Conduction system pacing and leadless pacemakers
His bundle pacing, left bundle branch area pacing, and single- and dual-chamber leadless pacemakers in adults, from rationale, anatomy, and capture criteria through implantation, evidence, guideline recommendations, complications, programming, and end-of-service management.
Conduction system pacing places the ventricular lead on the His bundle or in the left bundle branch area so that the ventricles are activated through the His-Purkinje system. Leadless pacemakers place the entire pacemaker inside the heart and remove the pocket and the transvenous lead. The 2023 Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society (HRS/APHRS/LAHRS) guideline rates cardiac physiologic pacing (biventricular, His bundle, or left bundle branch area) as reasonable (Class 2a) when substantial ventricular pacing is expected and left ventricular ejection fraction (LVEF) is 36% to 50%. Left bundle branch capture is diagnosed from the 12-lead ECG during threshold testing, and a leadless device is chosen when the risks of the pocket and lead outweigh its single- or dual-chamber limits.
Definitions: physiologic pacing, conduction system pacing, and leadless pacing
Conventional right ventricular pacing stimulates working myocardium at the apex or septum, and the impulse spreads slowly from cell to cell. Cardiac physiologic pacing is the 2023 HRS/APHRS/LAHRS term for pacing intended to restore or preserve ventricular synchrony. It includes biventricular pacing through a coronary sinus or epicardial left ventricular lead, His bundle pacing (HBP), and left bundle branch area pacing (LBBAP). HBP and LBBAP together are conduction system pacing (CSP). Both use a transvenous lead and a pectoral generator, so they change where the ventricle is activated but not the hardware that fails in conventional systems.
A leadless pacemaker is a self-contained capsule, containing battery, circuitry, and electrodes, that is delivered through a femoral (or, less often, jugular) venous sheath and fixed to the right ventricular or right atrial endocardium. It removes the pocket and the lead but, in current ventricular devices, paces right ventricular myocardium and so produces the same dyssynchronous activation as a transvenous right ventricular lead. The two technologies therefore solve different problems: CSP addresses the electrical consequences of pacing, and leadless pacing addresses the mechanical and infectious consequences of leads and pockets.
| Term | Definition | Characteristic ECG or device feature |
|---|---|---|
| Selective His bundle pacing | Capture of the His bundle alone | Isoelectric interval after the stimulus (approximating the His-to-ventricular [HV] interval); QRS identical to the conducted native QRS when distal conduction is normal |
| Nonselective His bundle pacing | Capture of the His bundle and adjacent septal myocardium | No isoelectric interval; pseudo-delta wave fused with a narrow QRS |
| Para-Hisian (right ventricular septal) pacing | Myocardial capture near the His bundle without conduction system capture | Wide QRS with slurring or notching; V6 R-wave peak time prolonged |
| Left bundle branch pacing (selective or nonselective) | Capture of the left bundle trunk or its fascicles, with or without septal myocardium | Right bundle conduction delay pattern in V1; short, constant stimulus to V6 R-wave peak time; output-dependent transitions |
| Left ventricular septal pacing | Capture of left ventricular septal subendocardium without proven left bundle capture | Deep septal lead, terminal r or R in V1, no left bundle capture criterion met |
| Deep septal pacing | Lead deep in the septum but short of the left ventricular subendocardium | QRS narrower than right ventricular septal pacing, no terminal r in V1 |
| Single-chamber leadless pacemaker | Ventricular capsule | VVI(R); paced QRS of right ventricular myocardial pacing |
| Accelerometer-based atrioventricular (AV)-synchronous leadless pacemaker | Ventricular capsule that senses atrial contraction mechanically | VDD; no atrial pacing |
| Dual-chamber leadless system | Atrial and ventricular capsules communicating beat to beat | DDD(R) with atrial pacing and sensing |
Capture types and leadless systems. Left bundle branch area pacing is the umbrella term that includes left bundle branch pacing and left ventricular septal pacing (2023 HRS/APHRS/LAHRS guideline, Table 7; 2023 European Heart Rhythm Association [EHRA] clinical consensus statement).
Development of the techniques. Permanent HBP in humans was first reported in 2000 in 18 patients with chronic atrial fibrillation, dilated cardiomyopathy, and a narrow QRS. The His bundle could be stimulated in 14, and a screw-in lead was placed successfully in 12, with acute thresholds of 2.4 ± 1.0 V at 0.5 ms; AV node ablation was performed in patients with a fast ventricular response, and LVEF rose from 20% to 31% during follow-up. Left bundle branch pacing was described in 2017, when a lead screwed through the septum captured the left bundle just beyond a site of block at low and stable output. The first leadless pacemaker trials were reported in 2015 and 2016, and the US Food and Drug Administration approved the first tine-fixed leadless pacemaker on April 6, 2016.
Why right ventricular pacing harms: dyssynchrony and pacing-induced cardiomyopathy
Right ventricular myocardial pacing activates the septum early and the lateral left ventricular wall late, in a sequence that resembles left bundle branch block. The result is mechanical dyssynchrony, wasted septal work, functional mitral regurgitation, and, in a subset of patients with frequent pacing, a fall in LVEF. Pacing-induced cardiomyopathy is usually defined as an LVEF decline of 10% or more from a baseline above 50%, with ventricular pacing of 20% or more and no other explanation. The 2023 HRS/APHRS/LAHRS guideline defines substantial ventricular pacing as an anticipated or documented burden of 20% to 40% or more. The incidence and predictors of pacing-induced cardiomyopathy and the mode-selection trials (MOST, DAVID, BLOCK HF) are presented under pacing-induced cardiomyopathy and on the bradycardia page under right ventricular pacing and physiologic pacing.
The harm depends on the dose of ventricular pacing and the substrate. Complete heart block and AV node ablation produce near-continuous ventricular pacing; sinus node dysfunction with intact AV conduction produces little when pacing-minimization algorithms are used. A wide native QRS, male sex, lower baseline LVEF, and a wide paced QRS predict cardiomyopathy. Pacing site alone does not prevent the problem: right ventricular septal and outflow tract pacing narrow the paced QRS modestly but still activate myocardium outside the conduction system. In the CSPACE trial, the comparator was right ventricular septal pacing, and conduction system pacing still reduced pacing-induced cardiomyopathy (evidence in bradycardia).
The same physiology applies to leadless devices. A ventricular leadless pacemaker on the septum produces a wide, left bundle branch block–like paced QRS and carries the same dyssynchrony as a transvenous lead at the same site. Long-term data on pacing-induced cardiomyopathy specific to leadless devices are lacking. In a patient expected to need near-continuous ventricular pacing, the choice of a leadless device weighs that physiology against the advantages of removing the lead.
Anatomy of the His bundle and left bundle branch for pacing
The His bundle continues from the compact AV node, penetrates the central fibrous body, and runs along the inferior border of the membranous septum on the crest of the muscular septum. A substantial length of the proximal His bundle lies on the atrioventricular part of the membranous septum above the tricuspid valve plane, so a lead on the atrial side of the annulus can capture it. Because the bundle is insulated by fibrous tissue, recording a His potential, even a sharp near-field one, does not guarantee capture. Proximal sites on the atrial side of the annulus give selective capture more often but carry low R-wave amplitudes and atrial or His oversensing; distal sites on the ventricular side give nonselective capture with better sensing and thresholds. The fiber organization, blood supply, and longitudinal dissociation within the His bundle are covered under anatomy of the His-Purkinje system.
The left bundle leaves the His bundle as a broad trunk that fans over the left septal surface into anterior, posterior, and septal fiber groups. This fan is the reason LBBAP succeeds: the target is centimeters wide, it lies beyond most proximal lesions, and it can be reached from the right ventricular septum by screwing a lead through the muscular septum. The original target was 1.5 to 2 cm from the His potential along a line toward the right ventricular apex. Most contemporary sites lie farther from the His bundle and correspond to fascicular capture instead of truncal capture. In MELOS, left bundle fascicular capture was the predominant capture type (69.5%), followed by left ventricular septal capture (21.5%) and proximal left bundle branch capture (9%).
Three nearby structures account for the specific complications of LBBAP. The right bundle runs subendocardially on the right ventricular septum, so lead manipulation produces transient right bundle branch block. Septal perforator branches of the left anterior descending artery, and septal veins draining to the coronary sinus, traverse the septum, so a helix can injure them. The tricuspid valve's septal leaflet and chordae lie near basal septal sites, so a lead placed close to the annulus can interfere with valve closure. Thin or scarred septa (prior infarction, sarcoidosis, myectomy, alcohol septal ablation) and thick or infiltrated septa (hypertrophic cardiomyopathy, amyloidosis) alter both the feasibility and the risk of transseptal lead placement.
His bundle pacing: capture types, criteria, and correction of bundle branch block
HBP activates both ventricles through the native conduction system when infra-Hisian conduction is intact. Selective capture produces a paced QRS identical to the native conducted complex after an isoelectric latency that equals the HV interval. Nonselective capture adds local septal myocardium, producing a pseudo-delta wave that merges with a narrow complex. Distinguishing nonselective HBP from para-Hisian myocardial pacing is the central interpretive task, because a pseudo-delta wave is present in both.
Threshold testing with a 12-lead ECG is the principal method: as output falls, a transition from nonselective to selective His capture, or from nonselective His capture to myocardial-only capture, identifies the thresholds of the two tissues. In up to 10% of patients, the His and myocardial thresholds are identical, and changing pulse width, programmed stimulation, or rapid pacing reveals the difference. In infranodal block, pacing at 120 beats per minute or faster confirms that conduction distal to the pacing site is reliable.
| Capture type | Electrocardiographic and electrogram criteria |
|---|---|
| Selective His capture, normal distal conduction | Stimulus-to-QRS interval equal to the native HV interval with an isoelectric segment; paced QRS identical to native QRS; one capture threshold (His) |
| Nonselective His capture, normal distal conduction | Stimulus-to-QRS interval shorter than the HV interval (usually zero) with a pseudo-delta wave; QRS duration and axis close to native; two capture thresholds (His and myocardium) |
| Nonselective His capture versus para-Hisian pacing | No slurring or notching in I, V1, or V4 to V6, and V6 R-wave peak time 100 ms or less, favor His capture. A difference of more than 12 ms between stimulus-to-V6 R-wave peak time and His-to-V6 R-wave peak time indicates absent His capture (99.1% sensitivity, 100% specificity) |
| His bundle pacing with correction of bundle branch block | Paced QRS narrower than native; up to three thresholds (His capture with correction, His capture without correction, myocardial capture) |
| His bundle pacing without correction of bundle branch block | Paced QRS retains the bundle branch block pattern; two thresholds (His with block, myocardium) |
| Bundle branch block correction not achieved despite His capture | Paced QRS with notching, slurring, or plateau in I and V4 to V6, or stimulus-to-V6 R-wave peak time above 110 ms, despite His capture |
Criteria for His bundle capture, adapted from the 2023 HRS/APHRS/LAHRS guideline (Table 6) and the 2023 EHRA clinical consensus statement. HV, His-to-ventricular interval.
