulse per scan line and combining the echoes. In pulse inversion, the second pulse is an inverted copy of the first; echoes from a linear scatterer are mirror images that cancel when summed, while the asymmetric bubble echoes leave a residual. In power (amplitude) modulation, the pulses differ in amplitude; the weaker echo is scaled and subtracted from the stronger one, which cancels linear echoes, whereas bubbles respond disproportionately to the stronger pulse and leave a residual. Many systems combine the two. These multipulse methods work at a very low MI and preserve bandwidth better than harmonic filtering alone. They have two drawbacks: several pulses per line lower the frame rate, and motion between pulses leaves incomplete cancellation of moving tissue.

The 2018 ASE guideline recommends very-low-MI multipulse imaging, with or without brief flash impulses, for regional wall motion analysis and quantification of ejection fraction (Class IIa), and recommends that manufacturers make contrast-specific presets and flash controls readily accessible. A contrast preset lowers the MI, disables spatial compounding, and redistributes gain. The focus is placed at or near the level of the mitral annulus, which spreads acoustic pressure away from the apex and reduces apical destruction. Gain is reduced once bubbles arrive, because the gain that suits unenhanced tissue saturates an opacified cavity and blooms over the endocardium.

Administration, monitoring, and laboratory protocol

  1. Indication and screening: the indication is documented from the unenhanced images, and the patient is asked about prior reactions to any ultrasound-enhancing agent and about hypersensitivity to PEG, including PEG-containing bowel preparations and laxatives (ASE 2021).
  2. Access: a peripheral intravenous catheter, preferably in an antecubital vein, permits a slow, controlled injection. The 2014 ASE sonographer guideline supports trained sonographers placing intravenous lines and administering agents under physician supervision.
  3. Preparation: the agent is activated, resuspended, or reconstituted according to its label and drawn up immediately before use; lipid agents settle and are resuspended before any bolus.
  4. Machine settings: the contrast preset is selected before injection, with the focus at the mitral annulus and depth set to include the whole ventricle.
  5. Injection: a small bolus with a slow saline flush is used for a resting study or a few stress views, and a titrated continuous infusion is used when steady opacification is needed for several views, serial stages, or perfusion.
  6. Acquisition: apical four-, two-, and three-chamber views, with parasternal views as needed, are acquired once the cavity is uniformly opacified and basal shadowing has cleared; flash impulses are used to clear myocardial contrast when the border is indistinct.
  7. Monitoring and readiness: resuscitation equipment and personnel trained to recognize and treat anaphylaxis are immediately available. Most serious reactions occur within 30 minutes of administration.
  8. Reporting: the agent, dose, route, and any adverse event are recorded, and the report states that measurements were obtained with contrast.

A bolus produces a rising and then falling concentration. The early peak can attenuate the far field, and the late tail can produce swirling, so the window for imaging lies between them. Boluses suit opacification at rest and during exercise or dobutamine stress, when a few views are acquired quickly. A continuous infusion maintains a steady, titratable concentration. Perfusion imaging requires it, because replenishment can be compared among segments and between rest and stress only when the concentration entering the myocardium is constant.

Indications for ultrasound-enhancing agents

The principal indication is inadequate endocardial border definition. The 2018 ASE guideline, repeating the 2008 consensus, recommends an enhancing agent at rest when two or more left ventricular segments cannot be visualized adequately for the assessment of ejection fraction and regional wall motion (Class I). The same Class I recommendation applies when the study indication requires accurate analysis of regional wall motion. The 2015 chamber quantification recommendations state the criterion as two or more contiguous segments not visualized in apical views. During stress, agents are used whenever adequate visualization of any coronary territory cannot be achieved at rest (Class I). The same criterion applies whether the poor window results from obesity, lung disease, mechanical ventilation, chest wall deformity, or postoperative dressings.

Application2018 ASE recommendationClass, level of evidence
Two or more segments not visualized for ejection fraction and wall motion at restAgent should be usedI, A
Quantitative ejection fraction important to prognosis or managementAgent should be used, with very-low-MI or low-MI harmonic imagingI, B-R
Stress echocardiography with any coronary territory not visualizedAgent should be usedI, A
Left ventricular thrombus that cannot be ruled in or out without contrastAgent should be usedI, B-NR
Left ventricular pseudoaneurysmAgent should be used to rule it in or outI, B-NR
Noncompaction, apical hypertrophy, or apical aneurysm not adequately assessed without contrastAgent should be consideredIIa, B-NR
Differential diagnosis of a cardiac mass by its vascularityAgent with very-low-MI imaging should be usedIIa, B-NR
Atrial appendage with dense spontaneous contrast or inadequate visualization on transesophageal imagingAgent should be consideredIIa, B-NR
Weak tricuspid regurgitant or stenotic Doppler signalEnhanced Doppler signals can be obtainedI, C-EO
Technically difficult studies in intensive care units and emergency departmentsAgent recommended; not withheld on the basis of any particular diagnosis or comorbidityI, B-NR
Suspected ischemia with a nondiagnostic electrocardiogram (ECG) in the emergency departmentRegional function assessment with contrast; perfusion assessment at experienced centersI, B-NR (function); IIa, B-NR (perfusion)
Myocardial perfusion with very-low-MI flash-replenishment imaging during stressAgent can be useful in trained handsIIa, B-R
Endoleak after endovascular aortic repairAgent recommended with low-MI imagingIIa, B-NR

Indications and recommendation classes from the 2018 ASE guideline. Only left ventricular opacification and endocardial border delineation are FDA-approved cardiac indications; the others are off-label uses supported by the guideline.

The European Association of Cardiovascular Imaging (EACVI) 2017 recommendations are broadly concordant with the ASE guideline, including the two-segment criterion, contrast during stress, and myocardial perfusion imaging in experienced centers.

Left ventricular opacification: volumes, ejection fraction, and wall motion

When unenhanced borders are poor, opacification converts a nondiagnostic or technically difficult study into a diagnostic one and improves the reproducibility of ejection fraction and wall motion analysis. In a prospective cohort of 632 consecutive patients with technically difficult studies, contrast lowered the proportion of uninterpretable studies from 11.7% to 0.3% and of technically difficult studies from 86.7% to 9.8%. The mean number of visualized segments rose from 11.6 to 16.8 of 17. Additional diagnostic procedures were avoided in 32.8% of patients, drug therapy was changed in 10.4%, and management changed overall in 35.6%. The impact was largest in intensive care units.

Opacification changes measured volumes. The opacified border lies at the true compacted endocardium, outside the trabeculations that unenhanced imaging often includes in the wall. The bright cavity also exposes foreshortening. Enhanced volumes are therefore usually larger than unenhanced volumes and closer to magnetic resonance values. The 2018 guideline notes that the normal range for ejection fraction does not appear to differ with contrast but that new reference ranges for enhanced end-diastolic and end-systolic volumes are needed. The 2015 chamber quantification limits are therefore applied to enhanced volumes with caution.

A patient with heart failure whose unenhanced study showed an end-diastolic volume of 118 mL and an end-systolic volume of 64 mL (ejection fraction 46%) may have 142 mL and 78 mL at follow-up with an enhancing agent (ejection fraction 45%). Both volumes are about 20% larger and the ejection fraction is unchanged, a pattern consistent with a change of technique; interval dilation is not reported, and the report states that the methods differ. Serial studies of the same patient, particularly during cancer therapy, use the same technique throughout.

Wall motion is read from the motion and thickening of the opacified endocardial border. Very-low-MI imaging with flash replenishment also shows the subendocardial layer, so that stress imaging can identify a subendocardial thickening abnormality that transmural analysis misses. Apical segments that are poorly seen without contrast are the most frequent beneficiaries, because near-field clutter obscures the apex in apical views.

Apical structure: thrombus, apical hypertrophy, noncompaction, aneurysm, and pseudoaneurysm

Left ventricular thrombus. An apical thrombus appears as a nonopacified filling defect within an opacified cavity, with a border distinct from the endocardium and usually in an akinetic or aneurysmal segment. In 121 patients at high risk after infarction or with heart failure, delayed-enhancement cardiac magnetic resonance served as the reference. Contrast echocardiography nearly doubled the sensitivity of noncontrast echocardiography for thrombus (61% versus 33%) and improved accuracy (92% versus 82%). Contrast echocardiography still missed small apical and mural thrombi detected by magnetic resonance, so a negative contrast study in a very high-risk patient does not exclude a small laminar thrombus. A statement that no thrombus is present is unsupported when the apex was not visualized; in that case the report states the limitation.

Apical hypertrophic cardiomyopathy. Unenhanced apical views often fail to show apical wall thickness because of near-field clutter. Opacification outlines the characteristic spade-shaped cavity at end-diastole, measures apical wall thickness, and shows apical systolic obliteration and any apical aneurysm with or without thrombus. An apical aneurysm in hypertrophic cardiomyopathy is a risk marker discussed in the hypertrophic cardiomyopathy hub.

Six apical echocardiographic views in two rows: the top row shows unenhanced apical four-, two-, and three-chamber views with an indistinct apex; the bottom row shows the same views after an ultrasound-enhancing agent, with a bright opacified cavity that narrows to a spade shape toward a thickened apex.
Apical hypertrophic cardiomyopathy in three apical views without contrast (top) and after an ultrasound-enhancing agent (bottom). The opacified cavity outlines the thickened apical walls that the unenhanced images do not define. Credit: Cotter B, Raisinghani A, DeMaria AN, Clin Cardiol 2022;45:1114–1122, Figure 4, via PubMed Central, CC BY 4.0.

