Cardiac MRI physics, pulse sequences, artifacts, and safety

The physics of cardiovascular magnetic resonance from nuclear magnetization and image formation through the clinical pulse sequences, artifacts, MR environment safety, gadolinium agents, and imaging of patients with implanted devices.

What cardiovascular magnetic resonance measures and what limits it

Cardiovascular magnetic resonance (CMR) builds images from the behavior of hydrogen nuclei in a strong magnetic field. It uses no ionizing radiation, places image planes at any orientation independent of acoustic windows, and produces contrast from several independent tissue properties: longitudinal relaxation (T1), transverse relaxation (T2 and T2*), proton density, flow velocity, and the distribution of an extracellular contrast agent. These properties allow one examination to measure ventricular volumes and mass, regional function, myocardial scar and infiltration, edema, iron, perfusion, and blood flow.

The same physics sets the limits of the test. Images are assembled over many heartbeats, so gating, rhythm, and breath-holding govern quality. Quantitative values depend on field strength, sequence, and vendor, so reference ranges are local. The examination takes place inside a static field tens of thousands of times stronger than Earth's magnetic field. Radiofrequency (RF) energy is deposited in tissue and the gradients switch rapidly, so screening for implants and ferromagnetic objects is part of all studies.

The clinical interpretation of late gadolinium enhancement for viability is covered under transmural extent of scar. CMR deformation imaging is covered under cardiac magnetic resonance strain, and stress protocols under pairing the stressor with cardiac magnetic resonance. The device-side protocol for magnetic resonance imaging (MRI) appears in the device follow-up page's MRI section. Disease-specific CMR findings appear in the hubs on cardiac amyloidosis, hypertrophic cardiomyopathy, and constrictive pericarditis.

Nuclear magnetization, precession, and the Larmor equation

The hydrogen nucleus is a single proton with spin and an associated magnetic moment. Hydrogen dominates clinical MRI because water and fat make up most of the body's mass and the proton gives a strong signal. Outside a magnet the moments point randomly and cancel. In a static field (B0) a very small excess of protons aligns with the field, and their sum is the net magnetization (M0), a vector along the field direction, called the z-axis. M0 grows in proportion to B0, which is the basic reason signal-to-noise ratio (SNR) rises with field strength.

The moments precess around the field direction at the Larmor frequency, given by the Larmor equation:

f0 = (γ/2π) × B0

where γ/2π is the gyromagnetic ratio expressed in frequency units, 42.58 MHz per tesla for the proton. The resonant frequency is therefore about 64 MHz at 1.5 tesla (T) and 128 MHz at 3 T, within the radio band (Ridgway 2010). Only an RF field oscillating at this frequency transfers energy to the protons, which is why the scanner's transmitter, receiver, and shielded room are tuned to the Larmor frequency of the system.

The local field experienced by a proton differs slightly from B0. Electron shielding in fat molecules lowers the resonant frequency of fat protons relative to water by roughly 210 Hz at 1.5 T and 420 Hz at 3 T, the chemical shift (Ferreira 2013). Differences in magnetic susceptibility between air, tissue, blood, and metal distort the field over millimeters to centimeters. These small frequency differences produce the fat-related and off-resonance artifacts described later and are exploited deliberately for fat suppression and T2* mapping.

Excitation, flip angle, and the free induction decay

An RF pulse at the Larmor frequency, applied by the transmit coil perpendicular to B0, tips the net magnetization away from the z-axis. The angle through which it is tipped, the flip angle, depends on the amplitude and duration of the pulse. A 90° pulse places all of M0 in the transverse (x-y) plane; a 180° pulse inverts it to point along −z; small flip angles of 10° to 70° leave most of the magnetization along z and are the basis of fast gradient-echo imaging.

The transverse component rotates at the Larmor frequency and induces a voltage in the receive coil, the free induction decay. Signal is available only from transverse magnetization, and the longitudinal component determines how much magnetization the next pulse can tip. Image contrast arises from how quickly a tissue loses transverse magnetization and regains longitudinal magnetization between pulses, and from how the sequence times its pulses and echoes.

T1, T2, and T2* relaxation and tissue contrast

T1 (longitudinal or spin-lattice) relaxation is the recovery of the z-component of magnetization toward M0 after excitation, as protons release energy to the surrounding molecular lattice. Recovery follows an exponential curve: after an inversion pulse, Mz(t) = M0 × (1 − 2e^(−t/T1)), and Mz passes through zero at t = T1 × ln 2, about 0.69 × T1. Energy exchange is most efficient when molecular tumbling is close to the Larmor frequency. Fat has a short T1; free water, with fast tumbling, has a long T1; water bound to macromolecules in muscle has an intermediate T1, and edema lengthens it (Ridgway 2010). T1 lengthens with field strength, so native T1 values at 3 T exceed those at 1.5 T.

T2 (transverse or spin-spin) relaxation is the irreversible loss of phase coherence among precessing protons from interactions between neighboring spins. Free water has a long T2, so edema raises myocardial T2.

T2* relaxation is the faster decay of transverse magnetization that combines T2 with dephasing from static field inhomogeneities. The inhomogeneity component is reversible by a 180° refocusing pulse in spin-echo sequences but not in gradient-echo sequences. Paramagnetic substances, including hemosiderin and ferritin iron and deoxyhemoglobin, increase local inhomogeneity and shorten T2*.

Contrast agents shorten relaxation times. Gadolinium chelates are paramagnetic and shorten T1 (and, at high concentration, T2 and T2*) of the water protons that come close to the gadolinium ion. Their effect is proportional to local concentration, which is the basis of enhancement imaging and extracellular volume measurement.

Relaxation propertyPhysical basisIncreased byDecreased byClinical CMR measurement
Native T1Recovery of longitudinal magnetization; energy exchange with the latticeEdema, acute injury, fibrosis, amyloid, higher field strengthFat, iron overload, glycosphingolipid in Anderson-Fabry disease, hemorrhageT1 mapping (inversion-recovery and saturation-recovery methods)
Post-contrast T1Gadolinium concentration in the extracellular spaceNormal myocardium with little extracellular space, washout over timeScar, interstitial expansion, amyloidLate gadolinium enhancement, post-contrast T1 map, ECV
T2Irreversible spin-spin dephasingFree water: edema, acute inflammation, acute infarctionIron; hemorrhage at the core of an infarctT2 mapping, T2-weighted STIR, or T2-prepared balanced steady-state free precession (bSSFP)
T2*T2 plus dephasing from local field inhomogeneityHomogeneous field, absence of paramagnetic materialIron, intramyocardial hemorrhage, metal, air-tissue interfacesMulti-echo gradient-echo T2* at 1.5 T

Relaxation properties and the tissue changes that alter them, after the 2017 mapping consensus of the Society for Cardiovascular Magnetic Resonance (SCMR) and the European Association of Cardiovascular Imaging (EACVI). Values depend on field strength, sequence, and vendor, so laboratories interpret measurements against their own reference ranges. ECV, extracellular volume fraction; STIR, short tau inversion recovery.

Spatial encoding: slice selection, phase encoding, and frequency encoding

The scanner localizes signal with three orthogonal gradient coils that add a small, position-dependent field to B0, so that the Larmor frequency varies linearly with position along the gradient.

Slice selection. A gradient is switched on during the RF pulse. Only protons in the plane whose Larmor frequency falls within the pulse's frequency band are excited. The bandwidth of the pulse and the gradient strength set the slice thickness, and the center frequency sets its position. An oblique cardiac plane is produced by combining gradients along two or three physical axes.

Phase encoding. After excitation, a gradient applied briefly along one in-plane direction makes protons precess faster or slower according to position. When it switches off, a column of tissue retains a phase offset proportional to its position. The sequence repeats with a different phase-encoding gradient amplitude per line of data, so the number of phase-encoding steps determines the number of pixels in that direction and much of the acquisition time (Ridgway 2010).

Frequency encoding. During signal readout a gradient along the other in-plane direction makes different positions emit at different frequencies. The receiver samples the echo, and the frequencies present are separated into positions. The range of frequencies across the field of view is the receiver bandwidth: a wide bandwidth shortens the readout and reduces chemical shift and susceptibility distortion but admits more noise.

The phase-encoding direction is the direction in which motion and flow produce ghosts and in which anatomy outside the field of view wraps into the image, so its orientation is chosen with these artifacts in mind.

k-space and image formation

A sampled echo fills one line of a raw data matrix called k-space, in which position corresponds to spatial frequency and not to anatomic location. A two-dimensional Fourier transform converts the filled matrix into the image. The center of k-space holds the low spatial frequencies, which determine overall signal and contrast. The periphery holds the high spatial frequencies, which define edges and fine detail. Two consequences follow. First, the time at which the center of k-space is acquired sets the effective contrast of the image, so the order in which lines are acquired (sequential, centric, or interleaved) matters for inversion-recovery and perfusion sequences. Second, motion during acquisition of the central lines degrades the whole image more than motion during the peripheral lines (Ferreira 2013).

The extent of k-space sampled sets spatial resolution, and the spacing between lines sets the field of view. Sampling too few peripheral lines blurs edges and produces Gibbs ringing; spacing lines too far apart shrinks the field of view and causes wrap-around. Partial-Fourier acquisition samples slightly more than half of k-space and synthesizes the rest from conjugate symmetry, saving time at a cost in SNR (Ferreira 2013).

Cardiac imaging divides k-space into segments. A segmented cine acquires a few lines (the views per segment) for all cardiac phases per heartbeat, so full images for all phases take a breath-hold of several heartbeats. Single-shot imaging fills all of k-space within one heartbeat, sacrificing spatial and temporal resolution to remove dependence on breath-holding and rhythm.

Repetition time, echo time, flip angle, and image weighting

The repetition time (TR) is the interval between successive excitations of the same tissue; the echo time (TE) is the interval between excitation and the center of the echo. Together with the flip angle and any preparation pulse, they determine which tissue property dominates image contrast.

A short TR does not allow full T1 recovery, so tissues with short T1 give more signal (T1 weighting). A long TE allows more T2 or T2* decay, so tissues with long T2 stay bright (T2 weighting). A long TR and short TE minimize both effects and reflect proton density. Preparation pulses add contrast independent of the readout: an inversion pulse for T1 weighting and myocardial nulling, a saturation pulse for perfusion imaging that is insensitive to heart rate, a T2-preparation module for edema imaging and coronary angiography, and spectral fat suppression.

ParameterRaising itMain trade-off
Repetition time (TR)More T1 recovery and signal per excitation; longer scanIn bSSFP, longer TR brings off-resonance bands closer together
Echo time (TE)More T2 or T2* weighting; more dephasing in jets and near metalLonger TE enlarges susceptibility voids and flow-related signal loss
Flip angleMore T1 weighting in spoiled GRE; more signal in bSSFP up to a limitSpecific absorption rate (SAR) rises with the square of the flip angle
Receiver bandwidthShorter readout, less chemical shift and distortionLower SNR
Matrix in the phase directionFiner spatial resolutionLonger scan time and lower SNR per voxel
Slice thicknessMore SNRThrough-plane partial volume, particularly at the apex and base
Views per segmentShorter breath-holdLonger acquisition window per phase and poorer temporal resolution
Parallel imaging factorShorter scan or higher resolutionSNR falls by at least the square root of the factor, with residual aliasing

Acquisition parameters and their trade-offs, after the CMR physics reviews of Ridgway (2010) and Biglands (2012). bSSFP, balanced steady-state free precession; GRE, gradient echo; SNR, signal-to-noise ratio.