Correction of left bundle branch block (LBBB). When block lies in the left-sided fibers within the His bundle or proximal left bundle, pacing distal to the lesion recruits latent fibers and normalizes the QRS. Correction often requires higher output than His capture alone, and the correction threshold is recorded at implantation because it governs programming. In the 2023 guideline's review, case series corrected LBBB with permanent HBP in 70% to 90% of patients, and in the His-Alternative trial 72% of patients assigned to HBP achieved correction. Block located more distally, or diffuse myocardial conduction delay, is not corrected by HBP.
Limitations of His bundle pacing: thresholds, sensing, revision, and battery
The His bundle is small and wrapped in fibrous tissue, so the electrode–tissue interface is less favorable than in trabeculated myocardium. Capture thresholds are higher at implantation and rise later in a proportion of patients. In one series cited by the 2023 guideline, HBP thresholds were 1.35 ± 0.9 V at 0.5 ms compared with 0.6 ± 0.5 V for right ventricular pacing. Of 294 patients with HBP, 15% had an increase in capture threshold, the largest share (41%) within the first 8 weeks, and 6% eventually needed lead revision. Late threshold rise requiring revision is reported in 7% to 11% of patients.
In the 5-year comparison of HBP with right ventricular pacing at sister hospitals, HBP succeeded in 75 of 94 attempts (80%). Lead revision was needed in 6.7% versus 3%, and generator change in 9% versus 1%, reflecting the battery drain of higher outputs. Sensed R waves are small at proximal sites, and atrial or His potentials can be oversensed, producing ventricular inhibition in a pacing-dependent patient or double counting. HBP also leaves the patient exposed to progression of disease distal to the pacing site: AV block below the lead, or a bundle branch block that develops after implantation, cannot be corrected by pacing above it.
These features explain two guideline positions. A backup ventricular lead may be reasonable in pacing-dependent patients receiving HBP (Class 2b). An HBP lead in a defibrillator system is not used for tachycardia detection, because small R waves and atrial oversensing can cause inappropriate shocks or undertreatment of ventricular arrhythmia. In practice, the operational difficulties of HBP have led many operators to adopt LBBAP as the default conduction system technique. HBP is retained for patients with a narrow QRS and nodal block, in whom it reproduces the native activation sequence exactly.
Left bundle branch area pacing: capture types and outcomes
LBBAP places the lead tip in the left ventricular septal subendocardium after it has crossed the septum from the right side. Three outcomes are possible at the final position. Left bundle branch capture (selective or nonselective) activates the left ventricle through its conduction system, giving rapid lateral wall activation and a short, constant left ventricular activation time. Left ventricular septal pacing activates the left side of the septum from its subendocardium without proven conduction system capture; activation of the lateral wall is slower, but still faster than with right ventricular pacing. Deep septal pacing, with the tip short of the subendocardium, produces a paced QRS narrower than right ventricular septal pacing but without the physiology of either.
In all three, the right ventricle is activated late by transseptal myocardial conduction unless septal myocardium is captured at the right side, so the paced QRS has a right bundle conduction delay pattern in V1: Qr, qR, rsR′, or R morphology. This pattern describes lead position and is not proof of conduction system capture. Selective left bundle capture produces a discrete potential or isoelectric interval after the stimulus in the local electrogram and a deeper terminal component in V1, while nonselective capture fuses septal myocardial activation with conduction system activation.
In MELOS RELOADED, among 3,382 propensity-matched patients with AV block, LVEF above 40%, and ventricular pacing above 20%, lack of confirmed left bundle branch capture was an independent predictor of mortality within the LBBAP group (hazard ratio [HR] 1.85). Confirmation of capture is therefore part of the procedure.
Criteria for left bundle branch capture
The 2023 HRS/APHRS/LAHRS guideline defines left ventricular septal pacing as deep septal lead placement, confirmed by fluoroscopy (the fulcrum sign), contrast septography, echocardiography, or CT, with a right bundle conduction delay pattern in V1. Left bundle branch pacing requires, in addition, at least one left bundle capture criterion. The 2023 EHRA consensus statement uses the same framework and adds validation data for the individual criteria. R-wave peak times are measured from the pacing stimulus at a high sweep speed (50 to 100 mm/s) on the recording system. An inferred QRS onset is not used as the starting point.
| Criterion | Threshold | Performance and notes |
|---|---|---|
| Output-dependent QRS transition | Nonselective to selective left bundle capture, or nonselective left bundle capture to left ventricular septal capture, during unipolar output decremented from a starting value of 5 to 10 V at 0.5 ms | Highly specific; regarded as the reference criterion. Seen in 26.4% of MELOS patients with left bundle capture, and in up to 75.4% at implantation when proximal sites are targeted; absence of a transition is nonconclusive |
| Stimulus to V6 R-wave peak time | Below 74 ms with narrow QRS or isolated right bundle branch block (RBBB); below 80 ms with LBBB, nonspecific intraventricular conduction delay, or wide escape rhythm | 100% specific at these values (EHRA: below 74 ms was 40% sensitive); EHRA uses practical cutoffs of below 75 and below 80 ms |
| V6–V1 interpeak interval | Above 44 ms | 100% specific; above 33 ms gives 71.8% sensitivity and 90.0% specificity for separating nonselective left bundle from septal capture |
| Abrupt change in V6 R-wave peak time with output | Shortening of 10 ms or more at high output during deep septal positioning (HRS); increase of 15 ms or more on loss of left bundle capture at lower output (EHRA) | The 15-ms increase is 82.6% sensitive and 100% specific; a change of about 10 to 12 ms lies within interobserver variability |
| Left bundle potential | Potential-to-V6 R-wave peak time equal to stimulus-to-V6 R-wave peak time within 10 ms; potential-to-QRS interval about 15 to 35 ms | Sensitivity 88.2%, specificity 95.4% for a potential-to-stimulus difference below 10 ms; potentials are not visible in all patients |
| Programmed deep septal stimulation | Selective conduction system response to a premature extrastimulus (refractory periods of septal myocardium and left bundle differ) | Diagnostic when present |
| Physiology-based criteria | Paced QRS onset to V6 R-wave peak time no longer than native (+10 ms); in LBBB, V6 R-wave peak time shorter by more than 8 ms than during corrective His bundle pacing | Require a native or His-paced reference |
Left bundle branch capture criteria from the 2023 HRS/APHRS/LAHRS guideline (Table 7) and the 2023 EHRA clinical consensus statement, with performance data from the V6–V1 interpeak validation study. One criterion, together with evidence of left ventricular septal pacing, establishes left bundle branch pacing.
Why the V6–V1 interpeak interval works. Capture of the left bundle shortens left ventricular activation and so moves the V6 R-wave peak earlier, while right ventricular activation, reflected by the V1 R′ peak, is unchanged. Capture of septal myocardium affects the V1 peak. In the validation study of 124 patients, the interval averaged 62.3 ms with selective left bundle capture, 41.3 ms with nonselective capture, and 26.5 ms with left ventricular septal capture. Loss of septal capture lengthened V1 R-wave peak time by 15 ms or more without changing V6, and loss of left bundle capture did the reverse.
Sources of misclassification. Anodal capture at the ring electrode during bipolar testing at high output can capture right ventricular myocardium and mimic a transition, which is why thresholds for capture classification are measured in unipolar configuration. A narrow paced QRS alone, a right bundle conduction delay pattern alone, and a low impedance alone do not establish capture. In LBBB, the native activation time is long, so the accepted stimulus-to-V6 R-wave peak time cutoff is below 80 ms; a value of 76 ms therefore meets the LBBB cutoff but not the narrow-QRS cutoff of below 74 ms.
Left bundle branch capture classifier
The classifier applies the 2023 HRS/APHRS/LAHRS definitions to measurements taken during implantation or at follow-up. It first establishes whether left ventricular septal pacing is present, then tests the left bundle capture criteria, and reports the V6–V1 interpeak interval from the two R-wave peak times.
Left bundle branch capture classifier
2023 HRS/APHRS/LAHRS and EHRA criteriaV6–V1 interpeak interval = stimulus to V1 R′-wave peak time minus stimulus to V6 R-wave peak time. The classifier uses the HRS cutoffs (below 74 ms without LBBB, below 80 ms with LBBB or wide escape) and treats an interpeak interval above 44 ms as specific. The output-dependent transition is the reference criterion; values near a cutoff are confirmed by repeated measurement at high sweep speed.
Implantation technique for His bundle pacing
HBP leads are placed through cephalic, axillary, or subclavian access, usually from the left. The lead is delivered through a fixed-curve or deflectable sheath whose posterior curve directs the tip to the septum. Mapping is unipolar with the pacing lead itself, connected to an electrophysiology recording system or pacing system analyzer; a separate diagnostic catheter is usually not needed. In a 20° to 30° right anterior oblique (RAO) view, the target is a sharp near-field His potential with a small or absent atrial electrogram and a larger ventricular electrogram. When no His signal is found, as in AV block without an escape rhythm, high-output pace mapping (5 V at 1 ms) for the narrowest paced QRS localizes the bundle. Contrast injection through the sheath can outline the tricuspid annulus and triangle of Koch.
In atrial fibrillation, high-frequency atrial signals mimic His potentials, and mapping begins from the ventricular side of the high membranous septum. When AV junction ablation is planned, a His site with a large atrial electrogram (above 0.5 mV) is avoided because ablation near the lead can raise the His threshold. In infranodal block, a proximal site may capture the His bundle only at high output because of block below it, whereas a slightly distal site may capture beyond the lesion at low output.