Hypertrabeculation and noncompaction. Contrast fills the deep recesses between trabeculations and separates the noncompacted from the compacted layer, which permits measurement of their ratio. A higher MI at real-time frame rates destroys the slow-moving bubbles within the trabecular layer before they replenish while faster cavity bubbles refill, and this sharpens the distinction between the layers. Prominent trabeculation is common in athletes, pregnancy, and dilated ventricles, so the finding is interpreted with ventricular size and function.

Aneurysm and pseudoaneurysm. A true aneurysm has a wide neck and a wall made of thinned myocardium that moves dyskinetically. A pseudoaneurysm is a contained rupture with a narrow neck, an abrupt discontinuity of the myocardium, and a sac lined by pericardium and thrombus. Contrast outlines the neck, shows flow into the sac, and distinguishes the two; the 2018 guideline gives a Class I recommendation for its use to rule a pseudoaneurysm in or out. Mechanical complications of infarction are covered in the cardiogenic shock hub.

Cardiac masses: vascularity of thrombus and tumor

A mass is characterized by the enhancement of its own microcirculation at a very low MI, compared with the adjacent myocardium, and by the rate of replenishment within it after a flash impulse. Thrombus is avascular and shows no enhancement. Malignant tumors are usually highly vascular and enhance as much as or more than the myocardium. Benign stromal tumors such as myxoma usually show partial enhancement. Quantitative comparison of mass and myocardial intensity adds objectivity to the visual reading. Enhancement characteristics guide the next test; tissue characterization by magnetic resonance and, when needed, biopsy confirm the diagnosis.

Doppler signal enhancement

A small dose of agent can complete a weak spectral envelope, most often a tricuspid regurgitant signal needed to estimate pulmonary artery systolic pressure, and occasionally the jet of a stenotic valve. Bubbles brighten and broaden the spectrum, and a faint speckled fringe can extend beyond the true envelope. The dense envelope is traced at low gain with the smallest effective dose. With a dense envelope at 2.9 m/s, a fringe at 3.3 m/s, and a right atrial pressure of 8 mm Hg, the dense envelope gives a right ventricular systolic pressure of 4 × 2.9² + 8 = 42 mm Hg and the fringe 52 mm Hg. Tracing the fringe overestimates the pressure by about 10 mm Hg. Color jets also enlarge after contrast, so color Doppler assessment of regurgitation is completed before the agent is given. Estimation of pulmonary pressures is covered in the Doppler hemodynamics hub.

Contrast in stress echocardiography

Stress echocardiography depends on visualizing all segments at rest and at peak stress, when tachycardia and hyperventilation degrade image quality. Enhancing agents raise the proportion of interpretable segments, improve reader confidence and agreement, and allow testing in patients who would otherwise be referred to another modality. The 2018 ASE guideline recommends an agent whenever any coronary territory cannot be adequately visualized at rest (Class I) and very-low-MI imaging with intermittent flash impulses for homogeneous opacification during stress (Class IIa); the 2020 ASE stress guideline gives the same criterion. Doses, settings, and the large safety cohorts in stress testing are described in the stress echocardiography hub.

Myocardial contrast echocardiography: destruction and replenishment

Myocardial contrast echocardiography images microbubbles within the myocardial microcirculation. Because the bubbles stay within vessels and most of the myocardial blood volume lies in capillaries, the steady-state myocardial signal reflects microvascular blood volume. A bright cavity reflects only the blood pool and gives no information about perfusion.

During a constant infusion at a very low MI, a flash impulse destroys the bubbles within the scan plane, and their return is recorded. The refilling signal follows y(t) = A × (1 − e−βt), in which A, the plateau, is proportional to microvascular blood volume and β, the rate constant, is proportional to microbubble velocity; myocardial blood flow is proportional to A × β. This relation, validated experimentally in 1998, underlies both visual reading and quantitative analysis. A segment with a normal plateau can still refill slowly, which indicates reduced flow, so plateau brightness alone is not read as flow.

PhaseSettingMyocardium
Steady stateVery low MI; constant infusionUniform signal in perfused segments
FlashMI 0.8 to 1.2 for several framesBubbles destroyed; myocardium dark (the intended starting point, with no implication of a perfusion defect)
ReplenishmentVery low MISignal returns at a rate set by microvascular velocity
PlateauVery low MIBrightness set by microvascular blood volume
ReadingEnd-systolic frames, two or more viewsDelayed refill or low plateau within a coronary territory

The destruction–replenishment sequence. In a two-dimensional imaging plane with a constant infusion, replenishment is uniform and complete within 5 s at rest and within 2 s during any form of stress (ASE 2018). Delayed replenishment confined to a coronary territory and reproduced in a second view indicates reduced flow reserve.

Perfusion abnormalities precede wall motion abnormalities in the ischemic cascade, so delayed replenishment during dobutamine or exercise stress can appear before an inducible wall motion abnormality and adds sensitivity. Vasodilators raise flow in normal territories much more than in territories supplied by a stenotic artery. The resulting heterogeneity of replenishment appears without a wall motion abnormality, because vasodilators seldom cause ischemia; flash-replenishment imaging is therefore required with vasodilator stress. A subendocardial delay with prompt refill of the outer wall is a recognized pattern. A multicenter study enrolled 516 patients who underwent rest and vasodilator perfusion imaging with SonoVue and gated single-photon emission computed tomography (SPECT), with quantitative angiography as the reference. For stenosis of 70% or more, contrast perfusion had higher sensitivity than SPECT (75.2% versus 49.1%) and lower specificity (52.4% versus 80.6%). In several cohorts, delayed replenishment without a wall motion abnormality independently predicted death and nonfatal infarction.

After reperfused ST-elevation myocardial infarction (STEMI), absence of myocardial contrast within the infarct zone identifies microvascular obstruction despite an open artery, and preserved contrast within dysfunctional myocardium indicates viability. The 2018 guideline supports very-low-MI perfusion imaging after STEMI at institutions with the necessary expertise (Class IIa).

Myocardial perfusion is not an FDA-approved indication. In the United States it is performed routinely at a limited number of experienced centers, and the 2018 guideline recommends that physicians who perform it obtain additional training at a high-volume center. Stress protocols, scoring, and the randomized comparison of perfusion with wall motion imaging are covered in the stress echocardiography hub.

Contrast in critical care and the emergency department

Critically ill patients have the poorest acoustic windows and the most to gain from a diagnostic study. Mechanical ventilation, positive end-expiratory pressure, chest tubes, dressings, and limited positioning commonly leave segments unseen. The 2018 guideline recommends enhancing agents in all technically difficult intensive care and emergency department studies and states that contrast echocardiography should not be withheld on the basis of any particular diagnosis or comorbidity (Class I). In a propensity-matched analysis of critically ill patients, those who underwent contrast echocardiography had lower 48-hour mortality than those who did not (1.70% versus 2.50%; odds ratio 0.66) and lower hospital mortality (14.85% versus 15.66%). Causation was not established; earlier accurate diagnosis is the proposed mechanism. A registry of more than 4.3 million hospitalized patients found no increase in acute mortality with contrast.

In the emergency department, regional function assessment with contrast in patients with suspected ischemia and a nondiagnostic ECG adds diagnostic and prognostic information beyond clinical and electrocardiographic evaluation (Class I), and perfusion assessment adds further value at experienced centers (Class IIa). Physicians trained in focused cardiac ultrasound encounter the same poor windows. The 2018 guideline identifies their training in contrast use and interpretation as a need for future multisociety documents.

Transesophageal and vascular applications and pediatric labeling

During transesophageal echocardiography, an enhancing agent helps distinguish thrombus from dense spontaneous echo contrast, sludge, or pectinate muscle in the left atrial appendage when the appendage has dense spontaneous contrast or is incompletely seen (ASE 2018, Class IIa). Appendage imaging before cardioversion is covered in the transesophageal echocardiography hub. Vascular uses include detection and classification of endoleak after endovascular aneurysm repair (Class IIa), carotid plaque neovascularization, and skeletal muscle perfusion, the last two still investigational. All three agents now carry pediatric labeling for left ventricular opacification; the 2018 guideline, written before those approvals, noted the absence of safety data in children younger than 5 years. The ASE has commissioned a unified guideline update on the clinical use of ultrasound-enhancing agents; until it is published, the 2018 guideline and current labels govern practice.

Contrast artifacts and their correction

Enhancing agents produce characteristic artifacts, cataloged in the 2026 ASE artifact guideline and described with the other artifact families in the echocardiography physics hub. These artifacts have known mechanisms and corrections, and any of them can be mistaken for disease.