Signal-to-noise ratio, spatial and temporal resolution, and scan time

Image quality is a balance among SNR, spatial resolution, temporal resolution, coverage, and acquisition time. SNR is proportional to voxel volume and to the square root of the number of signal samples. Halving the in-plane pixel size in both directions reduces voxel volume, and therefore SNR, by a factor of four. Doubling the number of averages increases SNR by only about 40%. All cardiac acquisitions are also limited by the time the patient can hold a breath and by the length of the quiescent period in the cardiac cycle.

Temporal resolution in cine imaging is the time between acquired frames, roughly views per segment × TR. The 2020 SCMR protocols specify 45 ms or less between cine phases for wall-motion evaluation and 50 ms or less for phase-contrast flow, calculated from frames actually acquired, because vendors can interpolate additional frames that add no information (Kramer 2020). In late gadolinium enhancement, the acquisition window per heartbeat is kept below 200 ms and shorter during tachycardia to avoid blurring.

Scanner hardware: magnet, gradients, radiofrequency coils, and shimming

A clinical system has three electromagnetic components (Ridgway 2010). The main magnet is a superconducting solenoid cooled by liquid helium to produce a static field that is always on, including when the scanner is not imaging. The gradient coils, inside the magnet bore, produce the switched linear fields used for spatial encoding; their maximum amplitude (mT/m) and slew rate (T/m/s) determine how short TR and TE can be. Rapid switching of gradient current in the static field causes the vibration that produces acoustic noise and induces electric fields in the body. The RF system includes the body transmit coil built into the bore and multi-element receive coils placed over the chest. SCMR recommends a multi-element cardiac surface coil, typically with 8 or more elements, because parallel imaging requires it. SCMR also recommends electrocardiogram (ECG) hardware that supports vectorcardiographic gating (Kramer 2020).

Shimming adjusts small correcting currents so that B0 is as uniform as possible over the heart. Global shimming cannot remove local inhomogeneity from susceptibility differences at the lung-heart and heart-liver interfaces and around the cardiac veins. At 3 T, SCMR advises careful shimming over the heart before balanced steady-state free precession imaging and, in some patients, a patient-specific frequency adjustment, for which a frequency scout is helpful (Kramer 2020).

Surface coil intensity. Receive sensitivity falls with distance from the coil elements, so anterior structures appear brighter than posterior ones on uncorrected images. Vendors apply intensity normalization, and coil placement centered on the heart limits shading. Uncorrected coil shading alters signal-intensity comparisons across the myocardium, which matters in visual perfusion and edema imaging.

Field strength: 0.55 T, 1.5 T, and 3 T

Most clinical CMR is performed at 1.5 T, with an increasing share at 3 T (Kramer 2020). SCMR requires a field strength of 1.0 T or more for clinical CMR. At 3 T, SNR is approximately doubled and RF power deposition is approximately quadrupled, because SAR rises with the square of the field strength (Ferreira 2013). Off-resonance from susceptibility and chemical shift doubles in frequency terms, B1 transmit inhomogeneity becomes significant, and the ECG is more distorted by the magnetohydrodynamic effect.

Low-field systems built with modern gradients and reconstruction have reopened interest in 0.55 T. In a 0.55-T system adapted from a 1.5-T platform, T1 was on average 32% shorter, T2 26% longer, and T2* 40% longer than at 1.5 T. Image distortion at air-tissue interfaces in the lungs, airway, sinuses, and bowel was reduced. In addition, 9 of 16 commercially available metallic interventional devices heated less than 1 °C at the tip during 2 minutes of continuous imaging (Campbell-Washburn 2019).

Feature0.55 T1.5 T3 T
Proton resonant frequencyAbout 23 MHzAbout 64 MHzAbout 128 MHz
Fat-water frequency differenceAbout 77 HzAbout 210 HzAbout 420 Hz
Relative SNRLowest; offset by efficient samplingReferenceAbout twice 1.5 T
RF power depositionLowestReferenceAbout four times 1.5 T
bSSFP cineRobust, with lower SNRDefault cine methodBanding and flow artifacts; careful shimming and frequency adjustment, or spoiled GRE
Perfusion and LGEFeasibleStandardBenefits from higher SNR and contrast
T2* iron quantificationNot standardizedValidated thresholdsShorter values; 1.5-T thresholds do not apply
DevicesLess heating and susceptibility artifactField strength of most device labeling and of the large legacy-device registriesOnly devices labeled for 3 T

Field-strength comparison drawn from the 2020 SCMR protocols, the 2017 SCMR/EACVI mapping consensus, Ferreira (2013), and Campbell-Washburn (2019). Fat-water frequency differences are approximate. bSSFP, balanced steady-state free precession; GRE, gradient echo; LGE, late gadolinium enhancement.

ECG gating: vectorcardiographic triggering, prospective triggering, and retrospective gating

Because most cardiac acquisitions take several heartbeats, data are synchronized to the cardiac cycle. Flowing blood in the static field generates a voltage (the magnetohydrodynamic effect) that adds to the surface ECG, mainly during systole, and can be larger than the T wave; gradient switching adds further noise. A conventional R-wave detector can trigger on the distorted T wave. Vectorcardiographic gating combines leads into a spatial vector and detects the R wave by its direction, which differs from that of the flow artifact, and reduces false triggering (Fischer 1999). Peripheral pulse gating, with an inherent delay after the R wave, serves for retrospectively gated cine when the ECG cannot be used (Kramer 2020).

Prospective triggering starts the acquisition at a set delay after the R wave and acquires for a fixed fraction of the RR interval, leaving a gap at end-diastole while the scanner waits for the next R wave. It is used for LGE, black-blood, perfusion, and mapping, where one phase per heartbeat is imaged. For cine, it omits late diastole, including atrial contraction.

Retrospective gating acquires data continuously, records the timing of individual lines relative to the R wave, and sorts the lines into phases after acquisition, normalizing the cycles to the average RR interval. It covers the entire cardiac cycle, including atrial systole, and is standard for cine and phase-contrast imaging. Because it assumes similar cycle lengths, beat-to-beat variation produces blurring and errors that can be hidden by the reconstruction (Kramer 2020).

Arrhythmia, arrhythmia rejection, and real-time imaging

Atrial fibrillation, frequent ectopy, and bigeminy disrupt segmented acquisitions. Lines from cycles of different length are combined into the same nominal phase, producing blurred endocardial borders, ghosting, and inaccurate end-systolic volumes. Arrhythmia rejection discards beats whose RR interval falls outside a window around the expected value and reacquires them, at the cost of a longer breath-hold; prospective triggering with a shortened acquisition window is an alternative.

Real-time cine acquires a frame within a fraction of a second without combining heartbeats, using accelerated k-space strategies. SCMR prefers a temporal resolution of 60 ms or less for real-time cine and notes that absolute volume quantification is less accurate and precise. It lists real-time imaging for irregular rhythm, inability to hold a breath, and assessment of ventricular interdependence during free breathing (Kramer 2020). Single-shot LGE with a bSSFP readout serves the same role for scar imaging. Saturation-recovery preparation, which resets magnetization to zero before any image is acquired, makes perfusion and some mapping sequences relatively insensitive to irregular rhythm (Ferreira 2013).

For T1 mapping, SCMR recommends diastolic acquisition with a regular rhythm; in atrial fibrillation, systolic acquisition may be more robust but requires its own normal values, and acquisitions are repeated and averaged (Messroghli 2017).

Respiratory motion: breath-holding and navigators

Respiration displaces the heart, mainly in the craniocaudal direction. Most two-dimensional cine, LGE, and mapping images are acquired during a breath-hold, and the slices require breath-holds of consistent depth so that the stack covers the ventricle without gaps or overlaps. End-expiratory breath-holds are more reproducible; inspiratory breath-holds are easier for some patients (Kramer 2020). Inconsistent breath-hold positions shift slices, so a short-axis stack may skip or double-count myocardium.

For free-breathing three-dimensional acquisitions, including whole-heart anatomy and coronary MR angiography, a navigator echo, a one-dimensional column image through the dome of the right hemidiaphragm, records diaphragm position beat by beat (Ferreira 2013). Data are accepted only when the diaphragm lies within an acceptance window at end-expiration, and slice tracking shifts the imaging volume to follow residual motion. The relationship between diaphragm and coronary motion varies among patients, so fixed correction factors are approximate. Image-based navigators, motion-corrected reconstruction, and multiple signal averages are alternatives for patients who cannot hold their breath (Kramer 2020).

Accelerated imaging: segmentation, parallel imaging, and compressed sensing

Parallel imaging uses the spatial sensitivity profiles of the separate receive-coil elements to replace part of the spatial encoding normally performed by phase-encoding steps (Pruessmann 1999; Griswold 2002). The scanner acquires only one line in two or three (an acceleration factor R of 2 or 3), which shrinks the field of view and produces aliasing. SENSE (sensitivity encoding) unfolds the aliased image in the image domain using a separately acquired coil sensitivity map. GRAPPA (generalized autocalibrating partially parallel acquisitions) synthesizes the missing k-space lines from neighboring acquired lines using calibration lines acquired with the data. SNR falls with the square root of the acceleration factor, about 30% for R = 2, and further by a geometry factor that depends on coil arrangement (Biglands 2012). A field of view set too small in the phase direction leaves residual aliasing artifacts in the middle of the image.

Compressed sensing acquires a pseudo-random undersampled subset of k-space and reconstructs the image iteratively, exploiting the sparsity of cardiac images in a transform domain. It permits single-breath-hold or free-breathing cine stacks and is listed with parallel imaging as a way to shorten acquisition in the SCMR cine protocol (Kramer 2020). Other strategies include view sharing and temporal undersampling with k-t methods. Accelerated methods trade SNR or reconstruction fidelity for speed, and highly accelerated images can show smoothing or temporal blurring at sharp edges.

Spin-echo and black-blood imaging

A spin-echo sequence follows the excitation pulse with a 180° refocusing pulse that reverses dephasing from static field inhomogeneities, so the echo reflects T2 and not T2*. Spin-echo sequences tolerate off-resonance and metal better than gradient-echo sequences but are slower and sensitive to motion (Ferreira 2013). Turbo (fast) spin-echo sequences acquire several echoes after a single excitation, filling multiple lines of k-space per TR.