The lead is fixed by rotation of the lead body, after which the sheath is withdrawn and slit and the lead slack is adjusted in the right atrium. A His current of injury, an ST-segment–like deflection on the His electrogram after fixation, is associated with low acute and 1-year thresholds. Lowering the high-pass filter of the recording system from 30 Hz to 0.5 to 1 Hz improves its recording. The 2023 HRS/APHRS/LAHRS guideline considers assessment of the His or left bundle current of injury with appropriate filter settings beneficial (Class 2a). Its suggested implant threshold for HBP is below 2.5 V at 1 ms. The EHRA statement considers 1.5 V or less at 0.5 ms optimal and accepts 2.0 to 2.5 V at 0.5 ms when HBP is the only means of resynchronization. When a current of injury is present, the EHRA statement advises re-evaluating the His capture threshold after 5 minutes.
Implantation technique for left bundle branch area pacing
Choosing the entry site. The His bundle potential, or the superior tricuspid annulus identified from atrial and ventricular electrograms and sheath movement, is marked in the 20° to 30° RAO view. The sheath is then advanced about 15 to 20 mm toward the apex and rotated counterclockwise to reach the basal to mid septum. Contrast-based methods refine this. One approach targets a sector 15 to 35 mm from the tricuspid annulus summit at an angle of −10° to 30° below the line joining the summit to the apex. Another divides the RAO image into a 3 × 3 grid. Leads placed more than 16 to 19 mm from the annulus were associated with less tricuspid regurgitation than more proximal sites.
Paced QRS at the entry site. Unipolar pacing from the right septal surface before screwing shows the expected final result. A favorable site shows a W-shaped QRS in V1 with a notch at its nadir, and discordant QRS polarity in lead II (positive or equiphasic) and lead III (equiphasic or negative). Sites with concordant polarity in II and III can still succeed and tend to capture the anterior or posterior fascicle.
Penetrating the septum. In a 30° to 40° left anterior oblique (LAO) view, the sheath is oriented perpendicular to the septum, with the lead usually 20° to 30° (range 10° to 40°) above the horizontal. Contrast septography outlines the septal thickness and shows the depth of the tip. The lead is advanced with rapid rotations while paced morphology, impedance, and current of injury are monitored between turns:
- The V1 notch migrates to the end of the QRS and becomes a terminal r or R as the tip approaches the left side.
- Fixation beats, premature ventricular complexes with a right bundle conduction delay pattern arising from the advancing helix, occur in 59% to 96% of implants and signal proximity to the left conduction system. A terminal r or R in V1 marks the point at which capture criteria are tested.
- Unipolar impedance usually rises during the early turns and falls as the tip nears the left ventricular endocardium. A fall below about 500 Ω, or a drop of 200 Ω, warns of impending perforation; a steep rise with a stylet-driven lead suggests helix retraction.
- The myocardial current of injury rises to about 20 to 35 mV in mid-septum and falls toward 10 to 12 mV near the subendocardium. A fall to about 5 mV warrants caution.
- A left bundle or fascicular potential indicates that the subendocardium has been reached, and any further rotation is made with extreme caution; a current of injury on the potential indicates that further rotation is contraindicated.
Accepting the position. The 2023 EHRA statement considers an LBBAP threshold below 1.5 V at 0.5 ms acceptable (ideally below 1 V) and a bipolar R wave above 4 mV desirable. In MELOS, mean threshold was 0.77 V and sensing 10.6 mV over 6.4 months. Capture is documented against the criteria in the table above. The thresholds for left bundle capture and for septal capture are both recorded, and a final 12-lead ECG at the programmed output becomes the reference for follow-up.
| Measure | His bundle pacing | Left bundle branch area pacing |
|---|---|---|
| Target capture threshold | ≤1.5 V at 0.5 ms optimal (EHRA); <2.5 V at 1 ms (HRS) | <1.5 V at 0.5 ms, ideally <1 V (EHRA) |
| Sensed R wave | Bipolar ideally >2.0 mV (unipolar >4 mV) | Bipolar ideally >4 mV |
| Reasons to abandon the site (HRS Table 8) | His bundle not located; lead instability; threshold >5 V at 1 ms; R wave <2 mV; atrial oversensing; paced QRS not ≤130 ms or not narrowed by >20% from a wide baseline; no selective or nonselective His capture | Septum not penetrated to the subendocardium; lead instability; septal perforation risk; bundle branch block not corrected; no right bundle conduction delay pattern or paced QRS not ≤130 ms; left ventricular activation time not <74 to 80 ms |
| Late concerns | Threshold rise, revision, battery drain, oversensing, distal disease progression | Late septal perforation, loss of left bundle capture, extraction of a deep septal lead |
Implantation targets for conduction system pacing from the 2023 EHRA clinical consensus statement and the 2023 HRS/APHRS/LAHRS guideline (Table 8). HBP and LBBAP failure rates of 10% to 40% are reported with current tools, and crossover to a coronary sinus lead or right ventricular lead is planned before the procedure.
Learning curve and failure. In MELOS, LBBAP lead implantation succeeded in 92.4% of patients with bradycardia indications and 82.2% with heart failure indications. The learning curve was steepest over the first 110 cases and plateaued after 250. Heart failure, a broad baseline QRS, and a larger left ventricular end-diastolic diameter independently predicted failure. Septal scar, very thick or fibrotic septa, and dilated right chambers that weaken sheath support are the common anatomic obstacles; reshaping the sheath or using a sheath-in-sheath technique gives additional reach.
Leads and delivery tools: lumenless and stylet-driven
The original CSP lead is a lumenless, 4.1 French, isodiametric lead with a fixed, electrically active helix, delivered through a dedicated sheath and fixed by rotating the lead body. Its small diameter and fixed helix favor deep septal penetration and make His bundle bail-out maneuvers simple. Stylet-driven leads with extendable–retractable helices, delivered through larger sheaths, are stiffer and transmit torque better, and their wider bodies (about 5.6 to 6.0 French) may increase septal and tricuspid trauma. Helix retraction during rotation is recognized by a rise in impedance. Nonrandomized LBBAP series with stylet-driven leads report success rates of 87% to 95%, and helix damage or fracture has been reported in 0.8% to 5.0% of implants across lead types. In the EHRA physician survey, a minority of operators used stylet-driven leads for HBP and more than half used them for LBBAP.
The lead connects to the ventricular port of a conventional generator or, in resynchronization configurations, to the left ventricular or right ventricular port. The port choice determines which capture management, sensing, and timing algorithms apply to the CSP lead.
Complications of conduction system pacing
The generic complications of transvenous implantation (pneumothorax, pocket hematoma, infection, venous occlusion, lead-related tricuspid regurgitation) apply to CSP and are covered under implantation complications. In MELOS, the overall complication rate was 11.7%, and complications specific to the transseptal route occurred in 8.3% of 2,533 patients.
| Complication | Reported frequency | Recognition and management |
|---|---|---|
| Acute septal perforation into the left ventricle | 0.0% to 14.1% during implantation | Fall in unipolar impedance, loss of current of injury, left ventricular endocardial paced morphology; lead withdrawn and repositioned at a new site |
| Delayed septal perforation | 0.1% to 0.3% | Loss of capture, low impedance, change in paced QRS; imaging confirms tip in the left ventricular cavity; revision, with attention to thromboembolism while the tip is intracavitary |
| Right bundle branch injury | 19.9% during implantation; 6.3% persistent | Caused by catheter or lead contact with the right septal surface; matters most with preexisting LBBB, in which it causes AV block |
| Complete heart block during implantation | 9.4% acute; 2.6% persistent | Backup pacing available during implantation in patients with preexisting conduction disease |
| Lead dislodgement | 3.0% during implantation; 0.3% to 1.5% after | Loss of capture or change in paced QRS; repositioning |
| Septal coronary injury | Acute coronary syndrome 0.4% to 0.7%; coronary artery fistula 1.4% to 2.0% | Chest pain, ST changes, or a fistula on imaging; most fistulas are small and close spontaneously |
| Helix damage or fracture | 0.8% to 5.0% | Impedance change, noise; retained fragments |
| Worsening tricuspid regurgitation | 7.3% to 32.6% | More frequent with basal septal sites close to the annulus |
| Threshold rise above 1 V (LBBAP) | 0.3% to 1.8% | Output reprogramming or revision |
| Loss of left bundle capture over time | 0.3% to 11.5% | Wider paced QRS and longer V6 R-wave peak time with preserved septal capture; often silent on device counters |
| His bundle pacing: loss of capture and revision | Loss of capture up to 17%; revision up to 11% | Threshold rise or dislodgement; reprogramming, revision, or conversion to LBBAP |
Complications specific to conduction system pacing, with ranges from the studies compiled in the 2023 EHRA clinical consensus statement. Rates vary with operator experience, lead type, and definitions.
Septal hematoma, septal vein injury with contrast staining, and pericardial effusion from perforation of the anterior septum into the right ventricular outflow or pericardium are rarer. The 2023 HRS/APHRS/LAHRS guideline identifies septal perforation, extraction of deep septal leads, and the long-term consequences of delayed right ventricular activation as areas without long-term data.