ArtifactMechanismAppearanceCorrection
Attenuation and shadowingDense near-field bubbles absorb and scatter the beamDark basal segments and mitral annulus in apical views, typically early after a bolusImaging after dispersal; lower dose or infusion rate; slower flush
SwirlingLow bubble concentration or apical bubble destructionMottled, rotating apical signal whose shape changes beat to beatHigher infusion rate; lower MI; focus at the mitral annulus
Near-field destructionMI above the very low range, shallow focus, high frame rateWeak apical signal mimicking thrombus or an apical defectLower MI; deeper focus; lower frame rate
BloomingExcess concentration or gainCavity signal spills over the endocardium; wall looks thinLower gain and dose
Doppler brighteningBubble signal broadens the spectrumFaint fringe beyond the true envelopeTracing of the dense envelope; lower Doppler gain
Apparent perfusion defectAttenuation, rib shadow, or lungBasal or lateral dark region outside a coronary patternConfirmation in a second view and against a coronary territory

Contrast artifacts. Swirling changes from beat to beat and clears with better opacification, whereas thrombus is a fixed defect with a stable shape. Attenuation lies in the far field of the view, deep to the dense cavity, and moves when the window changes.

Safety, the boxed warning, and contraindications

Serious reactions to enhancing agents are uncommon. They include anaphylactoid reactions and cardiopulmonary events such as hypotension, arrhythmia, and respiratory arrest, most within 30 minutes of administration. In a retrospective analysis of 78,383 doses of Definity and Optison, severe reactions occurred in 0.01% of doses and anaphylactoid reactions in 0.006%. The ASE estimates the rate of serious reactions at about 1 in 10,000 administrations. Many of these rare reactions are attributed to complement activation–related pseudoallergy, which cannot be distinguished from type I hypersensitivity without specialized testing. Treatment of both is the same (epinephrine, corticosteroids, antihistamines, bronchodilators, and supportive care). Minor effects include headache, nausea, flushing, altered taste, and back or flank pain with lipid agents.

YearFDA labeling action
2007Boxed warning added after reports of deaths temporally associated with the agents; new contraindications in pulmonary hypertension and unstable cardiopulmonary conditions; 30-minute monitoring required for all patients
2008Contraindications in pulmonary hypertension and unstable cardiopulmonary conditions changed to warnings; monitoring limited to those patients
2011–2012Requirement to monitor patients with pulmonary hypertension or unstable cardiopulmonary conditions removed for Definity (2011) and Optison (2012), along with the statement that safety in stress testing had not been established
2014Lumason approved in the United States for left ventricular opacification
2016–2017Contraindication in known or suspected right-to-left, bidirectional, or transient right-to-left shunts removed and replaced by a warning (Optison and Lumason in 2016, Definity thereafter)
2021FDA MedWatch alert on anaphylaxis in patients with PEG allergy; the ASE statement lists known PEG hypersensitivity as a contraindication to PEG-containing agents, and current Definity and Lumason labels list PEG among the components

History of FDA labeling of ultrasound-enhancing agents, from Table 6 of the 2018 ASE guideline and the 2021 ASE statement. Current labels for all three agents retain a boxed warning that serious cardiopulmonary reactions, including fatalities, have occurred uncommonly, most within 30 minutes. The warning requires assessment for conditions that preclude use and the ready availability of resuscitation equipment and trained personnel.

Current contraindications. The only labeled contraindication for all three agents is known or suspected hypersensitivity to the agent or its components: PEG for the lipid agents Definity and Lumason, and perflutren or albumin for Optison. A patient with a documented reaction to an agent does not receive it again; premedication and test doses do not remove the risk. The 2021 ASE statement on PEG allergy recommended asking all patients about PEG hypersensitivity and found no basis for changing monitoring policies. It estimated that reactions in patients with preexisting PEG allergy correspond to fewer than one per million lipid-stabilized doses.

Warnings that are not contraindications. The labels warn that the risk of cardiopulmonary reactions may be higher in patients with unstable cardiopulmonary conditions (acute myocardial infarction or acute coronary syndrome, worsening or unstable heart failure, serious ventricular arrhythmias); these patients are monitored, and the agent is not withheld when the study is needed. In patients with cardiac shunts, microspheres can bypass pulmonary filtration and enter the arterial circulation, and the labels advise assessment for embolic events. The 2018 ASE guideline states that the agents can be used safely in patients with pulmonary hypertension and with right-to-left shunts (Class I). The Definity label reports no clinically important hemodynamic changes in patients with pulmonary artery systolic pressure up to 75 mm Hg; higher pressures were not studied. After reports of vaso-occlusive crisis, the Definity label advises discontinuation if a patient with sickle cell disease develops new or worsening pain. All agents are for intravenous use only.

Pregnancy and lactation. The 2018 ASE guideline found no safety data in pregnancy and did not recommend use in pregnant patients until such data emerge. The current Definity label reports that published case reports have not identified a drug-associated risk of major birth defects, miscarriage, or adverse outcomes and that fetal exposure is not expected to be clinically relevant given the very short half-life; the Optison and Lumason labels report no human data and no adverse developmental outcomes in animal studies. Use in pregnancy is therefore limited to studies whose answer will change management and cannot be obtained without contrast.

Agitated saline compared with ultrasound-enhancing agents

Agitated saline and enhancing agents are both bubble contrast, but they answer different questions. Agitated saline bubbles are large, unstable, and filtered by the lungs, so they reach the left heart only through a right-to-left shunt. An enhancing agent crosses the lungs by design and opacifies the left heart in all patients, so it cannot identify a shunt.

FeatureUltrasound-enhancing agentAgitated saline
ContentsLow-solubility gas in a lipid or albumin shellAir, saline, and often a small volume of the patient's blood; no stable shell
Bubble sizeSimilar to or smaller than a red cellLarger and variable
Pulmonary passageCrosses the capillariesFiltered by the lungs
Left heart opacificationIn all patientsOnly through a right-to-left shunt or pulmonary arteriovenous malformation
Main usesEndocardial borders, apical structure, masses, perfusion, Doppler enhancementPatent foramen ovale and atrial septal defect, persistent left superior vena cava, pulmonary arteriovenous malformation
ImagingContrast-specific preset at very low MIStandard imaging with release of a Valsalva maneuver or cough

Enhancing agents and agitated saline. A patient with cryptogenic stroke and poor apical views may need both: agitated saline for the shunt and an enhancing agent for the ventricle. Bubble study technique and grading are described in the atrial septal defect and patent foramen ovale hub.

Sonothrombolysis, molecular imaging, and therapeutic ultrasound

Intermittent high-MI impulses from a diagnostic transducer, applied while microbubbles flow through a thrombosed vessel, produce inertial cavitation that disrupts thrombus and restores microvascular flow. In a randomized trial of 100 patients with a first STEMI, high-MI impulses were applied during an intravenous Definity infusion before and after primary percutaneous coronary intervention (PCI). Sonothrombolysis increased angiographic recanalization before PCI (48% versus 20%) and early ST-segment resolution, and it reduced infarct size on magnetic resonance (29 versus 40 g). Ejection fraction at 6 months was higher, and fewer patients met the ejection fraction threshold for a defibrillator (5% versus 18%). Sonothrombolysis is investigational; it is not an approved indication and is not part of guideline-directed STEMI care.

Microbubbles with ligands attached to the shell adhere to activated endothelium or to specific molecular targets and can image inflammation, angiogenesis, or thrombus. Microbubbles can also carry drugs or genes released by ultrasound at the target. These remain research applications.

Cost-effectiveness and downstream testing

The economic case for enhancing agents rests on converting nondiagnostic studies into diagnostic ones. A nondiagnostic echocardiogram usually leads to a second test, such as nuclear imaging, magnetic resonance, computed tomography (CT), or transesophageal echocardiography, with its own cost, delay, and in some cases radiation. In the prospective cohort described above, contrast avoided additional diagnostic procedures in one-third of patients and saved $122 per patient after the cost of the agent. Contrast also raises the positive predictive value of stress echocardiography by reducing false-positive wall motion readings from poorly seen segments, and it shortens the time to diagnosis in the emergency department. These benefits depend on laboratory workflow: standing protocols that let sonographers start an intravenous line and give the agent when the criterion is met avoid recalling the patient.

Reporting a contrast-enhanced study

The report states the indication for contrast (for example, two or more segments not visualized), the agent, the dose and route, bolus or infusion, the imaging mode, and any adverse reaction. Volumes and ejection fraction are identified as contrast-enhanced, and a comparison with a prior unenhanced study notes the difference in method. Thrombus or mass findings describe the enhancement pattern. Perfusion studies report segmental replenishment at rest and stress by coronary territory. Standardized reporting elements are set out in the 2019 ASE comprehensive transthoracic guideline.

Matrix-array transducers and volumetric beam formation

Three-dimensional echocardiography acquires a pyramidal volume of echo data in place of a single plane. A conventional phased array has one row of elements and steers only within its imaging plane. A matrix array arranges thousands of independently timed elements in a two-dimensional grid, and delay patterns across both directions of the grid steer and focus the beam in azimuth and elevation. The probe therefore sweeps a pyramid, or forms two or more live intersecting planes, without mechanical motion. Matrix arrays are built for transthoracic and transesophageal use.

A cable cannot carry a separate channel per element. Circuitry within the transducer combines groups of elements into subapertures, a process called micro-beamforming or subarray beamforming, and the system completes beam formation. Parallel receive processing is used far more extensively than in two-dimensional imaging: a transmitted pulse insonates a small block of the volume, and many receive lines are formed from it. The added dimension takes time, and all three-dimensional acquisitions balance field of view, line density, and volume rate against the patient's rhythm and breathing.

Three-dimensional imaging is chosen when the clinical question depends on spatial relationships or when a two-dimensional method depends on a geometric assumption. A biplane volume assumes a regular shape between two apical planes, and an apical view that misses the true apex underestimates volume. A volume that contains the whole cavity requires neither assumption. An en face view displays an entire irregular mitral lesion, septal defect, or paravalvular leak at once, which a series of planes cannot.