Black-blood imaging suppresses the blood pool so that walls, masses, and the pericardium stand out. The double inversion-recovery preparation applies a nonselective inversion to the whole body and an immediate slice-selective re-inversion that restores magnetization within the imaging slice. Blood in the slice is replaced by inverted blood from outside the slice, and imaging is timed to the null point of blood. A third, fat-selective inversion (triple inversion recovery, or short tau inversion recovery) suppresses fat for T2-weighted edema imaging. Black-blood suppression fails when blood moves slowly or remains in the slice, so stagnant blood along a dilated or akinetic wall or in an aneurysm can appear bright and mimic thrombus, mass, or edema. SCMR lists bright signal in areas of slow flow, signal dropout from motion, and low signal in regions of intramyocardial hemorrhage as limitations of black-blood T2-weighted STIR (Kramer 2020).

Half-Fourier single-shot turbo spin echo (HASTE in one vendor's nomenclature) acquires a black-blood image in a single heartbeat and is used for axial anatomic localizers, extracardiac findings, and imaging near metal, where it shows smaller susceptibility voids than bSSFP (Ferreira 2013).

Gradient-echo sequences and the spoiled gradient echo

Gradient-echo (GRE) sequences form the echo by reversing a readout gradient without a refocusing pulse. They permit very short TR (under 10 ms) and small flip angles, and they are weighted by T2* because field inhomogeneity is not refocused (Ridgway 2010).

Spoiled gradient echo destroys residual transverse magnetization after the readout with RF or gradient spoiling, so the next TR begins with longitudinal magnetization only. With a short TR and TE and a flip angle around 30°, the sequence behaves as a T1-weighted sequence; stationary blood and myocardium become partially saturated, and the blood pool is bright mainly because unsaturated blood flows into the slice (inflow enhancement). Blood-myocardium contrast is therefore weaker than in bSSFP and falls with slow or in-plane flow. Spoiled GRE is used as the readout for many LGE and perfusion sequences, for contrast-enhanced angiography, for phase-contrast flow, and as the cine sequence of choice when bSSFP is degraded by off-resonance at 3 T or near implanted devices (Kramer 2020). Multi-echo spoiled GRE with increasing TE provides T2* measurement.

Balanced steady-state free precession cine

Balanced steady-state free precession (bSSFP) is the default cine sequence at 1.5 T (Kramer 2020). Its gradients are fully balanced within a TR, so transverse magnetization is recycled from one TR to the next instead of being spoiled, and a steady state is reached in which signal is high. Contrast depends on the ratio T2/T1. Blood has a higher T2/T1 ratio than myocardium and appears bright regardless of inflow, which gives sharp endocardial borders even where flow is slow or in-plane (Ridgway 2010). Fat and fluid are also bright. The high signal allows a wide receiver bandwidth and a short TR of about 3 ms.

The price is sensitivity to off-resonance. Signal stays nearly constant over a range of frequency offsets but falls to near zero at regularly spaced offsets, with a spacing equal to 1/TR. At a TR of 2.2 ms, dark bands appear where the off-resonance reaches about ±227 Hz, half of 1/TR, and recur at 455-Hz intervals beyond it; doubling TR halves the spacing (Ferreira 2013). Field inhomogeneity across the heart therefore produces dark bands, most often near the interfaces with lung and liver and over the cardiac veins. Shimming, shortening TR, and adjusting the center frequency move the bands away from the myocardium (Kramer 2020). Flow artifacts in bSSFP cine appear as dark spots and streaks in rapidly moving blood, particularly in the outflow tracts at 3 T.

bSSFP also has a high RF duty cycle, which raises SAR; at 3 T the flip angle is often reduced and TR lengthened, increasing off-resonance sensitivity and reducing contrast (Ferreira 2013).

Standard imaging planes and the cine protocol

The 2020 SCMR protocol builds the ventricular examination from localizers (Kramer 2020). Transaxial, coronal, and sagittal single-shot scouts are followed by a two-chamber long-axis plane prescribed through the apex and the center of the mitral valve, and a four-chamber plane orthogonal to it. The short-axis stack is planned from these long-axis views, perpendicular to the long axis of the left ventricular body, a plane that is not always parallel to the mitral annulus. It extends from the base through the apex. Short-axis slices are 6 to 8 mm thick, with or without 2- to 4-mm gaps for a total of 10 mm per slice position, and temporal resolution is 45 ms or less. The three-chamber view passes through the apex, the center of the mitral valve, and the outflow tract.

For the right ventricle, the basal slice is placed immediately on the myocardial side of the tricuspid annulus, a right ventricular vertical long-axis view is aligned with tricuspid inflow, an outflow tract view crosses the pulmonary valve, and a transaxial cine stack is an alternative for right ventricular volumetry (Kramer 2020). For arrhythmogenic cardiomyopathy, SCMR recommends 5- to 6-mm transaxial or oblique transaxial bSSFP cine covering the right ventricle and outflow tract.

Ventricular volumes, ejection fraction, and mass are measured by summing contoured areas across the short-axis stack (Simpson's method). Basal slice selection, papillary muscle and trabeculation handling, and through-plane motion of the base are the principal sources of variability.

Inversion recovery and the late gadolinium enhancement sequence

Gadolinium-based contrast agents distribute in the extracellular space. They do not cross intact cell membranes, so normal myocardium, with its small extracellular space, holds little agent. Where the extracellular distribution volume is increased by membrane rupture in acute infarction, replacement fibrosis in chronic scar, or interstitial expansion from amyloid or edema, more gadolinium accumulates and washes out more slowly. Ten or more minutes after injection, these regions have a shorter T1 than normal myocardium.

The late gadolinium enhancement (LGE) sequence magnifies this T1 difference with an inversion pulse (Simonetti 2001). A nonselective 180° pulse is applied after the R wave; the inversion time (TI) is chosen so that the readout occurs when longitudinal magnetization of normal myocardium crosses zero. Normal myocardium appears black. Scar, with a shorter T1, has already recovered past zero and appears bright, and blood, which has the highest gadolinium concentration early after injection, is bright. Data are usually acquired on alternate heartbeats to permit recovery between inversions. SCMR modifies this to acquisition on consecutive heartbeats in bradycardia below 60 beats/min and on one heartbeat in three in tachycardia above 100 beats/min or arrhythmia (Kramer 2020).

The SCMR protocol for LGE specifies 2D segmented inversion-recovery GRE or bSSFP, phase-sensitive inversion recovery (PSIR), or 3D sequences when breath-holding and SNR allow. Single-shot bSSFP-readout imaging serves as backup for irregular rhythm or poor breath-holding. Imaging starts at least 10 minutes after contrast (shorter with lower doses, and about 5 minutes after the second injection of a stress-rest perfusion study). Planes and slice thickness match those of cine, in-plane resolution is about 1.4 to 1.8 mm, and acquisition occurs during diastolic standstill (Kramer 2020).

Inversion time selection, phase-sensitive reconstruction, and LGE variants

The TI scout. The null point of normal myocardium depends on its post-contrast T1, which in turn depends on dose, renal clearance, time since injection, and heart rate. A TI scout, a cine acquired after an inversion pulse at the R wave, displays myocardium at a range of inversion times and identifies the time at which it is darkest. SCMR notes that the TI scout does not use the same readout as the segmented LGE sequence, so the correct TI on the LGE images may differ by up to 50 ms (Kramer 2020). As gadolinium washes out, myocardial T1 lengthens and the null point moves later, so the TI is lengthened during the examination.

Errors in TI. With magnitude reconstruction, the displayed signal is the absolute value of longitudinal magnetization. If TI is too short, normal myocardium has not yet reached zero and its negative magnetization appears bright in magnitude images. Scar near its own null point appears dark or isointense. Together these effects can conceal or invert true enhancement. If TI is too long, normal myocardium is gray, contrast between scar and normal tissue falls, and diffuse enhancement can be simulated.

Phase-sensitive inversion recovery (PSIR) retains the sign of the magnetization by referencing the image to a low-flip-angle reference acquisition, so tissues with negative magnetization are displayed darker than zero and tissues with positive magnetization brighter (Kellman 2002). The image intensity order of normal myocardium, scar, and blood is preserved over a wide range of TI values, the window can be set retrospectively, and a precise TI is not required (Kramer 2020).

Four-panel figure. Panels a and b are diagrams of longitudinal magnetization against time after an inversion pulse triggered from the R wave, for normal myocardium and nonviable myocardium, with the segment acquisition window shaded; panel a shows magnitude reconstruction with both curves folded above zero, panel b shows phase-sensitive reconstruction with signed curves. Panels c and d are matching short-axis late gadolinium enhancement images in magnitude and phase-sensitive reconstruction, with an arrow pointing to enhancement in the inferolateral wall.
Inversion recovery for late gadolinium enhancement. After the inversion pulse, nonviable myocardium (blue) recovers faster than normal myocardium (red) because gadolinium has shortened its T1. Magnitude reconstruction (a) displays absolute values, so the order of signal depends on TI; phase-sensitive reconstruction (b) keeps the sign, so normal myocardium remains darker than scar across a wide range of TI. The short-axis images show the same enhancement (arrows) in magnitude (c) and phase-sensitive (d) reconstruction; TD is the trigger delay after the R wave. Credit: Ferreira PF, Gatehouse PD, Mohiaddin RH, Firmin DN. Cardiovascular magnetic resonance artefacts. J Cardiovasc Magn Reson. 2013;15:41, Figure 15, PubMed Central, CC BY 2.0.

Long- and short-TI imaging. With a long inversion time, about 550 ms at 1.5 T and 850 ms at 3 T, myocardium and blood are both bright, and avascular structures such as thrombus and no-reflow zones remain dark, which separates mural thrombus and microvascular obstruction from viable myocardium. A short TI of about 200 ms with PSIR helps separate subendocardial scar from the blood pool (Kramer 2020).

Early gadolinium enhancement. Imaging within 1 to 3 minutes after injection shows microvascular obstruction as a dark core within the infarct while contrast is still entering the tissue; SCMR lists it as an option in acute infarction and myocarditis protocols (Kramer 2020).

Dark-blood LGE combines inversion recovery with blood-pool suppression, so subendocardial scar is not lost against bright blood (Kramer 2020). Three-dimensional LGE acquires the whole ventricle in one breath-hold or with a navigator, with thinner slices, at the cost of a longer acquisition window. Wideband LGE for patients with devices is described in the device imaging section.

The clinical reading of enhancement patterns (subendocardial and transmural ischemic scar, midwall, subepicardial, and diffuse patterns) and the relation of transmural extent to functional recovery are covered in the viability hub under transmural extent of scar.

First-pass perfusion imaging

First-pass perfusion follows a gadolinium bolus through the left ventricular cavity and into the myocardium. The images are prepared with a saturation pulse, which sets magnetization to zero regardless of the preceding heartbeat, followed by a fixed delay and a rapid single-shot readout (bSSFP, spoiled GRE, or a hybrid GRE-echo-planar readout). SCMR specifies at least three short-axis slices per heartbeat for ischemia evaluation, 8- to 10-mm slices, in-plane resolution under about 3 mm, and a readout duration of about 100 to 125 ms or shorter. The contrast dose is 0.05 to 0.1 mmol/kg at 3 to 7 mL/s followed by at least 30 mL of saline. The breath-hold starts before contrast reaches the left ventricle, and typically 50 to 60 heartbeats are acquired (Kramer 2020).