Conduction system pacing in bradycardia and atrioventricular block: evidence
The evidence moved from single-center observational comparisons to propensity-matched registries and, from 2025, randomized trials with clinical end points. The two randomized trials enrolled patients with AV block or high expected pacing burden and showed that conduction system pacing reduces pacing-induced cardiomyopathy compared with right ventricular pacing. Neither showed a significant difference in mortality or in heart failure hospitalization considered separately, and CSPACE recorded more lead revisions with CSP.
| Study | Design and population | Main findings |
|---|---|---|
| Sister-hospital comparison of HBP with right ventricular pacing at 5 years (2018) | 94 HBP attempts versus 98 right ventricular pacing | HBP success 80%. Pacing-induced cardiomyopathy 2% versus 22%. With ventricular pacing above 40%, death or heart failure hospitalization 32% versus 53%. Lead revision 6.7% versus 3%; generator change 9% versus 1% |
| Consecutive-patient comparison of HBP with right ventricular pacing (2018) | 332 HBP versus 433 right ventricular pacing | HBP success 92%. Death, heart failure hospitalization, or biventricular upgrade 25% versus 32% (HR 0.71), driven by patients with ventricular pacing above 20% (HR 0.65) |
| Geisinger-Rush registry (2022) | 321 LBBAP versus 382 right ventricular pacing; mean LVEF 58% | Death, heart failure hospitalization, or biventricular upgrade 10.0% versus 23.3% (HR 0.46), driven by patients with ventricular pacing above 20% |
| CSPACE (2025) | Randomized; 202 patients with AV block and no resynchronization indication; CSP versus right ventricular septal pacing; mean follow-up 25.2 months | CSP success 88.1%. Composite of pacing-induced cardiomyopathy, biventricular upgrade, heart failure hospitalization, and death 7.17 versus 20.69 events per 100 person-years (HR 0.35), driven by pacing-induced cardiomyopathy (HR 0.31). No significant difference in heart failure hospitalization or death. Lead revision 7.9% versus 1.0% |
| LBBP-FAVOUR (2026) | Randomized; 160 patients with high pacing burden and high risk of cardiac dysfunction; median follow-up 36 months | Death, heart failure hospitalization, or pacing-induced cardiomyopathy 11.6% versus 33.9% (HR 0.31), driven by pacing-induced cardiomyopathy (6.5% versus 18.2%). LVEF improvement 5.34 points greater with left bundle branch pacing. No significant difference in mortality or heart failure hospitalization |
| MELOS RELOADED (2026) | Registry; 3,382 propensity-matched patients with AV block, LVEF above 40%, ventricular pacing above 20% | 4-year survival 11.8 percentage points higher with LBBAP; HR for death 0.53. Lack of confirmed left bundle capture predicted mortality within the LBBAP group (HR 1.85) |
Comparisons of conduction system pacing with right ventricular pacing in patients with bradycardia indications. HR, hazard ratio; HBP, His bundle pacing; LBBAP, left bundle branch area pacing.
Interpretation of the conduction system pacing trials and registries. The randomized trials confirm the underlying mechanism, which is prevention of dyssynchrony-induced remodeling, and support CSP as an initial strategy when substantial ventricular pacing is expected. The mortality difference in MELOS RELOADED is large for a pacing strategy and comes from an observational comparison subject to selection; randomized mortality data are lacking. The trade-offs are a longer, more technical procedure, a learning curve, more lead revisions in some series, and specific complications. In sinus node dysfunction with intact conduction, the expected ventricular pacing burden is low with pacing-minimization algorithms, and the incremental benefit of CSP is small.
Guideline recommendations for conduction system pacing
The 2023 HRS/APHRS/LAHRS guideline is the governing US document. It grades CSP by LVEF and by the expected ventricular pacing burden, and adds procedural and follow-up recommendations. The 2018 American College of Cardiology/American Heart Association/Heart Rhythm Society (ACC/AHA/HRS) bradycardia guideline preceded the LBBAP evidence. It considered more physiologic ventricular activation (resynchronization or HBP) reasonable in preference to right ventricular pacing when LVEF is 36% to 50% and ventricular pacing is expected more than 40% of the time.
| Patient or situation | Recommendation | Class (level of evidence, LOE) |
|---|---|---|
| Pacing indication, LVEF 36%–50%, substantial ventricular pacing expected | Cardiac physiologic pacing (biventricular, HBP, or LBBAP) to reduce pacing-induced cardiomyopathy | 2a (B-R for cardiac resynchronization therapy [CRT]; B-NR for HBP and LBBAP) |
| Pacing indication, normal LVEF, substantial ventricular pacing expected | Cardiac physiologic pacing may be reasonable | 2b (B-NR) |
| Pacing-dependent patient receiving HBP | Additional backup ventricular lead may be reasonable | 2b (C-LD) |
| LVEF >35%, less than substantial pacing expected | Right ventricular lead with programming to minimize ventricular pacing is reasonable | 2a (B-R) |
| LVEF 36%–50%, less than substantial pacing expected | HBP or LBBAP lead may be considered in place of a right ventricular lead | 2b (C-LD) |
| Normal LVEF, less than substantial pacing expected | LBBAP lead may be considered in place of a right ventricular lead; biventricular pacing not indicated | 2b (C-LD); 3: No Benefit (B-R) |
| Decline in left ventricular function or heart failure attributed to substantial ventricular pacing | Upgrade to biventricular pacing; revision to CSP can be beneficial | 1 (B-NR); 2a (B-NR) |
| AV junction ablation, any LVEF | HBP with or without a backup lead, or an LBBAP lead, may be reasonable | 2b (C-LD); 2b (C-LD) |
| High ventricular pacing burden | HBP or LBBAP may be reasonable to decrease the risk of atrial fibrillation | 2b (C-LD) |
| CSP implantation | 12-lead ECG during implantation to assess conduction system capture; documentation of conduction system and myocardial capture thresholds | 1 (C-EO); 1 (C-EO) |
| CSP follow-up | Multilead or 12-lead ECG in office follow-up to assess conduction system capture and bundle branch block correction | 1 (B-NR) |
| HBP with threshold rise above 1 V | More frequent office follow-up to judge need for revision | 2a (C-EO) |
| CRT or CSP with high threshold causing rapid battery drain | New lead may be considered at generator replacement after shared decision-making | 2b (C-EO) |
Recommendations on conduction system pacing in the 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing. Substantial pacing is defined as anticipated or actual ventricular pacing of 20% to 40% or more. Resynchronization indications are given in the section on cardiac resynchronization. LOE, level of evidence; B-R, randomized; B-NR, nonrandomized; C-LD, limited data; C-EO, expert opinion.
The 2021 European Society of Cardiology (ESC) pacing guideline, written before the larger LBBAP data sets, issued graded recommendations for HBP and anticipated that conduction system pacing would play a growing role once randomized trials confirmed safety and efficacy. The 2023 HRS/APHRS/LAHRS guideline makes no recommendation on the type of device after transcatheter aortic valve replacement (TAVR) beyond those for AV block or LBBB in other settings, because data were insufficient. Conduction disturbances after TAVR are covered under conduction disturbances after TAVR.
Conduction system pacing with atrioventricular junction ablation
AV junction ablation creates complete heart block and permanent ventricular pacing, so the pacing strategy determines whether the patient receives synchronous or dyssynchronous activation for the rest of life. The indications, timing, initial lower rate of 80 to 90 per minute, and APAF-CRT results are covered under AV node ablation and pacing. For LVEF of 50% or less, biventricular pacing is reasonable (Class 2a); HBP with or without a backup lead and LBBAP may be reasonable at any LVEF (both Class 2b).
His bundle pacing in pace and ablate. HBP succeeded in about 95% of patients undergoing AV junction ablation in observational series. In ALTERNATIVE-AF, 50 patients with persistent atrial fibrillation and LVEF of 40% or less received both HBP and biventricular leads before AV junction ablation and were randomized to 9 months of HBP and 9 months of biventricular pacing in crossover. Among the 38 who completed both phases, LVEF improved more with HBP (phase 1 change 21.3% versus 16.7%), while functional class and natriuretic peptide improved similarly with both. One series found acute His threshold rise after ablation in 7 of 15 patients, which explains why a His site with a small atrial electrogram is selected and why some operators implant a backup lead.
Left bundle branch area pacing in pace and ablate. The LBBAP lead lies farther from the compact node than an HBP lead, so ablation is technically easier and the risk of damaging the lead site or raising its threshold is lower. In a comparison of 98 patients undergoing AV junction ablation (48 HBP, 50 LBBAP), both preserved or improved LVEF, and LBBAP had higher success and fewer lead complications. The lead is implanted before or on the day of ablation, and capture is confirmed before the node is ablated.
Conduction system pacing for cardiac resynchronization
In LBBB with a proximal lesion, LBBAP places the lead beyond the block and activates the left ventricle through its conduction system, which is the rationale for CSP as a resynchronization strategy. In the 2023 HRS/APHRS/LAHRS guideline, biventricular pacing remains the Class 1 therapy for LVEF of 35% or less, LBBB, QRS of 150 ms or more, and New York Heart Association (NYHA) class II to IV symptoms. CSP with HBP (with LBBB correction) or LBBAP is reasonable when effective biventricular resynchronization cannot be achieved (Class 2a) and may be considered as an alternative to biventricular pacing (Class 2b). Crossover to CSP is reasonable when the coronary sinus lead is unsuccessful or suboptimal (Class 2a). The full indication table is under guideline indications for CRT and cardiac physiologic pacing.
Randomized comparisons with biventricular pacing have produced discordant results, ranging from a 5.6-point greater LVEF gain with left bundle branch pacing in 40 patients (LBBP-RESYNC) to larger trials with clinical end points. They are analyzed under conduction system pacing as an alternative to biventricular CRT. When LBBAP alone leaves the paced QRS wide, as in nonspecific intraventricular conduction delay, scar, or a dilated ventricle, a coronary sinus lead can be added (left bundle–optimized CRT, LOT-CRT). In an international series, LOT-CRT succeeded in 91 of 112 patients (81%) and narrowed the QRS from 182 ms to 144 ms, more than biventricular pacing in the same patients. The same principle with HBP is His-optimized CRT.
Special substrates: infranodal block, bundle branch block, infiltrative and hypertrophic disease, and congenital anatomy
Infranodal block and bundle branch block. HBP in infranodal block depends on capturing beyond the lesion, often at high output, and the lesion can progress. LBBAP lies distal to most proximal lesions and is generally preferred in infranodal block, bifascicular block, and LBBB. In patients with preexisting LBBB, catheter trauma to the right bundle during implantation produces complete heart block, so temporary pacing or backup pacing is available.
Cardiac amyloidosis and sarcoidosis. Infiltrated or scarred septa can raise thresholds, reduce R-wave amplitude, and impede septal penetration. Because these diseases also carry heart failure and ventricular arrhythmia risk, the device is chosen for the whole course. A defibrillator system with a CSP lead is used where indicated, and in sarcoidosis later ventricular tachycardia ablation and imaging are anticipated. Conduction disease in amyloidosis is covered under conduction disease in cardiac amyloidosis.
Hypertrophic cardiomyopathy and septal reduction. A thick septum lengthens the transseptal path, and septal reduction scar alters the target. AV block after alcohol septal ablation or myectomy is managed with pacing according to the indications on the bradycardia page, and the location of septal scar is reviewed on imaging before a transseptal lead is planned.