Volume rate, temporal resolution, and acquisition modes

A transmit event requires a round trip to the deepest displayed depth, about 13 µs per centimeter, and a volume contains the product of the scan lines in azimuth and in elevation. Volume time is therefore (lines in azimuth × lines in elevation ÷ receive lines per transmit) × 13 µs × depth in centimeters. A pyramid of 64 × 64 lines acquired to 16 cm with one receive line per transmit takes 4,096 × 208 µs, or 0.85 s, about one volume per second. With 16 parallel receive lines it takes 53 ms, about 19 volumes per second, or 15 volumes per cardiac cycle at a heart rate of 75 beats per minute. Even with extensive parallel processing, the volume rate is a fraction of a typical two-dimensional frame rate, so all modes limit volume size, line density, or the number of beats spanned. The same trade-off among depth, line density, and frame rate in two-dimensional imaging is described in the echocardiography physics hub.

ModeBeatsBest useMain limitation
Simultaneous biplane or multiplane imagingLiveAlignment, orifice planimetry, transseptal puncture, device guidanceTwo or three planes only; no volume
Live narrow-sector volumeSingleMoving focal anatomy; irregular rhythmTarget may not fit the pyramid
Three-dimensional zoomUsually singleValves, septa, defects, appendageAn oversized box lowers volume rate and line density
Multibeat full volumeTwo to six or more, ECG-gatedChamber volumes; large fields at high volume rateStitching with arrhythmia, breathing, or probe motion
Single-beat full volumeSingleChamber volumes in atrial fibrillation or poor breath-holdingLower volume rate and line density
Three-dimensional color DopplerSingle or severalRegurgitant orifice, paravalvular leak, and shunt geometryLowest volume rate; blooming; stitching

Three-dimensional acquisition modes, as standardized by the 2012 European Association of Echocardiography (EAE) and ASE recommendations. The 2015 chamber quantification recommendations specify a multibeat acquisition with minimal depth and sector angle for right ventricular volumes, to reach a temporal resolution above 20 to 25 volumes per second.

A multibeat full-volume acquisition divides the pyramid into subvolumes, acquires one per beat over several ECG-gated beats, and joins them. The narrow subvolumes are acquired at a high rate, so the stitched dataset combines a wide field with the volume rate of a single subvolume. The same 64 × 64 pyramid acquired in four beats contains 16 × 64 lines per beat, takes 13.3 ms per subvolume, and reaches about 75 volumes per second, but it spans 3.2 s at 75 beats per minute, all of which must fall within one breath-hold. In atrial fibrillation or when the patient cannot suspend breathing, the subvolumes misalign and single-beat acquisition is used. A three-dimensional zoom box sized closely to its target keeps the volume rate high: halving the box in azimuth and elevation quarters the line count and raises the volume rate fourfold.

Three-dimensional volume rate calculator

Time per volume—
Volume rate—
Volumes per cardiac cycle—
Acquisition span—
—

Volume time = (lines in azimuth × lines in elevation) ÷ (receive lines per transmit × beats) × 13 µs × depth; volume rate = 1 ÷ volume time. The model assumes one focal zone and ignores processing time, so displayed system rates are somewhat lower. A mitral zoom box of 48 × 48 lines to 12 cm with 16 receive lines per transmit gives about 45 volumes per second.

Stitching and other three-dimensional artifacts

Stitching artifact is a straight seam along a subvolume border, seen as a step through the myocardium or a valve. Before analysis, the dataset is inspected in a cross-section perpendicular to the seams, and it is reacquired if steps appear. Irregular rhythm or failed breath-holding calls for single-beat acquisition.

ArtifactCausePrevention or check
StitchingIrregular rhythm, breathing, or probe motion across beatsRegular rhythm and breath-hold; single-beat acquisition
DropoutLow gain; thin or beam-parallel tissue (fossa ovalis, thin leaflets)Gain adjustment; review of source planes and color Doppler
Blurring and bloomingHigh gain; low line density; smoothingLower gain; smaller volume; source planes
Cropping defectCrop plane placed through real tissueProgressive withdrawal of the crop plane
TruncationPyramid too narrow or shallow for the chamberConfirmation of the whole chamber in orthogonal planes before measurement
Shadowing and reverberationCalcium, prostheses, devicesOther windows and planes
Surface-mesh errorBorder tracking misses the true endocardiumReview and correction of tracking in the source slices

Three-dimensional artifact families and their checks, consistent with the 2026 ASE artifact guideline. A suspected perforation, cleft, or defect is reported only after it persists in orthogonal multiplanar planes and, where expected, carries a color Doppler jet.

A volume-rendered hole in an anterior mitral leaflet that closes when gain is raised, is absent in the multiplanar source planes, and carries no color flow is dropout. A true perforation persists in orthogonal planes and usually shows a jet. A cleft is a slit that reaches the free edge of the leaflet.

Display: volume rendering, surface rendering, and multiplanar reconstruction

A volume of voxels is displayed in several ways. Volume rendering assigns brightness and transparency to voxel intensity and shades by depth; it shows anatomic relationships and en face views but depends on gain. Surface rendering converts a traced boundary into a geometric mesh; it shows chamber shape and supports volume calculation, but the mesh is a model and smoothing removes trabeculations. Multiplanar reconstruction cuts any two-dimensional plane from the stored volume, which allows measurement in a true orthogonal plane and verification of a rendered finding; an oblique cut can be misread as a standard view. Multislice display shows parallel short-axis slices from base to apex for regional wall motion and lesion extent.

Cropping removes the part of the volume between the viewer and the structure of interest. The crop plane is advanced progressively, because a plane placed through a leaflet or septum removes real tissue and creates a false defect. Low gain erases thin tissue and produces holes; high gain thickens and blurs margins.

Two panels of three-dimensional transesophageal mitral valve analysis, each with three multiplanar reconstruction planes through the mitral annulus and a color-coded surface model of the valve viewed en face, with the anterior annulus, aortic valve, and anterolateral and posteromedial commissures labeled.
Three-dimensional transesophageal mitral valve analysis in Barlow disease (A) and fibroelastic deficiency (B). Three orthogonal multiplanar reconstruction planes are used to trace the annulus and leaflets, and the lower right panel in both shows the resulting surface-rendered model en face, with the aortic valve (Ao) at the anterior annulus and the anterolateral (AL) and posteromedial (PM) commissures; prolapsing scallops are coded red. Credit: Apor A, Nagy AI, Kovács A, et al., Cardiovasc Ultrasound 2016;14:32, Figure 1, via PubMed Central, CC BY 4.0.

Three-dimensional left ventricular volumes and ejection fraction

Three-dimensional left ventricular volumes are calculated from a surface that semiautomated software fits to the endocardium throughout the cardiac cycle. The enclosed volume is counted without an assumed shape, and the dataset contains the true apex, which removes foreshortening. A meta-analysis of 23 studies (1,638 echocardiograms) that used cardiac magnetic resonance as the reference found that 3D echocardiography underestimated end-diastolic volume by 19 mL and end-systolic volume by 10 mL, with an ejection fraction bias of −0.6 percentage points and limits of agreement of ±11.8 percentage points. In the nine studies that also measured two-dimensional (2D) volumes, the 2D biases were −48 mL and −28 mL, and 3D was both more accurate for volumes and more precise for all three measurements. The 2015 chamber quantification recommendations state that in laboratories with experience in 3D echocardiography, 3D measurement and reporting of left ventricular volumes is recommended when image quality permits.

PopulationEDV index, menEDV index, womenESV index, menESV index, womenEF, menEF, women
Scandinavian cohort (n = 166)66 (46–86)58 (42–74)29 (17–41)23 (13–33)57 (49–65)61 (49–73)
Japanese cohort (n = 410)50 (26–74)46 (28–64)19 (9–29)17 (9–25)61 (53–69)63 (55–71)
White European cohort (n = 978, combined)49 (31–67)42 (26–58)19 (9–29)16 (8–24)61 (49–73)62 (52–72)
Asian Indian cohort (same study)41 (23–59)39 (23–55)16 (6–26)15 (7–23)62 (52–72)62 (52–72)
White European cohort (n = 226)63 (41–85)56 (40–78)24 (14–34)20 (12–28)62 (54–70)65 (57–73)

Normal three-dimensional left ventricular values from Table 3 of the 2015 ASE/EACVI chamber quantification recommendations: mean (lower and upper limits of normal, mean ± 2 standard deviations). EDV, end-diastolic volume; ESV, end-systolic volume; EF, ejection fraction. Volumes are indexed in mL/m², and ejection fraction is in percent. Reference values vary with ethnicity, vendor, and software, and no single 3D threshold has been adopted.

Method determines accuracy. The whole ventricle, including the apex and lateral wall, must lie within the pyramid, and the endocardium must be visible throughout the cycle. The tracked surface is reviewed in the source slices and corrected, with trabeculations and papillary muscles included in the cavity as in two-dimensional tracing. The volume rate must capture true end-systole: at a fast heart rate, a sparse single-beat sequence can miss the smallest volume, so end-systolic volume is overestimated and ejection fraction underestimated.