Because the readout of a slice occupies 100 ms or more, slices are acquired at different points in the cardiac cycle, some in systole. The relationship between signal and gadolinium concentration is nonlinear at high concentration, which matters for quantitative perfusion; dual-bolus and dual-sequence approaches, with proton-density images for baseline correction, address it (Kramer 2020). Quantitative pixel-wise perfusion mapping estimates myocardial blood flow in mL/g/min from the arterial input and tissue curves.

Vasodilator and dobutamine protocols, doses, and contraindications for stress CMR are covered in the stress testing hub under pairing the stressor with cardiac magnetic resonance. The dark-rim artifact that mimics subendocardial ischemia is described with the Gibbs artifact below.

T1 mapping and extracellular volume fraction

Parametric mapping produces an image in which the pixels hold measured relaxation times instead of relative signal intensities (Messroghli 2017). Weighted images show focal differences between regions; maps also detect diffuse change that affects the whole myocardium, which LGE cannot show because LGE relies on a normal reference region to null.

Acquisition. Most T1 mapping sequences acquire several single-shot bSSFP images at different times after a preparation pulse and fit a recovery curve pixel by pixel. Inversion-based methods include the modified Look-Locker inversion recovery (MOLLI) sequence and its shortened variant (ShMOLLI); saturation-based methods include saturation recovery single-shot acquisition (SASHA); and hybrid methods combine the two (Messroghli 2017). MOLLI, introduced as a breath-hold sequence that merges images from several inversions into one data set (Messroghli 2004), is the most widely used. It is precise, but its readout perturbs recovery, so it requires a Look-Locker correction, and its sensitivity to magnetization transfer, T2, flip angle, inversion efficiency, heart rate, and off-resonance reduces its accuracy. The 5(3)3 scheme and ShMOLLI shorten the breath-hold with a modest noise penalty. Saturation-recovery methods largely remove sensitivity to magnetization transfer, T2, heart rate, and off-resonance at the price of reduced precision (Messroghli 2017).

Protocol. SCMR specifies native T1 mapping with MOLLI, ShMOLLI, or an equivalent Look-Locker method; diastolic acquisition with a regular rhythm (systolic acquisition may be preferred in atrial fibrillation); 6- to 8-mm slices with in-plane resolution of about 1.6 to 2.0 mm; and at least one short-axis map in all studies (Kramer 2020). Source images are inspected for motion, and the map is repeated when misregistration is present. For global and diffuse disease, a single region of interest is drawn in the septum of a mid-ventricular short-axis map, away from lung, liver, and veins that cause susceptibility artifact. Regions under 20 pixels are avoided, and the region is drawn without reference to LGE images but may exclude infarcts (Messroghli 2017).

Extracellular volume fraction (ECV) expresses the extracellular space as a percentage of myocardial volume. It is calculated from native and post-contrast T1 of myocardium and blood and the hematocrit:

ECV = (100 − hematocrit) × (1/T1 myocardium post − 1/T1 myocardium native) ÷ (1/T1 blood post − 1/T1 blood native)

The ratio of changes in relaxation rate (1/T1) between myocardium and blood estimates the partition coefficient of gadolinium, and the plasma fraction of blood converts it to an extracellular fraction. SCMR recommends a post-contrast T1 map at least once between 10 and 30 minutes after a bolus of 0.1 to 0.2 mmol/kg of an extracellular, non-protein-bound agent, with hematocrit measured, ideally within 24 hours of imaging (Kramer 2020; Messroghli 2017). Synthetic ECV estimates the hematocrit from native blood T1 using a sequence-specific relationship. ECV depends less on field strength and sequence than native T1, so published ranges from the same system and sequence may be used when a local range is not available; ECV is expressed as a percentage (Messroghli 2017). Diffuse fibrosis rarely raises ECV above 40%, and values above 55% occur within transmural LGE (Messroghli 2017).

Reference ranges. Native T1 and T2 depend on field strength, vendor, sequence, and post-processing. The consensus requires a local reference range, defined as the mean ± 2 standard deviations in healthy subjects scanned and analyzed with the same method, and states that quantitative native T1 and T2 should not be reported clinically when a local range is not available (Messroghli 2017). Sex and age shift normal values by about 0.5 standard deviation, which matters for subtle diffuse disease but little for amyloidosis or iron overload. Reports state the sequence, field strength, reference range, contrast agent, and dose, and give the measured value with its Z-score where available.

The patterns of change by disease, including the high native T1 and ECV of cardiac amyloidosis and the low native T1 of Anderson-Fabry disease, are covered in the cardiac amyloidosis CMR section and the hubs on individual cardiomyopathies.

T2 mapping, T2-weighted imaging, and T2* iron quantification

T2-weighted imaging and T2 mapping. Myocardial edema raises T2. Black-blood T2-weighted STIR was the original edema sequence; it is limited by surface-coil intensity variation, bright signal from slow-flowing blood along the endocardium, and signal loss from motion. Bright-blood T2-prepared single-shot bSSFP reduces some of these problems. T2 mapping, with T2-prepared bSSFP at several preparation durations, gradient- and spin-echo, or fast spin-echo methods, quantifies T2 with 6- to 8-mm slices and in-plane resolution of about 1.6 to 2.0 mm (Kramer 2020). The updated Lake Louise criteria for myocardial inflammation require at least one T2-based marker of edema together with at least one T1-based marker of injury, such as elevated native T1, elevated ECV, or nonischemic LGE (Ferreira VM 2018).

T2* mapping. Iron is paramagnetic and shortens T2. SCMR specifies a single-breath-hold, ECG-triggered multi-echo gradient-echo acquisition of a single mid-ventricular short-axis slice before contrast. The protocol uses 6 to 9 echoes from about 2 ms to about 18 ms spaced about 2 ms apart (or about 1 to 12 ms at 1-ms spacing in severe overload), an 8- to 10-mm slice thickness, and a trigger delay of 0 ms after the R wave (Kramer 2020). An ungated axial acquisition through the liver measures hepatic iron (Kramer 2020). T2 is measured in a full-thickness region of interest in the septum, away from the susceptibility artifact of the cardiac veins and lungs, because iron is deposited preferentially in the subepicardium (Messroghli 2017). At 1.5 T with a gradient-echo sequence of 8 or more echoes, myocardial iron is graded in three tiers: T2 above 20 ms, low risk; 10 to 20 ms, intermediate; and below 10 ms, high risk (Messroghli 2017). These thresholds derive from thalassemia cohorts in which all patients with ventricular dysfunction had a myocardial T2 below 20 ms (Anderson 2001), and they do not apply at 3 T. T2* of infarcted myocardium also falls with intramyocardial hemorrhage after reperfusion.

CMR physics calculator

The calculator applies four relations: the Larmor equation, the fat-water frequency difference and the first echo time at which fat and water signals are opposed, the null point of a tissue after an inversion pulse, and the extracellular volume fraction from native and post-contrast T1.

Larmor frequency—
Fat-water frequency difference—
First opposed-phase echo time—
Null-point inversion time—
Extracellular volume fraction—
—

Larmor frequency = 42.58 MHz/T × B0. Fat-water difference taken as 140 Hz per tesla (about 210 Hz at 1.5 T and 420 Hz at 3 T); first opposed-phase TE = 1 ÷ (2 × difference). Null point for an ideal inversion repeated at intervals of TR with negligible readout: TI = T1 × [ln 2 − ln(1 + e^(−TR/T1))], which approaches 0.69 × T1 when TR is long. ECV = (100 − hematocrit) × (1/T1 myocardium post − 1/T1 myocardium native) ÷ (1/T1 blood post − 1/T1 blood native). Default T1 values are illustrative and are not reference values; mapping results are interpreted against the local reference range for the sequence and field strength.

At the default inputs, the Larmor frequency at 1.5 T is 63.9 MHz, fat and water are first opposed at a TE of about 2.4 ms, a tissue with T1 of 450 ms inverted at 2,000-ms intervals nulls at about 307 ms, and the ECV is 26.2%. At 3 T, the opposed-phase TE falls to about 1.2 ms, matching the values reported for fat-water cancellation (Ferreira 2013).

Phase-contrast velocity mapping, VENC, and 4D flow

Moving protons in a magnetic field gradient accumulate a phase shift proportional to their velocity along the gradient. Phase-contrast imaging applies a bipolar gradient pair with two different flow sensitivities and subtracts the resulting phase images, which removes background phase and leaves a phase difference proportional to velocity (Biglands 2012). The velocity-encoding value (VENC) is the velocity that produces a 180° phase shift. Velocities above the VENC wrap around and appear as flow in the opposite direction (aliasing); a VENC far above the true peak velocity wastes dynamic range and makes the measurement noisier. Integrating velocity over the lumen area frame by frame and over the cardiac cycle gives forward and reverse flow volumes, regurgitant fraction, cardiac output, and the pulmonary-to-systemic flow ratio (Qp/Qs).

SCMR specifies a through-plane, retrospectively gated cine phase-contrast acquisition with the plane perpendicular to flow in two directions and centered in the vessel at the scanner isocenter. Slice thickness is 5 to 8 mm, in-plane resolution is at least one-tenth of the vessel diameter, and temporal resolution is 50 ms or less between acquired frames. The VENC is the lowest that avoids aliasing: the phase images are checked after the scan, and the VENC is raised if aliasing is present. The TE is the shortest available, particularly across stenoses (Kramer 2020). In the SCMR example, a VENC of 250 cm/s aliased at the sinotubular junction in aortic stenosis and a VENC of 350 cm/s did not.

Sources of error. Background phase offsets from eddy currents and concomitant (Maxwell) gradients add a small apparent velocity to all pixels. Over a large vessel and many frames, a small offset accumulates into a large volume error, and correction is applied in software or against a stationary phantom (Ferreira 2013). Through-plane motion of the aortic and pulmonary roots, partial volume at the vessel edge, intravoxel dephasing in turbulent jets, misalignment of the plane with flow, and arrhythmia also bias results. Phase-contrast flow is checked for internal consistency: aortic and pulmonary forward volumes agree with left and right ventricular stroke volumes from cine when there is no regurgitation or shunt.

4D flow acquires velocity in three directions over a three-dimensional volume throughout the cardiac cycle, so flow through any plane can be measured after the scan, along with jet direction, wall shear stress, and flow patterns. The 2023 consensus describes sequence options, acquisition parameters, post-processing, and minimum validation standards for clinical use (Bissell 2023).

Angiography, whole-heart imaging, and myocardial tagging

MR angiography. Contrast-enhanced MR angiography uses a 3D spoiled gradient-echo acquisition with very short TR and TE, timed to the arterial first pass of a gadolinium bolus, so that T1-shortened blood is bright regardless of flow. SCMR lists 0.1 to 0.2 mmol/kg at 2 to 3 mL/s for the thoracic aorta and 0.05 mmol/kg for time-resolved angiography (Kramer 2020). Timing uses a test bolus, bolus tracking, or time-resolved acquisition. Noncontrast techniques include ECG- and navigator-gated 3D bSSFP whole-heart imaging with T2 preparation and fat saturation, which depicts coronary origins and proximal courses, cardiac chambers, and great vessels for congenital heart disease and coronary anomalies (Biglands 2012).