Persistent left superior vena cava and congenital anatomy. Case reports describe CSP in a persistent left superior vena cava through a left-sided approach, a right-sided approach, and the femoral or iliac route. In adults with congenital heart disease, the conduction system may be displaced (as in congenitally corrected transposition or AV septal defects). The 2023 guideline addresses cardiac physiologic pacing in congenital heart disease separately, and mapping of the conduction system precedes lead placement.
Programming and follow-up of conduction system pacing devices
Programming follows from the type of capture documented at implantation. Output is set to secure conduction system capture, beyond myocardial capture alone, because loss of conduction system capture with retained septal capture is silent on device counters: the device records successful capture while the patient receives dyssynchronous activation.
| Parameter | His bundle pacing | Left bundle branch area pacing |
|---|---|---|
| Output | His capture (or correction) threshold plus at least 1 V | Left bundle capture threshold with a conventional safety margin; septal capture threshold also recorded |
| Automatic capture algorithms | Avoided unless His and myocardial thresholds are known to be similar | Can be used successfully |
| AV delay | 30 to 50 ms shorter than conventional | 20 to 30 ms shorter than conventional |
| Sensing | Atrial and His oversensing with proximal sites; ventricular undersensing; blanking and sensitivity adjusted | R waves generally large; standard settings |
| Unipolar sensing | Avoided in pacing-dependent patients | Avoided in pacing-dependent patients |
| Defibrillator systems | HBP lead not used for tachycardia detection | Detection by the right ventricular defibrillation lead or by the LBBAP lead according to configuration |
Programming of conduction system pacing, adapted from the 2023 HRS/APHRS/LAHRS guideline (Table 9). AV delays are shortened because conduction system capture inserts a latency from stimulus to QRS onset that right ventricular pacing does not.
Office follow-up of conduction system pacing. At office visits, a multilead or 12-lead ECG documents conduction system capture and bundle branch block correction (Class 1). A comprehensive assessment including His or left bundle capture, correction, and myocardial thresholds can be useful (Class 2a). The paced QRS duration and morphology, and the stimulus to V6 R-wave peak time, are compared with the implant reference. A wider QRS at the same output indicates loss of conduction system capture (from threshold rise or microdislodgement), progression of distal disease, drug effect, or hyperkalemia. Threshold testing with a 12-lead recording distinguishes these: a discrete widening at a defined output identifies loss of conduction system capture. In LBBAP, thresholds were stable during 3 to 6 months of follow-up, and in 618 patients a significant late threshold rise occurred in 1% and revision for dislodgement in 0.3%. A His threshold rise above 1 V leads to more frequent visits (Class 2a). Remote monitoring follows the standard for all devices, as summarized under device interrogation and follow-up; device-level troubleshooting is under troubleshooting leadless and conduction system pacing systems.
Generator replacement of conduction system pacing devices. CSP thresholds are tested at generator replacement because an unrecognized rise can leave little margin. When a high threshold has caused rapid battery depletion, a new lead may be considered at replacement (Class 2b).
Extraction and revision of conduction system pacing leads
Extraction of a lead whose helix lies deep in the septum was an early concern. The TECSPAM multicenter study reported 341 patients undergoing extraction of 224 HBP and 117 LBBAP leads (321 lumenless, 20 stylet-driven), with a mean dwell time of 22 ± 26 months. Complete procedural success was achieved in 99%, with retained distal fragments in 3 patients (1%), and minor complications in 2.1%. Among leads in place more than 6 months, manual traction succeeded in 87%, mechanical tools were needed in 10%, and laser in 3%. Reimplantation in the conduction system succeeded in 95% of attempts. The LBBAP leads in this series had dwell times shorter than 3 years, so extraction after a decade of septal encapsulation remains less well described.
Early revision of a lumenless lead within weeks of implantation is usually accomplished by counter-rotation and repositioning. A failing HBP lead can be converted to LBBAP through the same access, a backup right ventricular lead can be added to a free port, or the lead can be extracted. Indications for extraction (infection, malfunction, venous occlusion blocking an upgrade) are the same as for other leads and are summarized under cardiac implantable electronic device infection.
Leadless pacemakers: design and current systems
Leadless pacemakers integrate battery, circuitry, and a steroid-eluting tip electrode in a capsule implanted through a large femoral venous introducer. Two fixation designs exist: four self-expanding nitinol tines that engage trabeculated myocardium, and an active screw-in helix. Both designs carry a proximal docking feature used for delivery and retrieval.
| System | Fixation and size | Modes and sensor | Status and principal data |
|---|---|---|---|
| Nanostim (helix) | Screw-in helix; 42 mm, 1.0 cm³ | VVI(R); temperature sensor | LEADLESS II: implant success 95.8%. Implantation halted after premature battery depletion in 7 of 1,423 devices (0.5%) at 29 to 37 months |
| Micra VR (tines) | Four nitinol tines; 25.9 × 6.7 mm, 1.75 g; 23 French inner (27 French outer) introducer | VVI(R); 3-axis accelerometer | US approval 2016. Micra investigational device exemption (IDE) study: implant success 99.2%; 96.0% free of major complications at 6 months |
| Micra AV | Same capsule | VDD from accelerometer detection of atrial contraction; VVI(R) | US approval 2020. MARVEL 2: AV synchrony 89.2% at rest in complete heart block |
| Micra VR2 and AV2 | Same size, larger battery capacity | As above, with revised atrial sensing algorithms | US approval 2023; projected median longevity 16.7 years (VR2) and 15.6 years (AV2) |
| Aveir VR (helix) | Screw-in helix with mapping before fixation; 38 mm, 1.1 cm³; 25 French inner introducer | VVI(R); temperature sensor | LEADLESS II phase 2: implant success 98%; 96.0% free of serious device effects at 6 weeks |
| Aveir DR (Aveir AR plus Aveir VR) | Atrial capsule at the base of the right atrial appendage plus ventricular capsule | DDD(R) through beat-to-beat implant-to-implant conductive communication | Aveir DR i2i: implant success 98.3%; 90.3% free of complications at 90 days; ≥70% AV synchrony in 97.3% |
| Leadless pacemaker communicating with a subcutaneous implantable cardioverter-defibrillator (ICD) | Ventricular capsule | VVI pacing and antitachycardia pacing commanded by the subcutaneous ICD | MODULAR ATP: 97.5% free of pacemaker-related major complications; communication 98.8% |
Leadless pacing systems. Data from LEADLESS II, the Micra IDE study, the US Food and Drug Administration summary of safety and effectiveness for Micra, MARVEL 2, LEADLESS II phase 2, Aveir DR i2i, MODULAR ATP, and a 2025 device longevity model. The Micra capsule occupies less than 2% of right ventricular volume.
Communication and rate response. Tine-fixed devices use an accelerometer for rate response; helix-fixed devices use a blood temperature sensor. Programmers communicate with the capsules through skin-surface or wand-based telemetry. In the dual-chamber system, the two capsules exchange small conducted signals through blood and tissue beat by beat, so the ventricular device knows when the atrium has been sensed or paced.
Leadless pacemaker implantation
Implantation uses femoral venous access with ultrasound guidance, a large-bore introducer, intravenous heparin during the procedure, and fluoroscopy in RAO and LAO views. The delivery catheter crosses the tricuspid valve, and contrast injection outlines the right ventricle and septum. The capsule is deployed against the mid or apical septum with controlled forward pressure. For tine-fixed devices, a tug test confirms that at least two tines are engaged, and the electrical parameters are measured while the device is still tethered, so it can be recaptured and redeployed. For helix-fixed devices, the electrogram is mapped before fixation, the device is screwed in by rotation, and the tether is tested before release. Hemostasis after removal of the large-bore sheath is obtained with a figure-of-eight suture, compression, or a closure device.
Implantation site of leadless pacemakers. In the Micra IDE study, most devices were placed at the apex, and pericardial effusion occurred in 1.6%. Implantation guidance later shifted toward the septum, and pericardial effusion fell to 0.8% in the post-approval registry. Mid and high septal positions give a narrower paced QRS than apical septal sites with similar thresholds and sensing.
Jugular approach to leadless implantation. An internal jugular route gives a more natural angle to the septum, avoids femoral complications, and allows immediate ambulation. It has been used when the femoral route is unavailable or unsuccessful and in centers that prefer it routinely, with series of 100 jugular implants reporting results comparable with femoral implants.
Atrial and dual-chamber implantation. The atrial capsule of the dual-chamber system is fixed at the base of the right atrial appendage. In Aveir DR i2i, 4 patients (1.3%) had atrial capsule dislodgement during the procedure because of inadequate fixation, and 8 patients (2.7%) needed revision during follow-up for atrial dislodgement. The mean atrial capture threshold was 0.82 V and the mean P-wave amplitude 3.58 mV at 3 months; 21 patients (7%) had P waves below 1.0 mV, and none needed revision for sensing.
Candidates for leadless pacing and contraindications
The pacing indication itself is the same for leadless and transvenous systems. The choice of a leadless device rests on the risks the leadless design avoids, the functions it cannot provide, and the expected lifetime pacing needs. Leadless pacing is favored with absent or threatened upper-extremity venous access (bilateral occlusion, superior vena cava obstruction, or hemodialysis access that must be preserved), prior device infection or high infection risk (hemodialysis, diabetes, immunosuppression, frailty), and lead-related tricuspid regurgitation or a bioprosthetic tricuspid valve. It also suits frail patients with little subcutaneous tissue for a pocket and patients whose occupation or sport makes a pectoral device undesirable. The 2021 ESC guideline considers a leadless pacemaker an alternative to a transvenous system when no upper-extremity venous access exists or when the risk of pocket infection is particularly high, such as after previous infection and in patients on hemodialysis.
A ventricular leadless device fits patients who need ventricular pacing without atrial pacing: permanent atrial fibrillation with slow ventricular response, AV junction ablation, and intermittent AV block or reflex asystole with an expected low pacing burden. An accelerometer-based AV-synchronous device extends this to AV block with normal sinus function in less active patients. A dual-chamber leadless system adds atrial pacing for sinus node dysfunction, at substantially higher cost than a transvenous dual-chamber system.