A truncated apex is the most consequential error. In a dilated ventricle with an akinetic apex and true volumes of 160 mL and 100 mL (ejection fraction 37.5%), a dataset that excludes an apical cap of 18 mL at end-diastole and 14 mL at end-systole reports 142 mL and 86 mL, an ejection fraction of 39.4%. The ejection fraction looks credible while both volumes, and any remodeling on serial studies, are understated by 11% to 14%.

Serial measurement and cardio-oncology. In women receiving chemotherapy for breast cancer whose function was stable by global longitudinal strain, noncontrast 3D ejection fraction and volumes had the lowest temporal variability of all methods over a year. With 3D, a change in ejection fraction of about 6 percentage points could arise from measurement variability, compared with more than 10 points with two-dimensional methods. Three-dimensional ejection fraction is therefore preferred for serial surveillance when image quality permits. Cancer therapy–related dysfunction is covered in the transthoracic echocardiography hub, and strain surveillance during cancer therapy in the myocardial strain hub.

Three-dimensional strain. Speckle tracking applied to a 3D dataset yields longitudinal, circumferential, radial, and area strain from one acquisition and is not limited to motion within a plane. It is constrained by the lower volume rate and spatial resolution of 3D data, and normal values differ between vendors; two-dimensional global longitudinal strain remains the standard deformation measurement in clinical reports, as described in the myocardial strain hub.

Three-dimensional right ventricular and atrial volumes

The crescentic right ventricle, with a separate inflow and outflow tract, defeats simple geometric models. A valid 3D dataset contains the inflow, apex, free wall, and outflow tract and is analyzed with software validated for the right ventricle, with the moderator band and trabeculations included in the cavity. Clipping the outflow tract or free wall underestimates the volumes and invalidates the ejection fraction. Three-dimensional right ventricular ejection fraction is of particular value after cardiac surgery, when longitudinal indices such as tricuspid annular plane systolic excursion fall and no longer represent overall function.

ParameterNormalMildly abnormalModerately abnormalSeverely abnormal
3D RV end-diastolic volume index (mL/m²)<9090–103>103–115>115
3D RV end-systolic volume index (mL/m²)<4141–48>48–55>55
3D RV ejection fraction (%)>45≤45 to >39≤39 to ≥32<32

Three-dimensional right ventricular (RV) reference values and severity grades from the 2025 ASE right heart guideline. The 2015 chamber quantification recommendations gave sex-specific upper limits of 87 mL/m² in men and 74 mL/m² in women for end-diastolic volume index and 44 and 36 mL/m² for end-systolic volume index, with an ejection fraction below 45% usually abnormal.

Three-dimensional atrial volumes avoid geometric assumptions and are larger than biplane volumes, but current thresholds for left atrial enlargement and diastolic function rest on biplane left atrial volume index. The 2025 right heart guideline gives reference values for 3D right atrial volume indexed to body surface area and notes that 3D right atrial function indices, which differ by sex, age, and race and ethnicity, have not been widely adopted in clinical practice.

The mitral valve in three dimensions

A volume can be rotated to any angle, so anatomy is identified from landmarks, whatever the screen position. The standard en face display of the mitral valve, viewed from the left atrium (the surgical view), places the aortic valve at the top (12 o'clock), the left atrial appendage on the left (9 o'clock), and the interatrial septum on the right (3 o'clock), as recommended in 2012 by the EAE and the ASE. Viewed from the ventricle with the aortic valve still at the top, the valve appears as a mirror image and the appendage lies on the right. After any rotation, the aortic valve and appendage are located first and the scallops named from them.

Landmark or segmentPosition in the surgical view
Aortic valveTop, 12 o'clock
Left atrial appendageLeft, 9 o'clock
Interatrial septumRight, 3 o'clock
Anterior leaflet, A1 to A3Upper half, adjacent to the aortic valve
Posterior leaflet, P1 to P3Lower half, away from the aortic valve
A1 and P1, anterolateral commissureAppendage side
A3 and P3, posteromedial commissureSeptal side

Orientation of the mitral valve in the en face surgical view from the left atrium (EAE/ASE 2012).

Three-dimensional transesophageal imaging localizes prolapsing and flail scallops, commissural fusion, clefts, perforations, and calcification. It also measures the saddle-shaped annulus and quantifies leaflet area, tenting volume, and coaptation. These measurements guide surgical repair, annuloplasty ring sizing, and transcatheter edge-to-edge repair. Three-dimensional planimetry of the mitral orifice in rheumatic stenosis, with the plane placed at the leaflet tips in multiplanar reconstruction, avoids the oblique-plane error of two-dimensional planimetry. Localizations are confirmed in the source planes. Severity grading and the integrated approach to mitral regurgitation are covered in the mitral regurgitation hub.

Aortic and tricuspid valves and three-dimensional color Doppler

No single plane captures the whole tricuspid valve, whose leaflet number and arrangement vary. An en face view shows the leaflets, commissures, annulus, and any device lead that impinges on a leaflet, which matters in lead-related regurgitation and in planning transcatheter tricuspid repair or replacement. An en face view of the aortic valve defines cusp number, raphe, fusion, and orifice shape. Rapid motion, calcification, and shadowing limit leaflet definition, so 3D planimetry supplements Doppler assessment of aortic stenosis without replacing it. Three-dimensional imaging also measures the oval aortic annulus, although CT is the standard for transcatheter aortic valve sizing.

Three-dimensional color Doppler displays the shape of a regurgitant orifice or shunt. A plane cut perpendicular to the jet at its narrowest point allows planimetry of the vena contracta area. The advantage is greatest when a functional mitral or tricuspid orifice is elongated, because a two-dimensional vena contracta width measures only the diameter that lies in the imaging plane. An elliptical orifice with diameters of 1.2 cm and 0.5 cm has an area of π × 1.2 × 0.5 ÷ 4 = 0.47 cm²; a four-chamber width of 0.5 cm across the short diameter, treated as circular, implies 0.20 cm², less than half the true area. Severity thresholds follow the 2017 ASE native valve regurgitation recommendations and are applied in the valve hubs. A rendered color jet volume depends on gain, aliasing velocity, and driving pressure and is not a regurgitant volume.

Three-dimensional guidance of structural interventions, fusion imaging, and 3D printing

Live 3D transesophageal imaging, often with simultaneous biplane display, guides transseptal puncture (height above the mitral plane and anteroposterior position) and transcatheter edge-to-edge mitral and tricuspid repair. It also guides left atrial appendage occlusion (orifice and landing-zone sizing, device position, and peridevice leak), closure of atrial septal defects and paravalvular leaks, and transcatheter mitral valve-in-valve and valve-in-ring procedures. Echocardiography–fluoroscopy fusion overlays echocardiographic landmarks on the fluoroscopic image to coordinate the two views. Three-dimensional datasets from transesophageal echocardiography or CT can be segmented and printed as physical models to rehearse complex procedures, particularly paravalvular leak closure and adult congenital repairs. Procedural imaging is covered in detail in the transesophageal echocardiography hub.

Three-dimensional imaging has lower spatial and temporal resolution than two-dimensional imaging and depends more on image quality, rhythm, and breath-holding. During stress, tachycardia lowers the number of volumes per cycle and hinders breath-holding, so 3D stress imaging supplements standard acquisition without replacing it.

Focused cardiac ultrasound and cardiac POCUS: definitions and scope

Point-of-care ultrasound is the acquisition, interpretation, and immediate clinical integration of ultrasound imaging by the treating clinician. The definition does not depend on the location, the device, or the clinician's specialty. The 2013 ASE recommendations defined focused cardiac ultrasound (FCU) as a bedside adjunct to the physical examination that answers a limited set of questions about the heart with a limited number of views, usually in two-dimensional grayscale. They distinguished FCU from a limited echocardiogram, a study of full diagnostic quality restricted in scope, such as follow-up of a known effusion. A limited echocardiogram is performed and interpreted to laboratory standards, and its reader is accountable for all abnormalities visible in the views recorded. The 2013 document advised against hybrid terms such as "hand-held echocardiography" and "point-of-care echocardiography," because they blur what the examination can and cannot exclude.

The 2024 ASE nomenclature recommendations, endorsed by emergency, critical care, chest, thoracic, and ultrasound societies, set out the current vocabulary.

TermPerformed and interpreted byModalitiesDocumentation
Basic cardiac POCUS (focused cardiac ultrasound, FoCUS)Treating clinicianTwo-dimensional grayscale, addressing a defined set of pathologies in focused viewsProgress note; images archived
Advanced cardiac POCUS, including critical care echocardiographyTreating clinician with advanced trainingGrayscale plus M-mode, color and spectral Doppler, and in some settings strain, 3D, and agitated salineProcedure note; images archived
Consultative echocardiography, comprehensive or limitedSeparate consultative team, usually the echocardiography laboratoryFull range of modalities as the question requiresFormal structured report
Ultrasound-assisted physical examinationTreating clinicianBrief imaging to augment the physical examinationArchiving not universally required

Cardiac ultrasound nomenclature from the 2024 ASE recommendations. The intrinsic difference between cardiac POCUS and consultative echocardiography is who performs and interprets the study, whatever the device or location. A consultative examination may be comprehensive or limited, and "consultative" is not a synonym for "comprehensive." Use of artificial intelligence is designated "AI-assisted cardiac POCUS"; wearable ultrasound and robotic scanning fall outside the definition.