Tagging and feature tracking. Tagging applies spatially modulated saturation (stripes or a grid) at end-diastole; the tags move with the myocardium and fade with T1 recovery over the cardiac cycle, so deformation can be measured from tissue displacement. Feature tracking derives strain from the motion of borders and features on routine cine images, with no additional acquisition, and depends on image quality and software. Strain values, reproducibility, and their comparison with speckle-tracking echocardiography are covered under cardiac magnetic resonance strain.

Matching the sequence to the clinical question

SCMR builds examinations from modules matched to the clinical question (Kramer 2020).

SequenceSource of contrastClinical questionSCMR parameters
bSSFP cineT2/T1 ratio; bright blood without inflow dependenceVolumes, ejection fraction, mass, regional function, valve motion6–8 mm slices, gaps to 10 mm; ≤45 ms temporal resolution
Spoiled GRE cineT1 weighting with inflow enhancementFunction at 3 T or near devices when bSSFP is degradedSame planes as bSSFP
Real-time cinebSSFP or GRE, single heartbeat per frameArrhythmia, poor breath-holding, ventricular interdependence≤60 ms preferred
Black-blood fast spin echoDouble inversion nulls blood; T1 or T2 weightingAnatomy, masses, pericardium, fat (with fat suppression)Single-shot or segmented
T2-weighted STIR or T2-prepared bSSFPFree waterEdema in acute injury and inflammationArtifacts from slow flow and motion
Native T1 mapComposite tissue T1Diffuse fibrosis, edema, amyloid, Anderson-Fabry disease, ironMOLLI or equivalent; local reference range
T2 mapQuantitative T2Edema and active inflammationT2-prepared bSSFP, GraSE, or FSE
T2* mapField inhomogeneity from ironMyocardial and hepatic iron, hemorrhage6–9 echoes, ~2 to ~18 ms, 1.5 T
First-pass perfusionSaturation-recovery T1 weighting during the contrast bolusInducible ischemia, microvascular obstruction≥3 slices per beat; 0.05–0.1 mmol/kg at 3–7 mL/s
Early gadolinium enhancementT1 shortening 1–3 minutes after contrastMicrovascular obstruction, thrombusOptional in acute infarction
Late gadolinium enhancementInversion recovery nulling normal myocardiumInfarct scar, replacement fibrosis, infiltration, viability≥10 minutes; 0.1–0.2 mmol/kg; 1.4–1.8 mm in-plane
Post-contrast T1 map and ECVGadolinium partition with hematocritInterstitial expansion, amyloid burden10–30 minutes after contrast; hematocrit within 24 hours
Phase contrastVelocity-encoded phase shiftFlow, regurgitation, shunts, gradients5–8 mm; ≤50 ms; lowest VENC without aliasing
Contrast-enhanced MR angiographyT1 shortening of arterial bloodAorta, pulmonary and systemic vessels0.1–0.2 mmol/kg at 2–3 mL/s

Sequences and parameters from the 2020 SCMR standardized protocols and the 2017 SCMR/EACVI mapping consensus. FSE, fast spin echo; GraSE, gradient and spin echo; STIR, short tau inversion recovery.

Classification of CMR artifacts

An artifact is signal or signal loss in the image that does not correspond to tissue. Most CMR artifacts arise from five sources: motion (cardiac, respiratory, and flow), the finite sampling of k-space (Gibbs ringing and aliasing), chemical shift, B0 inhomogeneity from susceptibility and metal, and B1 inhomogeneity (Ferreira 2013). A true abnormality stays anatomically coherent across orthogonal planes, repeated acquisitions, and complementary sequences; an artifact changes with the phase-encoding direction, the sequence, or the timing.

ArtifactMechanismAppearanceRemedies
Respiratory ghosting and misregistrationMotion during acquisition of k-space, especially its centerRepeated copies of moving structures along the phase direction; slice shift between breath-holdsConsistent breath-holds, navigator gating, single-shot or real-time imaging
Arrhythmia blurringLines from cycles of different lengths combinedBlurred borders, ghosting, wrong end-systolic frameArrhythmia rejection, prospective triggering, real-time cine, single-shot LGE
Flow-related signal loss and ghostingIntravoxel dephasing in jets; pulsatile flowDark jets; ghosts of vessels along the phase directionShort TE, bSSFP, saturation bands, swapped phase direction
bSSFP dark bandsOff-resonance at multiples of 1/TRCurved dark bands crossing myocardium or bloodShimming, shorter TR, frequency adjustment, spoiled GRE
Susceptibility artifactLocal field distortion by metal, air, or ironSignal void with bright rim, distortion, bandingShort TE, spin echo, GRE in place of bSSFP, wideband inversion, wider bandwidth
Chemical shift misregistrationFat-water frequency difference read as positionBright and dark bands at fat-water interfaces along the frequency directionWider receiver bandwidth, fat suppression
Fat-water cancellation (India ink)Fat and water opposed in phase within one voxel at the echoDark line at fat-water boundaries, around the right ventricle and pericardiumChange TE; water-fat separated imaging
Aliasing (wraparound)Anatomy outside the field of view in the phase directionArms or back folded into the imageLarger field of view, phase oversampling, swapped phase direction
Parallel-imaging residual aliasingField of view too small for the unfoldingSignal fold-over in the center of the imageLarger reference field of view, lower acceleration
Gibbs ringing and dark rimTruncated sampling of high spatial frequencies; motion during readoutParallel lines at sharp edges; transient subendocardial dark rim in perfusionHigher spatial resolution, filtering, faster readout
Incorrect inversion timeNormal myocardium not at its null pointGray or bright myocardium; concealed or inverted enhancementTI scout, repeated TI adjustment, PSIR
B1 inhomogeneityWavelength effects in the body at 3 T; coil sensitivityRegional shading, uneven fat suppressionMulti-transmit RF, adiabatic pulses, intensity correction

Common CMR artifacts, after Ferreira (2013) and the 2020 SCMR protocols. GRE, gradient echo; LGE, late gadolinium enhancement; PSIR, phase-sensitive inversion recovery.

Motion, ghosting, and gating artifacts

The heart moves with contraction, rotation, and respiration, and blood moves faster still. When motion occurs during acquisition of peripheral k-space lines, edges blur in the phase-encoding direction; when it occurs during acquisition of central lines, the entire image degrades and discrete ghosts appear, spaced along the phase direction according to how the motion is synchronized with the acquisition (Ferreira 2013). Ghosts can project over the myocardium as a faint copy of the anterior chest wall or the pulsating aorta.

False triggering on a tall T wave makes the cine start in mid-systole and shortens the cycle. In a patient with ectopy, cine images can show an end-systolic frame that does not represent the true minimum volume. In LGE, a trigger delay that places the readout in systole blurs thin walls and can create apparent subendocardial enhancement from partial volume with the blood pool.

Because breath-holds are separate, slices of a short-axis stack can be offset from one another. Respiratory ghosting during free breathing is reduced by navigator gating or averaging.

Flow artifacts

In gradient-echo sequences, turbulent or accelerating jets contain a range of velocities within a voxel; the resulting phase dispersion cancels signal and produces a dark jet. The size of the jet depends on TE, voxel size, and the sequence, not only on the severity of the lesion, so signal void length in cine imaging is a qualitative sign of stenotic or regurgitant flow. bSSFP is less sensitive to flow-related signal loss than spoiled GRE (Ferreira 2013). Through-plane inflow brightens blood in spoiled GRE; in-plane slow flow saturates and darkens it.

Pulsatile flow in the aorta and ventricles produces ghosts along the phase direction because blood signal varies from one acquisition to the next.

Off-resonance artifacts: bSSFP banding and susceptibility from metal

Tissue magnetic susceptibility differs: most tissue is weakly diamagnetic, air differs sharply from tissue, and iron and metal are paramagnetic or ferromagnetic. Interfaces between them distort the local field, and the resulting off-resonance causes signal loss on gradient-echo images, banding on bSSFP images, and spatial distortion when the frequency shift is read as a change in position (Ferreira 2013).

bSSFP banding. Dark bands appear where the off-resonance reaches the null points of the bSSFP frequency response, spaced at 1/TR. At 3 T, bands cross the myocardium more often because frequency offsets double. Shimming over the heart, a shorter TR, and a frequency adjustment shift the bands out of the region of interest; when they persist, spoiled GRE cine replaces bSSFP (Kramer 2020).

Metal. Sternal wires, mechanical and biological valve prostheses, annuloplasty rings, septal occluders, stents, embolization coils, insertable cardiac monitors, and generators distort the field over a region much larger than the object. The void is larger on gradient-echo images with long TE than on spin-echo images, and bSSFP shows banding in the surrounding region; a half-Fourier single-shot turbo spin-echo image of the same plane shows the anatomy with less loss (Ferreira 2013). Susceptibility artifact around a mechanical valve prevents assessment of the occluder and adjacent jets, and flow measurement planes are placed away from the prosthesis. Within a metallic aortic stent, the lumen can appear dark on bright-blood images because the stent wall blocks RF penetration, and black-blood spin-echo imaging shows the wall outside the stent (Ferreira 2013).

Four cardiac MR images with arrows. Top row: two four-chamber views in the same patient with a septal occluder, the left image acquired with a gradient-echo sequence and the right with balanced steady-state free precession, each showing a dark signal void at the interatrial septum. Bottom row: two balanced steady-state free precession images with localized dark signal loss at a bileaflet mechanical aortic valve prosthesis (left) and a bileaflet mechanical mitral valve prosthesis (right).
Susceptibility artifact from cardiac implants. A septal occluder produces a signal void at the interatrial septum on gradient-echo (a) and balanced steady-state free precession (b) four-chamber views; the void is larger on the gradient-echo image in this patient. Bileaflet mechanical aortic (c) and mitral (d) prostheses produce localized signal loss on balanced steady-state free precession images that obscures the occluders and adjacent flow. Credit: Ferreira PF, Gatehouse PD, Mohiaddin RH, Firmin DN. Cardiovascular magnetic resonance artefacts. J Cardiovasc Magn Reson. 2013;15:41, Figure 35, PubMed Central, CC BY 2.0.

Mapping and T2*. Field inhomogeneity at the lateral wall near lung and over the great cardiac vein lowers apparent T2 and alters T1 and T2 maps, which is why both T1 and T2 regions of interest are placed in the septum (Messroghli 2017).

Chemical shift artifacts

Fat resonates about 210 Hz below water at 1.5 T and 420 Hz at 3 T (Ferreira 2013). Two artifacts follow.