Labeled contraindications. The Micra label contraindicates implantation in patients with an implanted inferior vena cava filter, a mechanical tricuspid valve, an implanted device that would interfere with implantation, or an implanted device providing active therapy that could interfere with sensing. Other labeled contraindications are femoral venous anatomy that cannot accommodate the 7.8-mm (23 French) introducer, morbid obesity that prevents telemetry within 12.5 cm, known intolerance to the device materials or heparin, sensitivity to contrast medium that cannot be adequately premedicated, and intolerance of a single 1.0-mg dose of dexamethasone acetate. Implantation through the internal jugular vein has been reported in patients with an inferior vena cava filter; such use falls outside the label. Leadless devices provide neither defibrillation nor biventricular pacing, so a patient likely to need either in the near future is better served by a system that can deliver it.
Outcomes of leadless pacing: trials and registries
All pivotal leadless studies were single-group studies compared against performance goals or historical transvenous cohorts. No randomized trial has compared leadless and transvenous pacing.
| Study | Population | Implant success | Safety and performance |
|---|---|---|---|
| LEADLESS II, Nanostim (2015) | 526 patients; primary cohort 300 | 95.8% | Device-related serious adverse events 6.7% at 6 months: dislodgement with retrieval 1.7%, perforation 1.3%, threshold rise requiring retrieval 1.3% |
| Micra IDE (2016) | 725 patients | 99.2% | 96.0% free of major complications at 6 months; no dislodgements; 98.3% with low, stable thresholds; major complications 51% lower than historical transvenous controls (HR 0.49) |
| Micra post-approval registry, 30 days (2017) | 795 patients, 96 centers | 99.6% | Major complications 1.51% at 30 days; effusion or perforation 0.13% |
| Micra post-approval registry, 5 years (2024) | 1,809 patients; median follow-up 51.1 months | — | Major complications 4.5% at 60 months; at 36 months 4.1% versus 8.5% with transvenous reference (HR 0.47). Revision 4.9% at 60 months, mostly upgrades (41.2%) or elevated thresholds (30.6%); no removals for infection |
| LEADLESS II phase 2, Aveir VR (2022) | 200 patients in the primary analysis | 98% | 96.0% free of serious adverse device effects at 6 weeks |
| Aveir DR i2i (2023) | 300 patients; sinus node dysfunction 63.3%, AV block 33.3% | 98.3% | 90.3% free of complications at 90 days; adequate atrial threshold and sensing 90.2%; ≥70% AV synchrony 97.3% |
| MODULAR ATP (2024) | 293 enrolled; 162 in 6-month end-point cohort; mean LVEF 33.1% | — | 97.5% free of pacemaker-related major complications; communication success 98.8%; 97.4% with threshold ≤2.0 V at 0.4 ms; antitachycardia pacing terminated 61.3% of episodes |
Principal studies of leadless pacemakers. HR, hazard ratio; IDE, investigational device exemption.
Complications of leadless pacing: perforation, vascular injury, and dislodgement
Cardiac perforation and tamponade. Perforation is the most serious complication of leadless implantation. It results from the stiff delivery system and capsule pressed against thin myocardium, and it is more frequent at the apex and free wall than at the septum. Across the pivotal studies, pericardial effusion occurred in 1.5% (LEADLESS II), 1.6% (Micra IDE), 0.8% (Micra post-approval registry), and 1.9% (LEADLESS II phase 2). In the Medicare comparison of 5,746 leadless and 9,662 transvenous single-chamber implants, pericardial effusion or perforation within 30 days occurred in 0.8% versus 0.4% after adjustment. An analysis of the US Food and Drug Administration adverse event database identified 363 major adverse events after Micra implantation from 2016 to 2020, including 96 deaths, with tamponade in 79.1% and rescue thoracotomy in 27.3%. Patients with myocardial tears survived more often with surgical repair. The authors of that analysis estimated the incidence at less than 1%. Implantation is performed where emergency pericardiocentesis and cardiac surgery are available.
Vascular complications of the large femoral introducer. The large femoral introducer causes hematoma, arteriovenous fistula, pseudoaneurysm, and, rarely, retroperitoneal bleeding or deep venous thrombosis; groin complications occurred in 0.6% to 1.2% of patients in the pivotal studies.
Dislodgement and embolization. Dislodgement occurred in 1.1% to 1.7% of patients with the original helix-fixed device, 0% in the Micra IDE study, and 1% in LEADLESS II phase 2. An embolized capsule travels to the pulmonary artery or, through a patent foramen ovale, to the left heart, and is retrieved with a snare or surgically.
Other complications of leadless pacing. Threshold rise requiring a new system, pacemaker syndrome with VVI pacing in sinus rhythm, and tricuspid valve injury from the delivery catheter or retrieval occur. Device infection is rare: in the Micra post-approval registry, no device was removed for infection during 5 years, and in the Medicare comparison infection at 3 years was below 0.2% with leadless devices.
Leadless versus transvenous pacing: complications, infection, and survival
The Micra Coverage with Evidence Development (CED) study linked US Medicare claims to device registration data and compared all leadless and transvenous single-chamber ventricular implants at the same facilities. Leadless recipients were sicker: end-stage kidney disease affected 12.0% versus 2.3%, and the Charlson comorbidity index was higher.
| Time point | Leadless (n) | Transvenous (n) | Adjusted findings |
|---|---|---|---|
| 30 days | 5,746 | 9,662 | Acute complications 7.7% versus 7.4% (not significant); pericardial effusion or perforation 0.8% versus 0.4% |
| 2 years | 6,219 | 10,212 | Reinterventions HR 0.62; chronic complications HR 0.69; all-cause mortality HR 0.97 (not significant) |
| 3 years | 6,219 | 10,212 | Reinterventions HR 0.59; chronic complications HR 0.68; infection below 0.2% versus 0.7%; heart failure hospitalization HR 0.90; all-cause mortality HR 0.97 (not significant) |
Leadless versus transvenous single-chamber ventricular pacing in US Medicare beneficiaries (Micra CED study). The comparison is observational, and residual confounding remains despite adjustment. HR, hazard ratio.
The pattern is consistent: leadless pacing trades a small excess of early perforation and access complications for fewer lead and pocket complications, reinterventions, and infections over time, without a difference in mortality. The comparator was a single-chamber transvenous device; the comparison with dual-chamber transvenous pacing in patients in sinus rhythm also involves AV synchrony and atrial pacing, which ventricular leadless devices do not provide.
AV-synchronous leadless pacing with an accelerometer
A ventricular capsule cannot sense atrial electrical activity reliably, so the AV-synchronous device detects atrial contraction mechanically. The accelerometer records the cardiac cycle as a sequence of signals: tricuspid and mitral closure, ventricular ejection, passive filling (A3), and the atrial contraction signal (A4). The device tracks the A4 signal and paces the ventricle after a programmed delay, providing VDD pacing without an atrial electrode and without atrial pacing.
In MARVEL 2, among 40 patients with sinus rhythm and complete AV block, AV synchrony at rest rose from 26.8% during VVI pacing to 89.2% during VDD pacing (median 94.3%), and 95% of patients had at least 70% synchrony. In AccelAV, 152 patients received the device; in 54 with complete AV block and normal sinus function, mean resting synchrony was 85.4% at 1 month and ambulatory synchrony 74.8%, rising to 82.6% in a subset with programming optimization. In a real-world analysis of 1,662 patients with ventricular pacing above 90%, the median proportion of atrial mechanical–sensed ventricular pacing was 74.7%. The median projected longevity across all devices in VDD mode was 10.5 years.
Limits of mechanical atrial sensing. Synchrony falls at higher sinus rates, when the A4 signal merges with passive filling, and in patients with weak atrial contraction, atrial fibrillation, or large respiratory and postural variation. Sinus rates above about 100 per minute are the principal limitation for active patients. Atrial pacing is not available, so chronotropic incompetence and sinus node dysfunction are not treated. Next-generation algorithms improved automatic synchrony at higher rates in simulation studies, but atrial sensing remains mechanical. Patients with AV block who are active or young, or who have sinus node disease, are better served by a dual-chamber transvenous or dual-chamber leadless system.
Dual-chamber leadless pacing with implant-to-implant communication
The dual-chamber system pairs an atrial capsule at the base of the right atrial appendage with a ventricular capsule. Beat by beat, the two devices signal sensed or paced events to one another through conducted communication, so AV timing behaves like a conventional DDD(R) pacemaker. In the Aveir DR i2i study of 300 patients, 63.3% had sinus node dysfunction and 33.3% had AV block as the primary indication. Implantation with established communication succeeded in 98.3%, 90.3% were free of complications at 90 days (performance goal 78%), and at least 70% AV synchrony while sitting was achieved in 97.3%.
The system offers atrial pacing and electrical atrial sensing, addressing the main limitations of the accelerometer-based device. Its limits are two femoral implants with two sets of perforation and dislodgement risks, higher cost, and a shorter battery life for the atrial capsule than for the ventricular capsule. Long-term data beyond the first years of follow-up are limited. Chronic retrieval of an atrial capsule from the thin appendage has been studied in an animal model and remains less well characterized in humans.
Tricuspid regurgitation with leadless and transvenous systems
Transvenous leads cause tricuspid regurgitation by impinging on, adhering to, or perforating leaflets and chordae, and the regurgitation can lead to right heart failure years after implantation. A leadless capsule leaves nothing across the valve, but the delivery catheter crosses it, the capsule sits near the subvalvular apparatus, and right ventricular pacing alters annular and papillary muscle function.
The data are mixed. In a 53-patient echocardiographic study, tricuspid regurgitation was graded more severe in 43% of patients 12 months after leadless implantation, similar to 38% in matched dual-chamber transvenous recipients. A septal capsule position was associated with more regurgitation than an apical one (odds ratio 5.20), and right ventricular function indexes and LVEF declined slightly. In a single-center study of 69 patients and a meta-analysis of 297 patients, significant tricuspid regurgitation did not change after leadless implantation (risk ratio 1.22, not significant). A comparison of 27 leadless and 24 transvenous recipients found less severe regurgitation with the leadless device.
For a patient with existing lead-related regurgitation, a leadless device avoids adding a second lead across the valve, and after transcatheter or surgical tricuspid valve repair or bioprosthetic replacement it avoids lead entrapment. CSP leads also cross the valve, and basal LBBAP sites close to the annulus are associated with more regurgitation.