The questions a basic focused examination answers are qualitative: Is left ventricular systolic function grossly normal, reduced, or severely reduced? Is the right ventricle grossly enlarged? Is there a pericardial effusion, and does it compromise filling? Is the inferior vena cava small and collapsing or large and fixed? The 2010 consensus of the ASE and the American College of Emergency Physicians (ACEP) built the core emergency applications on these questions, together with guidance of pericardiocentesis and confirmation of transvenous pacing capture. A focused scan directs immediate care, but it does not quantify, does not grade valve disease, and does not exclude regional dysfunction. When it raises a question it cannot answer, a diagnostic echocardiogram follows.

Hand-held and pocket-sized devices

Hand-held ultrasound devices range from pocket units that display grayscale and color Doppler to tablet- and phone-connected probes with pulsed- and continuous-wave Doppler and simple measurements. Compared with full platforms they have smaller apertures, lower output power, fewer transducers, and limited processing. Some lack continuous-wave Doppler, calibrated measurements, electrocardiographic gating, harmonic imaging, or connection to the laboratory archive.

FeatureFull echocardiography platformTypical hand-held device
Imaging2D, M-mode, harmonic imaging, 3D, strain, contrast presets2D grayscale; color Doppler on most current devices
Spectral DopplerPulsed, continuous-wave, and tissue DopplerAbsent on some devices; limited on others
MeasurementCalibrated linear, area, volume, Doppler, and strainNone or basic linear measurement
ECG gatingStandardUsually absent
ArchivingIntegrated with the laboratory systemVariable; local storage or cloud transfer
ReportDiagnostic reportDocumented as a focused examination

Capabilities of full platforms and hand-held devices. When the question requires a quantity the device cannot measure, such as aortic stenosis velocity without continuous-wave Doppler, the finding is described qualitatively and a diagnostic study is requested.

In the hands of experienced echocardiographers, image quality from current hand-held devices approaches that of full systems for two-dimensional questions. In 349 patients referred for echocardiography and scanned by experts with both a hand-held device and a high-end system, endocardial visibility was slightly lower with the hand-held device. Agreement on regional wall motion was good (κ 0.73), and ejection fraction differed by a mean of 1.8 percentage points. No pericardial effusion or valve stenosis was missed, and regurgitation missed by the hand-held device was minimal. These results describe expert operators only. The EACVI 2018 position statement on hand-held devices supports their use for focused examinations as an extension of the physical examination and requires training and storage of images used for decisions. It states that a hand-held scan does not replace a complete echocardiogram. Hand-held devices are also used by echocardiographers for rapid triage, follow-up of a known finding, screening in community and low-resource settings, and teaching.

The focused cardiac examination: views and target findings

A focused cardiac examination uses the parasternal long- and short-axis, apical four-chamber, and subcostal four-chamber and inferior vena cava views. The subcostal window is often the only one available in a supine, ventilated patient or during resuscitation, and the parasternal window is often best in patients with lung hyperinflation.

QuestionFocused findingWhat the focused scan does not establish
Left ventricular systolic functionVisual estimate: normal or hyperdynamic, reduced, severely reducedQuantitative ejection fraction; regional wall motion abnormality; diastolic function
Right ventricular sizeRight ventricle as large as or larger than the left ventricle; septal flatteningCause of dilation; pulmonary pressure; chronicity
Pericardial effusionEcho-free space, circumferential or loculated; right-sided chamber collapseDoppler criteria for tamponade; effusive-constrictive physiology
Inferior vena cavaSmall and collapsing, or dilated and fixedFluid responsiveness by itself; right atrial pressure in ventilated patients
Gross valve abnormalitySevere calcification, flail leaflet, large vegetation, prosthetic rockingSeverity of stenosis or regurgitation; exclusion of valve disease
Cardiac activity during arrestOrganized contraction versus standstillPrognosis by itself; decision to stop resuscitation

Target questions and limits of focused cardiac ultrasound, from the 2013 ASE recommendations and the 2010 ASE/ACEP consensus. Findings are integrated with the clinical picture and confirmed with diagnostic imaging when management depends on them.

Global left ventricular function by eye. Visual estimation of ejection fraction by trained readers correlates well with quantitative methods in broad categories. A small, hyperdynamic ventricle with end-systolic cavity obliteration suggests hypovolemia or vasodilation; a dilated, globally hypokinetic ventricle suggests cardiomyopathy; regional akinesis suggests ischemia, but a focused scan does not reliably exclude a regional abnormality. The finding directs fluid, vasopressor, inotrope, or diuretic decisions in shock and dyspnea.

Pericardial effusion and tamponade at the bedside

Pericardial fluid appears as an echo-free space between the epicardium and parietal pericardium, seen in the subcostal and parasternal views. In the parasternal long-axis view, pericardial fluid tracks anterior to the descending thoracic aorta, whereas a left pleural effusion extends posterior to it. Epicardial fat is echogenic, anterior, and moves with the heart. Tamponade is a clinical and hemodynamic diagnosis. Focused findings that support it are a moderate or large effusion with right atrial collapse lasting more than one-third of the cardiac cycle, right ventricular diastolic collapse, and a dilated inferior vena cava without respiratory collapse. Doppler respiratory variation of mitral and tricuspid inflow requires spectral Doppler and belongs to an advanced or diagnostic study. Focused ultrasound detects a large effusion in a hypotensive patient, guides pericardiocentesis, and, during cardiac arrest with pulseless electrical activity, identifies a reversible cause. Effusion grading, tamponade physiology, and drainage are covered in the pericardial effusion and tamponade hub.

Right ventricular dilation and pulmonary embolism

In the apical four-chamber view, the right ventricle is normally smaller than the left. A right ventricle as large as or larger than the left (a right-to-left ventricular diameter ratio of 1.0 or more), a flattened septum producing a D-shaped left ventricle in the parasternal short-axis view, and a dilated inferior vena cava indicate pressure or volume overload. In a hemodynamically unstable patient with suspected pulmonary embolism, absence of right ventricular dilation or dysfunction makes pulmonary embolism an unlikely cause of shock. Its presence, when CT is not immediately possible, supports reperfusion therapy in the high-risk category. Right ventricular dilation lacks sensitivity and specificity for pulmonary embolism in stable patients, because chronic pulmonary hypertension, right ventricular infarction, and lung disease produce the same picture. A thick right ventricular free wall suggests chronic pressure overload. McConnell's sign (free wall akinesis with apical sparing) and thrombus in transit are specific but infrequent. Echocardiographic risk stratification in pulmonary embolism is covered in the pulmonary embolism hub.

Inferior vena cava and volume status

The inferior vena cava is imaged in the subcostal long axis, with the diameter measured 1 to 2 cm from the right atrial junction, perpendicular to the long axis of the vessel. In a spontaneously breathing patient, the 2025 ASE right heart guideline assigns a right atrial pressure of 3 mm Hg (range 0 to 5) when the diameter is 2.1 cm or less and collapses 50% or more with a sniff. It assigns 15 mm Hg when the diameter exceeds 2.1 cm and collapses less than 50%, and an intermediate value of 8 mm Hg when the findings fit neither pattern. In mechanically ventilated patients, positive intrathoracic pressure dilates the cava, so a dilated cava does not indicate high right atrial pressure. A small cava (1.2 cm or less) in these patients suggests a right atrial pressure below 10 mm Hg. A dilated cava is normal in some young athletes and common in pregnancy.

Inferior vena cava size and collapse estimate right atrial pressure; they do not by themselves predict whether stroke volume will rise with fluid. Fluid responsiveness is assessed with dynamic tests such as passive leg raising with stroke volume measurement, described in the cardiac intensive care hub. Abdominal pressure, tricuspid regurgitation, and a small subcostal sweep that cuts the cava obliquely all distort the measurement.

Lung ultrasound: B-lines, lung sliding, and pleural effusion

Lung ultrasound complements focused cardiac imaging in dyspnea and shock. Normal aerated lung shows the pleural line sliding with respiration and horizontal A-lines, reverberations of the pleural line at regular intervals. B-lines are vertical, laser-like artifacts arising from the pleural line, extending to the bottom of the screen without fading, erasing A-lines, and moving with lung sliding. They indicate thickened interlobular septa or alveolar fluid. The 2012 international consensus defined a positive region as three or more B-lines in a longitudinal plane between two ribs, and diffuse interstitial syndrome as two or more positive regions bilaterally on an eight-region anterior and lateral scan. B-lines are not specific for cardiogenic edema: pneumonia, acute respiratory distress syndrome, contusion, and pulmonary fibrosis also produce them. A patchy distribution, an irregular or thickened pleural line, and subpleural consolidation point to a pulmonary cause.

In a meta-analysis of seven studies of 1,075 patients with acute dyspnea, B-lines had a sensitivity of 94.1% and a specificity of 92.4% for acute cardiogenic pulmonary edema. In a patient with low pretest probability, a scan without diffuse B-lines makes cardiogenic edema unlikely. B-lines also clear with decongestion and can be followed during treatment of heart failure. Absent lung sliding with A-lines suggests pneumothorax, and the lung point, where sliding and nonsliding lung meet, confirms it. A pleural effusion is an anechoic space above the diaphragm, often with compressed lung floating within it. The Bedside Lung Ultrasound in Emergency (BLUE) protocol combines these signs into profiles for pulmonary edema, pneumonia, obstructive airway disease, pulmonary embolism, and pneumothorax. The Fluid Administration Limited by Lung Sonography (FALLS) protocol uses the appearance of B-lines during fluid loading as a limit to volume administration.