Misregistration (chemical shift of the first kind). Frequency encoding assigns position from frequency, so fat is displaced relative to water along the frequency-encoding direction by the frequency difference divided by the receiver bandwidth per pixel. With a narrow bandwidth, epicardial and pericardial fat shift by one or more pixels, producing a bright band on one side of an interface and a dark band on the other, which can mimic or conceal pericardial thickening. In echo-planar perfusion readouts, the displacement occurs in the phase-blip direction and is larger; in one example, chest-wall fat was displaced about 4 pixels when fat saturation was off (Ferreira 2013).

Phase cancellation (chemical shift of the second kind, India ink artifact). In gradient-echo sequences, fat and water in the same voxel precess in and out of phase. At the echo times when they are opposed, about 2.4 ms at 1.5 T and 1.2 ms at 3 T, their signals subtract and dark lines outline fat-water boundaries (Ferreira 2013). bSSFP sequences, with TE close to these values, commonly outline the right ventricular free wall and pericardium in black. The same effect is used deliberately in in-phase and opposed-phase imaging and water-fat separated (Dixon) methods to identify fat in masses and in arrhythmogenic cardiomyopathy.

Aliasing and wraparound

When the body extends beyond the field of view in the phase-encoding direction, the excluded anatomy is undersampled and folds into the opposite side of the image (Ferreira 2013). In short-axis cardiac planes, the arms and posterior chest wall wrap over the heart. Frequency-direction aliasing is removed by oversampling at no cost in time; phase-direction aliasing is removed by enlarging the field of view, phase oversampling (at the cost of time), or placing the phase direction along the shorter body dimension. With parallel imaging, a field of view too small for the unfolding produces residual artifact in the middle of the image instead of at its edge.

Phase-contrast aliasing is a separate effect in which velocities exceeding the VENC wrap to the opposite sign; it is corrected by raising the VENC or by phase unwrapping.

Gibbs ringing and the dark-rim artifact

All MR images are reconstructed from a finite extent of k-space. Truncating the high spatial frequencies produces ringing, parallel lines of alternating brightness at sharp, high-contrast edges, called Gibbs or truncation artifact (Ferreira 2013). It is most visible at the blood-myocardium interface and in images with coarse resolution in the phase direction.

In first-pass perfusion, the contrast between the gadolinium-filled cavity and myocardium is highest at peak blood enhancement. Gibbs ringing, cardiac motion during the 100-ms readout of a slice, and susceptibility effects of concentrated gadolinium combine to produce a thin dark subendocardial rim (Ferreira 2013). The dark-rim artifact is thin, appears at the time of peak cavity signal before contrast reaches the myocardium, follows the endocardial contour across territories without respecting a coronary distribution, and fades within a few frames as cavity signal falls. True subendocardial hypoperfusion persists after the peak of myocardial enhancement, conforms to a coronary territory, and is wider. Higher spatial resolution, shorter readouts, and filtering reduce the artifact. Motion ringing from a sequential phase order is typically darker than pure Gibbs ringing (Ferreira 2013).

Late enhancement and mapping artifacts

Inversion time in diffuse disease. Diffusely abnormal myocardium, as in cardiac amyloidosis, may have a null point near or before that of blood, so a TI scout in which myocardium nulls before the blood pool indicates diffuse extracellular expansion; this finding is described in the cardiac amyloidosis CMR section. PSIR reduces dependence on TI.

Partial volume and blood pool. Bright blood adjacent to subendocardial scar obscures its border. Dark-blood LGE, PSIR with a short TI, and long-axis views separate the scar border from the blood pool.

Ghosting and phase-direction artifact. Bright ghosts from the blood pool or chest wall can project over myocardium and simulate enhancement; repeating the slice with the phase-encoding direction swapped distinguishes artifact, which moves, from scar, which does not (Ferreira 2013).

Device-related hyperintensity. Near an implantable cardioverter-defibrillator (ICD) generator, the inversion pulse fails to invert myocardium whose frequency is shifted outside its bandwidth, and that myocardium appears falsely bright (Rashid 2014). The remedy is the wideband inversion pulse described below.

Mapping. Motion between the source images of a map produces errors at borders; heart rate, off-resonance, and partial volume with blood or fat bias T1 and T2. Source images and quality maps are reviewed before values are reported (Messroghli 2017).

The three fields of the MR environment and their hazards

The American College of Radiology (ACR) Manual on MR Safety organizes MR safety around three physical fields with their own hazards and controls (ACR 2026).

The static field (B0) is always present, whether or not the scanner is imaging. It exerts translational force and torque on ferromagnetic objects, drawing them toward the bore as projectiles, and it can activate magnetically sensitive switches in implanted devices.

The radiofrequency field (B1) is present only during imaging. It deposits energy as heat in tissue and induces currents and voltages in conductive implants, leads, monitoring cables, and loops formed by the body itself, causing local burns and, in leads, heating at the electrode tip.

The gradient fields (dB/dt) are switched rapidly during imaging. They induce electric fields that can stimulate peripheral nerves, induce voltages in implanted leads, and generate acoustic noise.

Manufacturers and regulators limit RF and gradient output through operating modes defined by the International Electrotechnical Commission (IEC): Normal Operating Mode, in which output presents negligible risk and which is used for most clinical imaging; First Level Controlled Operating Mode, in which higher output is permitted under medical supervision; and Second Level Controlled Operating Mode, reserved for research protocols (ACR 2026).

Static field hazards: projectiles, force, torque, and MR safety zones

Force and torque. A ferromagnetic object is pulled toward the magnet with a force that depends on the spatial gradient of the field and, until the object is magnetically saturated, on the field strength itself; it also twists to align with the field. The force rises steeply over a short distance near the bore entrance, so an object that feels light at the door can become a projectile a step or two closer to the bore. Manufacturers publish maps of B0 and of the spatial field gradient around their scanners, which are used to judge forces on implants (ACR 2026). Nonferromagnetic conductors moving through the field experience opposing (Lenz) forces that can impede a cart or oxygen tank near the bore and are larger at 7 T. ACR notes this consideration for metallic aortic and mitral valve replacements in 7-T scanners (ACR 2026).

Projectile events. ACR records at least three deaths from oxygen cylinders that became projectiles (2001, 2018, and 2021) and one death in 2023 when a firearm discharged in the MR environment. All projectile events are preventable, and burns are the most frequently reported MRI injury (ACR 2026).

Zones. ACR divides the MR facility into four zones with progressively restricted access.

ZoneDefinitionAccess
Zone IAreas freely accessible to the public, outside the MR facilityUnrestricted
Zone IIInterface between public areas and the controlled zones: waiting, changing, preparation, and screening, including ferromagnetic detectionPatients and companions admitted by MR personnel
Zone IIIMR controlled access area with direct doorway access to Zone IV, including the operator console, and any space where the field exceeds 9 gaussScreened individuals only, under the supervision of Level 2 MR personnel
Zone IVThe scanner roomScreened patients and personnel; always within Zone III

MR safety zones from the ACR Manual on MR Safety (2026 edition). The cryogen venting area outside the building is also access controlled and labeled.

The 9-gauss line. The fringe field outside the bore falls with distance. The 5-gauss (0.5-mT) contour was historically the "pacemaker line." Device standards require cardiac implantable electronic devices to tolerate static fields up to 10 gauss (1.0 mT). The 2022 IEC standard, recognized by the US Food and Drug Administration (FDA) and endorsed by ACR, defines the static-field hazard area as the space within the 9-gauss (0.9-mT) line (ACR 2026). People with active implants, including pacemakers and ICDs, do not cross the 9-gauss line unless cleared in writing by the facility's MR medical director or a Level 2 MR physician.

Field strength limits. The FDA classifies a static field above 8 T as significant risk for adults, children, and infants older than 1 month, and above 4 T for neonates (FDA 2014).

Radiofrequency heating, specific absorption rate, and burns

Specific absorption rate (SAR) is the scanner's estimate of RF power absorbed per kilogram of tissue (W/kg). In Normal Operating Mode, whole-body SAR is limited to 2 W/kg or less. In First Level Controlled Operating Mode it is limited to 4 W/kg or less. This mode requires medical supervision and exclusion of patients who cannot tolerate the heat load (ACR 2026). The FDA considers whole-body SAR above 4 W/kg averaged over 15 minutes, or head SAR above 3.2 W/kg averaged over 10 minutes, significant risk (FDA 2014). SAR rises with the square of B0 and of the flip angle and with the duty cycle of RF pulses, so bSSFP at 3 T is the cardiac sequence most often limited by SAR.

B1+rms. SAR is a vendor-specific model estimate. The root-mean-square transmitted RF field, B1+rms (in microtesla), is a direct measure of RF exposure that does not depend on a patient model; whole-body SAR scales approximately with its square, and implant labeling increasingly specifies B1+rms limits (ACR 2026).

Burns. RF induces currents in conductive material within the transmitted field. Burns occur at points of high resistance in a conductive loop: skin-to-skin contact between hands and thighs, between the thighs, or between arms and trunk; looped ECG cables or monitoring leads; and dark or loop-shaped tattoos. Insulating pads, separation of the patient from the bore, MR-conditional ECG electrodes placed per labeling, and straight, unlooped cables prevent them (ACR 2026). Elongated conductors such as leads concentrate heating at their tips. A lead of resonant length, or an abandoned lead end without a generator, can heat more than a lead connected to a generator (Vigen 2021).

Gradient fields: peripheral nerve stimulation and acoustic noise

Rapidly switched gradients induce electric fields in tissue. Above a threshold that depends on slew rate and patient anatomy, they stimulate peripheral nerves, causing tingling or twitching at the torso or limbs. Scanners limit dB/dt to avoid painful stimulation, and FDA criteria classify any dB/dt sufficient to produce severe discomfort or painful nerve stimulation as significant risk (FDA 2014).

Lorentz forces on the gradient coils produce the knocking noise of the scanner. ACR and the IEC standard treat an A-weighted root-mean-square sound pressure above 99 dB(A) with hearing protection in place as a significant risk and require hearing protection that reduces exposure below that level (ACR 2026; FDA 2014). Earplugs or headphones, properly fitted, are provided to all patients.

Screening, MR labeling, emergencies, and pregnancy

Screening. All patients and all other persons entering Zone III complete a written screening form and a verbal interview reviewed by Level 2 MR personnel, and ferromagnetic detection systems supplement direct inspection (ACR 2026). Patients who cannot give a history are screened from the medical record, prior imaging, a family member's history, and a physical inspection for scars and devices; radiographs are obtained when an implant cannot otherwise be excluded. For patients with a history of device revision or removal, radiographic screening for abandoned leads is considered before MRI (ACR 2026).

Labeling. The ASTM International terms used in device labeling are: MR Safe, an item that is nonmetallic, nonconducting, and nonmagnetic and poses no known hazard in any MR environment; MR Conditional, an item with no known hazards in a specified MR environment under specified conditions of use; and MR Unsafe, an item that poses unacceptable risks in the MR environment (ACR 2026). Conditions on an MR-conditional label typically include static field strength, maximum spatial field gradient, maximum dB/dt, maximum SAR or B1+rms, permitted transmit coils, scan duration, and, for active devices, the configuration of components and required programming (Vigen 2021). A device scanned outside its labeled conditions, or a system combining MR-conditional and non-MR-conditional components, is not MR Conditional for that examination. A label from a regulatory authority outside the country of scanning does not confer MR-conditional status locally (Vigen 2021).