Battery longevity, end of service, retrieval, and abandonment
Battery longevity of leadless pacemakers. The original tine-fixed device has a manufacturer-estimated longevity of about 12 years in VVI mode with 100% pacing at typical thresholds, and a real-world analysis over 7 years supported that estimate. A device longevity model projected median longevity of 12.3 years for the original VR device and 16.7 years for the VR2, and 10.8 years for the original AV device and 15.6 years for the AV2. In the model, 91% of VR2 recipients and 80% of AV2 recipients would need only one device over their lifetime. For the helix-fixed ventricular device, the manufacturer projects longevity of 17.6 years, a figure not yet confirmed by clinical follow-up. In the dual-chamber system, the atrial capsule has the shorter projected longevity.
Options at end of service. A depleted capsule can be retrieved and replaced, switched off and left in place with a new capsule implanted nearby, or switched off with a transvenous system implanted. Cadaver studies showed that up to three tine-fixed capsules can sit in the right ventricular septum without interaction, and two capsules have coexisted in patients. When a second capsule is implanted beside an abandoned one, imaging confirms spatial separation, and the abandoned device is programmed off so that it neither paces nor interferes.
Retrieval of tine-fixed devices. The tine-fixed device was designed to be abandoned at end of life, but it can be retrieved. In a worldwide series of 40 successful retrievals (29 with procedural detail), the most common indication for immediate retrieval was an elevated threshold after tether release. Delayed retrieval (median 46 days, up to 95 days) was performed for elevated threshold, endovascular infection, or the need for a transvenous system. There were no serious complications. In the premarket and continued access studies (989 patients), the revision rate at 24 months was 1.4%, 75% lower than the 5.3% with transvenous pacemakers. In 64% of revisions the capsule was switched off and left in place, and percutaneous retrieval succeeded as late as 14 months. In a single-center series of 9 nonacute retrievals after 3.1 ± 2.8 years (up to 9.0 years), 8 succeeded (88.9%); the failure was a capsule embedded in surrounding tissue.
Retrieval of helix-fixed devices. The screw-in design was intended for chronic retrieval by counter-rotation with a dedicated catheter. With the first-generation helix-fixed device (Nanostim), 66 of 73 devices were retrieved at a median of 2.9 years; in 6 of the 7 failures, the docking button could not be reached. Chronic retrieval data for the current helix-fixed ventricular and atrial devices are limited. Retrieval is performed where pericardiocentesis and cardiac surgery are immediately available.
Infection. Endovascular infection involving a leadless capsule is rare. When device infection is established, removal follows the principles for any infected cardiac device.
Leadless pacing with a subcutaneous ICD and leadless resynchronization
The subcutaneous ICD avoids transvenous leads but cannot provide bradycardia pacing or antitachycardia pacing; its other limitations are covered under limitations of the subcutaneous ICD. A modular system pairs a leadless ventricular capsule with a subcutaneous ICD that commands it to deliver antitachycardia pacing or pacing after a shock. In MODULAR ATP, 293 patients at risk for sudden death were enrolled, with 162 in the 6-month end-point cohort. Freedom from pacemaker-related major complications was 97.5%, and communication succeeded in 98.8% of tests. Of the 151 patients with 6-month follow-up, 97.4% had thresholds of 2.0 V or less at 0.4 ms. Antitachycardia pacing terminated 61.3% of treated episodes, with no episode in which pacing failed because of communication. Eight of 162 patients died, none from arrhythmia or the procedure. When both a leadless pacemaker and a separate subcutaneous ICD are implanted without such a system, the ICD's sensing is screened during ventricular pacing, because the paced QRS and T wave can cause double counting.
Leadless resynchronization uses a wireless left ventricular endocardial electrode stimulated by ultrasound from a subcutaneous transmitter. The transmitter senses the right ventricular pacing output of a coexisting transvenous or leadless device, and the electrode paces the left ventricle almost simultaneously. Phase I of SOLVE-CRT described its initial clinical use in nonresponders to conventional resynchronization and in patients in whom coronary sinus pacing was not possible or carried high risk. Its results in SOLVE-CRT are given under epicardial, endocardial, and leadless left ventricular pacing. Leadless conduction system pacing requires smaller devices with fixation suited to the septum and has no clinical data.
MRI, electromagnetic interference, and surgery
Magnetic resonance imaging (MRI). Current leadless devices are labeled MRI conditional at 1.5 and 3 tesla when scanned under their labeled conditions, with the device programmed to an MRI mode before the scan. CSP leads are MRI conditional only as part of specified lead and generator combinations; a lead from one manufacturer connected to another's generator, an abandoned lead, or an adapter falls outside labeling and is managed with the protocol used for non-MR-conditional systems. Pacing-dependent patients are programmed to asynchronous pacing during the scan. The MagnaSafe experience and the protocol for non-conditional systems are summarized under MRI in patients with pacemakers.
Magnet response and electromagnetic interference. Leadless devices have no pocket to place a magnet over; their magnet response, where present, differs by manufacturer and is confirmed before a procedure. Responses are tabulated under magnet response of pacemakers, and the perioperative approach is detailed under perioperative management. Perioperative care relies on reprogramming to asynchronous pacing for pacing-dependent patients when monopolar electrosurgery is used near the heart, and on bipolar electrosurgery when possible. External defibrillation can raise thresholds at the electrode–tissue interface, so the device is interrogated afterward. For CSP systems, the generator behaves like any conventional pacemaker, but reprogramming to asynchronous pacing at an output that secures myocardial (not only conduction system) capture prevents loss of capture during the procedure.
Choosing among conduction system, leadless, and conventional pacing
The choice depends on the pacing indication, expected ventricular pacing burden, LVEF, venous access, infection risk, and age and life expectancy, and on any need for atrial pacing, AV synchrony, defibrillation, or resynchronization. A young patient with complete heart block will accumulate decades of ventricular pacing and several device changes, whereas a frail older patient with permanent atrial fibrillation and slow ventricular response needs reliable backup pacing with the fewest complications.
| Clinical situation | Usual first choice | Reasoning |
|---|---|---|
| Complete or high-grade AV block, LVEF 36%–50% | Conduction system pacing (LBBAP preferred), or biventricular pacing | Class 2a for physiologic pacing; randomized data show less pacing-induced cardiomyopathy than right ventricular pacing |
| Complete or high-grade AV block, normal LVEF | LBBAP when expertise is available; right ventricular pacing acceptable | Class 2b; benefit greatest when pacing burden is high |
| Sinus node dysfunction, intact AV conduction | Dual-chamber transvenous system with ventricular pacing minimized; dual-chamber leadless system when transvenous risk is high | Low ventricular pacing burden; atrial pacing needed |
| Permanent atrial fibrillation with slow ventricular response, low expected pacing | Ventricular leadless pacemaker or single-chamber transvenous system | No atrial function to restore; leadless avoids lead and pocket complications |
| AV junction ablation | Biventricular pacing if LVEF ≤50%; CSP (LBBAP or HBP) otherwise | Pacing dependence and near-100% ventricular pacing |
| Prior device infection, hemodialysis, venous occlusion | Leadless pacemaker | Avoids pocket and transvenous lead; ESC guidance supports it |
| Lead-related tricuspid regurgitation or tricuspid valve intervention | Leadless pacemaker, or coronary sinus or epicardial pacing | Avoids a lead across the valve; mechanical tricuspid valve contraindicates leadless implantation |
| Pacing-induced cardiomyopathy | Upgrade to biventricular pacing or revision to CSP | Class 1 and Class 2a |
| Need for defibrillation or resynchronization now or soon | Transvenous ICD or CRT system, with a CSP lead where appropriate | Leadless devices provide neither |
Typical choice of pacing system by clinical situation, derived from the 2023 HRS/APHRS/LAHRS guideline, the 2021 ESC guideline, and the device data above. Individual anatomy, operator experience, and patient preference modify the choice.
Pacing indications and mode selection are covered on the bradycardia and atrioventricular block page, and the conventional pacing modes under pacemaker code and pacing modes.
Sources of error in conduction system and leadless pacing
- Accepting a narrow paced QRS as conduction system capture. Deep septal and left ventricular septal pacing also narrow the QRS. Left bundle capture requires a specific criterion: an output-dependent transition, a short stimulus to V6 R-wave peak time, a V6–V1 interpeak interval above 44 ms, or a left bundle potential with matching timing.
- Applying the narrow-QRS cutoff in LBBB. A stimulus to V6 R-wave peak time below 74 ms applies with a narrow native QRS; with LBBB, intraventricular conduction delay, or a wide escape rhythm the cutoff is 80 ms.
- Measuring R-wave peak time from QRS onset. Capture criteria are measured from the pacing stimulus at high sweep speed; measurement from an inferred QRS onset shortens the interval and overcalls capture.
- Testing capture in bipolar configuration at high output. Anodal ring capture produces right ventricular myocardial activation and false transitions. Classification thresholds are measured in unipolar configuration.
- Treating nonselective His capture and para-Hisian pacing as equivalent. Both show a pseudo-delta wave. Slurring or notching in I, V1, or V4 to V6, or a V6 R-wave peak time above 100 ms, indicates myocardial capture only.
- Relying on device capture counters after CSP. Loss of His or left bundle capture with retained myocardial capture is recorded as successful capture. Only the 12-lead ECG at follow-up reveals it.
- Using automatic capture algorithms with HBP. The algorithm tracks myocardial capture and can program output below the His threshold unless the two thresholds are known to be similar.
- Advancing the helix after a left bundle potential or a sharp fall in impedance appears. Both signal that the tip is at the left ventricular subendocardium; further rotation risks perforation into the left ventricle.
- Implanting HBP in a pacing-dependent patient without a backup plan. Late threshold rise and progression of distal disease make loss of capture likely in some patients; a backup lead may be reasonable.
- Assuming a leadless ventricular device avoids pacing-induced cardiomyopathy. It paces right ventricular myocardium and carries the same dyssynchrony as a transvenous right ventricular lead at a similar site.
- Choosing an accelerometer-based AV-synchronous device for an active patient with AV block. AV synchrony falls at sinus rates above about 100 per minute, and atrial pacing is not available.
- Overlooking labeled contraindications. An implanted inferior vena cava filter and a mechanical tricuspid valve are labeled contraindications to implantation of the tine-fixed device.
- Implanting a leadless device at the apex or free wall. Apical and free wall positions carry more perforation risk than the septum, and in adverse event reports perforation after leadless implantation more often required thoracotomy or led to death than perforation by a transvenous lead.