Shock protocols: the RUSH and FAST examinations

Protocols structure bedside ultrasound in undifferentiated shock. The Rapid Ultrasound in Shock (RUSH) examination evaluates the "pump," "tank," and "pipes" in sequence. The Focused Assessment with Sonography in Trauma (FAST) looks for pericardial, peritoneal, and, in extended form, pleural fluid and pneumothorax after injury.

ComponentViewsHypovolemic shockCardiogenic shockObstructive shockDistributive shock
PumpParasternal, apical, subcostalSmall, hyperdynamic ventricleReduced contractility; dilated ventricleEffusion with tamponade; dilated right ventricleHyperdynamic early; reduced in septic cardiomyopathy
TankInferior vena cava, lungs, abdomen, pleuraSmall, collapsing cava; free fluid if hemorrhageDilated cava; diffuse B-lines; effusionsDilated cava; absent lung sliding in tension pneumothoraxSmall or normal cava; focal B-lines or consolidation in pneumonia
PipesAbdominal aorta, femoral and popliteal veinsAortic aneurysm or dissectionNormalDeep vein thrombosisNormal

The RUSH examination and its expected findings by shock type. Bedside findings narrow the differential diagnosis and are confirmed with diagnostic imaging and hemodynamic data. Shock classification and management are covered in the cardiogenic shock hub.

Ultrasound during cardiac arrest

During cardiac arrest, focused ultrasound can show tamponade, right ventricular dilation suggesting pulmonary embolism, an empty hyperdynamic ventricle, pneumothorax, and the presence or absence of organized cardiac motion. In a prospective study of 793 patients with out-of-hospital or emergency department arrest with pulseless electrical activity or asystole, cardiac activity on ultrasound was the variable most strongly associated with survival to admission (odds ratio 3.6) and discharge (odds ratio 5.7). Of patients without cardiac activity, 0.6% survived to discharge. Patients with effusion who underwent pericardiocentesis survived more often than others (15.4% versus 1.3%). The 2025 American Heart Association (AHA) advanced life support guideline states that point-of-care ultrasound by experienced clinicians may be considered to identify reversible causes during cardiac arrest when it does not interrupt resuscitation (Class 2b). The main harm is a longer pause in compressions; images are obtained during the planned rhythm check, recorded as a clip, and interpreted after compressions resume. Absence of cardiac motion is not used alone to stop resuscitation. Technique and evidence are covered in the advanced life support hub.

Diagnostic accuracy and limitations of point-of-care cardiac ultrasound

Focused ultrasound by trained clinicians detects left ventricular dysfunction, pericardial effusion, and right ventricular enlargement more accurately than physical examination. Its accuracy is lower than that of a comprehensive echocardiogram and depends on training, image quality, and pretest probability. Most published studies measure image quality and agreement with full studies; fewer measure changes in management or outcomes. The principal limitations are these:

Training, competency, credentialing, and image archiving

Competence in focused ultrasound depends on structured training in three components: didactic education in physics, anatomy, and the pathophysiology of the conditions in the clinician's scope; hands-on acquisition with real-time proctoring; and supervised interpretation of normal and abnormal studies. The 2013 ASE recommendations did not set numeric requirements and noted the weak correlation between the number of studies interpreted and competence. Emergency medicine, critical care, anesthesiology, internal medicine, and hospital medicine training programs have defined their own curricula, and the National Board of Echocardiography offers a special competence examination in critical care echocardiography. The 2020 ASE recommendations describe how echocardiography laboratories can support this training by providing supervised scanning, image review, and quality assurance.

For cardiologists, the Core Cardiology Training Symposium (COCATS) 4 Task Force 5 recommendations define Level I training in transthoracic echocardiography as at least 75 studies performed and 150 interpreted, and Level II, for independent interpretation, as at least 150 performed and 300 interpreted. Advanced competencies, including 3D imaging, contrast and perfusion echocardiography, and procedural guidance, are set out in the 2019 American College of Cardiology (ACC)/AHA/ASE advanced training statement; the 2018 contrast guideline recommends additional training at a high-volume center for perfusion imaging.

Archiving is part of the examination. The 2013 ASE recommendations state that images obtained to evaluate a symptomatic patient and direct management should be stored in a retrievable location, ideally the same digital system as the laboratory's studies. They also specify that all focused examinations are documented with its date, operator, indication, findings, and limitations. Archived images make quality review, comparison with later diagnostic studies, and tele-echocardiography possible: images acquired at a rural clinic or by a novice can be transmitted for remote expert interpretation.

Artificial intelligence guidance and automated measurement

Deep-learning software can guide probe movement in real time and assess image quality as images are acquired. In a prospective study at two academic centers, eight nurses with no prior echocardiography experience used an AI guidance system to acquire 10-view limited studies in 240 patients. Expert readers judged the studies of diagnostic quality for left ventricular size and function and for pericardial effusion in 98.8% and for right ventricular size in 92.5%. Results were similar to those of sonographers for most secondary measures. Automated tools also estimate ejection fraction, measure chamber volumes, and detect patterns of disease. Such tools follow the acquired image: they do not correct foreshortening, an oblique plane, or a missed end-systolic frame, and the interpreting clinician remains responsible for all reported values. The 2024 nomenclature recommendations designate these studies as AI-assisted cardiac POCUS. Automated measurement in the laboratory is covered in the transthoracic echocardiography hub.

Special populations: pregnancy, obesity, mechanical ventilation, and athletes

Pregnancy. Echocardiography involves no ionizing radiation and is the first imaging test for cardiac symptoms in pregnancy. Normal pregnancy increases chamber dimensions modestly and dilates the inferior vena cava, and a small pericardial effusion is common late in gestation. Ultrasound-enhancing agents are used only when the answer will change management and cannot be obtained otherwise, as described under safety.

Obesity and lung disease. Obesity and hyperinflation are the most common reasons for inadequate windows. Enhancing agents restore endocardial definition in most such patients, and the subcostal or parasternal window may be preferable for focused imaging. Three-dimensional datasets in these patients are more often limited by dropout.

Mechanical ventilation. Positive pressure limits windows, dilates the inferior vena cava, and alters right ventricular loading. Contrast is recommended in technically difficult ventilated patients, and transesophageal imaging is used when transthoracic images remain inadequate; critical care transesophageal echocardiography is covered in the transesophageal echocardiography hub.

Athletes. Physiologic enlargement of all chambers, prominent trabeculation, and a dilated inferior vena cava are normal findings in trained athletes. Contrast that shows hypertrabeculation in an athlete with normal function is interpreted with ventricular size, function, and family history; it does not by itself establish noncompaction cardiomyopathy.

Sources of error in contrast, three-dimensional, and point-of-care echocardiography