Emergencies. A patient who arrests in the scanner is removed immediately from Zone IV to a predetermined, magnetically safe location on an MR-conditional gurney while basic life support begins, because resuscitation teams and equipment that are not MR Safe or MR Conditional cannot enter Zone IV (ACR 2026). A quench, the deliberate or spontaneous loss of superconductivity, vents cryogen gas rapidly through the quench pipe, and the external vent area is access controlled because of frostbite and asphyxiation hazards. Emergency magnet-off (a deliberate quench), fire, and entrapment procedures are part of the facility's written emergency policy (ACR 2026).

Pregnancy. Most data show no harmful effect of MRI on the developing fetus at 3 T or below; a noncontrast study in Normal Operating Mode (whole-body SAR 2 W/kg or less) is performed when the medical benefit justifies it, with additional consideration of thermal effects in the first trimester (ACR 2026). In a population-based Ontario cohort of more than 1.4 million births, first-trimester MRI was not associated with a significantly increased risk of stillbirth or neonatal death, congenital anomalies, neoplasm, or vision or hearing loss (Ray 2016). Pregnant staff may work in Zones III and IV but do not remain in Zone IV during scanning (ACR 2026). Gadolinium in pregnancy is addressed below.

Passive implants and other devices in the MR environment

Most cardiovascular implants are passive metallic devices that present little hazard at 1.5 and 3 T. Their main effects are susceptibility artifact and, for some, heating.

ImplantMR safety positionImaging effect
Coronary stentsNo known coronary stent is ferromagnetic. Scanning at 3 T or less is acceptable at any time after implantation, including immediately, with whole-body SAR of 2 W/kg or less when the stent is in the transmit field (4 W/kg when outside it) and up to 15 minutes per sequence. No adverse event attributed to MRI has been reportedLocal signal void that obscures the stented lumen
Prosthetic valves and annuloplasty ringsScanned under labeled conditions at 1.5 and 3 T; Lenz forces on moving metallic leaflets are a consideration at 7 TSignal loss around the sewing ring and occluders; mechanical occluder motion not seen
Septal occluders, left atrial appendage occluders, embolization coils, inferior vena cava filtersScanned per device labelingSignal void, larger on gradient-echo images
Retained temporary epicardial pacing wire remnantsShort remnants have not been shown to be a barrier to scanning, with no reported adverse events; partially removed temporary wires are not considered abandoned pacing leadsUsually minimal
Insertable cardiac monitorsCurrent monitors are labeled MR Conditional at 1.5 and 3 T, several with no waiting period after insertion; the older Reveal XT required a 6-week wait, so models are scanned under their own labelingbSSFP signal loss and banding near the device; less on single-shot turbo spin echo
Leadless pacemakersLabeled MR Conditional at 1.5 and 3 T with programming to an MRI modeArtifact at the right ventricular apex or septum
Pulmonary artery pressure sensor (CardioMEMS)MR Conditional at 1.5 and 3 TSmall local artifact
Ventricular assist devices, neurostimulators, drug pumps, other active implantsScanned only within their own labeling; a device without MR-conditional labeling is treated as MR UnsafeVariable

Sources: coronary stent guidelines of the International Society for Magnetic Resonance in Medicine (ISMRM) (Shellock 2025); ACR Manual on MR Safety; Vigen (2021); and the library pages on leadless pacemakers and CardioMEMS. Implants are identified by make and model and scanned under their current labeling.

The MRI labeling of leadless and conduction system pacing systems is described under MRI, electromagnetic interference, and surgery, and that of CardioMEMS, cardiac contractility modulation, and baroreflex activation therapy under device interactions and MRI.

How MRI interacts with pacemakers and defibrillators

The three fields act on a cardiac implantable electronic device (CIED) through distinct mechanisms (Vigen 2021):

Newer devices contain less ferromagnetic material, have better filtering and sensing algorithms, and are more resistant to electromagnetic interference. A systematic review of 5,099 patients with non-MR-conditional devices who underwent 5,908 examinations found small changes in lead parameters and battery voltage. All 94 electrical resets (1.6% of scans) occurred in older generators, and 3 lead failures were reported, none directly attributable to MRI (Shah 2018; Kim 2024).

MR-conditional and legacy systems: evidence and the scan protocol

MR-conditional systems. Since 2011, manufacturers have offered pacemakers and ICDs labeled MR Conditional for the combination of generator and leads. The 2017 Heart Rhythm Society (HRS) consensus gives a Class 1 recommendation for MRI in MR-conditional systems scanned under their labeled conditions. It gives a Class 2a recommendation for systems that are not MR Conditional when there are no fractured, epicardial, or abandoned leads, MRI is the best test for the clinical question, and the institution has a protocol with a designated MR physician and CIED physician (Indik 2017; Vigen 2021).

Legacy systems. Two prospective registries at 1.5 T, MagnaSafe (1,500 nonthoracic examinations) and a study of 2,103 thoracic and nonthoracic examinations, found no long-term clinically significant adverse events under a standardized protocol (Russo 2017; Nazarian 2017). A series of 200 scans in patients with abandoned leads found no serious adverse events (Schaller 2021). In the United States, the Centers for Medicare & Medicaid Services (CMS) has covered MRI in patients with non-MR-conditional devices since 2018 when specified conditions are met, including a 1.5-T scanner in Normal Operating Mode; the coverage excludes patients with fractured, epicardial, or abandoned leads (Kim 2024).

The 2024 SCMR consensus groups patients into three categories (Kim 2024). The first is MR-conditional systems scanned within labeling. The second is non-MR-conditional systems, including an MR-conditional generator with leads outside its MR-conditional labeling, without abandoned, fractured, or epicardial leads, scanned at 1.5 T under an institutional protocol. The third is systems with abandoned or fractured intracardiac leads or surgically placed epicardial leads, which fall outside CMS coverage and are best scanned at experienced centers. The categories, registry details, and post-scan thresholds are presented on the device follow-up page under MRI in MR-conditional and non-MR-conditional systems. Pacemaker-specific and defibrillator-specific discussion appears under MRI in patients with pacemakers and MRI in patients with an ICD.

StepPhysical rationaleProtocol element
Identification of the generator and all leadsMR-conditional status applies only to a labeled combinationDevice records or manufacturer query; chest radiograph for abandoned or fractured leads
Time since implantationMost data exclude recently implanted leadsSCMR advises against scanning within 6 weeks of implantation unless the indication is compelling; ISMRM treats the 6-week wait as optional after a normal first device check
Pre-scan interrogationBaseline for threshold, sensing, impedance, batteryInterrogation of the generator and all leads before the scan; a battery near replacement and abnormal lead values both increase risk
Programming by pacing dependenceRF and gradients can be sensed and inhibit outputAsynchronous pacing (VOO or DOO) above the intrinsic rate if pacing dependent; nonpacing (OVO, ODO) or inhibited (VVI, DDI) mode otherwise
Tachytherapy, magnet response, rate response offPrevents inappropriate shocks and reed-switch effectsICD, CRT-D, and subcutaneous ICD therapies disabled
Field strength and RF limitsLead heating rises with RF exposure and field strengthLegacy systems at 1.5 T, Normal Operating Mode for SAR and dB/dt, body transmit coil; MR-conditional systems per label
Monitoring and personnelArrhythmia or loss of capture during scanningECG and pulse oximetry, voice and visual contact, staff trained in advanced cardiac life support; external defibrillator and programmer just outside Zone III
Post-scan interrogation and restorationDetects threshold rise, sensing change, resetInterrogation immediately after the scan and reprogramming to clinical settings

Physical basis of the CIED scan protocol, after the 2017 HRS consensus, the 2020 SCMR protocols, the 2024 SCMR CIED consensus, and the ISMRM recommendations (Vigen 2021). CRT-D, cardiac resynchronization therapy defibrillator.

Imaging the heart in patients with devices: wideband LGE and sequence choice

A generator distorts B0 over a large region, and the anterior and lateral left ventricle lie close to a left pectoral device. Several adjustments reduce artifact (Kramer 2020):

When artifact persists, the affected segments are reported as nondiagnostic, because signal loss can be misread as wall thinning and bright inversion artifact as enhancement.

Gadolinium-based contrast agents: chemistry, relaxivity, and dosing

Free gadolinium ions are toxic, so clinical agents bind gadolinium in a chelate. Linear agents wrap the ion in an open-chain ligand; macrocyclic agents cage it in a ring, which gives greater kinetic stability and less release of free gadolinium. Ionic and nonionic forms differ in osmolality. All agents used for cardiac imaging are extracellular: they distribute from plasma into the interstitium within minutes, do not enter intact cells, and are excreted mainly by glomerular filtration. Relaxivity, the shortening of T1 per unit concentration, differs among agents; agents with higher relaxivity achieve the same effect at lower dose (Kramer 2020). Gadobutrol is formulated at 1 mmol/mL, twice the concentration of most others, so its injection volume and rate are halved (Kramer 2020).

Dosing for CMR. SCMR gives 0.05 to 0.1 mmol/kg per perfusion injection, a total of 0.1 to 0.2 mmol/kg for LGE, 0.1 to 0.2 mmol/kg for thoracic angiography, and 0.05 mmol/kg for time-resolved angiography (Kramer 2020). The lowest dose that achieves adequate image quality is used, and noncontrast techniques are used where they can replace contrast (Kramer 2020). In the United States, gadobutrol carries a labeled indication for cardiac MRI to assess myocardial perfusion at stress and rest, and late gadolinium enhancement, in adults with known or suspected coronary artery disease. The labeled dose is 0.05 mmol/kg at peak pharmacologic stress followed by 0.05 mmol/kg at rest. Other agents are used for CMR off label (Gadavist prescribing information 2026).

AgentStructureACR NSF groupNotes for CMR
GadobutrolMacrocyclic, nonionicII1 mmol/mL; US label includes cardiac MRI perfusion and LGE in adults with known or suspected CAD
Gadoterate (gadoteric acid)Macrocyclic, ionicII0.5 mmol/mL
GadoteridolMacrocyclic, nonionicII0.5 mmol/mL
GadopiclenolMacrocyclicII* (provisional)Limited data when the estimated glomerular filtration rate (eGFR) is below 30 mL/min/1.73 m²; practices may treat it as group II or III
GadobenateLinear, ionicIIGroup II despite linear structure
GadoxetateLinear, hepatobiliaryIILiver-specific agent; not used for CMR
Gadodiamide, gadopentetate, gadoversetamideLinearIMost NSF cases; contraindicated in CKD stage 4 or 5 and on dialysis; avoided in known or suspected AKI (FDA label: contraindicated in AKI)

Classification of gadolinium-based agents by nephrogenic systemic fibrosis (NSF) risk from the ACR Manual on Contrast Media (2026 edition, NSF chapter updated 2025); no agent currently occupies group III. ACR risk grouping and chelate structure are distinct classifications. AKI, acute kidney injury; CAD, coronary artery disease; CKD, chronic kidney disease.