- Leaving an abandoned capsule active. A depleted or replaced capsule that is not programmed off can interfere with a new device placed beside it.
References
- Chung MK, Patton KK, Lau CP, et al. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm. 2023;20:e17–e91. doi:10.1016/j.hrthm.2023.03.1538
- Burri H, Jastrzebski M, Cano Ó, et al. EHRA clinical consensus statement on conduction system pacing implantation: endorsed by the Asia Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS), and Latin American Heart Rhythm Society (LAHRS). Europace. 2023;25:1208–1236 (erratum, reference numbering: Europace. 2023;25:1520). doi:10.1093/europace/euad043
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation. 2019;140:e382–e482. doi:10.1161/CIR.0000000000000628
- Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42:3427–3520. doi:10.1093/eurheartj/ehab364
- Deshmukh P, Casavant DA, Romanyshyn M, Anderson K. Permanent, direct His-bundle pacing: a novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation. 2000;101:869–877. doi:10.1161/01.cir.101.8.869
- Huang W, Su L, Wu S, et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can J Cardiol. 2017;33:1736.e1–1736.e3. doi:10.1016/j.cjca.2017.09.013
- Huang W, Chen X, Su L, et al. A beginner's guide to permanent left bundle branch pacing. Heart Rhythm. 2019;16:1791–1796. doi:10.1016/j.hrthm.2019.06.016
- Jastrzębski M, Burri H, Kiełbasa G, et al. The V6-V1 interpeak interval: a novel criterion for the diagnosis of left bundle branch capture. Europace. 2022;24:40–47. doi:10.1093/europace/euab164
- Jastrzębski M, Kiełbasa G, Cano O, et al. Left bundle branch area pacing outcomes: the multicentre European MELOS study. Eur Heart J. 2022;43:4161–4173. doi:10.1093/eurheartj/ehac445
- Jastrzębski M, Kiełbasa G, Cano O, et al. Left bundle branch area pacing vs right ventricular pacing for atrioventricular block: the MELOS RELOADED study. Eur Heart J. 2026;47:1541–1550. doi:10.1093/eurheartj/ehaf699
- Vijayaraman P, Naperkowski A, Subzposh FA, et al. Permanent His-bundle pacing: long-term lead performance and clinical outcomes. Heart Rhythm. 2018;15:696–702. doi:10.1016/j.hrthm.2017.12.022
- Abdelrahman M, Subzposh FA, Beer D, et al. Clinical outcomes of His bundle pacing compared to right ventricular pacing. J Am Coll Cardiol. 2018;71:2319–2330. doi:10.1016/j.jacc.2018.02.048
- Sharma PS, Patel NR, Ravi V, et al. Clinical outcomes of left bundle branch area pacing compared to right ventricular pacing: results from the Geisinger-Rush Conduction System Pacing Registry. Heart Rhythm. 2022;19:3–11 (corrigendum: Heart Rhythm. 2023;20:1100). doi:10.1016/j.hrthm.2021.08.033
- Chow CLD, Wong C, Sutherland N, et al. Clinical outcomes of conduction system pacing vs right ventricular septal pacing in atrioventricular block: the CSPACE randomized controlled trial. J Am Coll Cardiol. 2025;86:563–573. doi:10.1016/j.jacc.2025.06.043
- Qiu N, Liu X, Wang Z, et al. Randomized trial of left bundle branch pacing vs right ventricular pacing in vulnerable cardiac function. J Am Coll Cardiol. 2026;88:188–201. doi:10.1016/j.jacc.2026.03.161
- Huang W, Wang S, Su L, et al. His-bundle pacing vs biventricular pacing following atrioventricular nodal ablation in patients with atrial fibrillation and reduced ejection fraction: a multicenter, randomized, crossover study—the ALTERNATIVE-AF trial. Heart Rhythm. 2022;19:1948–1955. doi:10.1016/j.hrthm.2022.07.009
- Wang Y, Zhu H, Hou X, et al. Randomized trial of left bundle branch vs biventricular pacing for cardiac resynchronization therapy. J Am Coll Cardiol. 2022;80:1205–1216. doi:10.1016/j.jacc.2022.07.019
- Jastrzębski M, Moskal P, Huybrechts W, et al. Left bundle branch-optimized cardiac resynchronization therapy (LOT-CRT): results from an international LBBAP collaborative study group. Heart Rhythm. 2022;19:13–21. doi:10.1016/j.hrthm.2021.07.057
- Vijayaraman P, Trivedi RS, Koneru JN, et al. Transvenous extraction of conduction system pacing leads: an international multicenter (TECSPAM) study. Heart Rhythm. 2024;21:1953–1961. doi:10.1016/j.hrthm.2024.04.054
- Reddy VY, Exner DV, Cantillon DJ, et al. Percutaneous implantation of an entirely intracardiac leadless pacemaker. N Engl J Med. 2015;373:1125–1135. doi:10.1056/NEJMoa1507192
- Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374:533–541. doi:10.1056/NEJMoa1511643
- Roberts PR, Clementy N, Al Samadi F, et al. A leadless pacemaker in the real-world setting: the Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm. 2017;14:1375–1379. doi:10.1016/j.hrthm.2017.05.017
- El-Chami MF, Garweg C, Clementy N, et al. Leadless pacemakers at 5-year follow-up: the Micra transcatheter pacing system post-approval registry. Eur Heart J. 2024;45:1241–1251. doi:10.1093/eurheartj/ehae101
- Piccini JP, El-Chami M, Wherry K, et al. Contemporaneous comparison of outcomes among patients implanted with a leadless vs transvenous single-chamber ventricular pacemaker. JAMA Cardiol. 2021;6:1187–1195 (erratum: JAMA Cardiol. 2021;6:1344). doi:10.1001/jamacardio.2021.2621
- El-Chami MF, Bockstedt L, Longacre C, et al. Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. Eur Heart J. 2022;43:1207–1215. doi:10.1093/eurheartj/ehab767
- Crossley GH, Piccini JP, Longacre C, et al. Leadless versus transvenous single-chamber ventricular pacemakers: 3 year follow-up of the Micra CED study. J Cardiovasc Electrophysiol. 2023;34:1015–1023. doi:10.1111/jce.15863
- Hauser RG, Gornick CC, Abdelhadi RH, et al. Major adverse clinical events associated with implantation of a leadless intracardiac pacemaker. Heart Rhythm. 2021;18:1132–1139. doi:10.1016/j.hrthm.2021.03.015
- Reddy VY, Exner DV, Doshi R, et al. Primary results on safety and efficacy from the LEADLESS II–Phase 2 worldwide clinical trial. JACC Clin Electrophysiol. 2022;8:115–117. doi:10.1016/j.jacep.2021.11.002
- Steinwender C, Khelae SK, Garweg C, et al. Atrioventricular synchronous pacing using a leadless ventricular pacemaker: results from the MARVEL 2 study. JACC Clin Electrophysiol. 2020;6:94–106. doi:10.1016/j.jacep.2019.10.017
- Chinitz LA, El-Chami MF, Sagi V, et al. Ambulatory atrioventricular synchronous pacing over time using a leadless ventricular pacemaker: primary results from the AccelAV study. Heart Rhythm. 2023;20:46–54. doi:10.1016/j.hrthm.2022.08.033
- Garweg C, Piccini JP, Epstein LM, et al. Correlation between AV synchrony and device collected AM-VP sequence counter in atrioventricular synchronous leadless pacemakers: a real-world assessment. J Cardiovasc Electrophysiol. 2023;34:197–206. doi:10.1111/jce.15726
- Knops RE, Reddy VY, Ip JE, et al. A dual-chamber leadless pacemaker. N Engl J Med. 2023;388:2360–2370. doi:10.1056/NEJMoa2300080
- Knops RE, Lloyd MS, Roberts PR, et al. A modular communicative leadless pacing–defibrillator system. N Engl J Med. 2024;391:1402–1412. doi:10.1056/NEJMoa2401807
- Okabe T, Hummel JD, Bank AJ, et al. Leadless left ventricular stimulation with WiSE-CRT System—initial experience and results from phase I of SOLVE-CRT Study (nonrandomized, roll-in phase). Heart Rhythm. 2022;19:22–29. doi:10.1016/j.hrthm.2021.06.1195
- Beurskens NEG, Tjong FVY, de Bruin-Bon RHA, et al. Impact of leadless pacemaker therapy on cardiac and atrioventricular valve function through 12 months of follow-up. Circ Arrhythm Electrophysiol. 2019;12:e007124. doi:10.1161/CIRCEP.118.007124
- Haeberlin A, Bartkowiak J, Brugger N, et al. Evolution of tricuspid valve regurgitation after implantation of a leadless pacemaker: a single center experience, systematic review, and meta-analysis. J Cardiovasc Electrophysiol. 2022;33:1617–1627. doi:10.1111/jce.15565
- Leal MA, Sheldon T, Escalante K, et al. Device longevity of a leadless pacemaker family. Future Cardiol. 2025;21:753–758. doi:10.1080/14796678.2025.2527466
- Breeman KTN, Oosterwerff EFJ, Dijkshoorn LA, et al. Real-world long-term battery longevity of Micra leadless pacemakers. J Interv Card Electrophysiol. 2023;66:839–841. doi:10.1007/s10840-022-01447-y
- Afzal MR, Daoud EG, Cunnane R, et al. Techniques for successful early retrieval of the Micra transcatheter pacing system: a worldwide experience. Heart Rhythm. 2018;15:841–846. doi:10.1016/j.hrthm.2018.02.008
- Grubman E, Ritter P, Ellis CR, et al. To retrieve, or not to retrieve: system revisions with the Micra transcatheter pacemaker. Heart Rhythm. 2017;14:1801–1806. doi:10.1016/j.hrthm.2017.07.015
- Funasako M, Hála P, Janotka M, et al. Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker. Europace. 2024;26:euae256. doi:10.1093/europace/euae256
- Saleem-Talib S, Hoevenaars CPR, Molitor N, et al. Leadless pacing: a comprehensive review. Eur Heart J. 2025;46:1979–1990. doi:10.1093/eurheartj/ehaf119
- US Food and Drug Administration. Summary of safety and effectiveness data: Micra Transcatheter Pacing System, Model MC1VR01 (PMA P150033). Approved April 6, 2016. accessdata.fda.gov