References

  1. Porter TR, Mulvagh SL, Abdelmoneim SS, et al. Clinical applications of ultrasonic enhancing agents in echocardiography: 2018 American Society of Echocardiography guidelines update. J Am Soc Echocardiogr. 2018;31:241–274. doi:10.1016/j.echo.2017.11.013
  2. Porter TR, Abdelmoneim S, Belcik JT, et al. Guidelines for the cardiac sonographer in the performance of contrast echocardiography: a focused update from the American Society of Echocardiography. J Am Soc Echocardiogr. 2014;27:797–810. doi:10.1016/j.echo.2014.05.011
  3. Mulvagh SL, Rakowski H, Vannan MA, et al. American Society of Echocardiography consensus statement on the clinical applications of ultrasonic contrast agents in echocardiography. J Am Soc Echocardiogr. 2008;21:1179–1201. doi:10.1016/j.echo.2008.09.009
  4. Senior R, Becher H, Monaghan M, et al. Clinical practice of contrast echocardiography: recommendation by the European Association of Cardiovascular Imaging (EACVI) 2017. Eur Heart J Cardiovasc Imaging. 2017;18:1205–1205af. doi:10.1093/ehjci/jex182
  5. Lindner JR, Belcik T, Main ML, et al. Expert consensus statement from the American Society of Echocardiography on hypersensitivity reactions to ultrasound enhancing agents in patients with allergy to polyethylene glycol (PEG). J Am Soc Echocardiogr. 2021;34:707–708. doi:10.1016/j.echo.2021.05.002
  6. Lantheus Medical Imaging. DEFINITY and DEFINITY RT (perflutren lipid microsphere) injectable suspension: US prescribing information. Revised March 2024. DailyMed, US National Library of Medicine. dailymed.nlm.nih.gov
  7. GE Healthcare. OPTISON (perflutren protein-type A microspheres) injectable suspension: US prescribing information. Revised May 2025. DailyMed, US National Library of Medicine. dailymed.nlm.nih.gov
  8. Bracco Diagnostics. LUMASON (sulfur hexafluoride lipid-type A microspheres) for injectable suspension: US prescribing information. Revised July 2024. DailyMed, US National Library of Medicine. dailymed.nlm.nih.gov
  9. Kurt M, Shaikh KA, Peterson L, et al. Impact of contrast echocardiography on evaluation of ventricular function and clinical management in a large prospective cohort. J Am Coll Cardiol. 2009;53:802–810. doi:10.1016/j.jacc.2009.01.005
  10. Main ML, Ryan AC, Davis TE, et al. Acute mortality in hospitalized patients undergoing echocardiography with and without an ultrasound contrast agent (multicenter registry results in 4,300,966 consecutive patients). Am J Cardiol. 2008;102:1742–1746. doi:10.1016/j.amjcard.2008.08.019
  11. Main ML, Hibberd MG, Ryan A, et al. Acute mortality in critically ill patients undergoing echocardiography with or without an ultrasound contrast agent. JACC Cardiovasc Imaging. 2014;7:40–48. doi:10.1016/j.jcmg.2013.08.012
  12. Wei K, Mulvagh SL, Carson L, et al. The safety of Definity and Optison for ultrasound image enhancement: a retrospective analysis of 78,383 administered contrast doses. J Am Soc Echocardiogr. 2008;21:1202–1206. doi:10.1016/j.echo.2008.07.019
  13. Weinsaft JW, Kim RJ, Ross M, et al. Contrast-enhanced anatomic imaging as compared to contrast-enhanced tissue characterization for detection of left ventricular thrombus. JACC Cardiovasc Imaging. 2009;2:969–979. doi:10.1016/j.jcmg.2009.03.017
  14. Wei K, Jayaweera AR, Firoozan S, et al. Quantification of myocardial blood flow with ultrasound-induced destruction of microbubbles administered as a constant venous infusion. Circulation. 1998;97:473–483. doi:10.1161/01.cir.97.5.473
  15. Senior R, Moreo A, Gaibazzi N, et al. Comparison of sulfur hexafluoride microbubble (SonoVue)-enhanced myocardial contrast echocardiography with gated single-photon emission computed tomography for detection of significant coronary artery disease: a large European multicenter study. J Am Coll Cardiol. 2013;62:1353–1361. doi:10.1016/j.jacc.2013.04.082
  16. Mathias W Jr, Tsutsui JM, Tavares BG, et al. Sonothrombolysis in ST-segment elevation myocardial infarction treated with primary percutaneous coronary intervention. J Am Coll Cardiol. 2019;73:2832–2842. doi:10.1016/j.jacc.2019.03.006
  17. Cotter B, Raisinghani A, DeMaria AN. Established and emerging roles for ultrasound enhancing agents (contrast echocardiography). Clin Cardiol. 2022;45:1114–1122. doi:10.1002/clc.23924
  18. Pellikka PA, Arruda-Olson A, Chaudhry FA, et al. Guidelines for performance, interpretation, and application of stress echocardiography in ischemic heart disease: from the American Society of Echocardiography. J Am Soc Echocardiogr. 2020;33:1–41.e8. doi:10.1016/j.echo.2019.07.001
  19. Saric M, Sadeghpour A, Alizade L, et al. Recommendations for the identification and mitigation of cardiac ultrasound artifacts: a guideline from the American Society of Echocardiography. J Am Soc Echocardiogr. 2026;39:435–475. Correction: J Am Soc Echocardiogr. 2026;39:1102. doi:10.1016/j.echo.2026.01.007
  20. Lang RM, Badano LP, Tsang W, et al. EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography. J Am Soc Echocardiogr. 2012;25:3–46. doi:10.1016/j.echo.2011.11.010
  21. Lang RM, Badano LP, Mor-Avi V, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28:1–39.e14. doi:10.1016/j.echo.2014.10.003
  22. Mukherjee M, Rudski LG, Addetia K, et al. Guidelines for the echocardiographic assessment of the right heart in adults and special considerations in pulmonary hypertension: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2025;38:141–186. Correction: J Am Soc Echocardiogr. 2025;38:641 (doi:10.1016/j.echo.2025.05.001). doi:10.1016/j.echo.2025.01.006
  23. Dorosz JL, Lezotte DC, Weitzenkamp DA, Allen LA, Salcedo EE. Performance of 3-dimensional echocardiography in measuring left ventricular volumes and ejection fraction: a systematic review and meta-analysis. J Am Coll Cardiol. 2012;59:1799–1808. doi:10.1016/j.jacc.2012.01.037
  24. Thavendiranathan P, Grant AD, Negishi T, Plana JC, Popović ZB, Marwick TH. Reproducibility of echocardiographic techniques for sequential assessment of left ventricular ejection fraction and volumes: application to patients undergoing cancer chemotherapy. J Am Coll Cardiol. 2013;61:77–84. doi:10.1016/j.jacc.2012.09.035
  25. Zoghbi WA, Adams D, Bonow RO, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American Society of Echocardiography developed in collaboration with the Society for Cardiovascular Magnetic Resonance. J Am Soc Echocardiogr. 2017;30:303–371. doi:10.1016/j.echo.2017.01.007
  26. Apor A, Nagy AI, Kovács A, Manouras A, Andrássy P, Merkely B. Three-dimensional dynamic morphology of the mitral valve in different forms of mitral valve prolapse: potential implications for annuloplasty ring selection. Cardiovasc Ultrasound. 2016;14:32. doi:10.1186/s12947-016-0073-4
  27. Mitchell C, Rahko PS, Blauwet LA, et al. Guidelines for performing a comprehensive transthoracic echocardiographic examination in adults: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2019;32:1–64. doi:10.1016/j.echo.2018.06.004
  28. Spencer KT, Kimura BJ, Korcarz CE, Pellikka PA, Rahko PS, Siegel RJ. Focused cardiac ultrasound: recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2013;26:567–581. doi:10.1016/j.echo.2013.04.001
  29. Labovitz AJ, Noble VE, Bierig M, et al. Focused cardiac ultrasound in the emergent setting: a consensus statement of the American Society of Echocardiography and American College of Emergency Physicians. J Am Soc Echocardiogr. 2010;23:1225–1230. doi:10.1016/j.echo.2010.10.005
  30. Kirkpatrick JN, Panebianco N, Díaz-Gómez JL, et al. Recommendations for cardiac point-of-care ultrasound nomenclature. J Am Soc Echocardiogr. 2024;37:809–819. doi:10.1016/j.echo.2024.05.001
  31. Kirkpatrick JN, Grimm R, Johri AM, et al. Recommendations for echocardiography laboratories participating in cardiac point of care cardiac ultrasound (POCUS) and critical care echocardiography training: report from the American Society of Echocardiography. J Am Soc Echocardiogr. 2020;33:409–422.e4. doi:10.1016/j.echo.2020.01.008
  32. Cardim N, Dalen H, Voigt JU, et al. The use of handheld ultrasound devices: a position statement of the European Association of Cardiovascular Imaging (2018 update). Eur Heart J Cardiovasc Imaging. 2019;20:245–252. doi:10.1093/ehjci/jey145
  33. Prinz C, Voigt JU. Diagnostic accuracy of a hand-held ultrasound scanner in routine patients referred for echocardiography. J Am Soc Echocardiogr. 2011;24:111–116. doi:10.1016/j.echo.2010.10.017
  34. Volpicelli G, Elbarbary M, Blaivas M, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38:577–591. doi:10.1007/s00134-012-2513-4
  35. Volpicelli G, Gargani L, Zieleskiewicz L, et al. International evidence-based recommendations for point-of-care lung ultrasound: 2025 focused update of the 2012 recommendations. Intensive Care Med. 2026;52:1415–1446. doi:10.1007/s00134-026-08487-2
  36. Al Deeb M, Barbic S, Featherstone R, Dankoff J, Barbic D. Point-of-care ultrasonography for the diagnosis of acute cardiogenic pulmonary edema in patients presenting with acute dyspnea: a systematic review and meta-analysis. Acad Emerg Med. 2014;21:843–852. doi:10.1111/acem.12435
  37. Lichtenstein DA. BLUE-protocol and FALLS-protocol: two applications of lung ultrasound in the critically ill. Chest. 2015;147:1659–1670. doi:10.1378/chest.14-1313
  38. Perera P, Mailhot T, Riley D, Mandavia D. The RUSH exam: Rapid Ultrasound in SHock in the evaluation of the critically ill. Emerg Med Clin North Am. 2010;28:29–56. doi:10.1016/j.emc.2009.09.010
  39. Gaspari R, Weekes A, Adhikari S, et al. Emergency department point-of-care ultrasound in out-of-hospital and in-ED cardiac arrest. Resuscitation. 2016;109:33–39. doi:10.1016/j.resuscitation.2016.09.018
  40. Wigginton JG, Agarwal S, Bartos JA, et al. Part 9: adult advanced life support: 2025 American Heart Association guidelines for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2025;152(16 suppl 2):S538–S577. doi:10.1161/CIR.0000000000001376
  41. Narang A, Bae R, Hong H, et al. Utility of a deep-learning algorithm to guide novices to acquire echocardiograms for limited diagnostic use. JAMA Cardiol. 2021;6:624–632. doi:10.1001/jamacardio.2021.0185
  42. Ryan T, Berlacher K, Lindner JR, Mankad SV, Rose GA, Wang A. COCATS 4 Task Force 5: training in echocardiography. J Am Coll Cardiol. 2015;65:1786–1799. doi:10.1016/j.jacc.2015.03.035
  43. Wiegers SE, Ryan T, Arrighi JA, et al. 2019 ACC/AHA/ASE advanced training statement on echocardiography (revision of the 2003 ACC/AHA clinical competence statement on echocardiography). J Am Soc Echocardiogr. 2019;32:919–943. doi:10.1016/j.echo.2019.04.002