Nephrogenic systemic fibrosis and kidney disease

Nephrogenic systemic fibrosis (NSF) is a fibrosing disorder of the skin and subcutaneous tissues that can involve the lungs, esophagus, heart, and skeletal muscle, causing skin thickening, contractures, and in some patients death. It was first described in 2000 and linked to gadolinium exposure in patients with advanced kidney disease in 2006 (ACR Manual on Contrast Media 2026). Patients with CKD stage 4 or 5 have an estimated 1% to 7% risk after one or more exposures to group I agents. Most cases occurred in patients on dialysis, and 12% to 20% of confirmed cases occurred in patients with acute kidney injury (ACR Manual on Contrast Media 2026).

With group II agents the risk is very low or absent. A meta-analysis of 4,931 patients with CKD stage 4 or 5 who received a group II agent found no cases of NSF (upper bound of the 95% confidence interval, 0.07%) (Woolen 2020).

The ACR recommendations follow from these data (ACR Manual on Contrast Media 2026):

The US label of gadobutrol carries a boxed warning that NSF risk is highest with a glomerular filtration rate below 30 mL/min/1.73 m² or acute kidney injury (Gadavist prescribing information 2026).

Gadolinium retention, acute reactions, and regulatory actions

Retention. Gadolinium is retained for months to years in brain, bone, skin, and other organs after exposure, more with linear than macrocyclic agents. No adverse health effect from retention has been confirmed (ACR Manual on Contrast Media 2026). In 2017 the European Medicines Agency suspended the marketing authorizations of multipurpose linear agents, and the FDA retained all agents with class warnings and a Medication Guide describing greater retention with linear agents (Kramer 2020). This is the main difference between European and US regulation. Repeated contrast studies are limited to those with a clinical indication, and prior exposure alone does not justify withholding an indicated contrast examination.

Acute reactions. Allergic-like reactions to gadolinium agents occur in about 0.004% to 0.7% of injections and resemble reactions to iodinated contrast; severe life-threatening reactions occur in about 0.001% to 0.01% (ACR Manual on Contrast Media 2026). Patients with a prior allergic-like reaction to a gadolinium agent are at higher risk of recurrence. Recognition and treatment follow the ACR contrast reaction protocols, and the treatment equipment and medications are kept outside Zone IV.

Intrathecal administration of gadolinium agents can cause death, coma, encephalopathy, and seizures, and is not an approved route (Gadavist prescribing information 2026).

Gadolinium in pregnancy and breastfeeding

Gadolinium agents are likely to cross the placenta, are excreted by the fetus into the amniotic fluid, and are swallowed, so a small amount may recirculate (ACR Manual on Contrast Media 2026). In the Ontario cohort, gadolinium-enhanced MRI at any time in pregnancy (397 exposures) was associated with a higher risk of a broad group of rheumatologic, inflammatory, or infiltrative skin conditions in childhood (adjusted hazard ratio 1.36) and with stillbirth or neonatal death. The number of deaths was small, and the comparison group had no MRI at all (Ray 2016). ACR recommends gadolinium in pregnancy only when its use is considered critical and the potential benefit justifies the unknown fetal risk. The decision follows discussion among the referring clinician, radiologist, and patient, and a group II agent is given at the lowest diagnostic dose (ACR Manual on Contrast Media 2026). Noncontrast CMR answers most cardiac questions in pregnancy.

Less than 0.0004% of a maternal dose of gadolinium agent is estimated to reach the infant's circulation through breast milk, and ACR states that breastfeeding need not be interrupted for 24 hours after maternal administration; the decision about expressing and discarding milk rests with the informed patient (ACR Manual on Contrast Media 2026).

Noncontrast CMR when gadolinium is withheld

When gadolinium is contraindicated, declined, or best avoided, much of the examination remains: bSSFP cine for volumes, mass, and function; native T1 and T2 mapping for diffuse fibrosis, edema, amyloid, and Anderson-Fabry disease; and T2* for iron. The examination also includes phase-contrast flow for regurgitation, shunts, and cardiac output; black-blood imaging for masses and the pericardium; real-time cine for ventricular interdependence; noncontrast 3D whole-heart imaging for anatomy and coronary origins; and tagging or feature tracking for deformation (Kramer 2020). Scar cannot be imaged directly without contrast, but native T1 mapping detects many myocardial abnormalities in the same regions. The lack of LGE and ECV is stated in the report, since a normal noncontrast study does not exclude scar.

Sources of error in CMR acquisition and interpretation

References

  1. Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. doi:10.1186/s12968-020-00607-1
  2. Messroghli DR, Moon JC, Ferreira VM, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19:75. doi:10.1186/s12968-017-0389-8. Correction: J Cardiovasc Magn Reson. 2018;20:9. doi:10.1186/s12968-017-0408-9
  3. Ridgway JP. Cardiovascular magnetic resonance physics for clinicians: part I. J Cardiovasc Magn Reson. 2010;12:71. doi:10.1186/1532-429X-12-71
  4. Biglands JD, Radjenovic A, Ridgway JP. Cardiovascular magnetic resonance physics for clinicians: part II. J Cardiovasc Magn Reson. 2012;14:66. doi:10.1186/1532-429X-14-66
  5. Ferreira PF, Gatehouse PD, Mohiaddin RH, Firmin DN. Cardiovascular magnetic resonance artefacts. J Cardiovasc Magn Reson. 2013;15:41. doi:10.1186/1532-429X-15-41
  6. Simonetti OP, Kim RJ, Fieno DS, et al. An improved MR imaging technique for the visualization of myocardial infarction. Radiology. 2001;218:215–223. doi:10.1148/radiology.218.1.r01ja50215
  7. Kellman P, Arai AE, McVeigh ER, Aletras AH. Phase-sensitive inversion recovery for detecting myocardial infarction using gadolinium-delayed hyperenhancement. Magn Reson Med. 2002;47:372–383. doi:10.1002/mrm.10051
  8. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52:141–146. doi:10.1002/mrm.20110
  9. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. PubMed 10542355
  10. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202–1210. doi:10.1002/mrm.10171
  11. Fischer SE, Wickline SA, Lorenz CH. Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions. Magn Reson Med. 1999;42:361–370. PubMed 10440961
  12. Campbell-Washburn AE, Ramasawmy R, Restivo MC, et al. Opportunities in interventional and diagnostic imaging by using high-performance low-field-strength MRI. Radiology. 2019;293:384–393. doi:10.1148/radiol.2019190452
  13. Ferreira VM, Schulz-Menger J, Holmvang G, et al. Cardiovascular magnetic resonance in nonischemic myocardial inflammation: expert recommendations. J Am Coll Cardiol. 2018;72:3158–3176. doi:10.1016/j.jacc.2018.09.072
  14. Anderson LJ, Holden S, Davis B, et al. Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J. 2001;22:2171–2179. doi:10.1053/euhj.2001.2822
  15. Bissell MM, Raimondi F, Ait Ali L, et al. 4D flow cardiovascular magnetic resonance consensus statement: 2023 update. J Cardiovasc Magn Reson. 2023;25:40. doi:10.1186/s12968-023-00942-z. Corrigendum: J Cardiovasc Magn Reson. 2026;28:101912. doi:10.1016/j.jocmr.2025.101912
  16. American College of Radiology Committee on MR Safety. ACR Manual on MR Safety. 2026 edition. acr.org
  17. American College of Radiology Committee on Drugs and Contrast Media. ACR Manual on Contrast Media. 2026 edition (nephrogenic systemic fibrosis chapter updated 2025). acr.org
  18. US Food and Drug Administration. Criteria for significant risk investigations of magnetic resonance diagnostic devices: guidance for industry and Food and Drug Administration staff. June 20, 2014. fda.gov
  19. Indik JH, Gimbel JR, Abe H, et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017;14:e97–e153. doi:10.1016/j.hrthm.2017.04.025
  20. Kim D, Collins JD, White JA, et al. SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device. J Cardiovasc Magn Reson. 2024;26:100995. doi:10.1016/j.jocmr.2024.100995
  21. Vigen KK, Reeder SB, Hood MN, et al; ISMRM Safety Committee. Recommendations for imaging patients with cardiac implantable electronic devices (CIEDs). J Magn Reson Imaging. 2021;53:1311–1317. doi:10.1002/jmri.27320
  22. Russo RJ, Costa HS, Silva PD, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376:755–764. doi:10.1056/NEJMoa1603265
  23. Nazarian S, Hansford R, Rahsepar AA, et al. Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med. 2017;377:2555–2564. doi:10.1056/NEJMoa1604267
  24. Shah AD, Morris MA, Hirsh DS, et al. Magnetic resonance imaging safety in nonconditional pacemaker and defibrillator recipients: a meta-analysis and systematic review. Heart Rhythm. 2018;15:1001–1008. doi:10.1016/j.hrthm.2018.02.019
  25. Schaller RD, Brunker T, Riley MP, Marchlinski FE, Nazarian S, Litt H. Magnetic resonance imaging in patients with cardiac implantable electronic devices with abandoned leads. JAMA Cardiol. 2021;6:549–556. doi:10.1001/jamacardio.2020.7572
  26. Rashid S, Rapacchi S, Vaseghi M, et al. Improved late gadolinium enhancement MR imaging for patients with implanted cardiac devices. Radiology. 2014;270:269–274. doi:10.1148/radiol.13130942
  27. Rashid S, Rapacchi S, Shivkumar K, Plotnik A, Finn JP, Hu P. Modified wideband three-dimensional late gadolinium enhancement MRI for patients with implantable cardiac devices. Magn Reson Med. 2016;75:572–584. doi:10.1002/mrm.25601
  28. Shellock FG. Guidelines for the management of patients with coronary artery stents referred for MRI examinations. ISMRM Society for MR Radiographers and Technologists safety resources; 2025. ismrm.org
  29. Woolen SA, Shankar PR, Gagnier JJ, MacEachern MP, Singer L, Davenport MS. Risk of nephrogenic systemic fibrosis in patients with stage 4 or 5 chronic kidney disease receiving a group II gadolinium-based contrast agent: a systematic review and meta-analysis. JAMA Intern Med. 2020;180:223–230. doi:10.1001/jamainternmed.2019.5284
  30. Ray JG, Vermeulen MJ, Bharatha A, Montanera WJ, Park AL. Association between MRI exposure during pregnancy and fetal and childhood outcomes. JAMA. 2016;316:952–961. doi:10.1001/jama.2016.12126
  31. Bayer HealthCare Pharmaceuticals. Gadavist (gadobutrol) injection, prescribing information. Revised February 2026. DailyMed
  32. GE Healthcare. Omniscan (gadodiamide) injection, prescribing information. Revised March 2025. DailyMed
  33. Medtronic. Reveal LINQ insertable cardiac monitor: MRI conditions of use. medtronic.com